The MicroScan® Dried Gram Negative Panel is designed for use in the Confirmation of Extended-Spectrum Beta-Lactamase (ESBL) production in Escherichia coli, Klebsiella spp. and Proteus mirabilis. After inoculation, panels are incubated for 16-20 hours at 35°C +/- 1°C in a non-CO2 incubator and read either visually or with MicroScan® instrumentation, according to the Package Insert. This particular submission is for the use of cefotaxime (2, 16 µg/ml), cefotaxime/clavulanic acid (0.5/4, 4/4 µg/ml), ceftazidime (1, 8 µg/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 µg/ml) for ESBL confirmatory testing on MicroScan® Dried Gram Negative MIC/Combo Panels and the addition of automated read methods. The Gram Negative organisms which may be used for confirmation of ESBL production in this panel are: Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis
Device Story
MicroScan® Dried Gram-Negative MIC/Combo Panels utilize dehydrated microdilutions of antimicrobial agents in Mueller Hinton Broth; panels rehydrated and inoculated with bacterial suspension (turbidity or Prompt® method); incubated 16-20 hours at 35°C in non-CO2 environment; growth visually observed or measured by autoSCAN®-4 or WalkAway® systems; device confirms ESBL production by comparing growth inhibition in wells with/without clavulanic acid; results inform antimicrobial resistance profiles for clinicians; aids in selection of appropriate antibiotic therapy.
Clinical Evidence
Bench testing only. Study evaluated 95 molecularly characterized isolates (42 ESBL positive) of E. coli, K. oxytoca, K. pneumoniae, and P. mirabilis. Compared turbidity and Prompt® inoculation methods across manual, autoSCAN-4, and WalkAway reading systems. Turbidity method showed 93% sensitivity and 100% specificity. Prompt® method showed 95% sensitivity and 96-98% specificity. Reproducibility demonstrated across 3 sites.
Technological Characteristics
Dried microdilution panels containing Mueller Hinton Broth and nutrients; growth-based detection; non-CO2 incubation; manual or automated (autoSCAN-4, WalkAway) reading; turbidity or Prompt® inoculation methods; compliant with CLSI M7 (M100-S16) standards.
Indications for Use
Indicated for confirmation of Extended-Spectrum Beta-Lactamase (ESBL) production in E. coli, K. oxytoca, K. pneumoniae, and P. mirabilis. For prescription use in clinical laboratory settings.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Predicate Devices
MicroScan® Dried ESBL plus ESBL Confirmation Panel (K020037)
Related Devices
K020037 — DRIED GRAM NEGATIVE MIC/COMBO PANELS · Dade Microscan, Inc. · Mar 22, 2002
K193536 — MicroScan Dried Gram Negative MIC/Combo Panels with Ciprofloxacin (Cp) (0.004 - 8µg/mL) · Beckman Coulter, Inc. · Mar 18, 2020
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary
K061687
JUL 21 2006
#### 510(k) Submission Information:
| Device Manufacturer: | Dade Behring Inc. |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact name: | Maureen Mende, Regulatory Affairs Group Manager |
| Fax: | 916-374-3144 |
| Date prepared: | June 14, 2006 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan® Dried Gram Negative MIC/Combo Panels |
| Intended Use: | To be used in the confirmation of Extended-Spectrum<br>Beta-Lactamase production in Escherichia coli, Klebsiella spp. and<br>Proteus mirabilis. |
| 510(k) Notification: | Modification to k020037- ESBL Confirmation |
| Predicate device: | MicroScan® Dried ESBL plus ESBL Confirmation Panel |
#### 510(k) Summary:
MicroScan® Dried Gram Negative MIC/Combo Panels are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility for Gram Negative organisms and screening for suspected ESBL production in E. coli, Klebsiella spp. and P. mirabilis.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in broth and dehydrated. Various antimicrobial agents are diluted in broth to concentrations bridging the range of clinical interest. Panels are rehydrated with water, after inoculation with a standardized suspension of the organism. After incubation in a non-CO2 incubator at 35 °C for a minimum of 16 hours, the minimum inhibitory concentration (MIC) for the test organism is determined by observing the lowest antimicrobial concentration showing inhibition of growth.
The antimicrobics: cefotaxime/clavulanic acid, ceftazidime and ceftazidime/clavulanic acid have been cleared for confirmation of suspected extended-spectrum beta-lactamases with E. coli, K. oxytoca and K. pneumoniae via Premarket Notification submission (k020037). Confirmation of suspected ESBL with P. mirabilis was submitted to the FDA and is pending clearance.
This Premarket Notification 510(k) presents data in support of a request for confirmatory testing of suspected extendedspectrum beta-lactamase producing E. coli, K. oxytoca, K. pneumoniae and P. mirabilis using streamlined dilutions of the antimicrobial agents recommended for confirmation testing described in the CLSI document M100-S16 for cefotaxime and ceftazidime, alone and in combination with clavulanic acid. In addition, this Premarket Notification 510(k) presents data for automated (WalkAway® and autoSCAN®-4 instruments) reads of the MicroScan® Dried Gram Negative panel with streamlined dilutions.
Challenge studies with the MicroScan® Dried Gram Negative panel with cefotaxime (2, 16 ug/ml), cefotaxime/clavulanic acid (0.5/4, 4/4 µg/ml), ceftazidime (1, 8 µg/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 µg/ml) were conducted using stock challenge strains. The Design Validation studies were designed to confirm the acceptability of streamlined dilutions of these antimicrobial agents for confirmation of suspected ESBLproducing organisms by comparing the panel confirmation result against the molecular characterization result. The dried Test panel with streamlined dilutions of the antimicrobial agents demonstrated an overall Agreement of > 93% with ESBL and non-ESBL-producing strains.
Inoculation method and instrument reproducibility testing demonstrated acceptable reproducibility with cefotaxime (2, 16 µg/ml), cefotaxime/clavulanic acid (0.5/4, 4/4 µg/ml), ceftazidime (1, 8 µg/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 µg/ml) regardless of which inoculation method (i.e. Turbidity and read mathod (Manual, WalkAway® and autoSCAN®-4 instruments) used.
The MicroScan® Dried Gram Negative panel with cefotaxime/clavulanic acid, ceflazidime and ceftazidime/clavulanic acid antimicrobial agents demonstrated acceptable Quality Control throughout each phase of the ESBL evaluation.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Maureen Mende Regulatory Affairs Group Manager Dade Behring Inc. 2040 Enterprise Blvd. West Sacramento, CA 95691
JUL 2 1 2006
Re: k061687
Trade/Device Name: MicroScan® Dried Gram Negative MIC/Combo Panels with cefotaxime (2, 16 µg/ml), cefotaxime/clavulanic acid (0.5/4, 4/4 µg/ml), ceftazidime (1, 8 µg/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 µg/ml)
Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Regulatory Class: Class II Product Code: JWY, LRG Dated: June 14, 2006 Received: June 15, 2006
Dear Ms. Mende:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR). Parts 800 to 895. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA`s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Paris 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitts your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0450. Also, please one the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its tolli dic
(900) 638-0041 - 1981) 11-12-2557 - 1991) 11-12-275 (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Sally, attorn
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known): K 0411687
#### Device Name:
MicroScan® Dried Gram Negative MIC/Combo Panels with cefotaxime (2, 16 ug/ml), cefotaxime/clavulanic acid (0.5/4, 4/4 µg/ml), ceftazidime (1, 8 µg/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 ug/ml)
#### Indications For Use:
The MicroScan® Dried Gram Negative Panel is designed for use in the Confirmation of Extended-Spectrum Beta-Lactamasc (ESBL) production in Escherichia coli, Klebsiella spp. and Proteus mirabilis.
After inoculation, panels are incubated for 16-20 hours at 35°C +/- 1°C in a non-CO2 incubator and read either visually or with MicroScan® instrumentation, according to the Package Insert.
This particular submission is for the use of cefotaxime (2, 16 µg/ml), cefotaxime/clavulanic acid (0.5/4, 4/4 µg/ml), ceftazidime (1, 8 µg/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 µg/ml) for ESBL confirmatory testing on MicroScan® Dried Gram Negative MIC/Combo Panels and the addition of automated read methods.
The Gram Negative organisms which may be used for confirmation of ESBL production in this panel arc:
> Escherichiu coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
Fuddhih t. Patil.
Page 1 of __ 1 __
Vision Sign-Off
Office of In Vitre Diagnostic Device Evaluation and Saicty
510(k) K061687
viii
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): k061687
#### Device Name:
MicroScan® Dried Gram Negative MIC/Combo Panels with cefotaxime (2, 16 µg/m), cefotaxime/clavulanic acid (0.5/4, 4/4 ug/ml), ceftazidime (1, 8 ug/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 ug/ml)
#### Indications For Use:
The MicroScan® Dried Gram Negative Panel is designed for use in the Confirmation of Extended-Spectrum Beta-Lactamase (ESBL) production in Escherichia coli, Klebsiella spp. and Proteus mirabilis.
After inoculation, panels are incubated for 16-20 hours at 35°C +/- 1°C in a non-CO2 incubator and read either visually or with MicroScan® instrumentation, according to the Package Insert.
This particular submission is for the use of cefotaxime (2, 16 µg/ml), cefotaxime/clavulanic acid (0.5/4, 4/4 µg/ml), ceftazidime (1, 8 µg/ml) and ceftazidime/clavulanic acid (0.25/4, 2/4 µg/ml) for ESBL confirmatory testing on MicroScan® Dried Gram Negative MIC/Combo Panels and the addition of automated read methods.
The Gram Negative organisms which may be used for confirmation of ESBL production in this panel are:
> Escherichia coli Klebsiella oxytoca Klebsiella pneumoniae Proteus mirabilis
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Advisor Sign-Off
Office of
Evaluation an
agnostic Device
Kiloliz
iii
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.