← Product Code [GZJ](/productcode/GZJ) · K061828

# LORENZ BIOTECH APTIVA, VERSION 2.13 (K061828)

_Lorenz Biotech S.P.A. · GZJ · Aug 24, 2007 · Neurology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K061828

## Device Facts

- **Applicant:** Lorenz Biotech S.P.A.
- **Product Code:** [GZJ](/productcode/GZJ.md)
- **Decision Date:** Aug 24, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The AptivaTM is a multifunctional transcutaneous electrical nerve stimulator (TENS) equipped with a surface electromyography and electroneurography. AptivaTM is indicated for symptomatic relief and management of chronic pain. intractable pain and/or as an adjunctive treatment for the management of posttraumatic pain. The dynamic surface electromyography and electroneurography system is used for biofeedback, relaxation training, and muscle re-education as well as monitoring and display of bioelectric signals produced by muscles, stimulate peripheral nerves, and monitor and display the electrical activity produced by nerves.

## Device Story

Aptiva™ v2.13U is a multifunctional console-based system (mobile floor stand) providing TENS, EMG, and electroneurography (ENG) capabilities. Input: bioelectric signals from muscles and nerves via surface electrodes. Processing: system monitors, displays, and processes these signals for biofeedback, relaxation training, and muscle re-education; also provides electrical stimulation to peripheral nerves. Output: visual display of bioelectric activity and electrical nerve stimulation. Used in clinical settings by healthcare providers. Benefits: pain management, muscle re-education, and nerve activity monitoring.

## Clinical Evidence

No clinical data provided. Performance testing consisted of bench testing, including electrical safety, electromagnetic compatibility, and design verification and validation testing. Results confirmed the device fulfills prospectively defined performance specifications.

## Technological Characteristics

Console-based system with mobile floor stand, power supply, battery charger, and patient remote. Uses commercially available surface electrodes. Functions include TENS, EMG, and ENG. Models vary by feature set (Ballet, Flamenco, Hip-Hop, Jazz). Connectivity includes patient remote and printer (select models).

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Apex Medical Digital TENS ([K021755](/device/K021755.md))
- GEMS-HV TENS ([K032994](/device/K032994.md))
- Bioscope SD TENS ([K970249](/device/K970249.md))
- Promax TENS ([K022405](/device/K022405.md))
- NeuroTrain III ([K003433](/device/K003433.md))
- EMG Retrainer Dual Channel ([K972487](/device/K972487.md))
- MES-9000/EMG System ([K013399](/device/K013399.md))
- Synergy System ([K981405](/device/K981405.md))
- Brevio ([K010269](/device/K010269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

# 510(k) Summary Lorenz Biotech S.p.A Aptiva™ v2.13U

#### 1. SPONSOR

AUG 2 4 2007

Lorenz Biotech S.p.A. Via Statale 12, 109/B/C 41036 Medolla Italy

| Contact Person: | Mr. Riccardo Isani              |
|-----------------|---------------------------------|
| Email address:  | Riccardo.isani@lorenzbiotech.it |
| Date Prepared:  | August 6, 2007                  |

#### 2. DEVICE NAME

| Proprietary Name:    | Aptiva™ v2.13U                                                                                                                                |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:   | Transcutaneous nerve stimulator (TENS)<br>Biofeedback device<br>Electromyography device (EMG)<br>Nerve conduction velocity measurement device |
| Classification Name: | Transcutaneous nerve stimulator (TENS)<br>Biofeedback device<br>Electromyography device (EMG)<br>Nerve conduction velocity measurement device |

#### 3. PREDICATE DEVICES

| Functionality                      | Predicate Device                       |
|------------------------------------|----------------------------------------|
| TENS                               | • Apex Medical Digital TENS (K021755)  |
|                                    | • GEMS-HV TENS (K032994)               |
|                                    | • Bioscope SD TENS (K970249)           |
|                                    | • Promax TENS (K022405)                |
|                                    | • NeuroTrain III (K003433)             |
| Electromyography (EMG)/Biofeedback | • EMG Retrainer Dual Channel (K972487) |
|                                    | • MES-9000/EMG System (K013399)        |
|                                    | • Synergy System (K981405)             |
| Nerve conduction velocity (ENG)    | Brevio (K010269)                       |

{2}------------------------------------------------

#### 4. DEVICE DESCRIPTION

The Aptiva™ v2.13U consists of a console on a mobile floor stand, power cord, battery charger, patient remote control, cables, and Manuals (User and Application). The Aptiva™ v2.13U is provided in four functional models (Ballet, Flamenco, Hip-Hop, and Jazz). EMG features are provided in the Ballet and Hip Hop models. One version of the Ballet model (Option Etoile) includes electroneurography (ENG) functions and a printer. Aptiva™ v2.13U uses commercially available electrodes. Models designated "Light" versions do not include the full treatment databases.

#### 5. INTENDED USE

The AptivaTM is a multifunctional transcutaneous electrical nerve stimulator (TENS) equipped with a surface electromyography and electroneurography.

AptivaTM is indicated for symptomatic relief and management of chronic pain. intractable pain and/or as an adjunctive treatment for the management of posttraumatic pain.

The dynamic surface electromyography and electroneurography system is used for biofeedback, relaxation training, and muscle re-education as well as monitoring and display of bioelectric signals produced by muscles, stimulate peripheral nerves, and monitor and display the electrical activity produced by nerves.

#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The Aptiva™ v2.13U is substantially equivalent to specified predicate devices based on intended use, indications for use, operational characteristics, and fundamental technological characteristics. Detailed side-by-side comparisons of the Aptiva™ v2.13U with cited predicate devices were included in the premarket notification and associated amendments.

#### 7. PERFORMANCE TESTING

Testing included in this premarket notification includes electrical safety, electromagnetic compatibility, and design verification and validation testing. No materials of the Aptiva™ v2.13U are in direct contact with treatment sites and therefore no biocompatibility testing was provided. Test results demonstrate that the Aptiva™ v2.13U fulfills the prospectively defined performance specifications.

{3}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 4 2007

Lorenz Biotech S.P.A. C/O Rosina Robinson, RN, MEd, RAC Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760

Re: K061828

Trade Name: Aptiva 10 v2.13 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: GZJ, HCC, IKN, JXE Dated: August 6, 2007 Received: August 7, 2007

Dear Ms. Robinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

### Page 2 - Rosina Robinson, RN, MEd, RAC

CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson, M.S.

Mark N. Melkerson, M.S Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known): K061828

Device Name: Lorenz Biotech Aptiva™ v2.13U

Indications for Use:

The Aptiva™ v2.13U is a multifunctional transcutaneous electrical nerve stimulator (TENS) equipped with a surface electromyography and electroneurography.

Aptiva™ is indicated for symptomatic relief and management of chronic pain, intractable pain and/or as an adjunctive treatment for the management of post-traumatic pain.

The dynamic surface electromyography and electroneurography system is used for biofeedback, relaxation training, and muscle re-education as well as monitoring and display of bioelectric signals produced by muscles, stimulate peripheral nerves, and monitor and display the electrical activity produced by nerves.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

sion Sign-Off Division of General. Restorative. and Neurological Devices

510(k) Number K061828

---

**Source:** [https://fda.innolitics.com/device/K061828](https://fda.innolitics.com/device/K061828)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
