The 21.3inch Monochrome LCD Monitor is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians. The 21.3inch Monochrome LCD Monitor must not be used for primary diagnostic in mammography.
Device Story
21.3-inch monochrome LCD monitor; utilizes Super Fine TFT (SA-SFT) panel technology; features low-reflection overcoat for improved contrast and black representation. Device supports portrait or landscape orientation via rotation; includes tilt, swivel, and height-adjustable stand. Input: digital image data. Processing: 10-bit gamma correction for grayscale tuning; internal power supply conversion. Output: high-resolution grayscale images (up to 1024 simultaneous grayscales from 3061 palette). Used by physicians in clinical settings for diagnostic review. Benefits: improved diagnostic accuracy through precise grayscale reproduction and wide viewing angles (178°).
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
21.3-inch monochrome LCD; Super Fine TFT (SA-SFT) panel; 10-bit gamma correction; 178° viewing angle; 1600x1200 (landscape) or 1200x1600 (portrait) resolution; internal power supply; UL60601 safety standard compliance.
Indications for Use
Indicated for displaying and viewing digital images for diagnosis by trained physicians. Contraindicated for primary diagnostic use in mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Barco Coronis 2MP Medical Flat Panel Display System (K023322)
Submission Summary (Full Text)
{0}------------------------------------------------
K06/947
#### ·Section E -
## 510(k) Summary
E-1. 510(k) Application Date
March, 04. 2006
JUL 25 2006
E-2. Manufacturer
Submitter Name : NEC Display Solutions, Ltd.
Address : 4-13-23 Shibaura, Minato-ku, Tokyo, 108-0023 Japan
E-3. Contact Person
Name : Shuichi Kino (Mr.)
Manager
Engineering Administration Section, Planning and Administration Department R&D and Product Supply Division
Address : 686-1, Nishioi, Oi-machi, Ashigarakami-gun, Kanagawa, 258-8533 Japan Tel : +81-465-85-2376
Fax : +81-465-85-2378
E-4. Device
Device Name : 21.3inch Monochrome LCD Monitor Model : NEC MD21GS-2MP-CB
NEC MD21GS-2MP-BB
MITSUBISHI MD211GS2P-CB
## MITSUBISHI MD211GS2P-BB
These are identical each other, except for mode! designation..
E-5. Common Name
Monochrome LCD Monitor
## E-6. Registration Number
Manufacturer: NEC Display Solutions, Ltd. Registration Number: 3003623028 Factory: NPG DISPLAY (DONG GUAN) CO., LTD. Registration Number: 3002808782
{1}------------------------------------------------
## E-7. Factory
Name : NPG DISPLAY (DONG GUAN) Co., LTD. Address : Jin Xing Industrial Zone, Qing Xi Zhen, Dong Guan, Guang Dong Sheng 511746, P.R. China
## E-8. Classification
## Device Class : Class II
Classification Name : System, image processing, Radiological Regulation Number : 21 CFR Part 892 Radiology Devices Subpart B Section 892.2050 Device Description : Picture Archiving and Communications System Product code : 90 LLZ
## E-9. Reason for Submission
First time submission to market this LCD Monitor as Medical Device in the USA.
## E-10. Description of Marketed Device
Device Name : Barco Coronis 2MP Medical Flat Panel Display System
510(k) number : K023322
Intended use :
The Barco Coronis 2MP Medical Flat Panel Display System is intended to be used in displaying and viewing digital images for review by trained medical practitioners.
#### Device description :
The monitor combines a TFT(thin film transistor) liquid crystal display panel structure and a built-in backlight with inverter for a better picture quality.
The monitor can be used in portrait or landscape version, simply by turning the panel. The tilt and swivel foot allows ideal positioning of the panel, in height and viewing angle. The image brightness can be adjusted by means of a control wheel on the monitor.
The 21.3inch Monochrome LOD Monitor has the same intended use, technical characteristics and performance as the legally marketed device (K023322) and thus Substantial Equivalence is given.
{2}------------------------------------------------
E-11. Substantial Equivalence Comparison
Device Description :
The LCD panel adopts Super Fine TFT(SA-SFT) for the MD21GS-2MP-CB, BB and MD211GS2P-CB, BB.
Additionally, a low reflection overcoat is applied on top of the bead layer, which yields a low, uniform reflection over the surface of the glass.
This process also provides for a truer representation of black and a higher contrast ratio.
The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB can select the use of the landscape or the portrait only by rotating the LCD panel.
The tilt and swivel foot allows ideal positioning of the panel, in height and viewing angle,
In addition, because our device has the height adjustment function of the stand, a more ideal positioning can be offered.
Please refer to "Table E-11".
Intended Use :
The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB are intended to be used for displaying and viewing of digital images for diagnosis by trained physicians.
Barco Coronis 2MP is intended to be used in displaying and viewing digital images for review by trained medical practitioners.
It is intended to be used to support the diagnosis to either device.
Technical Characteristics and performance :
The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB can select the use of 1600*1200 of the landscape modes or the portrait modes of 1200*1600 only by rotating the panel equally to the legally marketed device.
Barco Coronis 2MP has the viewing angle of 170°.
The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB greatly decreases the shift of the Gamma characteristic by the view corner by the wide viewing angle of 178° by the adoption of the latest SA-SFT panel.
The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB provides support for improved diagnostic accuracy with simultaneous reproduction of up to 1024 from a palette of 3061 possible grayscales.
The 10-bit gamma correction ensures precise and smooth grayscale tuning and better representation of just noticeable difference.
Up to 1024 grayscales can be depicted at the same time for film-less diagnosis.
{3}------------------------------------------------
The MD21GS-2MP-C8, BB and MD211GS2P-CB, BB are different in the following points compared with the legally marketed device.
Barco Coronis 2MP in difference of built-in circuit of power supply instead of AC adapter.
In Barco Coronis 2MP, the AC voltage converts into the DC voltage with the AC adaptor and is supplied to the device.
In The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB, the AC voltage converts into the DC voltage in the device.
Therefore, the monitor is operated by the dc voltage, and is substantial equivalent to the legally marketed device.
The Barco Coronis 2MP has the insulation structure which meets the standard of UL60950.
However, The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB has the insulation structure which meets the standard of UL60601.
The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB are applied a severer safety standard compared with the legally marketed device.
Therefore, our monitor safety is more excellent than the legally marketed device.
Therefore, The MD21GS-2MP-CB, BB and MD211GS2P-CB, BB are device with higher safety compared with the legally marketed device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol featuring a staff with two snakes entwined around it. The caduceus is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular fashion around the symbol. The text is in all caps and is smaller than the caduceus symbol.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
JUL 25 2006
NEC Display Solutions, Ltd. % Mr. Tamas Borsai Division Manager, Medical Division TUV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470
Re: K061947
Trade/Device Name: 21.3 inch Monochrome LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 8 2006 Received: July 10, 2006
Dear Mr. Borsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/10 description: The image shows a circular logo with the text "1906-2006 FDA Centennial" inside. The letters "FDA" are prominently displayed in the center of the logo. Three stars are arranged below the word "Centennial".
Protecting and Promoting Public Health
{5}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.
Sincerely yours,
Nancy Brogdon
Nancy C. Brogdon
Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### ·Section D - Indications for use
D-1. Indications for use
# INDICATIONS FOR USE
510(K)Number : Not Known
Device Name : 21.3inch Monochrome LCD Monitor
Model : NEC MD21GS-2MP-CB
NEC MD21GS-2MP-BB MITSUBISHI MD211GS2P-CB MITSUBISHI MD211GS2P-B8
Indications for Use :
The 21.3inch Monochrome LCD Monitor is intended to be used for displaying and viewing of digital images for diagnosis by trained physicians.
The 21.3inch Monochrome LCD Monitor must not be used for primary diagnostic in mammography.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Marcel Burdin
(Division Sign-Off) ിന്നും of Reproductive, Abdominal and Radiological Devices 5 10(k) Number
DI
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.