The Medic4All Telemedicine System Model VMS-01 is intended to collect and transmit physiological information, such as glucose levels, via standard telephone lines or the internet to local or remote computer systems for use by the patient or a healthcare professional.
Device Story
VMS-01 is a wireless telephonic/internet-based physiologic monitoring system; collects physiological data (e.g., glucose levels); transmits data via standard telephone lines or internet to local or remote computer systems. Used for interactive, remote measurement, processing, storage, and display of physiological data. Intended for care of patients requiring periodic monitoring. Healthcare professionals or patients use the system to review transmitted data; facilitates remote clinical management and monitoring of patient health status.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation activities included EMC/electrical safety, unit testing, integration testing, data integrity testing, usability testing, environmental testing, compatibility testing, and load/performance testing. The system functioned as intended in all instances.
Technological Characteristics
Wireless telephonic/internet-based physiologic monitoring system. Supports remote measurement, transmission, processing, storage, and display of physiological data. Connectivity via standard telephone lines or internet. System architecture includes data collection and transmission components for remote monitoring.
Indications for Use
Indicated for patients requiring periodic monitoring of physiological information, such as glucose levels, for transmission to healthcare professionals or for patient use.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Predicate Devices
Philips Medical Systems M3810A TeleMonitoring System with M3812B TeleStation (K023749)
BL-Healthcare, Inc. Remote Care Management System (K051470)
Submission Summary (Full Text)
{0}------------------------------------------------
K061502
page 1 of 2
SEP 15 2006
#### 5 510(K) SUMMARY
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. § 807.92.
# Submitter:
# Medic4All (Israel) LTD.
10 Hamefalsim Street, POB 4222 Petach Tikva, Israel 49000 Phone number: +972-3-9226610 Fax number: +972-3-9226615
## Contact Person:
Jonathan S. Kahan Hogan & Hartson L.L.P. Columbia Square 555 Thirteenth Street, NW Washington DC 20004-1109
Phone: (202) 637-5794
Date Prepared: May 30, 2006
### Name of the Device:
Medic4All Telemedicine System VMS-01
### Predicate Devices:
The VMS-01 is substantially equivalent to:
- . The Philips Medical Systems M3810A TeleMonitoring System with M3812B TeleStation (K023749).
- The BL-Healthcare, Inc. Remote Care Management System (K051470). .
### Intended Use / Indications for Use
The Medic4All Telemedicine System Model VMS-01 is intended to collect and transmit physiological information, such as glucose levels, via standard telephone lines or the internet to local or remote computer systems for use by the patient or a healthcare professional.
{1}------------------------------------------------
K041502
page 2 of 2
### Technological Characteristics
Medic4All's VMS-01 telemedicine system (VMS-01) is a wireless telephonic/internet-based physiologic monitoring system, designed for care of patients requiring periodic monitoring. It is designed for interactive, remote, physiological data measurement, transmission, processing, storage and display of data.
### Performance Data
Verification and validation testing activities were conducted to establish the performance and reliability characteristics of the telemedicine system. Testing involved EMC/Electrical Safety; unit testing; integration testing; data integrity testing; usability testing; environmental testing; compatibility testing; performance testing, system testing, load/performance testing for the whole telemedicine system. In all instances, the VMS-01 functioned as intended.
### Substantial Equivalence
The VMS-01 system is substantially equivalent to the Philips Medical Systems M3810A TeleMonitoring System with M3812B TeleStation (K023749) and the BL-Healthcare, Inc. Remote Care Management System (K051470). The VMS-01 has the same intended uses and substantially similar indications, technological characteristics, and principles of operation as these predicate devices. The minor differences in indications for use and technology between the VMS-01 and the predicate devices raise no new issues of safety or effectiveness, as demonstrated by performance test data.
{2}------------------------------------------------
Public Health Service
Image /page/2/Picture/11 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 1 5 2006
Medic4All (Israel) LTD. c/o Mr. Jonathan S. Kahan Hogan & Hartson L.L.P. Columbia Square 555 Thirteenth Street, NW Washington DC 20004-1109
Re: K061502
Trade Name: Telemedicine System Model VMS-01 Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: Class II (two) Product Code: DRG Dated: August 30, 2006 Received: August 30, 2006
Dear Mr. Kahan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
## Page 2 – Mr. Jonathan S. Kahan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic . product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmmer for
Bram B. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### 4 INDICATIONS FOR USE STATEMENT
14041502 510(k) Number (if known):
Device Name: Telemedicine System Model VMS-01
Indications for Use:
The Medic4All Telemedicine System Model VMS-01 is intended to collect and transmit physiological information, such as glucose levels, via standard telephone lines or the internet to local or remote computer systems for use by the patient or a healthcare professional.
Prescription Use
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of
Ondell
(Division Sian-Off) Division of Cardiovascular Devices 510(k) Number KO6 502