QUANTA LITE INTRINSIC FACTOR ELISA

K061841 · Inova Diagnostics, Inc. · LIG · Dec 22, 2006 · Clinical Chemistry

Device Facts

Record IDK061841
Device NameQUANTA LITE INTRINSIC FACTOR ELISA
ApplicantInova Diagnostics, Inc.
Product CodeLIG · Clinical Chemistry
Decision DateDec 22, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1810
Device ClassClass 2

Indications for Use

The QUANTA Lite™ Intrinsic factor ELISA is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative detection of Intrinsic factor antibodies in human serum. The presence of Intrinsic factor antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of pernicious anemia.

Device Story

QUANTA Lite™ Intrinsic Factor ELISA is an in vitro diagnostic test for human serum samples. It utilizes enzyme-linked immunosorbent assay (ELISA) technology to detect intrinsic factor antibodies. The device is intended for use in clinical laboratory settings by trained laboratory personnel. The assay provides semi-quantitative results, which are interpreted by healthcare providers alongside clinical symptoms and other diagnostic tests to support the diagnosis of pernicious anemia. The device assists in identifying autoimmune-mediated vitamin B12 deficiency.

Clinical Evidence

No clinical data provided in the document; the submission relies on the device's performance as an ELISA for the detection of intrinsic factor antibodies.

Technological Characteristics

Polystyrene microwell ELISA plate coated with recombinant human intrinsic factor; goat anti-human IgG horseradish peroxidase conjugate; TMB chromogen; stop solution. Semi-quantitative spectrophotometric measurement at 450nm/620nm. Serum matrix. Manual/automated plate reader required.

Indications for Use

Indicated for the semi-quantitative detection of IgG class Intrinsic Factor antibodies in human serum to aid in the diagnosis of pernicious anemia. For prescription use only.

Regulatory Classification

Identification

A vitamin B12 test system is a device intended to measure vitamin B12 in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of anemias of gastrointestinal malabsorption.

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k061841 B. Purpose for Submission: New device C. Measurand: Intrinsic factor antibodies D. Type of Test: Semi-quantitative Enzyme-linked immunosorbent assay (ELISA) E. Applicant: INOVA Diagnostics, Inc. F. Proprietary and Established Names: QUANTA Lite™ Intrinsic Factor Antibody ELISA G. Regulatory Information: 1. Regulation section: CFR 862.1810 Vitamin B₁₂ test system 2. Classification: Class II 3. Product code: LIG Radioassay, intrinsic factor blocking antibody 4. Panel: CH 75 H. Intended Use: 1. Intended use: QUANTA Lite™ Intrinsic Factor Antibody ELISA is an enzyme-linked immunosorbent assay (ELISA) for the semi-quantitative detection of Intrinsic Factor antibodies of the IgG class in human serum. The presence of Intrinsic Factor antibodies can be used in conjunction with clinical findings and other laboratory tests to aid in the diagnosis of pernicious anemia. 2. Indication(s) for use: Same as intended use 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: Microwell plate reader capable of measuring OD at 450nm (and 620nm for dual wavelength readings) I. Device Description: The QUANTA Lite Intrinsic Factor Antibody ELISA consists of a polystyrene microwell ELISA plate coated with recombinant human intrinsic factor antigen; ELISA negative, low positive and high positive controls; sample diluent; wash concentrate; goat anti-human IgG horseradish peroxidase conjugate; TMB chromogen; and stop solution. {1} 2 J. Substantial Equivalence Information: 1. Predicate device name(s): INOVA QUANTA Lite Gastric Parietal Cell Antibody ELISA 2. Predicate 510(k) number(s): k010558 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | Intrinsic Factor Antibody ELISA | Gastric Parietal Cell Antibody ELISA | | Indications for Use | To aid in the diagnosis of pernicious anemia | To aid in the diagnosis of pernicious anemia | | Method | ELISA | Same | | Calculation of results | Compared to a cut-off control | Same | | Result interpretation | ≤20 U/mL = negative, 20.1-24.9 = equivocal, ≥25 = positive | Same | | Units | Arbitrary ELISA units | Same | | Test matrix | Serum | Same | | Solid phase | Coated polystyrene microwell plates | Same | | Sample diluent | Tris-buffered saline, Tween 20, absorbents and protein stabilizers | Same | | Wash concentrate | Tris-buffered saline and Tween 20 | Same | | HRP IgG conjugate | Goat anti-human IgG | Same | | Controls | Negative, low positive, and high positive | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | Intrinsic Factor antibody ELISA | Gastric Parietal Cell Antibody ELISA | | Analyte measured | Anti-intrinsic factor antibodies | Anti-gastric parietal cell antibodies | | Capture antigen | Recombinant human intrinsic factor | Purified H+/K+ ATPase isolated from pig gastric mucosa | K. Standard/Guidance Document Referenced (if applicable): None referenced {2} L. Test Principle: The assay utilizes plastic microwells as a solid phase for attachment of recombinant human intrinsic factor. Pre-diluted controls and diluted patient sera are added to separate wells, allowing any intrinsic factor antibodies present to bind to the immobilized antigen. Unbound sample is washed away and an enzyme labeled goat anti-human IgG antibody is added to each well. A second incubation allows the enzyme labeled anti-human IgG to bind to any patient antibodies which have become attached to the microwells. After washing away any unbound enzyme labeled anti-human IgG, the remaining enzyme activity is measured by adding a chromogenic substrate and measuring the intensity of the color that develops. The assay is evaluated by spectrophotometrically measuring and comparing the color intensity that develops in the patient wells with the color in the control wells. Results determined with the assay are interpreted as negative, equivocal, or positive and are reported in arbitrary units. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Intra-assay performance for the assay was evaluated by testing 7 specimens a total of 5 times each. The samples tested ranged from 10.6 to 115.2 Units with %CVs ranging from 1.8 to 8.0%. | | A (HPC) | B | C | D | E | F | G | H | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Mean units | 108.1 | 45.4 | 115.2 | 108.7 | 21.3 | 17.1 | 10.6 | 13.4 | | SD | 2.1 | 0.8 | 2.5 | 3.4 | 1.7 | 0.8 | 0.4 | 0.7 | | CV% | 1.9 | 1.8 | 2.2 | 3.2 | 8.0 | 5.0 | 4.0 | 5.3 | Inter-assay variation was assessed by testing, in duplicate, a panel of 5 specimens including 1 negative, 1 low positive, 1 moderate, 2 high positive and the kit controls twice daily (once in the morning and once in the afternoon) for 3 days. Percent CVs ranged from 3.6 to 14.4 Units. | | HPC | NC | Spec. 1 | Spec. 2 | Spec. 3 | Spec. 4 | Spec. 5 | | --- | --- | --- | --- | --- | --- | --- | --- | | Mean units | 109.1 | 1.1 | 107.9 | 47.4 | 13.9 | 108.0 | 29.2 | | SD | 5.96 | 0.16 | 4.82 | 1.71 | 0.69 | 5.45 | 2.41 | | CV% | 5.5 | 14.4 | 4.5 | 3.6 | 5.0 | 5.0 | 8.2 | b. Linearity/assay reportable range: No claims were made regarding linearity for the assay. It is a semi-quantitative assay with results reported out as negative: 0.0 – 20.0 Units, positive as ≥ 25 Units, or equivocal: 20.1 – 24.9 Units when results are interpreted by comparison to the low positive control value of 25 Units. Specimens giving OD readings above the readable range of the plate reader {3} may be reported as greater than the highest measurable OD. Alternatively, samples may be serially diluted, re-run and a calculated value obtained. However, reactivity is related to the quantity of antibody present in a non-linear fashion. c. Traceability, Stability, Expected values (controls, calibrators, or methods): There is no recognized standard or reference material for anti-intrinsic factor antibodies. Accelerated stability studies were conducted on 3 lots of antigen coated plates and 3 lots of low and high positive kit controls. Data supported a shelf life of one year. Real time stability studies are currently being performed. d. Detection limit: The determination of detection limit/analytical sensitivity was not performed or relevant for this assay. e. Analytical specificity: Microbially contaminated, heat-treated, samples with visible particulate grossly hemolyzed or lipemic specimens should not be used in the assay. f. Assay cut-off: A panel of 476 asymptomatic, healthy individuals was tested. Ages ranged from 14-78 (median 43). Of the 276 specimens with age and sex information available, 150 were from males and 126 from females. Of the 476 specimens, one was a strong positive at 103.2 Units, one was borderline positive at 25.5 and one was equivocal. Excluding the equivocal result, the specificity of the assay was 99.6%. Excluding the one strong positive result, the mean was 5.7 Units and the median value was 5.0 Units. 2. Comparison studies: a. Method comparison with predicate device: Testing was performed on 278 specimens that included 69 specimens from healthy individuals and 209 specimens submitted to 2 reference laboratories for gastric parietal cell antibody (GPA) and/or intrinsic factor antibody testing. | QUANTA Lite Intrinsic Factor Antibody ELISA | QUANTA Lite GPA ELISA | | | | | --- | --- | --- | --- | --- | | | Positive | Equivocal* | Negative | Total | | Positive | 22 | 0 | 5 | 27 | | Equivocal* | 0 | 0 | 0 | 0 | | Negative | 52 | 6 | 193 | 251 | | Total | 74 | 6 | 198 | 278 | *Equivocal results were considered negative in the calculations Positive percent agreement: $22 / 74 = 29.7\%$ {4} Negative percent agreement: $199 / 205 = 97.1\%$ Overall agreement: $221 / 278 = 79.5\%$ The agreement between the assays was low because 1) the assays measure different analytes; and 2) not all pernicious anemia (PA) patients have both antibodies. Gastric parietal cell antibodies present earlier in the progression of chronic atrophic gastritis to pernicious anemia and are found in approximately $90\%$ of PA patients. Intrinsic factor antibodies, while specific for PA, are present in only $50 - 60\%$ of PA patients. The company also compared to another legally marketed device, the DPC IF bAb $\left[{ }^{57} \mathrm{Co}\right]$ Radioimmunoassay (k832726). One hundred seventy four specimens submitted to a clinical reference laboratory for IF antibody evaluation were tested. Additional clinical information about these subjects was not available. | QUANTA Lite Intrinsic Factor Antibody ELISA | DPC RIA for Intrinsic Factor Antibody | | | | --- | --- | --- | --- | | | Positive | Negative | Total | | Positive | 65 | 0 | 65 | | Negative | 56 | 53* | 109 | | Total | 121 | 53 | 174 | *2 equivocal results with the ELISA assay were considered negative in the calculations Positive percent agreement: $65 / 121 = 53.7\%$ Negative percent agreement: $53 / 53 = 100\%$ Overall agreement: $118 / 174 = 67.8\%$ There is a distinct possibility of interference from therapeutic doses of vitamin $\mathrm{B}_{12}$ in the RIA assay. # b. Matrix comparison: Both assays use serum as the test matrix. # 3. Clinical studies: # a. Clinical Sensitivity: Clinical sensitivity was established by testing sera from 177 subjects which included 43 with confirmed PA, 55 with presumed PA and 79 suspected PA. In the 43 confirmed PA, the assay interpreted $95.3\%$ (41/43) as positive. | Clinical Status | N = | Intrinsic factor ELISA positive | Intrinsic factor ELISA equivocal | Intrinsic factor ELISA negative | | --- | --- | --- | --- | --- | | Confirmed PA | 43 | 41 | 1 | 1 | | Presumed PA | 55 | 54 | 1 | 0 | | Suspected PA | 79 | 0 | 0 | 79 | | Total | 177 | 92 | 2 | 1 | {5} The definitions for the categories are: Confirmed Pernicious Anemia: laboratory measures consistent with pernicious anemia including gastric parietal cell antibody positive, low vitamin B₁₂, mean corpuscular volume (MCV) increased (typical of pernicious anemia), hematological confirmation of megaloblastic anemia Presumed Pernicious Anemia: laboratory measures consistent with pernicious anemia including gastric parietal cell antibody positive, low vitamin B₁₂, mean corpuscular volume (MCV) increased (typical of pernicious anemia) Suspected Pernicious Anemia: one or more tests positive: gastric parietal cell antibody, low vitamin B₁₂, but intrinsic factor antibody assay results were discrepant ## b. Clinical specificity: A total of 499 sera from patients with other autoimmune or infectious disease (n=23), and normal subjects (n=476) were tested to assess potential cross-reactivity of other disease sera with the assay. | Other diseases and normal subjects | N = | IF ELISA positive | IF ELISA equivocal | IF ELISA negative | Total | | --- | --- | --- | --- | --- | --- | | Other (H. pylori, mitochondrial M2, cytomegalovirus, herpes simplex virus, ASCA, RNP, SSA, SSB, Scl-70, DNA, tissue transglutaminase, glomerular basement membrane) | 23 | 0 | 0 | 23 | 23 | | Normal | 476 | 2 | 1 | 473 | 476 | | Total | | 2 | 1 | 496 | 499 | * Sm (1), RNP (1), SSB (1), Histone (3), Scl-70 (1), ribosome P (1), chromatin (2), centromere (1), ASCA (2), GBM (2), Jo-1 (1) Clinical sensitivity and specificity calculations: | Groups | N | | --- | --- | | Confirmed pernicious anemia | 43 | | Presumed pernicious anemia | 55 | | Suspected pernicious anemia | 79 | | Disease controls | 23 | | Healthy controls | 476 | | Total | 676 | {6} Suspected PA included as disease positive | | Pernicious anemia | | | | --- | --- | --- | --- | | IF antibody result | + | - | Total | | + | 92 | 2 | 94 | | - | 85 | 497 | 582 | | Total | 177 | 499 | 676 | Clinical sensitivity: 92/177 = 52% Clinical specificity: 497/499 = 99.6% Suspected PA included as disease negative | | Pernicious anemia | | | | --- | --- | --- | --- | | IF antibody result | + | - | Total | | + | 92 | 2 | 94 | | - | 6 | 576 | 582 | | Total | 98 | 578 | 676 | Clinical sensitivity: 92/98 = 93.9% Clinical specificity: 576/578 = 99.6% c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: See assay cut-off and expected values. 5. Expected values/Reference range: The expected result in the normal population is negative (≥ 20 Units). N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...