ARX SPINAL SYSTEM

K061600 · Life Spine · MNI · Aug 22, 2006 · Orthopedic

Device Facts

Record IDK061600
Device NameARX SPINAL SYSTEM
ApplicantLife Spine
Product CodeMNI · Orthopedic
Decision DateAug 22, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing. The ARX™ Spinal System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization and immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities. When used as a posterior spine thoracic/lumbar system, the ARXTM Spinal System is indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, kyphosis, and/or lordosis), (4) spinal tumor, (5) failed previous fusion (6) pseudarthrosis, (7) spinal stenosis, (8) spondylolisthesis.

Device Story

ARX Spinal System consists of single-use titanium alloy bone screws and rods; creates rigid structure for spinal stabilization and fusion. Used in posterior thoracic/lumbar spine; installed by surgeons using Class I surgical instruments. Provides load-sharing support to maintain alignment during bone healing. Benefits patients by immobilizing spinal segments to facilitate fusion in cases of trauma, deformity, or degenerative disease.

Clinical Evidence

Bench testing only. Biomechanical testing conducted in accordance with ASTM F1717 to demonstrate substantial equivalence.

Technological Characteristics

Materials: Medical grade titanium alloy (ASTM F136). Components: Bone screws and rods. Design: Top-loading thoracolumbar spinal system. Energy source: None (mechanical). Sterilization: Not specified.

Indications for Use

Indicated for skeletally mature patients requiring posterior pedicle screw fixation of the non-cervical spine for stabilization and immobilization as an adjunct to fusion. Conditions include degenerative disc disease, trauma (fracture/dislocation), curvatures (scoliosis, kyphosis, lordosis), spinal tumor, failed previous fusion, pseudarthrosis, spinal stenosis, and spondylolisthesis.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the words "Life Spine" in a bold, sans-serif font. The word "Life" is on the left, and the word "Spine" is on the right. There is a curved line underneath the word "Spine". There is a registered trademark symbol next to the word "Spine". AUG 2 2 2006 K061600 pg 1 of 2 # 510(k) Summary of Safety and Effectiveness 510(k) Summary - ARXTM Spinal System | Submitted By: | Life Spine<br>2400 Hassell Road, Suite 370<br>Hoffman Estates, IL 60195 | |----------------------|-----------------------------------------------------------------------------------------------------| | | Telephone: 847-884-6117<br>Fax: 847-884-6118 | | 510(k) Contact: | Erin Malloy<br>Life Spine<br>2400 Hassell Road, Suite 370<br>Hoffman Estates, IL 60195 | | Date Prepared: | June 7, 2006 | | Trade Name: | ARX <sup>TM</sup> Spinal System | | Common Name: | Appliance, Fixation, Spinal Interlaminal | | Classification: | 888.3070 Pedicle screw spinal system | | Device Product Code: | MNH, CFR 888.3070, Class II<br>MNI, CFR 888.3070, Class II | | Predicate Device: | DePuy AcroMed <sup>TM</sup> Moss® Miami Spinal System<br>Stryker Spine TRIO® Spinal Fixation System | ### Device Description: The ARX™ Spinal System includes various types and sizes of single use implantable components. When assembled, the components create a rigid structure providing stabilization and promoting spinal fusion. The system is comprised of bone screws and rods. Class I surgical instruments are utilized for the installation of the implant. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Life Spine. The words "Life Spine" are written in a bold, sans-serif font. A curved line is underneath the word "Spine". The registered trademark symbol is next to the word "Spine". K061600 ps 2 of 2 #### Intended Use of the Device: Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing. The ARX™ Spinal System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization and immobilization of spinal segments as an adjunct to fusion. When used as a posterior spine thoracic/lumbar system, the ARX™ Spinal System is indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, kyphosis, and/or lordosis), (4) spinal tumor, (5) failed previous fusion (6) pseudarthrosis, (7) spinal stenosis, (8) spondylolisthesis. #### Material: The ARX™ Top-Loading Thoracolumbar Spinal System is manufactured from medical grade titanium alloy described by ASTM F136. #### Performance Data: Biomechanical testing in accordance with ASTM F1717 was conducted to demonstrate substantial equivalency. #### Substantial Equivalence: The ARX™ Spinal System was shown to be substantially equivalent to previously cleared devices, the DePuy AcroMedTM Moss® Miami Spinal System and the Stryker Spine TRIO® Spinal Fixation System, in indications for use, design, function, and materials used. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three figures instead of one. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 . ## AUG 2 2 2006 Life Spine % Ms. Erin Malloy Project Engineer 2400 Hassell Road, Suite 370 Hoffman Estates, Illinois 60195 Re: K061600 Trade/Device Name: ARX Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle Screw Spinal System Regulatory Class: Class II Product Code: MNH, MNI Dated: June 7, 2006 Received: June 8, 2006 Dear Ms. Malloy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Erin Malloy This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Barbara buchlun for Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K∅616∅∅ # Life S 1 ## Indications for Use 510(k) number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: ARX™ Spinal System Internal fixation implants are load-sharing devices intended to stabilize and maintain alignment until normal healing occurs. Implants are not intended to replace normal body structures or bear the weight of the body in the presence of incomplete bone healing. The ARX™ Spinal System, when properly used, is intended for posterior pedicle screw fixation of the non-cervical posterior spine in skeletally mature patients. It provides stabilization and immobilization of spinal segments as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities. When used as a posterior spine thoracic/lumbar system, the ARXTM Spinal System is indicated for one or more of the following: (1) degenerative disc disease (is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) trauma (i.e. fracture or dislocation), (3) curvatures (scoliosis, kyphosis, and/or lordosis), (4) spinal tumor, (5) failed previous fusion (6) pseudarthrosis, (7) spinal stenosis, (8) spondylolisthesis. Prescription Use x x (Part 21 CFR 801 Subpart D) And/Or Over-the-Counter Use _ (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Barbave Buehns Division of General. Restorative. and Neurological Devices 510(k) Number K061600
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...