← Product Code [HRS](/productcode/HRS) · K061973

# SYNTHES 3.5MM SPRING PLATE (K061973)

_Synthes (Usa) · HRS · Sep 11, 2006 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K061973

## Device Facts

- **Applicant:** Synthes (Usa)
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Sep 11, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Synthes 3.5mm Spring Plate is intended for pelvic and acetabular reconstructive surgery and fracture fixation of the distal fibula.

## Device Story

The Synthes 3.5mm Spring Plate is a metallic bone fixation device used in orthopedic surgery. It is a variation of the one-third tubular plate, featuring two sharp spikes on the bottom surface and a pre-bent convex shape to facilitate the reduction of small bone fragments and conform to uneven bony surfaces. The plate is available in 1-10 hole configurations with lengths from 19.5mm to 132mm and accepts 3.5mm cortex or pelvic screws. It is used by surgeons in clinical settings to stabilize fractures or assist in reconstruction. The device provides mechanical support to bone segments, aiding in the healing process.

## Clinical Evidence

bench testing only

## Technological Characteristics

Materials: Stainless Steel and Commercially Pure Titanium. Design: 1-10 hole plate, 19.5mm-132mm length, pre-bent convex shape with two sharp spikes. Fixation: Accepts 3.5mm cortex or pelvic screws. Sterilization: Provided sterile and non-sterile.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes One-Third Tubular LCP Plates, with collar

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "SYNTHES" in a bold, sans-serif font. To the left of the word is a circular logo with a stylized design. A horizontal line is present beneath the word, adding a sense of completion to the design. The overall impression is clean and professional.

SEP 1 2006

## 3.0 510(k) Summary

:

Page _________of _____________________________________________________________________________________________________________________________________________________________

| Sponsor:                    | Synthes (USA)<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>(610) 719-6500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |  |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Device Name:                | Synthes 3.5mm Spring Plate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |  |  |
| Classification:             | Class II, §888.3030 - Plate, Fixation, Bone (HRS)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |  |
| Predicate Device:           | Synthes One-Third Tubular LCP Plates, with collar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |  |
| Device Description:         | The Synthes 3.5mm Spring Plate is a variation of the Synthes<br>one-third tubular plate with collar which utilizes two sharp spikes<br>at the bottom surface and a pre-bent convex shape to aid in the<br>reduction of small bone fragments while conforming to bony<br>uneven surfaces. The plate incorporates a 1 - 10 hole design in<br>lengths ranging from 19.5mm - 132mm and accepts either 3.5mm<br>cortex or pelvic screws. In addition the plates are manufactured<br>from Stainless Steel and Commercially Pure Titanium and<br>provided STERILE and NON STERILE. |  |  |
| Intended Use:               | The Synthes 3.5mm Spring Plate is intended for pelvic and<br>acetabular reconstructive surgery and fracture fixation of the distal<br>fibula.                                                                                                                                                                                                                                                                                                                                                                                                                               |  |  |
| Substantial<br>Equivalence: | Comparative information presented supports substantial<br>equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |  |

:

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized representation of an eagle, which is a common symbol of the United States. The eagle is depicted with three lines forming its body and wings, giving it a modern and abstract appearance. 

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## SEP 1 1 2006

Synthes (USA) % Ms. Deborah Jackson Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380

Re: K061973 Trade/Device Name: Synthes (USA) 3.5mm Spring Plate Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: July 7, 2006 Received: July 12, 2006

Dear Ms. Jackson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 -- Ms. Deborah Jackson

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Adubare Buelum

Mark N. Melkerson Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A horizontal line runs underneath the word and logo.

ﺑ

| 2.0 | Indications for Use |
|-----|---------------------|
|-----|---------------------|

| 510(k) Number (if known): | K061973                                                                                                                                 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:              | Synthes (USA) 3.5mm Spring Plate                                                                                                        |
| Indications for Use:      | The Synthes 3.5mm Spring Plate is intended for pelvic and acetabular reconstructive surgery and fracture fixation of the distal fibula. |

| Prescription Use     | X |
|----------------------|---|
| (Per 21 CFR 801.109) |   |

AND/OR

| Over-The-Counter Use   |  |
|------------------------|--|
| (21 CFR 807 Subpart C) |  |

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Barbara Inelus

Division Sign-Off Division of General, Restorative and Neurological Devices

510(k) Number 406/973

---

**Source:** [https://fda.innolitics.com/device/K061973](https://fda.innolitics.com/device/K061973)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
