K061856 · Atom Medical, Inc. · FMZ · Sep 28, 2006 · General Hospital
Device Facts
Record ID
K061856
Device Name
V-2200 INFANT INCUBATOR
Applicant
Atom Medical, Inc.
Product Code
FMZ · General Hospital
Decision Date
Sep 28, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The V-2200 Infant Incubator is a device to keep premature infants or neonatal infants in a warm environment which is covered by a hood and isolated from ambient air and of which internal air temperature and humidity, are controlled. It is intended for inpatient use in maternity nurseries, and neonatal care environments of hospitals or other healthcare facilities.
Device Story
V-2200 Infant Incubator provides controlled thermal/humidified environment for premature/sick neonates; isolates infant from ambient air via hood. Components: hood, mattress platform, conditioning/humidity chambers, heater, fan, oxygen supply/filter, pulse oximeter, oxygen controller, weight monitor. Operation: air temperature control via heater/fan; skin temperature control via skin probe; humidity control via vapor generation. Optional pulse oximetry (Masimo technology) monitors SpO2/pulse rate non-invasively; oxygen controller regulates O2 concentration; weight monitor facilitates weighing without moving infant. Used in hospital maternity nurseries/neonatal care units by clinicians. Output: thermal/humidity regulation, SpO2/pulse data, O2 concentration, infant weight. Benefits: maintains stable physiological environment for vulnerable neonates; facilitates clinical observation/treatment.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Incubator with hood, conditioning/humidity chambers, heater, fan, and sensors. Optional integrated Masimo pulse oximeter, oxygen controller, and weight monitor. Connectivity: standalone. Sterilization: not specified. Software: embedded control logic for temperature, humidity, and O2 regulation.
Indications for Use
Indicated for premature or neonatal infants requiring a controlled thermal and humidified environment. Intended for inpatient use in hospital maternity nurseries and neonatal care units. Not for home use or transport. Prescription only.
Regulatory Classification
Identification
A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a date and a signature. The date reads "SEP 28 2006". The signature is illegible but appears to be a combination of letters and numbers, possibly "Ka01856".
#### 510(k) Summary of Safety and Effectiveness Section I
Applicant:
Atom Medical Inc lwakata Bldg 3rd Floor 3-18-15, Hungo Tokoyo, Bunkyo-ku 1130033 Japan Registration No: In process
Contact Person:
Neoforce Group, Inc 35 Commerce Drive lvyland, Pa. 18974
Mary Staniewicz Ph 215-672-6800 Fax 215-672-1123
Device trade/proprietary name:
V-2200 Infant Incubator
Device common/usual/classification name:
Infant Incubator
Classification:
General Hospital 21 CFR 880.5400 Infant Incubator, FMZ, Class II
Anesthesiology 21 CFR 870.2700 Oximeter, DQA, Class II
Performance Standards:
None applicable
Predicate Device:
10 10
. 8
:
K021809 V-2100G Infant Incubator K002355 V-505 Atom Infa Warmer K990966 Masimo Set Pulse Oximeter
{1}------------------------------------------------
## Device Description
This product consists of a hood, a sensor module, a mattress platform, a middle deck section, a conditioning chamber, a humidity chamber, an operation control section, a power source, a relay box, an oxygen supply/filter box, a pulse oximeter, an oxygen controller and a weight monitor. It is equipped with an incubator air temperature control function to circulate the air containing the heat energy generated by the heater attached to the conditioning chamber inside the hood by means of a fan in order to maintain the incubator air temperature at a fixed level. The device is also equipped with a skin temperature control function to maintain the infant's skin temperature at a fixed level in response to the patients temperature as measured by the skin probe. There is a humidity control function to adjust the amount of vapor generated in the humidity chamber. The following additional features can be installed into the incubator: a pulse oximeter function to monitor Sp02 concentration and pulse rate non-invasively; an oxygen control function to draw oxygen and outside air and control the oxygen concentration delivered to the patient compartment to the desired level; and a weight monitor to facilitate taking the infants weight without moving them from the incubator. This device is designed to be used in treatment, procedures and observation of low-birth-weight and sick neonates, providing heat to the neonate when the body temperature is low and also to provide humidity if desired
#### Intended Use
The V-2200 Infant Incubator is a device to keep premature infants or neonatal infants in a warm environment which is covered by a hood and isolated from ambient air and of which internal air temperature and humidity, are controlled. It is intended for inpatient use in maternity nurseries, and neonatal care environments of hospitals or other healthcare facilities.
### Description of Modifications
The V-2200 Infant Incubator now incorporates an optional pulse oximeter which was not the subject of the original V-2200 submission. The Pulse Oximetry SpO2 is provided by Masimo and is also a currently marketed device. The V-2100G Infant Incubator and the V-505 Infa Warmer are both currently marketed devices which incorporate this pulse oximetry technology from Masimo.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 8 2006
Atom Medical, Incorporated C/O Ms. Mary Staniewicz Neoforce Group, Incorporated 35 Commerce Drive Ivyland, Pennsylvania 18974
Re: K061856
Trade/Device Name: V-2200 Infant Incubator Regulation Number: 21 CFR 880.5400 Regulation Name: Neonatal Incubator Regulatory Class: II Product Code: FMZ, DQA Dated: June 5, 2006 Received: June 30, 2006
Dear Ms. Staniewicz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Staniewicz
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clare
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: V-2200 Infant Incubator
Indications for Use:
The V-2200 Infant Incubator is a device to keep premature infants or neonatal infants in a warm rne v-2200 miant modbated is a active tand isolated from ambient air and of which internal air temperature and humidity, are controlled. It is intended for inpatient use in maternity nurseries, temperature and hamilaty, are other healthcare facilities. Optional functions include pulse oximetry and oxygen delivery.
This device is not intended for home use.
This device is not intended as a transport incubator.
This is a prescription device.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
in Birnis
Jon Sign-Off) on of Anesthesiology, General Hospital, ction Control, Dental Devices
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Number KY 61356
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.