DIMENSION VISTA SYSTEM LIPID CALIBRATOR, MODEL KC220
K061750 · Dade Behring, Inc. · JIX · Aug 11, 2006 · Clinical Chemistry
Device Facts
Record ID
K061750
Device Name
DIMENSION VISTA SYSTEM LIPID CALIBRATOR, MODEL KC220
Applicant
Dade Behring, Inc.
Product Code
JIX · Clinical Chemistry
Decision Date
Aug 11, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The LIPID CAL is an in vitro diagnostic product for the calibration of high density lipoprotein cholesterol (HDLC) and low density lipoprotein cholesterol (LDLC) methods on the Dimension Vista™ System.
Device Story
LIPID CAL is a liquid, multi-analyte calibrator used to calibrate HDLC and LDLC assays on the Dimension Vista™ System. Product consists of six vials (three of Calibrator A, three of Calibrator B) containing human albumin, human lipoproteins, and bovine gamma globulins. Human source material tested negative for HIV-1/2, HBsAg, and HCV. Device is used in clinical laboratory settings by professional staff to ensure accuracy of cholesterol measurements. Value assignment performed via comparison to NCEP reference laboratory methods using 90 measurements across multiple instruments.
Clinical Evidence
Bench testing only. Stability verified via real-time and stress testing supporting 12-month refrigerated shelf life and 30-day open-vial stability. Value assignment validated through comparison to NCEP reference laboratory methods using 90 measurements across multiple instruments.
Technological Characteristics
Liquid, multi-analyte calibrator. Matrix: human albumin, human lipoproteins, bovine gamma globulins. Traceability: NCEP reference methods. Stability: 12-month shelf life at 4°C; 7-day on-board stability; 30-day open-vial stability. No software or electronic components.
Indications for Use
Indicated for the calibration of HDLC and LDLC assays on the Dimension Vista™ System in a clinical laboratory setting.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Dimension® HDL Calibrator (k983850)
Related Devices
K983850 — DIMENSION AHDL CALIBRATOR · Dade Behring, Inc. · Nov 18, 1998
K042550 — MULTI-LIPID CALIBRATOR · Wako Chemicals USA, Inc. · Nov 8, 2004
K993098 — HDL-C/LDL-C CALIBRATOR, MODEL 1913-003 · Data Medical Associates, Inc. · Oct 8, 1999
K020723 — DIMENSION ALDL CALIBRATOR, MODEL DC131 · Dade Behring, Inc. · May 8, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
K061750
AUG 1 1 2006
# 510(k) Summary for the Dimension Vista™ System Lipid Calibrator (LIPID CAL - KC220)
## A. 510(k) Number:
| B. Analytes: | High density lipoprotein cholesterol (HDLC) and low density<br>cholesterol lipoprotein (LDLC). |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| C. Type of Test: | Calibrator Material |
| D. Applicant: | Dade Behring Inc., P.O. Box 6101, Newark, DE 19714-6101<br>Victor M. Carrio, Regulatory Affairs and Compliance Manager<br>Office: (302) 631-0376 Fax: (302) 631-6299 |
## E. Proprietary and Established Names:
Dimension Vista™ System Lipid Calibrator (LIPID CAL - KC220)
## F. Regulatory Information:
- 1. Regulation section: 21 CFR § 862-1150 Calibrator
- 2. Classification: Class II
- 3. Product Code: JIX Calibrator, Multi-Analyte Mixture
- 4. Panel: Clinical Chemistry
- The LIPID CAL is an in vitro diagnostic product for the calibration G. Intended Use: of high density lipoprotein cholesterol (HDLC) and low density lipoprotein cholesterol (LDLC) methods on the Dimension Vista™ System.
### H. Device Description:
LIPID CAL is a liquid, multi-analyte, human albumin based product containing human lipoproteins and bovine gamma globulins. The kit consists of six vials, three vials of Calibrator A, and three vials of Calibrator B which are ready for use (no preparation is required). The volume per vial is 1.5 mL.
{1}------------------------------------------------
| Ch Analaman |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| |
| |
| |
| |
| |
| |
| |
| |
| |
| |
| 31 55 206 7 49 4 9 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4 4<br>A BAR BAR AND BARRE AN AND |
| |
| |
| |
| |
| |
| |
| CARD F |
| |
| |
| |
| |
| |
| |
| |
| 1000 100 100 100 100 100 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000000000000000000 |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| 1 |
| |
| |
| |
-
.
| Item | Device | Predicates | Predicates |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Dimension Vista™ System<br>Lipid Calibrator | Dimension® HDL Calibrator<br>K983850 | Dimension® ALDL Calibrator<br>K020723 |
| Intended Use | The LIPID CAL is an in vitro<br>diagnostic product for the calibration<br>of high density lipoprotein cholesterol<br>(HDLC) and low density lipoprotein<br>cholesterol (LDLC) methods on the<br>Dimension Vista™ System. | The HDL Cholesterol Calibrator is an in vitro<br>diagnostic product to be used to calibrate the<br>Dimension® clinical chemistry system for the<br>HDL Cholesterol (HDL) method. | The ALDL Calibrator is an in vitro diagnostic<br>product intended to be used to calibrate the<br>automated LDL (ALDL) method for the<br>Dimension® clinical chemistry system. |
| Analytes | High density lipoprotein cholesterol<br>and low density lipoprotein<br>cholesterol. | High density lipoprotein cholesterol. | Low density lipoprotein cholesterol. |
| Form | Liquid. | Liquid. | Lyophilized. |
| Traceability | National Cholesterol Education<br>Program (NCEP) reference method. | NBS SRM 911 a. | NCEP beta-quantification reference method for<br>LDL-C b. |
| Matrix | Human albumin based product<br>containing human lipoproteins and<br>bovine gamma globulins. | Stabilized aqueous solution containing<br>cholesterol. | Lyophilized buffered bovine serum albumin-<br>based product supplemented with human low<br>density lipoprotein cholesterol. |
| Number of<br>Levels | Two levels. | Three levels. | Three levels. |
a - NBS-SRM: National Bereau of Standards - Standard Reference Material.
b -- NCEP - National Cholesterol Education Program
- B. EDXA
- C. Optical Microscopy
- D. X-Ray Diffraction
- E. Thermogravimetric Analysis, Differential Scanning Calorimetry
- F. Brunauer-Emmett-Teller (BET) N2 Adsorption
:
{2}------------------------------------------------
# J. Standard/Guidance Document Referenced:
medical devices
| 1. Guidance: | Guidance for Industry - Abbreviated 510(k) Submissions for In<br>Vitro Diagnostic Calibrators; Final, 02/22/1999<br>Guidance for Industry and FDA Staff - Use of Symbols on Labels<br>and in Labeling of In Vitro Diagnostic Devices Intended for<br>Professional Use, 11/30/2004 |
|---------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Standards: | CEN 13640 Stability testing of In-Vitro Diagnostic Devices<br>ISO 14971:2000 Medical devices -Application of risk management to |
# K. Performance Characteristics:
| 1. Stability: | Target shelf life for the Dimension Vista™ Lipid Calibrator is 12 months. Calibrator shelf life is determined by comparing results of the product stored at 4°C with control stored at -20°C. The method is calibrated from this stored material. The 4°C material values are recovered versus the calibration. Recovery versus time is monitored and percent change over time is determined. Percent change should be less than or equal to 3%. Shelf-life stability (expiration) dating assignment at commercialization reflects the real-time data on file at Dade Behring, Inc. A vial punctured by the instrument and stored on board has a seven day claim. An open vial not stored on board of instrument, but recapped and stored in a refrigerator has a stability claim of 30 days. For testing, vials are opened /punctured on day zero. A quantity sufficient for multiple calibrations is removed and the vials are recapped and stored at 2 - 8 °C. Opened/punctured vials are tested on days 8, 15, 22, and 32 versus freshly opened vials. |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Traceability: | The assigned values of the Dimension Vista™ Lipid Calibrator are |
| Constituent | Traceability |
|-------------|-------------------------------------------------------------------|
| HDLC | National Cholesterol Education Program (NCEP)<br>reference method |
| LDLC | National Cholesterol Education Program (NCEP)<br>reference method |
standardized to the enclosed table of assigned values:
{3}------------------------------------------------
#### 3. Bottle Value Assignment:
Human HDL (bovine based) and LDL (human serum based) are weighed protein containing solutions. The HDL Master Pool and the LDL Anchor Pool are stored at -70 °C.
HDL human serum samples with Abel-Kendall values are used to produce a standard curve for the assignment of the Master Pool. LDL human serum pools with values from reference laboratories (CRMNL) are used to produce a standard curve for assignment of the Anchor Pool.
The commercial lot is made by adding human HDL and LDL to base matrix in appropriate concentrations for upper calibrator levels. The concentration of each level is verified by using an instrument calibrated with the Master Pool / Anchor Pools. The final bottle values for each level of the commercial lot is assigned and verified using multiple instruments by testing N = 90 replicates.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird-like figure.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. Victor M. Carrio RA/QS Compliance Manager Dade Behring, Inc. 500 GBC Dr. PO Box 6101, M/S 514 Newark, DE 19714-6101
# AUG 1 1 2006
Re: k061750
Trade/Device Name: Dimension Vista™ Lipid Calibrator (KC220) Regulation Number: 21 CFR§862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIX Dated: June 30, 2006 Received: June 21, 2006
Dear Mr. Carrio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto G
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications For Use Statement
Kolo 1750 510(k) Number (if known):
## Device Name:
Dimension Vista™ Lipid Calibrator (KC220)
#### Indications for Use:
The LIPID CAL is an in vitro diagnostic product for the calibration of high density lipoprotein cholesterol (HDLC) and low density lipoprotein cholesterol (LDLC) methods on the Dimension Vista™ System.
Prescription Use (Per 21 CFR 801 Subpart D) AND/OR
Over-the-counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of -In Vitro Diagnostic Devices (OIVD)
CFC
K061750
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.