The ACON Urinalysis Reagent Strips are for the qualitative and semi-quantitative detection of one or more of the following analytes in urine: glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, leukocytes and ascorbic acid. It is intended for professional use only.
Device Story
ACON Urinalysis Reagent Strips consist of plastic strips with multiple reagent pads; pads react with urine to produce visible color changes. User performs test by dipping strip in urine; compares resulting pad colors to reference blocks on bottle label. No instrumentation required. Used by healthcare professionals in point-of-care or laboratory settings. Results assist in screening for kidney, liver, metabolic, and urinary tract conditions; help determine if further microscopic analysis is necessary. Provides rapid, qualitative/semi-quantitative diagnostic data to support clinical decision-making.
Clinical Evidence
No clinical trials performed. Evidence consists of analytical bench testing (precision, linearity, detection limits, interference studies) and a method comparison study. Consumer study with 125 professionals compared ACON strips to Bayer Multistix 10 SG (visual and instrument read), showing high agreement within +/- one color block across all analytes.
Indicated for professional use in point-of-care or laboratory settings for qualitative/semi-quantitative detection of urine analytes (Glucose, Bilirubin, Ketone, Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes, Ascorbic Acid) to assist in diagnosing kidney function, UTI, carbohydrate metabolism disorders, liver function, acid-base balance, and urine concentration in patients.
Regulatory Classification
Identification
A urinary urobilinogen (nonquantitative) test system is a device intended to detect and estimate urobilinogen (a bile pigment degradation product of red cell hemoglobin) in urine. Estimations obtained by this device are used in the diagnosis and treatment of liver diseases and hemolytic (red cells) disorders.
Predicate Devices
Multistix 10 SG Reagent Strips for Urinalysis (K905396)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k061559
B. Purpose for Submission:
Clearance of a new device
C. Measurand:
Urinary Glucose, Occult Blood, Leukocytes, pH, Nitrite, Protein, Ketones, Urobilinogen, Bilirubin, Specific Gravity and Ascorbic Acid
D. Type of Test:
Qualitative and semi-quantitative urine tests
E. Applicant:
ACON Laboratories, Inc.
F. Proprietary and Established Names:
ACON Urinalysis Reagent Strips
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1340 Urinary Glucose (Non-Quantitative) Test System
21 CFR § 862.1115 Urinary Bilirubin and its Conjugates (Non-Quantitative) Test System
21 CFR § 862.1435 Ketones (Non-Quantitative) Test System
21 CFR § 864.6550 Occult Blood Test
21 CFR § 862.1550 Urinary pH (Non-Quantitative) Test System
21 CFR § 862.1645 Urinary Protein or Albumin (Non-Quantitative) Test System
21 CFR § 862.1785 Urinary Urobilinogen (Non-Quantitative) Test System
21 CFR § 862.1510 Nitrite (Non-Quantitative) Test System
21 CFR § 864.7675 Leukocyte Peroxidase Test
21 CFR § 862.1095 Ascorbic Acid Test System
2. Classification:
Class II: Urinary Glucose and Occult Blood
Class I: Urinary Leukocytes, Urinary pH, Nitrite, Urinary Protein, Ketones, Urinary Urobilinogen, Urinary Bilirubin, Specific Gravity and Ascorbic Acid
3. Product code:
JIL Urinary Glucose (non-quant.) test system
JIO Blood, Occult, Colorimetric, in urine
{2}
LJX Test, Urine Leukocyte
CEN Urinary, pH (non-quant.)
JMT Nitrite (urinary, non-quant.) test system
JIR Protein or Albumin (urinary, non-quant.) test system
JIN Ketones (urinary, non-quant.) test system
CDM Urinary Urobilinogen (non-quant.) test system
JJB Urinary Bilirubin & its conjugates (urinary, non-quant.) test system
JMA Acid, Ascorbic, 2, 4-Dinitrophenylhydrazine (Spectrophotometric)
4. Panel:
Chemistry (75)
Hematology (82)
## H. Intended Use:
1. Intended use(s):
The ACON Urinalysis Reagent Strips are for the qualitative and semi-quantitative detection of one or more of the following analytes in urine: glucose, bilirubin, ketone (acetoacetic acid), specific gravity, blood, pH, protein, urobilinogen, nitrite, leukocytes and ascorbic acid. It is intended for professional use only.
2. Indication(s) for use:
The ACON Urinalysis Reagent Strips (Urine) are for the qualitative and semi-quantitative for detection of one or more of the following analytes in urine: Glucose, Bilirubin, Ketone (Acetoacetic acid), Specific Gravity, Blood, pH, Protein, Urobilinogen, Nitrite, Leukocytes and Ascorbic Acid.
The ACON Urinalysis Reagent Strips (Urine) are for single use in professional near-patient (point-of-care) and centralized laboratory locations. The strips are intended for use in screening at-risk patients to assist diagnosis in the following areas:
- Kidney function
- Urinary track infections
- Carbohydrate metabolism (e.g. diabetes mellitus)
- Liver function
- Acid-base balance
- Urine concentration
The results can be used along with other diagnostic information to rule out certain disease states and to determine if microscopic analysis is needed. The test is to be read visually. It is intended for professional use only.
3. Special conditions for use statement(s):
For prescription use
4. Special instrument requirements:
{3}
Not applicable. The device is a visually read single-use device.
I. Device Description:
The ACON Urinalysis Reagent Strips are urine test strips of which glucose, bilirubin, ketone, specific gravity, blood, pH, protein, urobilinogen, nitrite and leukocyte reagent pads are affixed onto the plastic strips. The reagent pads react with the urine and provide a visible color reaction. Results are obtained by direct comparison of the test strip with the color blocks printed on the bottle label. The product is packaged with a drying agent in a plastic bottle. The entire reagent strip is disposable when the disposal directions are followed exactly. Laboratory instrumentation is not required. These tests are intended for professional use with human urine.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Bayer Multistix 10 SG Reagent Strips for Urinalysis
2. Predicate 510(k) number(s):
k905396
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | Professional use in point-of-care urine testing | Same |
| Target Population | Patients of physicians, hospitals, and clinics | Same |
| Intended Specimen | Urine | Same |
| Material Provided | Plastic strips affixed with several separate reagent areas. | Same |
| Visual read test time | Varies from 30 Seconds to 2 Minutes | Same |
| Glucose Methodology | Same | Same |
| Bilirubin Methodology | Same | Same |
| Ketone Methodology | Same | Same |
| Specific Gravity Methodology | Same | Same |
| Blood Methodology | Same | Same |
| pH Methodology | Same | Same |
| Protein Methodology | Same | Same |
| Urobilinogen Methodology | Same | Same |
| Nitrite Methodology | Same | Same |
| Leukocyte Methodology | Same | Same |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Ascorbic Acid Methodology | This test involves decolorization of Tillmann’s reagent. The presence of ascorbic acid causes the color of the test field to change from blue-green to orange. | This test is not included on Bayer test strip. |
| Storage | 2 to 30°C | 15 to 30°C |
K. Standard/Guidance Document Referenced (if applicable):
None were identified by the applicant.
L. Test Principle:
Glucose: This test is based on the enzymatic reaction that occurs between glucose oxidase, peroxidase and chromogen. Glucose if first oxidized to produce gluconic acid and hydrogen peroxide in the presence of glucose oxidase. The hydrogen peroxide reacts with potassium iodide chromogen in the presence of peroxidase. The extent to which the chromogen is oxidized determines the color which is produced, ranging from green to brown. Low amounts of glucose are normally excreted in urine. Glucose concentrations as low as 100 mg/dL, read at either 10 or 30 seconds, may be considered abnormal if results are consistent. At 10 seconds, results should be interpreted qualitatively. For semi-quantitative results, read at 30 seconds only.
Bilirubin: This test is based on azo-coupling reaction of bilirubin with diazotized dichloroaniline in a strongly acidic medium. Varying bilirubin levels will produce a pinkish-tan color proportional to its concentration in urine. In normal urine, no bilirubin is detectable by even the most sensitive methods. Even trace amounts of bilirubin require further investigation. Atypical results (colors different from the negative or positive color blocks shown on the color chart) may indicate that bilirubin-derived bile pigments are present in the urine specimen, and are possibly masking the bilirubin reaction.
Ketone: This test is based on ketones reacting with nitroprusside and acetoacetic acid to produce a color change ranging from light pink for negative results to a darker pink or purple color for positive results. Ketones are normally not present in urine. Detectable ketone levels may occur in urine during physiological stress conditions such as fasting, pregnancy and frequent strenuous exercise. In starvation diets, or in other abnormal carbohydrate metabolism situations, ketones appear in the urine in excessively high concentration before serum ketones are elevated.
Specific Gravity: This test is based on the apparent pKa change of certain pretreated polyelectrolytes in relation to ionic concentration. In the presence of an indicator,
{5}
colors range from deep blue-green in urine of low ionic concentration to green and yellow-green in urine of increasing ionic concentration. Randomly collected urine may vary in specific gravity from 1.003-1.035. Twenty-four hour urine from healthy adults with normal diets and fluid intake will have a specific gravity of 1.016-1.022. In cases of severe renal damage, the specific gravity is fixed at 1.010, the value of the glomerular filtrate.
**Blood:** This test is based on the peroxidase-like activity of hemoglobin which catalyzes the reaction of cumene-hydroperoxide and 3,3',5,5'-tetramethylbenzidine. The resulting color ranges from orange to green to dark blue. Any green spots or green color development on the reagent area within 60 seconds is significant and the urine specimen should be examined further. Blood is often, but not invariably, found in the urine of menstruating females.
**pH:** This test is based on a double indicator system which gives a broad range of colors covering the entire urinary pH range. Colors range from orange to yellow and green to blue. The expected range for normal urine specimens from newborns is pH 5-7. The expected range for other normal urine specimens is pH 4.5-8, with an average result of pH 6.
**Protein:** This reaction is based on the phenomenon known as the "protein error" of pH indicators where an indicator that is highly buffered will change color in the presence of proteins (anions) as the indicator releases hydrogen ions to the protein. At a constant pH, the development of any green color is due to the presence of protein. Colors range from yellow to yellow-green for negative results and green to green-blue for positive results. 1-14 mg/dL of protein may be excreted by a normal kidney. A color matching any block greater than trace indicates significant proteinuria.
**Urobilinogen:** This test is based on a modified Ehrlich reaction between p-diethylaminobenzaldehyde and urobilinogen acid in strongly acidic medium to produce a pink color. Urobilinogen is one of the major compounds produced in heme this test is 0.2-1.0 mg/dL (3.5-17 μmol/L). A result of 2.0 mg/dL (35 μmol/L) may be of clinical significance, and the patient specimen should be further evaluated.
**Nitrite:** This test depends upon the conversion of nitrate to nitrite by the action of Gram negative bacteria in the urine. In an acidic medium, nitrite in the urine reacts with p-arsanilic acid to form a diazonium compound. The diazonium compound in turn couples with 1 N-(1-naphthyl)-ethylenediamine to produce a pink color. Nitrite is not detectable in normal urine. The nitrite area will be positive in some cases of infection, depending on how long the urine specimens were retained in the bladder prior to collection. Retrieval of positive cases with the nitrite test ranges from as low as 40% in cases where little bladder incubation occurred, to as high as approximately 80% in cases where bladder incubation took place for at least 4 hours.
**Leukocytes:** This test reveals the presence of granulocyte esterases. The esterases cleave a derivatized pyrazole amino acid ester to liberate derivatized hydroxy
5
{6}
pyrazole. This pyrazole then reacts with a diazonium salt to produce a beige-pink to purple color. Normal urine specimens generally yield negative results. Trace results may be of questionable clinical significance. When trace results occur, it is recommended to retest using a fresh specimen from the same patient. Repeated trace and positive results are of clinical significance.
Ascorbic acid: This test involves decolorization of Tillmann's reagent. The presence of ascorbic acid causes the color of the test field to change from blue-green to orange.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
This reproducibility study employed two Bio-Rad Urinalysis Controls: Bio-Rad Urinalysis Control Level 1 and Bio-Rad Urinalysis Control Level 2. The specifications of these two controls are listed in the following tables.
#### Bio-Rad Urinalysis Control, Level 1 values
| Analyte | Control (Level 1) |
| --- | --- |
| Glucose | Neg |
| Bilirubin | Neg |
| Ketone | Neg |
| Specific Gravity | 1.010 – 1.020 |
| Blood | Neg |
| pH | 5.0 – 6.0 |
| Protein | Neg |
| Urobilinogen | 0.2 – 1.0 mg/dL |
| Nitrite | Neg |
| Leukocytes | Neg |
| Ascorbic Acid | Neg |
#### Bio-Rad Urinalysis Control, Level 2 values
| Analyte | Control (Level 2) |
| --- | --- |
| Glucose | 250 - ≥ 1,000 mg/dL |
| Bilirubin | Mod. - Large |
| Ketone | 5 – 40 mg/dL |
| Specific Gravity | 1.015 – 1.025 |
| Blood | Mod. - Large |
| pH | 6.5 – 7.5 |
| Protein | 30 - ≥ 300 mg/dL |
| Urobilinogen | 4.0 - 8.0 mg/dL |
| Nitrite | Positive |
| Leukocytes | Small – Large |
| Ascorbic Acid | Neg |
{7}
Three lots of the ACON Urinalysis Reagent Strips were used. Tests in duplicates were performed with 10 repeats by 5 operators per day for 10 days. A total of 3,000 strips were used for testing (3 x 2 x 10 x 5 x 10 = 3,000 strips). Results were read by comparing color reaction to the color blocks on the canister label according to the instructions on the package insert. The results indicate that the values obtained were all within the expected values for the control materials.
As the urinalysis controls did not have an ascorbic acid level, the sponsor used two ascorbic acid standard positive solutions, $10\mathrm{mg / dL}$ and $20\mathrm{mg / dL}$ . Testing was performed according to the package insert and the results were determined by visually comparing the reaction colors with the color chart. The ACON Urinalysis Reagent Strips from 3 lots were used and tests were performed in replicates of 10 for 2 runs by 10 operators per day for 10 days. The results indicate that the values obtained were all within the expected values for the control materials.
b. Linearity/assay reportable range:
| Reagent | Range |
| --- | --- |
| Glucose (GLU) | 50-100 mg/dL (2.5-5 mmol/L).Results may be read at 10 seconds for qualitative results or at 30 seconds for semi-quantitative results. |
| Bilirubin (BIL) | 0.4-1.0 mg/dL (6.8-17 μmol/L). |
| Ketone (KET) | 2.5-5 mg/dL (0.25-0.5 mmol/L). |
| Specific Gravity (SG) | Range: 1.000-1.030 |
| Blood (BLO) | 0.018-0.06 mg/dL or5-10 Ery/μL in urine specimens with ascorbicacid content of <50 mg/dL. |
| pH | Range: 5.0-9.0 |
| Protein (PRO) | 7.5-15 mg/dL (0.075-0.15 g/L). |
| Urobilinogen (URO) | 0.2-1.0 mg/dL (3.5-17 μmol/L). |
| Nitrite (NIT) | 0.05-0.1 mg/dL in urine with a low specificgravity and less than 30 mg/dL ascorbic acid. |
| Leukocytes (LEU) | 9-15 white blood cells Leu/μL in clinical urine. |
| Ascorbic Acid (ASC) | 5-10 mg/dL (0.28-0.56 mmol/L). |
{8}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The sponsor did not indicate any degree of traceability for their devices. Control materials are not specifically identified in the labeling, but the sponsor indicates in the labeling that for best results, performance of reagent strips should be confirmed by testing known positive and negative specimens/controls whenever a new test is performed, or whenever a new canister is first opened.
The recommended storage temperature range for the ACON Urinalysis Reagent Strips is between 2 – 30°C. Opened canister stability claim of the strips is that the strips are stable for at least 3 months. Unopened canister stability claim of the strips is that the strips are stable for 24 months from the date of manufacture.
d. Detection limit:
The sensitivity of the assay was validated by spiking positive urine samples of known concentrations for each analyte. These positive samples of each analyte were then diluted to the lowest "positive" concentrations indicated on the ACON color chart. For each analyte, aliquots of the lowest positive samples were further diluted to 4:1 or 80%, 3:2 or 60%, 1:1 or 50% and 1:3 or 25% of the originals with negative urine. Urine samples including negative and all spiked and diluted samples were then tested with three lots of the ACON Urinalysis Reagent Strips. For three consecutive days, each sample was tested 21 times by three operators (7 strips per operator per day). A total of 189 strips were used for each concentration tested (3 operators x 3 days x 7 strips x 3 lots = 189 strips). The minimum sensitivity level for each analyte of the ACON Urinalysis Reagent Strips is defined by the sponsor as the lowest level at which over 67% (2/3) of the test results are positive when the diluted positive samples for an analyte of known concentrations were tested. The results of each analyte sensitivity pad are summarized below. Gray blocks identify the low range (sensitivity) of each analyte.
Glucose Pad -
| Results | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Glucose Conc. | N | Neg. | 100 mg/dL* | 250 mg/dL | 500 mg/dL | % Positive |
| 25 mg/dL | 189 | 189 | 0 | 0 | 0 | 0.0% |
| 50 mg/dL | 189 | 63 | 125 | 0 | 0 | 67.2% |
| 60 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% |
| 80 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% |
| 100 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% |
* Lowest Positive Concentration
{9}
Bilirubin Pad -
| Results | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Bilirubin Conc. | N | Neg. | 1 mg/dL* | 2 mg/dL | 4 mg/dL | % Positive |
| 0.25 mg/dL | 189 | 189 | 0 | 0 | 0 | 0.0% |
| 0.4 mg/dL | 189 | 62 | 127 | 0 | 0 | 67.2% |
| 0.6 mg/dL | 189 | 62 | 127 | 0 | 0 | 67.2% |
| 0.8 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% |
| 1.0 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% |
* Lowest Positive Concentration
Ketone Pad -
| Results | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Ketone Conc. | n | Neg. | 5 mg/dL* | 15 mg/dL | 40 mg/dL | 80 mg/dL | % Positive |
| 1.25 mg/dL | 189 | 189 | 0 | 0 | 0 | 0 | 0.0% |
| 2.5 mg/dL | 189 | 63 | 126 | 0 | 0 | 0 | 67.2% |
| 3 mg/dL | 189 | 0 | 189 | 0 | 0 | 0 | 100.0% |
| 4 mg/dL | 189 | 0 | 189 | 0 | 0 | 0 | 100.0% |
| 5 mg/dL | 189 | 0 | 189 | 0 | 0 | 0 | 100.0% |
* Lowest Positive Concentration
Blood Pad -
| Results | | | | | |
| --- | --- | --- | --- | --- | --- |
| Blood Conc. | n | Neg. | 5-10 Ery/μL* | 50 Ery/μL | % Positive |
| 2.5 Ery/μL | 189 | 94 | 95 | 0 | 50.0% |
| 5 Ery/μL | 189 | 0 | 189 | 0 | 100% |
| 6 Ery/μL | 189 | 0 | 189 | 0 | 100% |
| 8 Ery/μL | 189 | 0 | 189 | 0 | 100% |
| 10 Ery/μL | 189 | 0 | 189 | 0 | 100% |
* Lowest Positive Concentration
| Results | | | | | |
| --- | --- | --- | --- | --- | --- |
| Hemoglobin Conc. | n | Neg. | +/- Color Block | + Positive Block | % Positive |
| 0.0075 mg/dL | 189 | 189 | 0 | 0 | 0.0% |
| 0.015 mg/dL | 189 | 127 | 62 | 0 | 33.3% |
| 0.018 mg/dL | 189 | 62 | 127 | 0 | 66.7% |
| 0.024 mg/dL | 189 | 0 | 189 | 0 | 100% |
| 0.03 mg/dL | 189 | 0 | 189 | 0 | 100% |
| 0.06 mg/dL | 189 | 0 | 0 | 189 | 100% |
{10}
Protein Pad -
| Results | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Protein Conc. | n | Neg. | 15 mg/dL* | 30 mg/dL | 100 mg/dL | % Positive |
| 3.75 mg/dL | 189 | 189 | 0 | 0 | 0 | 0.0% |
| 7.5 mg/dL | 189 | 63 | 126 | 0 | 0 | 67.2% |
| 9.0 mg/dL | 189 | 9 | 180 | 0 | 0 | 95.2% |
| 12.0 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% |
| 15.0 mg/dL | 189 | 0 | 189 | 0 | 0 | 100% |
* Lowest Positive Concentration
Urobilinogen Pad -
| Results | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Urobilinogen Conc. | n | 0.2 mg/dL | 1.0 mg/dL | 2 mg/dL | 4 mg/dL | % Agreement |
| 0.2 mg/dL | 189 | 189 | 0 | 0 | 0 | 100.0% (first color block) |
| 0.25 mg/dL | 189 | 189 | 0 | 0 | 0 | 100.0% (first color block) |
| 0.5 mg/dL | 189 | 189 | 0 | 0 | 0 | 100.0% (first color block) |
| 0.6 mg/dL | 189 | 105 | 84 | 0 | 0 | 44.4% (second color block) |
| 0.8 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% (second color block) |
| 1.0 mg/dL | 189 | 0 | 189 | 0 | 0 | 100.0% (second color block) |
Urine urobilinogen concentration above 1 mg/dL is considered "Abnormal". The minimum sensitivity level of urobilinogen was selected at 0.2 mg/dL. The test is not a reliable method for the detection of Porphobilinogen. The absence of Urobilinogen cannot be detected by the test.
Nitrite Pad -
| Results | | | | |
| --- | --- | --- | --- | --- |
| Nitrite Conc. | n | Neg. | Pos. | % Positive |
| 0.025 mg/dL | 189 | 189 | 0 | 0.0% |
| 0.05 mg/dL | 189 | 5 | 184 | 97.4% |
| 0.06 mg/dL | 189 | 0 | 189 | 100% |
| 0.08 mg/dL | 189 | 0 | 189 | 100% |
| 0.1 mg/dL | 189 | 0 | 189 | 100% |
{11}
Leukocytes Pad -
| Results (Cells/μL) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Leukocytes Conc. | n | Neg. | 15* | 70 | 125 | % Positive |
| 3.75 Cells/μL | 189 | 189 | 0 | 0 | 0 | 0.0% |
| 7.5 Cells/μL | 189 | 130 | 59 | 0 | 0 | 31.2% |
| 9 Cells/μL | 189 | 63 | 126 | 0 | 0 | 67.2% |
| 10 Cells/μL | 189 | 0 | 189 | 0 | 0 | 100.0% |
| 12 Cells/μL | 189 | 0 | 189 | 0 | 0 | 100.0% |
| 15 Cells/μL | 189 | 0 | 189 | 0 | 0 | 100.0% |
* Lowest Positive Concentration
Ascorbic Acid Pad -
| Results (mg/dL) | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Ascorbic Acid Conc. | n | Neg. | 10* | 20 | 40 | % Positive |
| 2.5 mg/dL | 189 | 189 | 0 | 0 | 0 | 0.0% |
| 5 mg/dL | 189 | 0 | 189 | 0 | 0 | 100% |
| 6 mg/dL | 189 | 0 | 189 | 0 | 0 | 100% |
| 8 mg/dL | 189 | 0 | 189 | 0 | 0 | 100% |
| 10 mg/dL | 189 | 0 | 189 | 0 | 0 | 100% |
* Lowest Positive Concentration
e. Analytical specificity:
Negative urine was spiked one at a time with the possible interfering substances with concentrations listed in the following table. Each urine sample was tested with 5 replicates of the ACON Urinalysis Reagent Strips. Results were read by comparing color reaction to the color blocks on the canister label according to the Instructions on the package insert.
The following potentially interfering substances at concentrations indicated were added to the urine:
{12}
| Interference Substance | Interference substances Conc. (mg/dL) | Testing Results | |
| --- | --- | --- | --- |
| | | Negative urine without spiked interference substance | Negative urine with spiked interference substance |
| Acetoacetate | 250 | Negative | Positive for Ketone but Negative for others |
| Ammonium Chloride | 200 | Negative | Negative |
| Albumin | 1,000 | Negative | Positive for Protein but Negative for others |
| Ascorbic Acid | 200 | Negative | Positive for Ascorbic acid but Negative for others |
| Bilirubin | 4 | Negative | Positive for Bilirubin but Negative for others |
| Calcium Chloride | 80 | Negative | Negative |
| Citric Acid | 65 | Negative | Negative |
| Creatinine | 600 | Negative | Negative |
| Glucose | 4,100 | Negative | Positive for Glucose but Negative for others |
| Glycine | 450 | Negative | Negative |
| Hemoglobin | 5 | Negative | Positive for Blood but Negative for others |
| KCl | 1,200 | Negative | SG ↑ but Negative for others |
| NaCl | 1,800 | Negative | SG ↑ but Negative for others |
| Oxalic Acid | 70 | Negative | Negative |
| Phenolphthalein | 1,200 | Negative | Negative |
| Sodium Nitrate | 0.3 | Negative | Negative |
| Sodium Nitrite | 10 | Negative | Positive for Nitrite but Negative for others |
| Sodium Phosphate | 500 | Negative | pH ↑ but Negative for others |
| Theophylline | 100 | Negative | Negative |
| Urea | 4,000 | Negative | Negative |
| Fructose | 1.2 | Negative | Negative |
| Galactose | 0.5 | Negative | Negative |
| Lactose | 1 | Negative | Negative |
The sponsor also performed interference studies that were analytes that the urinalysis strips detect in order to determine if an analyte is detected by the strip is abnormally high will it affect the results of other test pads on the strip. Urine samples were spiked with different analytes to the known concentrations which were then confirmed as positive urine samples by Bayer Multistix 10 SG Reagent Strips and Clinitek Status Analyzer. These positive urine samples were then spiked with possible interfering substances, one at a time to the levels listed in the following tables. The urine samples were tested with the ACON Urinalysis Reagent Strips in 5 replicates. Results were read by comparing color reaction
{13}
to the color blocks on the canister label according to the Instructions on the package insert.
## Acceptance Criteria
The reagent pad must produce expected color-reaction when testing with the corresponding spiked samples.
The Effect of Ascorbic Acid on ACON Glucose Reagent Strip
| Ascorbic Acid Concentration (mg/dL) | Glucose Concentration (mg/dL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | |
| | Negative | 100 | 250 | 500 | 1,000 |
| 0 | N | 100* | 250 | 500 | 1,000 |
| 10 | N | 100 | 250 | 500 | 1,000 |
| 25 | N | N(1)** - 100(4)** | 250 | 500 | 1,000 |
| 30 | N | N | 100 | 500 | 1,000 |
* 5 Replicates
** 1 tested at negative level and 4 tested at 100 mg/dL level
Ascorbic acid concentrations of 25 mg/dL or greater may reduce the sensitivity of the test.
The Effect of Acetoacetic Acid on ACON Glucose Reagent Strip
| Acetoacetic Acid Concentration (mg/dL) | Glucose Concentration (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | | |
| | Negative | 100 | 250 | 500 | 1,000 | 2,000 |
| 0 | N | 100 | 250 | 500 | 1,000 | 2,000 |
| 80 | N | 100 | 250 | 500 | 1,000 | 2,000 |
| 100 | N | N(1) - 100(4) | 100 | 250 | 500 | 1,000 |
| 120 | N | N | 100 | 250 | 500 | 1,000 |
| 140 | N | N | 100 | 250 | 500 | 1,000 |
Acetoacetic acid concentration of 100 mg/dL or higher decreases the sensitivity of the test.
13
{14}
The Effect of Specific Gravity on ACON Glucose Reagent Strip
| Specific Gravity | Glucose Concentration (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | | |
| | Negative | 100 | 250 | 500 | 1,000 | 2,000 |
| 1.005 | N | 100 | 250 | 500 | 1,000 | 2,000 |
| 1.010 | N | 100 | 250 | 500 | 1,000 | 2,000 |
| 1.015 | N | 100 | 250 | 500 | 1,000 | 2,000 |
| 1.020 | N | 100 | 250 | 500 | 1,000 | 2,000 |
| 1.025 | N | N(1) - 100(4) | 100 | 500 | 1,000 | 2,000 |
| 1.030 | N | N(1) - 100(4) | 100 | 500 | 1,000 | 2,000 |
As the Specific Gravity increases to 1.025, the sensitivity of the Glucose reagent pad decreases.
The Effect of Specific Gravity on ACON Blood Reagent Strip
| Specific Gravity | Blood Concentration (Cells/uL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | |
| | Negative | 10 (Trace) | 25 (Small) | 80 (Mod) | 200 (Large) |
| 1.005 | N | 10 | 25 | 80 | 200 |
| 1.010 | N | 10 | 25 | 80 | 200 |
| 1.015 | N | 10 | 25 | 80 | 200 |
| 1.020 | N | 10 | 25 | 80 | 200 |
| 1.025 | N | 10 | 25 | 80 | 200 |
| 1.030 | N | 10 | 25 | 80 | 200 |
Specific Gravity does not influence the sensitivity of the blood reagent pad.
{15}
The Effect of Ascorbic Acid on ACON Blood Reagent Strip
| Ascorbic Acid Concentration (mg/dL) | Blood Concentration (Cells/uL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test results on 5 Replicates | | | | |
| | Negative | 10 (Trace) | 25 (Small) | 80 (Mod) | 200 (Large) |
| 0 | N | 10 | 25 | 80 | 200 |
| 20 | N | 10 | 10 | 80 | 200 |
| 30 | N | N | 10 | 80 | 200 |
| 35 | N | N | 10 | 80 | 200 |
| 50 | N | N | N | 80 | 200 |
In high levels of Ascorbic Acid, the sensitivity of the blood reagent pad decreases.
The Effect of pH on ACON Blood Reagent Strip
| pH | Blood Concentration (Cells/uL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test results on 5 Replicates | | | | |
| | Negative | 10 (Trace) | 25 (Small) | 80 (Mod) | 200 (Large) |
| 5.0 | N | +/- | + | 2+ | 3+ |
| 6.0 | N | +/- | + | 2+ | 3+ |
| 7.0 | N | +/- | + | 2+ | 3+ |
| 7.5 | N | +/- | + | 2+ | 3+ |
| 8.0 | N | +/- | +/- | 2+ | 3+ |
| 9.0 | N | N | +/- | 2+ | 3+ |
As the pH increases to 8.0, the sensitivity of the blood reagent pad decreases.
The Effect of Ascorbic Acid on ACON Bilirubin Reagent Strip
| Ascorbic Acid Concentration (mg/dL) | Bilirubin Concentration | | | |
| --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | |
| | Negative | Small | Moderate | Large |
| 0 | N | S | M | L |
| 25 | N | S | M | L |
| 30 | N | N | S | L |
| 40 | N | N | S | L |
{16}
As the Ascorbic Acid increases to 30 mg/dL, the sensitivity of the bilirubin reagent pad decreases.
The Effect of Acetone on ACON Ketone Reagent Strip
| Acetone Concentration (mg/dL) | Ketone |
| --- | --- |
| | Test Results on 5 Replicates |
| 100 | No Color Changed on Strips |
| 500 | No Color Changed on Strips |
| 10,000 | No Color Changed on Strips |
Acetone does not influence the Ketone reagent pad.
The Effect of Specific Gravity on ACON Nitrite Reagent Strip
| Standard Material
Specific Gravity | Nitrite Concentration (mg/dL) | | | |
| --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | |
| | Negative | 0.05 | 0.1 | 1.0 |
| 1.005 | N | P | P | P |
| 1.010 | N | P | P | P |
| 1.015 | N | P | P | P |
| 1.020 | N | P | P | P |
| 1.025 | N | P | P | P |
| 1.030 | N | N | P | P |
As the Specific Gravity increases to 1.030 or over, the sensitivity (accuracy) of the Nitrite reagent pad decreases.
The Effect of Ascorbic Acid on ACON Nitrite Reagent Strip
| Ascorbic Acid
(mg/dL) | Nitrite Concentration (mg/dL) | | | |
| --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | |
| | Negative | 0.05 | 0.1 | 0.5 |
| 0 | N | P | P | P |
| 25 | N | P | P | P |
| 30 | N | N | P | P |
| 35 | N | N | P | P |
In high levels of Ascorbic acid, the sensitivity of the Nitrite reagent pad decreases.
16
{17}
The Effect of pH on ACON Nitrite Reagent Strip
| pH | Nitrite Concentration (mg/dL) | | | |
| --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | |
| | Negative | 0.05 | 0.1 | 1.0 |
| 5.0 | N | P | P | P |
| 6.0 | N | P | P | P |
| 7.0 | N | P | P | P |
| 7.5 | N | P | P | P |
| 8.0 | N | P | P | P |
| 9.0 | N | P | P | P |
The pH value does not influence the sensitivity of the Nitrite reagent pad.
The Effect of Specific Gravity on ACON Protein Reagent Strip
| Specific Gravity | Protein Concentration (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | | |
| | Negative | 15 | 30 | 100 | 300 | 2,000 |
| 1.005 | N | 15 | 30 | 100 | 300 | 2,000 |
| 1.010 | N | 15 | 30 | 100 | 300 | 2,000 |
| 1.015 | N | 15 | 30 | 100 | 300 | 2,000 |
| 1.020 | N | 15 | 30 | 100 | 300 | 2,000 |
| 1.025 | N | N | 30 | 100 | 300 | 2,000 |
| 1.030 | N | N | 30 | 100 | 300 | 2,000 |
As the Specific Gravity increases to 1.025, the sensitivity of the Protein reagent pad decreases.
{18}
The Effect of pH on ACON Protein Reagent Strip
| pH | Protein Concentration (mg/dL) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | | |
| | Negative | 15 | 30 | 100 | 300 | 2,000 |
| 5.0 | N | 15 | 30 | 100 | 300 | 2,000 |
| 6.0 | N | 15 | 30 | 100 | 300 | 2,000 |
| 7.0 | N | 15 | 30 | 100 | 300 | 2,000 |
| 7.5 | N | 15 | 30 | 100 | 300 | 2,000 |
| 8.0 | N | 15 | 30 | 100 | 300 | 2,000 |
| 9.0 | N | 30 | 100 | 300 | 300 | 2,000 |
As the pH increases to 9.0, the accuracy of the Protein reagent pad decreases.
The Effect of pH on ACON Leukocyte Reagent Strip
| pH | Leukocyte Concentration (Leu/uL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | |
| | Negative | 15 | 70 | 125 | 500 |
| 5.0 | N | 15 | 70 | 125 | 500 |
| 6.0 | N | 15 | 70 | 125 | 500 |
| 7.0 | N | 15 | 70 | 125 | 500 |
| 7.5 | N | 15 | 70 | 125 | 500 |
| 8.0 | N | 15 | 125 | 500 | 500 |
| 9.0 | N | 70 | 125 | 500 | 500 |
As the pH increases to 9.0, the sensitivity of the Leukocyte reagent pad increases.
The Effect of Glucose on ACON Leukocyte Reagent Strip
| Glucose Concentration (mg/dL) | Leukocyte Concentration (Leu/uL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | |
| | Negative | 15 | 70 | 125 | 500 |
| 0 | N | 15 | 70 | 125 | 500 |
| 1,000 | N | 15 | 70 | 125 | 500 |
| 2,000 | N | N | 70 | 125 | 500 |
| 4,000 | N | N | 70 | 125 | 500 |
{19}
As the concentration of Glucose increases to 2,000 mg/dL, the sensitivity of the Leukocyte reagent pad decreases.
The Effect of Specific Gravity on ACON Leukocyte Reagent Strip
| Specific Gravity | Leukocyte Concentration (Leu/uL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | |
| | Negative | 15 | 70 | 125 | 500 |
| 1.005 | N | 15 | 70 | 125 | 500 |
| 1.010 | N | 15 | 70 | 125 | 500 |
| 1.015 | N | 15 | 70 | 125 | 500 |
| 1.020 | N | 15 | 70 | 125 | 500 |
| 1.025 | N | N | 70 | 125 | 500 |
| 1.030 | N | N | 15 | 125 | 500 |
As the specific gravity increases to 1.025 and over, the accuracy of the Leukocyte reagent pad decreases.
The effect of Protein (Human Albumin) on ACON Leukocyte Reagent Strip
| Human Albumin Concentration (mg/dL) | Leukocyte Concentration (Leu/uL) | | | | |
| --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | |
| | Negative | 15 | 75 | 125 | 500 |
| 0 | Neg | 15 | 75 | 125 | 500 |
| 500 | Neg | 15 | 75 | 125 | 500 |
| 1,000 | Neg | Neg | 75 | 125 | 500 |
| 1,500 | Neg | Neg | 75 | 125 | 500 |
As protein concentration increases to 1,000 mg/dL, the sensitivity of the Leukocyte reagent pad decreases.
19
{20}
The Effect of Nitrite on ACON Urobilinogen Reagent Strip
| Nitrite Concentration (mg/dL) | Urobilinogen Concentration (mg/dL ) | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | | |
| | 0.2 | 1.0 | 2.0 | 4.0 | 8.0 | 12 |
| 0 | 0.2 | 1.0 | 2.0 | 4.0 | 8.0 | 12 |
| 1 | 0.2 | 1.0 | 2.0 | 4.0 | 8.0 | 12 |
| 10 | 0.2 | 1.0 | 2.0 | 4.0 | 8.0 | 12 |
| 20 | 0.2 | 1.0 | 2.0 | 4.0 | 8.0 | 12 |
The concentrations of Nitrite tested do not influence the accuracy of the Leukocyte reagent pad.
The Effect of pH on ACON Specific Gravity Reagent Strip
| pH | Specific Gravity | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | | | |
| | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 4.5 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 | 1.030 |
| 5 | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 6 | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 6.5 | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 7 | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 7.5 | 1.000 | 1.005 | 1.010 | 1.010 | 1.015 | 1.020 | 1.025 |
| 8 | 1.000 | 1.005 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 |
| 8.5 | 1.000 | 1.005 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 |
| 9 | 1.000 | 1.000 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 |
As the pH is higher than 7.5, the sensitivity of the Specific Gravity reagent pad decreases 0.005. As the pH is less than 5, the sensitivity of the Specific Gravity reagent pad increases 0.005.
{21}
The Effect of Protein on ACON Specific Gravity Reagent Strip
| Protein Concentration (mg/dL) | Specific Gravity | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| | Test Results on 5 Replicates | | | | | |
| | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 0 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 80 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 100 | 1.005 | 1.010 | 1.015 | 1.020 | 1.025 | 1.030 |
| 300 | 1.010 | 1.015 | 1.015 | 1.020 | 1.025 | 1.030 |
| 500 | 1.015 | 1.015 | 1.015 | 1.020 | 1.025 | 1.030 |
| 1,000 | 1.015 | 1.015 | 1.015 | 1.020 | 1.025 | 1.030 |
| 1,500 | 1.015 | 1.015 | 1.015 | 1.020 | 1.025 | 1.030 |
As the concentration of protein increases to 300 mg/dL, the accuracy of the Specific Gravity reagent pad decreases.
The Effect of Glucose on ACON Specific Gravity Reagent Strip
| Glucose Concentration (mg/dL) | Specific Gravity |
| --- | --- |
| | Test Results on 5 Replicates |
| 0 | 1.015 |
| 1,000 | 1.015 |
| 2,000 | 1.015 |
| 4,000 | 1.015 |
The test results on the Specific Gravity reagent pad are not affected by Glucose levels tested.
The sponsor also examined the effect of excess urine sample "runover" phenomenon between pH and Protein pads when the excess urine was not removed immediately. The pH standard solutions, 5.0, 6.0, 7.0, 8.0 and 9.0, prepared with fresh negative urine were used in the study.
Testing protocol #1: Dip the ACON Urinalysis Reagent Strips into standard urine solutions and take out immediately, remove excess urine from the strips right away according to the package insert. Read results in the scheduled time. Tests were repeated 10 times for each solution.
Testing protocol #2: Dip the strips into standard urine solutions and take out immediately, did not remove excess urine from the strips. Read results in the
{22}
scheduled time. Tests were repeated 10 times for each solution.
Testing was performed and results were determined visually by comparing the reaction colors with the color chart. Results were summarized in the following table.
| pH | Expected Results | Test Results with Excess Urine Sample Removed | | | Test Results without Excess Urine Sample Removed | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Testing Results | Percent Agreement | Percent Disagreement | Testing Results | Percent Agreement | Percent Disagreement |
| 5.0 | 5.0 | 10 @ 5.0 | 100% | 0% | 10 @ 5.0 | 100% | 0% |
| 6.0 | 6.0 | 10 @ 6.0 | 100% | 0% | 10 @ 6.0 | 100% | 0% |
| 7.0 | 7.0 | 10 @ 7.0 | 100% | 0% | 7 @ 7.0
3 @ 6.5 | 70% | 30% |
| 8.0 | 8.0 | 10 @ 8.0 | 100% | 0% | 6 @ 8.0
3 @ 7.0
1 @ 5.0 | 60% | 40% |
| 9.0 | 9.0 | 10 @9.0 | 100% | 0% | 7 @ 9.0
2 @ 8.0
1 @ 5.0 | 70% | 30% |
From the testing results, if the excess urine was not removed immediately, the edge of pH reagent pad became red as the result of the acidic component runover from the Protein pad onto the pH pad.
The sponsor also indicates that urinalysis reagent strips may be affected by substances that cause abnormal urine color such as drugs containing azo dyes (e.g. Pyridium®, Azo Gantrisin®, Azo Gantanol®), nitrofurantoin (Microdantin®, Furadantin®), and riboflavin. The color development on the test pad may be masked or a color reaction may be produced that could be interpreted as false results. In addition, the sensitivity of the Nitrite pad is reduced for urine specimens with highly buffered alkaline (e.g. Sodium Bicarbonate).
f. Assay cut-off: See Detection Limit above.
{23}
23
2. Comparison studies:
a. Method comparison with predicate device:
The sponsor performed a consumer study consisting of 125 professionals (the intended users of the device) to determine how results recoded by users correlated with results read on the predicate device. The study was performed at 3 different clinical sites, and the 125 samples were tested among the 3 sites, 72 samples at the first, 42 samples at the second, and 11 samples from the third. Results from 125 urine specimens were tested and recorded with the results outlined below.
| Comparison of Total Agreements Within +/- One Color Block | Leukocyte | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 121 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 122 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 122 | 125 |
| Comparison of Total Agreements Within +/- One Color Block | Nitrite | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 124 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 125 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 125 | 125 |
| Comparison of Total Agreements Within +/- One Color Block | Urobilinogen | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 124 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 125 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 123 | 125 |
{24}
| Comparison of Total Agreements Within +/- One Color Block | Protein | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 118 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 110 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 118 | 125 |
| Comparison of Total Agreements Within +/- One Color Block | pH | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 103 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 89 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 109 | 125 |
| Comparison of Total Agreements Within +/- One Color Block | Blood | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 121 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 120 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 118 | 125 |
| Comparison of Total Agreements Within +/- One Color Block | Specific Gravity | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 117 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 98 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 104 | 125 |
| Comparison of Total Agreements Within +/- One Color Block | Ketone | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 125 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 123 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 124 | 125 |
{25}
| Comparison of Total Agreements Within +/- One Color Block | Bilirubin | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 124 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 123 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 125 | 125 |
| Comparison of Total Agreements Within +/- One Color Block | Glucose | |
| --- | --- | --- |
| | Observed # of Correct Results Within +/- 1 Color Block | Expected # of Correct Results Within +/- 1 Color Block |
| ACON (Visual) vs. Bayer (Visual) | 124 | 125 |
| ACON (Visual) vs. Bayer (Instrument) | 124 | 125 |
| Bayer (Visual) vs. Bayer (Instrument) | 124 | 125 |
The sponsor also compared the pH of the ACON Urinalysis Reagent Strips to the Bayer Multistix 10 SG Urine Strips using two different methods:
Method #1: Seven pH standard solutions (pH from 5.02 to 9.0) were tested with the Bayer Multistix 10 SG Urine Strips and the ACON Urinalysis Reagent Strips, respectively, according to their package inserts. Results are summarized in the table below.
Method #1:
| pH Meter | 5.02 | 6.01 | 6.49 | 7.0 | 7.51 | 8.0 | 9.0 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| ACON | 5.0 | 6.0 | 6.5 | 7.0 | 7.5 | 8.0 | 9.0 |
| Bayer | 5.0 | 6.0 | 7.0 | 7.0 | 8.0 | 8.0 | 8.5* |
* Bayer color chart doesn't have the color block for pH 9.0.
Method #2: 47 clinical urine samples were tested with the Bayer Multistix 10 SG Urine Strips, the ACON Urinalysis Reagent Strips according to their package inserts together with a pH meter. Results are summarized in the table below.
{26}
# Method #2:
The differences between urine strips and pH meter results are presented below:
| Result Differences Between pH Meter and Urine Strips | < 0.5 | > 0.5 | Mean ± SD (Differences Between pH Meter and Strip Results) |
| --- | --- | --- | --- |
| ACON Urinalysis Reagent Strips | 47/47 | 0/47 | 0.180 ± 0.162 |
| Bayer Multistix 10 SG Reagent Strips for Urinalysis | 35/47 | 12/47 | 0.334 ± 0.242 |
b. Matrix comparison: Not applicable.
3. Clinical studies:
a. Clinical Sensitivity: Not applicable.
b. Clinical specificity: Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable): See Method Comparison section above.
4. Clinical cut-off: Not applicable.
5. Expected values/Reference range:
The package insert goes into detail on interpretation of expected values with citations of literature. The statements made are summarized below.
Urobilinogen: $0.2 - 1.0\mathrm{mg / dL}$ (3.5-17 $\mu \mathrm{mol} / \mathrm{L}$ )
Glucose: $<$ Trace
Bilirubin: $<$ Trace
Ketones: $0\mathrm{mg / dL}$
pH: The expected range for other normal urine specimens is pH 4.5-8, with an average result of pH 6. For newborns pH 5-7 is expected.
Blood: $<$ Trace
Specific Gravity (SG): 1.016-1.022.
Protein: $1 - 14\mathrm{mg / dL}$
Nitrite: $0\mathrm{mg / dL}$
Leukocyte: Negative
Ascorbic acid: urinary output of $2 - 10\mathrm{mg / dL}$ daily
{27}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
27
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.