SELECTSITE C304 DEFLECTABLE CATHETER SYSTEM, MODELS C304-S59, C304-L69, C304-XS59 & C304-XL74
K061416 · Medtronic Vascular · DQY · Oct 25, 2006 · Cardiovascular
Device Facts
| Record ID | K061416 |
| Device Name | SELECTSITE C304 DEFLECTABLE CATHETER SYSTEM, MODELS C304-S59, C304-L69, C304-XS59 & C304-XL74 |
| Applicant | Medtronic Vascular |
| Product Code | DQY · Cardiovascular |
| Decision Date | Oct 25, 2006 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1250 |
| Device Class | Class 2 |
Intended Use
The deflectable catheter system is indicated to provide a pathway though which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
Device Story
SelectSite™ C304 Deflectable Catheter System is a percutaneous catheter kit including a deflectable catheter, dilator, slitter, guidewire, needle, and syringe. Used by physicians in clinical settings to access heart chambers and coronary vasculature. The needle and syringe establish venous access; the guidewire facilitates entry; the deflectable catheter provides a pathway for transvenous devices (e.g., balloon catheters, leads); the dilator assists passage; the slitter allows catheter removal. The system enables precise navigation within the heart, facilitating lead placement or diagnostic procedures. Benefits include improved access to the coronary sinus and left heart vessels.
Clinical Evidence
Bench testing only. Evaluation included mechanical, functional, and biocompatibility testing consistent with ISO 10993-1. The device met all specified design and performance requirements.
Technological Characteristics
Percutaneous catheter system; includes deflectable catheter, dilator, slitter, guidewire, needle, syringe. Materials are biocompatible per ISO 10993-1 (external communicating, limited exposure <24 hours). Sterilized via Ethylene Oxide (EtO).
Indications for Use
Indicated for patients requiring transvenous access to heart chambers, coronary vasculature, or the coronary sinus for the introduction of diagnostic/therapeutic devices, balloon catheters, or left heart leads.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
- SelectSite™ C304-S5901 and C304-L6901 Deflectable Catheter Systems (K033989)
- Medtronic Attain® 6226DEF Deflectable Catheter Delivery System (K032312)
Related Devices
- K013517 — 10600 DEFLECTABLE CATHETER SYSTEM, MODEL 10600 · Medtronic Vascular · Apr 25, 2002
- K033989 — MEDTRONIC SELECTSITE DEFLECTABLE CATHETER SYSTEM, MODELS C304-S59 AND C304-L69 · Medtronic Inc. Cardiac Rhythm Management · Jan 22, 2004
- K061480 — ATTAIN 6227DEF DEFLECTABLE CATHETER DELIVERY SYSTEM · Medtronic, Ireland · Aug 21, 2006
- K223178 — SelectSite C304 Deflectable Catheter System, C315 Delivery System · Medtronic · Nov 8, 2022
- K183493 — SelectSite C304-HIS Deflectable Catheter System · Medtronic, Inc. · Mar 15, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the Medtronic logo. On the left side of the logo, there is a circular graphic with three figures in different positions, suggesting movement or rotation. To the right of the graphic, the word "Medtronic" is written in a bold, sans-serif font. The overall design is simple and professional, conveying a sense of innovation and reliability.
| 510(k) SUMMARY OF SUBSTANTIAL EQUIVALENCE | k061416 | |
|-------------------------------------------|------------------------------------------------------------------------------------------|-------------|
| Date Prepared: | May 19, 2006 | Page 1 of 2 |
| Applicant: | Medtronic Ireland<br>Parkmore Business Park West<br>Galway<br>Ireland | OCT 2 5 |
| Submission Correspondent: | Michelle Nivala<br>Regulatory Affairs Specialist<br>Medtronic, Inc.<br>1015 Gramsie Road | |
| Telephone: | (763) 505 7863 |
|------------------------|----------------------------------------------|
| Fax: | (763) 505 7877 |
| E-Mail | michelle.d.nivala@medtronic.com |
| Proprietary Name: | SelectSite™ C304 Deflectable Catheter System |
| Common Name: | Catheter, Percutaneous |
| Device Classification: | Class II, 21 CFR 870.1250 |
| Product Code: | DQY |
Shoreview MN 55126-3082
USA
# Device Description
The SelectSite™ C304 Deflectable Catheter System contains 1 deflectable catheter, 1 deflectable catheter dilator, 1 universal slitter, 1 guidewire, 1 needle and 1 syringe. The SelectSite™ C304 Deflectable Catheter System is designed to access the coronary sinus and the chambers of the heart. The percutaneous needle and syringe are used to access the venous insertion site, the guidewire to access the vein, the introducer valve to reduce blood loss during the implant procedure, the deflectable catheter to introduce a transvenous device, the deflectable catheter dilator to facilitate deflectable catheter passage and the guide catheter slitter to remove the deflectable catheter. The SelectSite™ C304 Deflectable Catheter System is available in four models which are the 2006
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the Medtronic logo. The logo consists of a circular symbol on the left and the word "Medtronic" in bold, sans-serif font on the right. The circular symbol contains three stylized figures arranged in a circular pattern.
K06 1416
page 2 of 2
C304-S59, C304-L69, C304-XS59 and C304-XL74. All components except the deflectable catheter and dilator are identical in each model.
# Indications for Use
The deflectable catheter system is indicated to provide a pathway though which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
## Substantially Equivalent Devices:
The SelectSite™ C304 Deflectable Catheter System uses similar technology and has similar intended uses, function, materials and method of operation to the following predicate devices:
- SelectSite™ C304-S5901 and C304-L6901 Deflectable Catheter Systems . (K#033989, cleared January 22, 2004)
- Medtronic Attain® 6226DEF Deflectable Catheter Delivery System (K#032312, . cleared November 3, 2003)
### Summary of Studies:
Device integrity testing was performed to support the equivalency of the SelectSite™ C304 Deflectable Catheter System to the predicate devices. Testing included mechanical, functional, and biocompatibility testing. The SelectSite met all specified design and performance requirements.
### Biocompatibility Information
The biocompatibility evaluation completed for the SelectSite™ C304 Deflectable Catheter Delivery System verifies that the SelectSite™ C304 Deflectable Catheter System is biocompatible. The testing which supports the biocompatibility of the SelectSite™ C304 Deflectable Catheter System is consistent with International Standard ISO 10993-1:, "Biological Evaluation of Medical devices- Part 1: Evaluation and Testing." When classified according to this standard, the catheter and dilator included in the SelectSite™ C304 Deflectable Catheter System are external communicating devices with limited exposure (<24 hours) to circulating blood.
#### Sterilization Validation
The SelectSite™ C304 Deflectable Catheter System will be sterilized using a validated Ethylene Oxide (EtO) sterilization process.
#### Conclusion
Through the data and information presented, Medtronic Ireland considers the SelectSite™ C304 Deflectable Catheter System to be substantially equivalent to legally marketed predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 5 2006
Medtronic Ireland c/o Ms. Michele Nivala Regulatory Affairs Specialist Medtronic, Inc. 1015 Gramsie Road Shoreview, MN, 55126-3082
Re: K061416
Trade/Device Name: Selectsite C304 Deflectable Catheter System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: October 4, 2006 Received: October 10, 2006
Dear Ms. Nivala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Michele Nivala
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-_. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Bhimmar for
Bram D. Zuckerman, MD Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known):_KOG 14 / 14 / 14 / 14 /
Device Name: SelectSite C304 Deflectable Catheter System
Indications For Use: The deflectable catheter system is indicated to provide a pathway though which diagnostic and therapeutic transvenous devices are introduced within the chambers and coronary vasculature of the heart, and for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blummenmo
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K061416