GE VOLUSON E8 ULTRASOUND SYSTEM
K061682 · General Electric Co. · IYO · Jul 12, 2006 · Radiology
Device Facts
| Record ID | K061682 |
| Device Name | GE VOLUSON E8 ULTRASOUND SYSTEM |
| Applicant | General Electric Co. |
| Product Code | IYO · Radiology |
| Decision Date | Jul 12, 2006 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 892.1560 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Intended Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (TV); Transrectal (TR); and Intraoperative (abdominal, PV and neurological).
Device Story
Voluson E8 is a mobile, general-purpose diagnostic ultrasound system; provides digital acquisition, processing, and display of ultrasound images. System includes mobile console, keyboard, specialized controls, 10-inch LCD touch screen, and 15-inch LCD display. Operates by transmitting ultrasonic pulses into the body and processing returning echoes to generate images or flow data. Used by qualified physicians in clinical settings for various anatomical regions and exam types. Output displayed on integrated monitor for diagnostic evaluation; assists in clinical decision-making for patient management. Benefits include non-invasive diagnostic imaging and biopsy guidance.
Clinical Evidence
No clinical data required. Substantial equivalence supported by non-clinical testing, including acoustic output measurements, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety testing in conformance with applicable standards.
Technological Characteristics
Mobile console (57x96x149 cm) with digital acquisition/processing. Includes 10-inch LCD touch screen and 15-inch LCD display. Supports multiple transducer types (e.g., RAB, IC, 9L, PA, M12LW, 3S, P2D, P6D, M6C, 11L series). Operating modes: B, M, PW/CW Doppler, Color/Power Doppler, Color M, Harmonic Imaging, Coded Pulse, and 3D/4D imaging. Conforms to 21 CFR 820, ISO 9001, and ISO 13485.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in fetal/OB, abdominal, pediatric, small organ, neonatal/adult cephalic, cardiac, peripheral vascular, musculoskeletal, transrectal, transvaginal, and intraoperative applications. For use by qualified physicians. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- GE Voluson 730 Diagnostic Ultrasound System (K003535, K003090, K003525, K032620, K041688)
Related Devices
- K103649 — VOLUSON S6, VOLUSON S8 · GE Healthcare · Mar 9, 2011
- K122387 — VOLUSON P6, VOLUSON P8 · GE Healthcare · Sep 11, 2012
- K192159 — Voluson E6, Voluson E8, Voluson E10 · GE Healthcare · Aug 29, 2019
- K122327 — VOLUSON E6/E8/E8EXPERT/E10 · GE Healthcare · Sep 21, 2012
- K082185 — GE LOGIQ E9 DIAGNOSTIC ULTRASOUND SYSTEM, MODELS 5205000, 5205000-2 · Ge Medical Systems Ultrasound and Primary Care Dia · Aug 15, 2008
Submission Summary (Full Text)
{0}------------------------------------------------
K061682
# Attachment B:
Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c). JUL 1 2 2006
GE Healthcare
General Electric Company P.O. Box 414, Milwaukee, WI 53201
#### Section a):
| 1. Submitter: | GE Medical Systems, Ultrasound and Primary Care Diagnostics, LLC<br>PO Box 414<br>Milwaukee, WI 53201 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Allen Schuh,<br>Manager, Safety and Regulatory Engineering<br>Telephone: 414-721-3992; Fax: 414-721-3899 |
| Date Prepared: | June 14, 2006 |
| 2. Device Name: | GE Voluson E8 Diagnostic Ultrasound System<br>Ultrasonic Pulsed Echo Imaging System, 21 CFR 892.1560, 90-IYO<br>Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN<br>Diagnostic Ultrasonic Transducer, 21 CFR 892.1570, 90-ITX |
| 3. Marketed Device: | GE Voluson 730 Diagnostic Ultrasound System, K003535, K003090, K |
GE Voluson 730 Diagnostic Ultrasound System - K003525, K032620, K041688 3. Marketed Device: A device currently in commercial distribution.
4. Device Description: The Voluson E8 is a full featured, general purpose diagnostic ultrasound system which consists of a mobile console approximately 57 cm wide, 96 cm deep and 149 cm high that provides digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls, 10 inch LCD touch screen and color 15 inch LCD image display. This modification will provide users with additional probe options, improved user interface and overall quality and image enhancement.
5. Indications for Use: The device is intended for use by a qualified physician for ultrasound evaluation of Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (TV); Transrectal (TR); and Intraoperative (abdominal, PV and neurological).
6. Comparison with Predicate Devices: The GE Voluson E8 is of a comparable type and substantially equivalent to the current GE Voluson 730 Pro/Expert. It has the same technological characteristics, key safety and effectiveness features, and is similar in physical design, construction and materials and has the same intended uses and basic operating modes as the predicate device.
#### Section b):
1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.
#### 2. Clinical Tests: None required.
3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with 21 CFR 820, ISO 9001and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Healthcare that the GE Voluson E8 Diagnostic Ultrasound is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### JUL 1 2 2006
Mr. Allen Schuh Manager, GE Ultrasound Product Safety and Regulatory Engineering GE Medical Systems, Ultrasound and Primary Care Diagnostics, LLC 9900 Innovation Drive WAUWATOSA WI 53226
Re: K061682
Trade Name: GE Voluson E8 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: June 14, 2006 Received: June 15, 2006
#### Dear Mr. Schuh:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the GE Voluson E8 Ultrasound System, as described in your premarket notification:
Image /page/1/Picture/11 description: The image shows a logo for the FDA Centennial 1906-2006. The logo is circular in shape and contains the letters FDA in a bold, stylized font. Below the letters, the word Centennial is written in a smaller font. The logo is surrounded by text that reads "A Century of Public Health".
*Protecting and Promoting Public Health*
{2}------------------------------------------------
#### Transducer Model Number
| RAB2-5-D | IC5-9-D | 9L-D |
|-----------|-----------|-------|
| RAB4-8-D | PA6-8-D | M12LW |
| RIC5-9-D | SP10-16-D | 3S-D |
| RNA5-9-D | RSP6-16-D | P2D |
| RRE6-10-D | RIC6-12-D | P6D |
| AB2-7-D | RAM3-8 | M6C |
| 4C-D | RSM5-14 | 11L-D |
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
{3}------------------------------------------------
Page 2 -- Mr. Schuh
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Ewa Czerska at (301) 594-1212.
Sincerely yours,
David Ch. Leverson
Image /page/3/Picture/6 description: The image shows the word "for" written in cursive. The word is written in black ink on a white background. The letter "f" is larger than the other letters and has a loop at the top. The letters "o" and "r" are connected.
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{4}------------------------------------------------
# Attachment E Indications for Use Forms
The following forms represent indications with clinical applications and The following forms roprocent indicant operation for the Voluson E8 exam types along with the mobelmode combinations. Combinations identified by "N" are new while "P" represents those previously cleared with the unmodified Voluson 730. In a similar manner, "E" represents with the unmodified Voluson Toodified Voluson 730 via Appendix E of Confinitions added to the subject modification does not alter the the 510(k) Cleared system level indications, clinical applications or modes of operation.
{5}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
### GE Voluson E8 Ultrasound System
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Pediatric | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Small Organ[2] | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | [5] |
| Adult Cephalic | P | P | P | P | P | P | P | P | P | P | |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | [5] |
| Peripheral Vascular | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Superficial | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | P | P | P | | P | P | P | P | P | P | |
| Intraoperative Neurological | P | P | P | | P | P | P | P | P | P | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic.
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Adult and Pediatric.
[5] 3D/4D Imaging Mode.
[6] Includes imaging of guidance of biopsy (2D/3D/4D).
[7] Includes infertility monitoring of follicle development.
[8] Includes urology/prostate.
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Ingram
{6}------------------------------------------------
### GE Voluson E8 with RAB2-5-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic
[5] 3D/4D Imaging Mode
[6] Includes imaging of quidance of biopsy (3D/4D)
[7] Includes infertility monitoring of follicle development
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Davidh. Seymon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number 206/682
{7}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
### GE Voluson E8 with RAB4-8-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Pediatric | E | E | E | E | E | E | E | E | E | E | [5,6] |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | P | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology
[5] 3D/4D Imaging Mode
[6] Includes imaging of guidance of biopsy (3D/4D)
[7] Includes infertility monitoring of follicle development
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
510(k) Number _
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
{8}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
### GE Voluson E8 with RIC5-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [5,6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [5] 3D/4D Imaging Mode
[6] Includes imaging of quidance of biopsy (3D/4D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal,
Prescription User (Per 21 CFR 801.109) and Radiological Devices 12061682
510(k) Number
{9}------------------------------------------------
### GE Voluson E8 with RNA5-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Abdominal[1] | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Pediatric | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Small Organ[2] | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Neonatal Cephalic | P | P | P | E | P | P | P | P | P | P | [5] |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | P | P | P | E | P | P | P | P | P | P | [5] |
| Peripheral Vascular | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Conventional | P | P | P | E | P | P | P | P | P | P | [5,6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal is Neonatal and pediatric
[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients [3] Cardiac is Neonatal and Pediatric.
[5] 3D/4D Imaging Mode
[6] Includes imaging of guidance of biopsy (3D/4D)
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel A. Seymore
(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number _
{10}------------------------------------------------
### Diagnostic Ultrasound Indications for Use Form
### GE Voluson E8 with RRE6-10-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | | | | | | | | | | | | |
| Abdominal[1] | | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | | [5,6] |
| Transvaginal | E | E | E | | E | E | E | E | E | E | | [5,6] |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [5] 3D/4D Imaging Mode
[6] Includes imaging of guidance of biopsy (3D/4D)
[8] Includes urology/prostate
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription User (Per 21 CFR 801.109)
Prescription User (Per 21 CFR 801.109)
{11}------------------------------------------------
### GE Voluson E8 with AB2-7-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>(Notes) |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] |
| Abdominal[1] | P | P | P | | P | P | P | P | P | P | [6] |
| Pediatric | P | P | P | | P | P | P | P | P | P | [6] |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | P | P | P | | P | P | P | P | P | P | [6] |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology
[6] Includes imaging of guidance of biopsy (2D)
[7] Includes infertility monitoring of follicle development
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Seggerman
(Division Sign-Off) Division of Reproductive, Abdon and Radiological Devices 510(k) Number _
{12}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
### GE Voluson E8 with 4C-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | E | P | P | P | P | P | P | [6] | |
| Abdominal[1] | P | P | P | E | P | P | P | P | P | P | [6] | |
| Pediatric | | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | | |
| Peripheral Vascular | P | P | P | E | P | P | P | P | P | P | [6] | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology
[6] Includes imaging of guidance of biopsy (2D)
[7] Includes infertility monitoring of follicle development
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Swanson
(Division Sign-Off) Division of Reproductive, Abdomir and Radiological Devices 510(k) Number
{13}------------------------------------------------
### GE Voluson E8 with IC5-9-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic | | | | | | | | | | | |
| Fetal / Obstetrics[7] | P | P | P | | P | P | P | P | P | P | [6] |
| Abdominal[1] | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ[2] | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac[3] | | | | | | | | | | | |
| Peripheral Vascular | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal[8] | P | P | P | | P | P | P | P | P | P | [6] |
| Transvaginal | P | P | P | | P | P | P | P | P | P | [6] |
| Transuretheral | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: {6} Includes imaging of guidance of biopsy (2D)
[7] Includes infertility monitoring of follicle development
[8] Includes urology/prostate
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel h. Sprom
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number __
{14}------------------------------------------------
### GE Voluson E8 with PA6-8-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] | |
| Ophthalmic | | | | | | | | | | | | |
| Fetal / Obstetrics[7] | | | | | | | | | | | | |
| Abdominal[1] | E | E | E | E | E | E | E | E | E | E | | |
| Pediatric | E | E | E | E | E | E | E | E | E | E | | |
| Small Organ[2] | | | | | | | | | | | | |
| Neonatal Cephalic | P | P | P | P | P | P | P | P | P | P | | |
| Adult Cephalic | | | | | | | | | | | | |
| Cardiac[3] | P | P | P | P | P | P | P | P | P | P | | |
| Peripheral Vascular | | | | | | | | | | | | |
| Musculo-skeletal Conventional | | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | | |
| Other[4] | | | | | | | | | | | | |
| Exam Type, Means of Access | | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | | |
| Transrectal[8] | | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | | |
| Transuretheral | | | | | | | | | | | | |
| Intraoperative | | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | | | |
N = new indication; P = previously cleared by FDA; E = added under Appendix E
Notes: [1] Abdominal is Neonatal , pediatric and obstetrics
[3] Cardiac is Adult and Pediatric.
[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.
(Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David R. Ingram
(Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number __
{15}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
### GE Voluson E8 with SP10-16-D Transducer
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | Mode of Operation | | | | | | | | | | |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler |…