SURPASS™ is a universal adhesive system that when properly employed, can be used to seal enamel/dentin prior to restoring with light-cured composite materials. Also, in situations of indirect restorations, the bonding system can be used to seal a preparation when using a light-cured, self-cured or dual-cured composite cement or glass ionomer or resin-modified glass ionomer cement. In addition, the system allows the dentist to bond a post and core, along with the use to treat hypersensitive and/or exposed root surfaces.
Device Story
SURPASS™ is a light-cured, self-etching dental adhesive system used by dentists in clinical settings. It consists of a multi-step application: an acidic conditioner (SURPASS™ 1), an ethanol-based adhesive resin (SURPASS™ 2), and a tie layer (SURPASS™ 3). The system is applied to prepared dentin/enamel surfaces to facilitate bonding of restorative materials. The conditioner prepares the substrate, the adhesive resin ensures proper moisture at the interface, and the tie layer acts as an interface between the adhesive and the filling material. The final layer is light-cured for 20 seconds. The device simplifies bonding protocols, reduces technique sensitivity, and provides predictable clinical results for direct and indirect restorations, post and core bonding, and root desensitization.
Clinical Evidence
Bench testing only. Evidence includes physical testing observations and analysis of shear bond strength to dentin and enamel compared to the predicate device. No clinical data provided.
Technological Characteristics
Light-cured, self-etching dental adhesive system. Components: acidic conditioner, ethanol-based adhesive resin, and unfilled bonding resin tie layer. Polymerization via free radical mechanism using photo initiator and chemical activator. Light-cure requirement: 20 seconds. Form factor: multi-component liquid system.
Indications for Use
Indicated for sealing enamel/dentin prior to light-cured or self-cured composite restorations; sealing preparations for indirect restorations using light-cured, self-cured, or dual-cured composite, glass ionomer, or resin-modified glass ionomer cements; desensitization of hypersensitive/exposed root surfaces; and bonding post and core. Prescription use only.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
107 4
KD6/981
JUL 3 1 2006
# Abbreviated 510 (K) SUMMARY
### As Required by the Safe Medical Devices Act of 1990
Apex Dental Materials, Inc. 23329 Mallard Court Deer Park, IL 60010 Phone: (877) 273-9123
#### July 12th, 2006 Abbreviated 510 (K) Submission Date:
II KLE
Chris Kulton Contact Person:
Device Name:
Trade Name: Common Name: Classification Name: SURPASS™ Dental Bonding Adhesive Resin Tooth Bonding Agent, per 21 CFR parts 872.3200
#### Classification:
| Regulatory Class: | |
|-------------------|--|
| Product Code: | |
# IDENTIFICATION OF THE LEGALLY MARKETED PREDICATE DEVICE
### PREDICATE DEVICE
ECLIPTOMER (510K number K945604, Bisco, Inc.) is a universal dental adhesive system that
is designed to bond composite to dentin, enamel, cast metals, treated porcelain and set amalgam.
This system is also designed for indirect techniques as well as amalgam bonding.
ECLIPTOMER (510K number K945604, Bisco, Inc.) is an acetone-based primer that is
dependent on a clean etched dentin/enamel surface that is visibly moist. Its physical properties
are similar to the applicant device and uses are identical. Like the applicant device,
ECLIPTOMER (510K number K945604, Bisco, Inc.) is one component used in conjunction with
Abbrevaited 510K Submission for SURPASSTM Apex Dental Materials, 23329 Mallard Court Deer Park, IL 60010
Page 12 of 15
{2}------------------------------------------------
K06/981
2 of 4
a complete dental bonding system. It hardens by a light cure polymerization mechanism employing a photo initiator, and a chemical activator.
Summary continued:
### DESCRIPTION OF APPLICATION DEVICE
SURPASS™ is a light cured (free radical polymerization), self-etching dental adhesive system. designed to provide a dentist with an increased ease of use that is less technique sensitive. The bonding system combines both the etchant and primer applications into one simple step followed by a thin layer of unfilled bonding resin. Thus, allowing for more predictable clinical results without sacrificing bond strength integrity.
SURPASS™ 2 is an ethanol-based adhesive that is dependent on the prepared surface of its conditioner (SURPASS™ 1). This acidic conditioner application is a procedure that is very similar to the phosphoric acid etching process of dentin/enamel substrate in the ECLIPTOMER (510K number K945604, Bisco, Inc.) system. Applying the SURPASS™ 2 (adhesive resin) directly to the conditioner prepared surface, assures proper moistness on the bonding interface, resulting in a reliable simplified bonding protocol. Once SURPASS™ 2 has been dried a thin layer of SURPASS™ 3 is applied to the SURPASS™ 2 surface and light cured for 20 seconds. The SURPASS™ 3 acts as a superior tie layer between SURPASS™2 and the filling material of choice.
Abbrevaited 510K Submission for SURPASSTM Apex Dental Materials. 23329 Mallard Court Deer Park, IL 60010
Page 13 of 15
{3}------------------------------------------------
K06/98/
3074
Summary continued:
### INTENDED USES OF APPLICANT DEVICE
SURPASS™ is a universal adhesive system that when properly employed, can be used to seal enamel/dentin prior to restoring with light-cured composite materials. Also, in situations of indirect restorations, the bonding system can be used to seal a preparation when using a light-cured, self-cured or dual-cured composite cement or glass ionomer or resinmodified glass ionomer cement. In addition, the system allows the dentist to bond a post and core, along with the use to treat hypersensitive and/or exposed root surfaces.
# PERFORMANCE CHARACTERISTICS and CONCEPTS
SURPASS™ has similar handling to the ECLIPTOMER (510K number K945604, Bisco, Inc.) adhesive system. From the physical testing observations and analysis, including shear bond strength to dentin and enamel, we suggest that SURPASS™ is substantially equivalent to ECLIPTOMER (510K number K945604, Bisco, Inc.). Along with this we would suggest the individual components of SURPASS™ are long-time industry standards and are utilized in numerous dental bonding systems currently marketed in the United States (see Confidential Formulation Details on page 5). In addition, SURPASS™ is essentially our Simplicity® Adhesive System (510K: K020570) with an added bonding resin (SURPASS™ 3). SURPASS™ has the same indication for use and same base composition as the Simplicity® Adhesive System
(510K: K020570).
Abbrevaited 510K Submission for SURPASSTM Apex Dental Materials, 23329 Mallard Court Deer Park, IL 60010
Page 14 of 15
{4}------------------------------------------------
K06/98/
4084
# Indications for Use
510(K) Number (if known):
SURPASSTM Device name:
Indications For Use:
- . Restorations: to seal enamel/dentin prior to restoring with light-cured or self-cured composite materials.
- Indirect Restorations: to seal a preparation when using a light-cured, self-cured or dual-cured . composite cement or glass ionomer or resin-modified glass ionomer cement.
- . Desensitization: to treat hypersensitive and/or exposed root surfaces.
- To bond in a post and core .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRL, Office of Device Evaluation (ODE)
Prescription use X (Per 21 CFR 801.109 OR
Over- The- Counter Use
(Optional Format 1-2-96)
Page 4 of 15
Abbrevaited 510K Submission for SURPASSTM Apex Dental Materials, 23329 Mallard Court Deer Park, IL 60010
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services, USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. In the center of the seal is an abstract symbol that resembles an eagle or bird with three stylized wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUL 3 1 2006
Mr. Chris Kulton Apex Dental Materials, Incorporated 23329 Mallard Court Deer Park, Illinois 60010
Re: K061981
Trade/Device Name: Surpass™M Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: II Product Code: KLE Dated: July 12, 2006 Received: July 17, 2006
Dear Mr. Kulton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2 - Mr. Kulton
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K061981
# Indications for Use
510(K) Number (if known):
SURPASS™ Device name:
Indications For Use:
- · Restorations: to seal enamel/dentin prior to restoring with light-cured or self-cured composite materials.
- Indirect Restorations: to seal a preparation when using a light-cured, self-cured or dual-cured . composite cement or glass ionomer or resin-modified glass ionomer cement.
- Desensitization: to treat hypersensitive and/or exposed root surfaces. ◆
- To bond in a post and core .
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRL, Office of Device Evaluation (ODE)
Prescription use X (Per 21 CFR 801.109
Apex Dental Materials,
23329 Mallard Court
Deer Park, IL 60010
Abbrevaited 510K Submission for SURPASSTM
OR
Over- The- Counter Use
(Optional Format 1-2-96)
Mubey for MSK Page 4 of 15
..un Sign-Off)
on of Anesthesiology, General Hospital.
cuon Control, Dental Devices
K06198)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.