← Product Code [LRC](/productcode/LRC) · K061903

# RELIEVA SINUS BALLOON CATHETER (K061903)

_Acclarent, Inc. · LRC · Aug 18, 2006 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K061903

## Device Facts

- **Applicant:** Acclarent, Inc.
- **Product Code:** [LRC](/productcode/LRC.md)
- **Decision Date:** Aug 18, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 874.4420
- **Device Class:** Class 1
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Indications for Use

To provide a means to access the sinus space and to dilate the sinus ostia and spaces within the paranasal sinus cavities for diagnostic and therapeutic procedures.

## Device Story

Sinus Balloon Catheter—Integrated Wire combines access and dilation functions into single device; eliminates need for separate guidewire and catheter exchange. Used by physicians in clinical settings to access sinus space and dilate sinus ostia/paranasal sinus cavities. Device facilitates diagnostic and therapeutic procedures by providing mechanical dilation of narrowed or obstructed sinus passages. Benefits include simplified procedural workflow and reduced device handling.

## Clinical Evidence

Bench testing only. Device met all performance testing acceptance criteria.

## Technological Characteristics

Manual surgical instrument; integrated sinus balloon catheter and guidewire. Class I device. Mechanical dilation principle. No software or electronic components.

## Regulatory Identification

An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.

## Predicate Devices

- Relieva Sinus Balloon Catheter ([K043527](/device/K043527.md))
- Relieva Sinus Guidewire ([K043445](/device/K043445.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top right of the "t", arcing over the word. The font is black and the background is white.

:

K061903

## APPENDIX A: 510(k) SUMMARY

| Sponsor/Submitter:                        | Acclarent, Inc.<br>1525-B O'Brien Drive<br>Menlo Park, California 94025                                                                                                                                                                                  | AUG 18 2006 |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact Person:                           | Keri Yen<br>Quality Engineer<br>Phone: (650) 687-5874<br>Fax: (650) 687-5889                                                                                                                                                                             |             |
| Date of Submission:                       | June 30, 2006                                                                                                                                                                                                                                            |             |
| Device Trade Name:                        | To be determined                                                                                                                                                                                                                                         |             |
| Common Name:                              | Sinus Balloon Catheter—Integrated Wire                                                                                                                                                                                                                   |             |
| Device Classification:                    | Class I                                                                                                                                                                                                                                                  |             |
| Regulation Number:                        | 21 CFR 874.4420                                                                                                                                                                                                                                          |             |
| Classification Name:                      | ENT Manual Surgical Instrument                                                                                                                                                                                                                           |             |
| Product Code:                             | LRC                                                                                                                                                                                                                                                      |             |
| Predicate Device:                         | Relieva Sinus Balloon Catheter (K043527)<br>Relieva Sinus Guidewire (K043445)                                                                                                                                                                            |             |
| Device Description:                       | The Sinus Balloon Catheter—Integrated Wire is a sinus balloon catheter that<br>has an integrated guidewire. The Sinus Balloon Catheter—Integrated Wire<br>allows access to and dilation of the sinus ostia and paranasal spaces with a<br>single device. |             |
| Indications for Use:                      | To provide a means to access the sinus space and to dilate the sinus ostia and<br>spaces within the paranasal sinus cavities for diagnostic and therapeutic<br>procedures.                                                                               |             |
| Technological<br>Characteristics          | The Sinus Balloon Catheter—Integrated Wire is a device that allows for the<br>capability to access and to dilate the sinus ostia with the same device.                                                                                                   |             |
| Performance Data                          | The Sinus Balloon Catheter—Integrated Wire met all performance testing<br>acceptance criteria.                                                                                                                                                           |             |
| Summary of<br>Substantial<br>Equivalence: | The Sinus Balloon Catheter—Integrated Wire is substantially equivalent to the<br>predicate devices as confirmed through relevant performance tests.                                                                                                      |             |

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## AUG 1 8 2006

Acclarent, Inc. c/o Keri Yen Quality Assurance Engineer 1525-B O'Brien Drive Menlo Park, CA 94025

Re: K061903

Trade/Device Name: Sinus Balloon Catheter – Integrated Wire Regulation Number: 21 CFR 874.4420 Regulation Name: ENT Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: July 31, 2006 Received: August 1, 2006

Dear Ms. Yen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

M.B. Eychlenis IWD

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nosc and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top of the "r" to the right, creating a visual element above the word. The text is black against a white background.

Sinus Balloon Catheter-Integrated Wire

APPENDIX B: INDICATIONS FOR USE STATEMENT

510(k) Number (if known):

K061903

Trade Name:

To be determined

Common Name:

Sinus Balloon Catheter--Integrated Wire

Indications For Use:

To provide a means to access the sinus space and to dilate the sinus ostia and spaces within the paranasal sinus cavities for diagnostic and therapeutic procedures.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

on of Ophthalmic Ear.

510(k) Number K061903

Page / of /

(Posted November 13, 2003)

Prescription Use (Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/device/K061903](https://fda.innolitics.com/device/K061903)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
