K061466 · Dlc Laboratories, Inc. · NUC · Dec 4, 2006 · Obstetrics/Gynecology
Device Facts
Record ID
K061466
Device Name
INTIMOL
Applicant
Dlc Laboratories, Inc.
Product Code
NUC · Obstetrics/Gynecology
Decision Date
Dec 4, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
INTIMOLTM Liquid Personal lubricant is principally intended as personal lubricant to supplement the body's natural lubricating fluids, and to enhance the ease and comfort of intimate sexual activity with or without a latex condom.
Device Story
INTIMOL™ is a non-sterile, aqueous-based personal lubricant; formulated as a clear, non-irritating, non-greasy, non-staining gel-like liquid. Intended for over-the-counter use to supplement natural body fluids and enhance comfort during intimate sexual activity. Compatible with latex condoms. Packaged in a plastic bottle with a flip-top cap. Device does not contain contraceptives or spermicides.
Clinical Evidence
Bench testing only. Independent laboratory testing confirmed condom compatibility, biocompatibility, and preservative effectiveness in accordance with established industry standards.
Technological Characteristics
Aqueous-based gel; ingredients include purified water, glycerin, cellulose thickeners, methylparaben, propylparaben, propylene glycol, and aloe vera. Ingredients meet USP/NF standards. Non-sterile. Packaged in plastic bottles. Water-soluble, non-staining, and condom-compatible.
Indications for Use
Indicated for individuals requiring supplemental lubrication during intimate sexual activity. Compatible with latex condoms. Not a contraceptive or spermicide.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
1001466
### 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS (21 CFR 807.92) [21 CFR 807.87(H)]
OEC - 4 2006
### INTIMOL™ Liquid Personal Lubricant
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.87(h), this information serves as a Summary of Safety and Effectiveness for the INTIMOL™ Liquid Personal Lubricant
- Submitted By: DLC Laboratories, Inc. 7008 Marcelle Street Paramount, CA 90723
- Date: May 23, 2006
- Contact Person: Juan Manzur Operations Manager 562-602-2184 or 800-858-3889 Telephone: Fax: 562-602-1931
- Proprietarv Name: INTIMOL TM
- Personal Lubricant Common Name:
- Classification Name: The General Hospital and Personal Use Device section of the General Medical Devices Panel within the FDA's Center for Medical Device & Radiological Health considers patient lubricants (21 CFR 8880.6375. Class I devices) to be Class II devices when promoted as being compatible for use with condoms (21 CFR 8884.5300)
- Predicate Device: Wet® Light Personal Moisturizer (K013086) Astroglide® Personal Lubricant (K935299) K-Y® Liquid Personal Lubricant (K955648)
- Intended Use: INTIMOLTM Liquid Personal lubricant is principally intended as personal lubricant to supplement the body's natural lubricating fluids, and to enhance the ease and comfort of intimate sexual activity with or without a latex condom.
#### Device Description
INTIMOLTM is a non-sterile, aqueous-based personal lubricant designed to supplement the body's own natural lubrication fluids. It is specifically formulated to be a clear, non-irritating, non-greasy, non-staining, gel-like liquid and is compatible
DLC Laboratories, Inc. 7008 Marcelle Street Paramount, CA 90723
Confidential Information
{1}------------------------------------------------
for use with or without a latex condom during intimate sexual activity as evidenced by condom compatibility test results. This device is not a contraceptive or spermicide, nor does it contain any such component.
The product is packaged in a plastic bottle with a flip-top cap.
#### Summary of Technological Characteristics
INTIMOL™ Liquid Personal lubricant contains ingredients that are substantially the same as those used in the manufacture of the predicate devices. The ingredients meet specifications defined in the United State Pharmacopoeia (USP) or National Formulary (NF). and are "generally recognized as safe for their intended use" (21 CFR 172.515).
#### Summary of Substantial Equivalence Information:
The intended use, ingredients, and application of the proposed device are substantially equivalent to those of the predicate devices. In determining substantial equivalence, the INTIMOL™ has been compared with the following legally marketed device to which the Sponsor claims substantial equivalence.
The table below compares the technological characteristics of INTIMOL™ Liquid Personal Lubricant to the predicate devices, Wet® Light Personal Moisturizer, K-Y® Liquid Personal Lubricant and Astroglide®.
| Feature | Intimol™<br>Personal<br>Lubricant | Wet Light<br>Personal<br>Moisturizer | K-Y®<br>Personal<br>Lubricant | Astroglide® |
|--------------------------------|-----------------------------------|--------------------------------------|-------------------------------|--------------|
| Manufacturer | DLC<br>Laboratories,<br>Inc. | Trigg<br>Laboratories,<br>Inc. | McNeil-PPC,<br>Inc. | BioFilm Inc. |
| Contains purified water | Yes | Yes | Yes | Yes |
| Contains glycerin | Yes | Yes | Yes | Yes |
| Contains Cellulose thickeners | Yes | Yes | Yes | No |
| Contains Methylparaben | Yes | Yes | Yes | Yes |
| Contains Propylparaben | Yes | No | No | Yes |
| Contains Propylene glycol | Yes | Yes | Yes | Yes |
| Contains Aloe Vera | Yes | Yes | No | No |
| Over-the-Counter Use | Yes | Yes | Yes | Yes |
| Labeled water soluble | Yes | Yes | Yes | Yes |
| Labeled non-staining | Yes | Yes | Yes | Yes |
| Labeled condom compatible | Yes | Yes | Yes | Yes |
| Contains alcohol and fragrance | No | No | No | No |
| Container material | Plastic | Plastic | Plastic | Plastic |
| Sterile | No | No | No | No |
The product was tested by independent laboratories for condom compatibility, biocompatibility and preservative effectiveness. Final results from these tests demonstrate that the device meets established acceptance criteria in accordance with the identified industry standards.
{2}------------------------------------------------
## Conclusion
INTIMOL™ Liquid Personal Lubricant is substantially equivalent to its predicate devices, Wet® Light Personal Moisturizer, K-Y® Liquid Personal Lubricant and Astroglide®. All of these products have the same intended use and similar technological characteristics. Therefore, no new safety and effectiveness issues are expected to be raised.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
Mr. Juan Manzur Operations Manager DLC Laboratories, Inc. 7008 Marcelle Street PARAMOUNT CA 90723
DEC - 4 2006
Re: K061466
Trade/Device Name: INTIMOL™ Liquid Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: November 16, 2006 Received: November 20, 2006
Dear Mr. Manzur:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a circular logo with the letters "FDA" in large, bold font. Above the letters, the years "1906-2006" are printed. Below the letters, the word "Centennial" is printed in a smaller font. Three stars are printed below the word "Centennial". The logo is surrounded by a circular border.
Protecting and Promoting Public Health.
{4}------------------------------------------------
Page 2 -
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150
or at its Internet address http://www.ida.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy Crogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications For Use
Page 1_ of 1_
510(k) Number (if known): K061466
Device Name: INTIMOL™ Liquid Personal Lubricant
Indications for Use:
INTIMOL™ Liquid Personal lubricant is principally intended as personal lubricant to supplement the body's natural lubricating fluids, and to enhance the ease and comfort of intimate sexual activity with or without a latex condom.
Prescription Use _ (Per 21 CFR 801.109) AND/OR
Over-The Counter Use __ X (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-Off) Division of Reproductive, Abde and Radiological Devices 510(k) Number _
DLC Laboratories, Inc.
7008 Marcelle Street
Paramount, CA 90723
Confidential Information
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.