VITROS Chemistry Products HCY Reagent: For in vitro diagnostic use only. VITROS Chemistry Products HCY Reagent is used to quantitatively measure total homocysteine (HCY) concentration in human serum and plasma. Serum and plasma homocysteine levels can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria. VITROS Chemistry Products Calibrator Kit 27: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 27 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of homocysteine (HCY). VITROS Chemistry Products HCY Performance Verifiers I, II & III: For in vitro diagnostic use only. VITROS Chemistry Products HCY Performance Verifiers are assayed controls used to monitor performance of VITROS HCY Reagents on VITROS 5,1 FS Chemistry Systems.
Device Story
VITROS Chemistry Products HCY Reagent, Calibrator Kit 27, and Performance Verifiers are used on VITROS 5,1 FS Chemistry Systems to measure total homocysteine in human serum/plasma. Disulfide-linked homocysteine is reduced by TCEP; reduced homocysteine reacts with serine via cystathionine β-synthase (CBS) to form L-cystathionine. L-cystathionine is cleaved by cystathionine β-lyase (CBL) to produce homocysteine, pyruvate, and ammonia. Pyruvate is reduced to lactate by lactate dehydrogenase (LDH) using NADH. NADH conversion to NAD is measured spectrophotometrically at 340 nm; concentration is directly proportional to NADH conversion. System uses stored calibration curves to determine sample concentration. Used in clinical laboratory settings by trained personnel. Results assist clinicians in diagnosing and treating hyperhomocysteinemia and homocystinuria.
Clinical Evidence
No clinical diagnostic studies performed. Analytical performance established via bench testing: precision (within-lab CV 1.1-3.4%), linearity (0.5-71.0 μmol/L), and method comparison against predicate (n=93, y=0.98x + 1.0, r=0.97). Reference intervals established using 205 healthy adults (106 female, 99 male). Interference testing conducted for various endogenous and exogenous substances.
Indicated for the quantitative measurement of total homocysteine in human serum and plasma to assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
Regulatory Classification
Identification
A urinary homocystine (nonquantitative) test system is a device intended to identify homocystine (an analogue of the amino acid cystine) in urine. The identification of urinary homocystine is used in the diagnosis and treatment of homocystinuria (homosystine in urine), a heritable metabolic disorder which may cause mental retardation.
Predicate Devices
Bio-Rad Homocysteine by HPLC Test (k993107)
VITROS Chemistry Products AAT Performance Verifiers I, II, & III (k052819)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k061588
B. Purpose for Submission:
New device
C. Measurand:
Homocysteine
D. Type of Test:
Quantitative, homogeneous enzymatic assay
E. Applicant:
Ortho-Clinical Diagnostics, Inc.
F. Proprietary and Established Names:
VITROS Chemistry Products HCY Reagent, VITROS Chemistry Products Calibrator Kit 27, VITROS HCY Performance Verifiers I, II, and III
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| Urinary Homocysteine (Nonquantitative) Test System (LPS) | Class II | 21 CFR 862.1377, Urinary homocysteine (nonquantitative) test system. | 75 Clinical Chemistry (CH) |
| Product Code | Classification | Regulation Section | Panel |
| Calibrators, Secondary (JIT) | Class II | 21 CFR 862.1150, Calibrator | 75 Clinical Chemistry (CH) |
| Product Code | Classification | Regulation Section | Panel |
| Single (Specified) analyte controls (assayed and unassayed) (JJX) | Class I | 21 CFR 862.1660, Quality Control Material (assayed and unassayed) | 75 Clinical Chemistry (CH) |
{2}
H. Intended Use:
1. Intended use(s):
VITROS Chemistry Products HCY Reagent: For in vitro diagnostic use only. VITROS Chemistry Products HCY Reagent is used to quantitatively measure total homocysteine (HCY) concentration in human serum and plasma. Serum and plasma homocysteine levels can assist in the diagnosis and treatment of patients suspected of having hyperhomocysteinemia and homocystinuria.
VITROS Chemistry Products Calibrator Kit 27: For in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 27 is used to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of homocysteine (HCY).
VITROS Chemistry Products HCY Performance Verifiers I, II & III: For in vitro diagnostic use only. VITROS Chemistry Products HCY Performance Verifiers are assayed controls used to monitor performance of VITROS HCY Reagents on VITROS 5,1 FS Chemistry Systems.
2. Indication(s) for use:
See Intended Use above.
3. Special conditions for use statement(s):
Prescription use only
4. Special instrument requirements:
The device is intended for use with the VITROS 5,1 FS Chemistry System. The VITROS 5,1 FS Chemistry System was cleared for market use in k031924.
I. Device Description:
The VITROS Chemistry Products HCY Reagent consists of two dual chambered reagent packs containing three ready-to-use liquid reagents. VITROS Chemistry Products FS Diluent Pack 2 (Saline/BSA) is a common reagent that is used by multiple assays on the VITROS 5,1 FS Chemistry System. This is a dual chambered package containing two ready-to-use liquid diluents. Diluent 1 is prepared from processed water to which inorganic salt has been added. Diluent 2 is prepared from processed water to which bovine serum albumin, inorganic salts and preservatives have been added. Diluent 1 contains saline (154 mM NaCl), which is used to dilute elevated homocysteine (HCY) samples on the analyzer.
VITROS Chemistry Products Calibrator Kit 27 is a two level fluid prepared from an aqueous solution containing amino acids and inorganic acid. These standards are used
{3}
to calibrate VITROS 5,1 FS Chemistry Systems for the quantitative measurement of homocysteine (HCY).
VITROS HCY Performance Verifiers I, II and III are prepared from processed human serum to which amino acid and preservative have been added. These are assayed controls used to monitor performance of VITROS HCY Reagent on VITROS 5,1 FS Chemistry Systems.
J. Substantial Equivalence Information:
| Predicate | k993107 Bio-Rad Homocysteine by HPLC Test |
| --- | --- |
| Describe the item being compared | |
| The Bio-Rad Homocysteine by HPLC test is based on precolumn derivatization and a 5 minute chromatography. The Bio-Rad Homocysteine by HPLC test is intended for the quantitative determination of total homocysteine in human plasma or serum. | |
Device Similarities
| Device Characteristic | VITROS HCY (New device)
BIO-RAD HOMOCYSTEINE by HPLC test (Predicate device) |
| --- | --- |
| Intended use | For in-vitro diagnostic use. Quantitative determination of total homocysteine in human serum or plasma |
| Analyte measured | Homocysteine |
| Sample type | Serum or Plasma |
| Measurement | Quantitative |
Device Differences
| Device Characteristics | VITROS HCY (New device) | BIO-RAD HOMOCYSTEINE by HPLC test (Predicate device) |
| --- | --- | --- |
| Reportable range | 1.0 – 50.0 μmol/L | 0.5 – 100 μmol/L |
| Sensitivity | 1.0 μmol/L | 0.5 μmol/L |
| Calibrator levels | Two levels (0, 27 μmol/L) | Single level (15-20 μmol/L) |
| Calibrator format | Liquid | Lyophilized |
| Calibrator matrix | Aqueous solution containing amino acids and inorganic acid | Human Serum |
| Instrumentation | Automated clinical chemistry analyzer | Isocratic HPLC System |
{4}
4
| Reference Interval | Males: 6.6 – 14.8 μmol/L
Females: 4.7 – 12.6 μmol/L | < 15 μmol/L |
| --- | --- | --- |
| Method | Homogeneous Enzymatic | Chromatography |
| Predicate | k052819 VITROS Chemistry Products AAT
Performance Verifiers I, II, & III |
| --- | --- |
Describe the item being compared
The VITROS Chemistry Products AAT Performance Verifiers I, II and III are prepared from processed human serum to which inorganic salts, buffers, and preservatives have been added. These are assayed controls used to monitor performance of VITROS AAT Reagent on VITROS 5,1 FS Chemistry Systems.
Device Similarities
| Indications for use | Assayed controls used to monitor assay performance on VITROS 5,1 FS Chemistry Systems |
| --- | --- |
| Matrix | Prepared from processed human serum with preservatives added. |
| Product type | Assayed control |
| Format | Liquid |
| Number of levels | Three |
Device Differences
| Device Characteristic | VITROS HCY Performance Verifiers (new device) | VITROS AAT Performance Verifiers (predicate device) |
| --- | --- | --- |
| Analyte measure | Homocysteine | alpha 1-Antitrypsin (AAT) |
K. Standard/Guidance Document Referenced (if applicable):
STANDARDS
Title and Reference Number
Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline (EP5-A)
Interference Testing in Clinical Chemistry; Approved Guideline (EP 7-A)
Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline (EP09-A2)
How to Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline - Second Edition (C28-A2)
Other Standards
{5}
5
| None | | | |
| --- | --- | --- | --- |
| GUIDANCE | | | |
| Document Title | Office | Division | |
| None | | | |
## L. Test Principle:
The quantitative measurement of homocysteine (HCY) is performed using the VITROS Chemistry Products HCY Reagent in conjunction with the VITROS Chemistry Products Calibrator Kit 27 and on the VITROS 5,1 FS Chemistry Systems. The VITROS Chemistry Products HCY Reagent consists of two dual chambered reagent packs containing three ready-to-use liquid reagents. Disulfide linked homocysteine (oxidized forms) in the sample is reduced by Tris (2-Carboxyethyl) phosphine hydrochloride (TCEP) to form reduced homocysteine. Reduced homocysteine reacts with serine in the presence of cystathionine b-synthase (CBS) to form L-cystathionine. L-cystathionine is broken down by cystathionine b-lyase (CBL) to produce homocysteine, pyruvate and ammonia. Pyruvate is reduced to lactate by lactate dehydrogenase (LDH) using NADH as coenzyme. The concentration of homocysteine is directly proportional to the amount of NADH converted to NAD⁺ and is measured spectrophotometrically at 340 nm. Once a calibration has been performed, the homocysteine concentration in each unknown sample can be determined using the stored calibration curve and the measured absorbance obtained in the assay of the sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
Precision was evaluated with quality control materials and a human serum pool on the VITROS 5,1 FS Chemistry System following NCCLS Protocol EP5.
{6}
VITROS Chemistry Products HCY Reagent Precision (Serum)
| System | SI Units (μmol/L) | | | Within-Lab %CV | No. Days | No. Obser |
| --- | --- | --- | --- | --- | --- | --- |
| VITROS 5,1 FS | Mean Conc. (μmol/L) | Within-Day SD* | Within-Lab SD** | | | |
| | 7.0 | 0.21 | 0.24 | 3.4 | 20 | 80 |
| | 11.9 | 0.23 | 0.27 | 2.3 | 20 | 80 |
| | 41.8 | 0.33 | 0.44 | 1.1 | 20 | 80 |
| | 23.7 | 0.24 | 0.30 | 1.3 | 20 | 80 |
| | 48.0 | 0.52 | 0.66 | 1.4 | 20 | 80 |
*Within Day precision was determined using two-runs/ day with two replicates per run
** Within Lab precision was determined using a single lot of reagents on a single analyzer, calibrating once a week (4 calibrations).
b. Linearity/assay reportable range:
The evaluation of the linearity of the VITROS HCY assay was performed based on NCCLS EP6-A. Three lots of VITROS Chemistry Products HCY Reagents were tested using a series of admixtures with concentrations of 0.50 μmol/L to 71.0 μmol/L. The VITROS Chemistry Products HCY Reagent linear range was determined to be 0.50 μmol/L to 71.0 μmol/L and the reportable range was chosen by the applicant to be 1.0 μmol/L to 50.0 μmol/L.
An on-board dilution study of patient samples with saline was performed using 10 samples with a dilution factor of 2 and 10 samples with a dilution factor of 3. The recovery ranged from 96.9% to 100.0% using the dilution factor of 2 and ranged from 92.4% to 98.5% using the dilution factor of 3.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Traceability**
Both the Calibrators and Verifiers are value assigned by performing multiple determinations using the VITROS 5,1 FS Chemistry Systems and the VITROS Chemistry Products HCY Reagent. The VITROS Chemistry Products Calibrator Kit 27 is traceable to NIST SRM 1955.
**Stability**
The long term (shelf life) stability of the VITROS HCY reagents, calibrators and verifiers were tested on a VITROS 5,1 FS Chemistry System using a real time study extending at least one month beyond the selected expiration date. Testing was done at intervals of 0, 1, 2, 3 months and is still ongoing. The
{7}
VITROS HCY reagents, calibrators and verifiers are stored at refrigerated (2- $8^{\circ}\mathrm{C}$ ).
The opened vial (in use) stability of the VITROS HCY reagents, calibrators and verifiers were tested on a VITROS 5,1 FS Chemistry System. Testing was done at intervals of 0, 7, 14 days and is still ongoing.
# d. Detection limit:
The LOB specimen tested was a homocysteine-free serum pool obtained from Golden West Biologics Inc, Temecula, CA. The LOD specimen tested was consisted of two homocysteine serum pools spiked with homocysteine to nominal concentrations of 0.01 and $0.48\mu \mathrm{mol} / \mathrm{L}$ . The Limit of Blank (LOB), Limit of Detection (LOD), and Limit of Quantitation (LOQ) were determined following NCCLS EP17-A. The lower linear range of the VITROS HCY assay was determined to be $0.5\mu \mathrm{mol} / \mathrm{L}$ based on the results of the linearity evaluation. The value that will be the claimed analytical limit (reportable range) at low levels is $1.0\mu \mathrm{mol} / \mathrm{L}$ based on linearity, LOB, LOD, and LOQ evaluations.
# e. Analytical specificity:
The specificity of the VITROS HCY reagent was evaluated by testing the following substances at homocysteine concentrations of approximately 15.0 to $38.0~\mu \mathrm{mol / L}$ using protocols based on NCCLS Protocol EP7-A and found not to interfere, bias $< 13.6\%$ , at the concentrations shown below.
| Compound | Concentration | |
| --- | --- | --- |
| Acetaminophen | 20 mg/dL | Acetaminophen |
| N-Acetyl L-Cysteine | 326 μg/L | N-Acetyl L-Cysteine |
| Adenosine | 6.7 g/dL | Adenosine |
| Amoxicillin | 20 μg/mL | Amoxicillin |
| Ampicillin | 100 mg/dL | Ampicillin |
| Ascorbic Acid (L) | 10 mg/dL | Ascorbic Acid (L) |
| Aspirin | 50 mg/dL | Aspirin |
| Atorvastatin | 300 ng/mL | Atorvastatin |
| Bilirubin | 30 mg/dL | Bilirubin |
| Caffeine | 10 mg/dL | Caffeine |
| Clopidogrel hydrogensulfate | 50 μg/mL | Clopidogrel hydrogensulfate |
| Creatinine | 30 mg/dL | Creatinine |
| L-Cysteine | 2.4 g/dL | L-Cysteine |
{8}
| Compound | Concentration | |
| --- | --- | --- |
| Doxycyclin | 5 mg/dL | Doxycyclin |
| D-penicillamine | 8 μg/mL | D-penicillamine |
| Enalapril Maleate | 120 ng/mL | Enalapril Maleate |
| Gemfibrozil | 50 μg/mL | Gemfibrozil |
| Gentamycin Sulfate | 120 μg/mL | Gentamycin Sulfate |
| Glutathione | 0.6 g/L | Glutathione |
| DL-Homocysteine thiolactone | 0.77 mg/dL | DL-Homocysteine thiolactone |
| Hemoglobin | 500 mg/dL | Hemoglobin |
| Hydrochlorothiazide | 2 μg/mL | Hydrochlorothiazide |
| Ibuprofen | 40 mg/dL | Ibuprofen |
| Intralipid | 500 mg/dL | Intralipid |
| L-Methionine | 29.8 mg/dL | L-Methionine |
| S-(5.-Adenosyl)-L-Methionine p-toluene sulfonate | 79.9 mg/dL | S-(5.-Adenosyl)-L-Methionine p-toluene sulfonate |
| Metronidazole | 20 mg/dL | Metronidazole |
| Nicotine | 2 mg/dL | Nicotine |
| Nifedipine | 200 ng/mL | Nifedipine |
| Procainamide | 100 μg/mL | Procainamide |
| Propanolol | 5 μg/mL | Propanolol |
| Protein (Serum) | 9.3 g/dL | Protein (Serum) |
| Pyruvic Acid | 17.6 mg/dL | Pyruvic Acid |
| Rheumatoid Factor | 373 IU/mL | Rheumatoid Factor |
| Theophylline | 250 μg/mL | Theophylline |
| Triglycerides | 908 mg/dL | Triglycerides |
| Urea | 500 mg/dL | Urea |
| Warfarin | 10 μg/mL | Warfarin |
Known Interferences
- Patient samples containing cystathionine (reference interval: 0.065 to 0.301 $\mu$mol/L) show positive equimolar interference with the VITROS Chemistry
{9}
Products HCY assay. For example, a sample containing 10.0 μmol/L homocysteine and 0.3 μmol/L cystathionine will generate a VITROS HCY result of 10.3 μmol/L. Elevated cystathionine levels can be observed in patients with certain conditions, such as renal disease, folate deficiency, and cystathionine beta-lyase deficiency.
- For samples that generate a Sample Integrity T-index flag, refer to the Sample Dilution section in the package insert.
## Other Limitations
Results should be interpreted with caution. Certain drugs and clinical conditions are known to alter homocysteine concentration *in vivo*. For example, patients who are taking methotrexate, carbamazepine, phenytoin, nitrous oxide, anticonvulsants, or 6-azuridine triacetate, may have higher levels of homocysteine due to metabolic interference with homocysteine metabolism. For additional information, refer to one of the published summaries cited in the package insert.
f. Assay cut-off:
None
2. Comparison studies:
a. Method comparison with predicate device:
A method comparison study was performed on 93 human serum samples on the VITROS 5,1 FS Chemistry Systems and a commercially available method, based on NCCLS Protocol EP9. Three reagent lots were used for the testing and all samples were analyzed in triplicate. A least squares linear regression analysis demonstrated the following relationship: y=0.98 x + 1.0 μmol/L, with a correlation coefficient = 0.97 where y= results obtained using the VITROS HCY assay and x= results obtained with the commercially available system BIO-RAD homocysteine by HPLC assay.
b. Matrix comparison:
Matrix comparison studies were performed by collecting fresh whole blood from a total of 95 donors. Samples were collected into plain glass, lithium-heparin plasma, K2EDTA plasma, K3EDTA plasma, and serum separator tubes at variable fill volumes and pairings and then centrifuged and the serum separated within 1 hr of collection. For each test condition (specimen type or stability), assay value difference were calculated from the baseline (serum in plain glass for specimen collection and fresh sample for stability). The mean and range of the % differences, across all samples, for each condition were calculated. For each storage condition, means were calculated and the bias from fresh results was compared to the applicant’s acceptance criteria for
9
{10}
determination of acceptability. All the specimen types tested were acceptable at the specimen matrix and stability claims except the plasma samples must be collected in plasma tubes that are at least half full. The information that smaller sample volumes can result in negative bias is presented in the package insert.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
None
4. Clinical cut-off:
None
5. Expected values/Reference range:
Reference intervals were determined for the VITROS HCY assay following NCCLS Guideline C28-A2. The reference interval is based on the central 95 percent of homocysteine values from an Ortho Clinical Diagnostics, Inc. study of 106 healthy female and 99 healthy male adults. Testing was performed on three VITROS 5,1 FS Chemistry System analyzers using 3 lots of VITROS HCY reagents. The reference range for males is 6.6-14.8 μmol/L and for females is 4.7-12.6 μmol/L.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.