Synapse Image Visualization Software MIP/MPR Obliquus enables the display of 3D (MIP/MPR) visualization of CT and MR studies. Typical users are radiologists, technologists and clinicians.
Device Story
Obliquus is an integrated software package for use with Fuji Synapse Workstation (v3.1.1+); processes DICOM-compliant CT and MRI studies. Inputs: CT/MRI image series. Processing: performs Maximum Intensity Projection (MIP) and Multi-Planar Reformatting (MPR) to generate orthogonal (coronal/sagittal) and oblique plane views; includes Compare viewer for synchronous multi-series display and Stereo viewer for MRA visualization. Output: 3D/2D visualizations displayed on standard PC hardware. Used in clinical settings by radiologists/clinicians for image review, storage, and distribution. Output assists in diagnostic interpretation and clinical decision-making by providing enhanced spatial visualization of anatomical structures.
Clinical Evidence
Bench testing only. Device functionality verified against Software Requirements Specifications via traceability matrix and functional testing during product development. No clinical data provided.
Technological Characteristics
Software-based image visualization; runs on standard PC hardware (Windows 2000/XP); requires 1-2 GB RAM. Integrates with Synapse medical network system. Implements MIP and MPR algorithms for DICOM image processing. No specific hardware materials or sterilization required as it is a software-only package.
Indications for Use
Indicated for radiologists, technologists, and clinicians to display 3D (MIP/MPR) visualizations of CT and MR studies.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Sectra IDS5 Radiology Workstation with Sectra 3D and MPR packages (K040376)
Submission Summary (Full Text)
{1}------------------------------------------------
Kobl672
JUN 2 9 2006
## 510(k) Summary
#### Date Prepared
May 12, 2006
## Submitter's Information
FUJIFILM Medical Systems U.S.A., Inc. 419 West Avenue Stamford, Connecticut 06902
Telephone: (203) 602-3774 Facsimile: (203) 363-3813 Contact: Debra A. Peacock
#### Device Trade Name,
Synapse 3D MIP/MPR Image Visualization Software OBLIQUUS
#### Device Common Name:
Image processing, management and 3D (MIP/MPR) visualization system
#### Regulation Number:
892.2050
#### Device Classification: Class II
Device Classification Name: Picture Archiving Communication System (PACS)
#### Panel:
Radiology
## Product Code:
90 LLZ
Date Received: TBD
## Decision Date:
TBD
## Decision:
TBD
#### Predicate Device
Sectra IDS5 Radiology Workstation with Sectra 3D and MPR packages 510(k) #: K040376, cleared by CDRH on 05/04/04
{2}------------------------------------------------
#### Description of the Device
Obliquus is an integrated Maximum Intensity Projection (MIP) and Multi-Planar Reformatting (MPR) software package intended to be used with Fuji Synapse Workstation Software Version 3.1.1 and higher. This software is designed to be used by radiologists, clinicians and referring physicians to acquire, process, review, store, print and distribute DICOM compliant image studies utilizing standard PC hardware. Obliquus offers similar user interface and image manipulation functionality as was cleared in Synapse Workstation software version 3.1.0.
The Obliquus software enables the user to display MPR visualization of CT or MRI studies in Synapse. Once opened in Synapse, the images can be presented in the following viewers:
MPR Viewer: Visualizes a study image of the selected series in the following four types of plane images:
(1) Original view is the original CT or MRI study image.
(2) Reformatted A view represents visualization of the original view orthogonal to the original study plane (i.e., if original image is axial, then Reformatted A will be of the coronal plane).
(3) Reformatted B view represents visualization of the original view orthogonal to the original study plane (i.e., if the original image is axial, then Reformatted B will be of the sagittal plane).
(4) Oblique view shows a plane image from any direction specified by the user.
Compare Viewer: Compare viewer consists of four windows and up to four series that can be displayed for visual comparison. The Compare viewer functions are as follows:
- . Synchronously display images of each series
- Make synchronous display setting which includes adjusting slice position and changing . synchro unit
- . Switch among 3 orthogonal plane images
- Set rendering mode and slab thickness .
- Display patient study and image information .
- . Toggle with the MPR viewer
Stereo Viewer: Presents images for stereopsis, particularly for Magnetic Resonance Angiography (MRA) MIP images. The Stereo viewer's functions include:
- . Rotating and displaying plane images horizontally
- . Setting image display format
- . Setting rendering mode, slab thickness and projection angle interval
- . Display patient, study and image information
- . Toggle with the MPR viewer
Images can be saved either as bitmap files or to the Synapse server.
{3}------------------------------------------------
The maximum allowable number of slices for which images can be visualized depends on the memory capacity of the computer being used. With main memory of 1G bytes, approximately 600 slices of ordinary CT images can be processed.
#### Software Specifications
Obliquus is optional software for use with the medical network system SYNAPSE version 3.1.1 and higher. To allow Obliguus to run on a PC, it is essential that SYNAPSE is running on that same PC. Specifications for operating this Software are congruous with the SYNAPSE specifications. Below are notes specific to this Software:
## (1) Supported operating systems
Windows 2000 and Windows XP.
## (2) Measure of required memory capacity
1 to 2 Gbytes.
#### Intended Use
## Obliquus:
Synapse Image Visualization Software MIP/MPR Obliquus enables the display of 3D (MIP/MPR) visualization of CT and MR studies. Typical users are radiologists, technologists and clinicians.
## Predicate Device:
The Sectra IDS5 device is intended for the manipulation and displaying of x-ray images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.
Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3D and angiography, etc .; and teleconferencing.
Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
#### Technological Characteristics
The optional Synapse Image Visualization Software (MIP/MPR) OBLIQUUS is comparable and technically equivalent to the visualization software contained in the Sectra IDS5 device with Sectra 3D and MPR software package in that both are integrated into their PACS workstation and can view 2D images .
#### Safety Information
Synapse Image Visualization Software (MIP/MPR) Obliquus introduces no new safety or efficacy issues other than those already identified with the predicate device. The results of the Hazard Analysis combined with the appropriate preventive measures taken indicate that the device is of minor concern as per the May 11, 2005 issue of the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
{4}------------------------------------------------
## Testing
Obliquus is tested with reference to its Software Requirements Specifications, as documented in Section 16.3, Traceability Matrix Management included in this 510(k) filing. Functional testing is a part of the Product Development process, also included in Section 16 of this filing.
## Conclusion
This 510(k) premarket notification submission has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health. We conclude the subject device to be as safe and effective as the predicate devices.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle or bird-like symbol, depicted with three curved lines that suggest wings or feathers.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
JUN 2 9 2006
Fujifilm Medical Systems USA, Inc. % Mr. Casey Conry Project Engineer Underwriters Laboratory, Inc. 1285 Wait Whitman Rd. MELVILLE NY 11747
Re: K061672
Trade/Device Name: SYNAPSE 3D Visualization Software OBLIOUUS Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: June 13, 2006 Received: June 14, 2006
Dear Mr. Conry:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/4/Picture/10 description: The image shows a circular logo with the letters "FDA" in the center. Above the letters, the years "1906 - 2006" are displayed. Below the letters, the word "Centennial" is written. The logo is surrounded by text, but the text is too blurry to read.
*Protecting and Promoting Public Health*
{6}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
| 510(k) Number (if known): | |
|---------------------------|--------------------------------------------|
| Device Name: | Synapse 3D Visualization Software Obliquus |
| Indications for Use: | |
Synapse Image Visualization Software MIP/MPR Obliquus enables the display of 3D (MIP/MPR) visualization of CT and MR studies. Typical users are radiologista, technologists and clinicians.
Prescription Use (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter-Use (Part 21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IS NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
---
(Division Sign-Off)
()
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ K 1. 6. 1672
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.