K061317 · Trumpf Kreuzer Medizin Systeme GmbH + Co.Kg · FSY · Jun 22, 2006 · General, Plastic Surgery

Device Facts

Record IDK061317
Device NameILED
ApplicantTrumpf Kreuzer Medizin Systeme GmbH + Co.Kg
Product CodeFSY · General, Plastic Surgery
Decision DateJun 22, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4580
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The iLED lighting system is for illuminating an examination and surgical site on the patient in the clinic and doctor's office.

Device Story

The iLED is a surgical lamp system consisting of hexagonal LED modules (iLED3 or iLED5 configurations) mounted on swivel arms. It provides high-intensity, shadow-free, "cold" illumination for surgical and examination sites in clinics and doctor's offices. The device features adjustable light intensity (30-100%) and color temperature (3500 K-5000 K). Each LED module includes integrated optical devices to illuminate the entire light field. The system supports optional CCD-video-camera integration and ceiling-mounted support systems. By using multiple LEDs (36 or 37 per module), the device ensures that failure of a single light source has minimal impact on overall intensity. The primary benefit is high-intensity illumination with reduced heat radiation and significantly longer durability compared to traditional gas-discharging lamps.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Surgical lamp utilizing LED light sources. Features hexagonal modules with integrated optics. Adjustable intensity (30-100%) and color temperature (3500-5000 K). Includes shadow control functionality. Modular design (iLED3/iLED5). Ceiling-mounted with swivel arms. Optional CCD-video-camera integration.

Indications for Use

Indicated for local illumination of the operating site on a patient's body with high-intensity, shadow-free, "cold" light.

Regulatory Classification

Identification

A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.

Special Controls

*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K06/3/7 JUN 2 2 2 2006 ## 5. 510 (k) Summary | Submitter: | TRUMPF KREUZER Medizin Systeme GmbH + Co. KG<br>Benzstraße 26<br>82178 Puchheim<br>Germany | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Willibald Hiemer<br>Manager Regulatory Affairs<br>Phone: ++49 89 80 90 7 - 0<br>Fax: ++49 89 80 90 7 - 20 | | Preparation Date: | April 28, 2006 | | Trade Name: | iLED | | Common Name: | Surgical lamp | | Classification Name: | Light, Surgical, Ceiling Mounted | | Device Description: | The iLED surgical light is suitable for all types of surgical<br>procedures.<br>The iLED light heads consists of several hexagonal modules,<br>which contain the LEDs with their optical devices. Each LED<br>with its optical device illuminates the complete light field. The<br>iLED5 has one center module and four edge modules; the<br>iLED3 has three edge modules for the illumination. Each light<br>head has its own control to adjust the illumination parameters.<br>The light intensity is adjustable between 30% and 100%, the<br>color temperature between 3500 K and 5000 K. A dimming of<br>the light to the point of an application at endoscopical working<br>is possible and a adaptability to different situations (shadow<br>control) to improve the illumination of the surgical field.<br>With the use of LEDs as light sources, TRUMPF realizes a<br>high illumination intensity with a low heat radiation and a<br>durability of at least 4 times compared with gas-discharging<br>lamps. | 11:40 PM IST 11: PAGE 11 of 137 PAGES : {1}------------------------------------------------ The light incorporates easy-to-operate swivel arms and can be combined to systems with one, two or three light heads of the type iLED3 or iLED5. An optional CCD-video-camera is available and the light could be combined with TRUMPF KREUZER ceiling mounted support systems. Intended Use of The iLED lighting system is for illuminating an examination the Device: and surgical site on the patient in the clinic and doctor's office. Indication for use: The surgical light iLED is intended to illuminate locally the operating site on the patient's body with a high intensity, shadow fee, "cold" light. Predicate Device: CHROMOPHARE X 65 K# 024132 Substantial Equivalence: The iLED is substantially equivalent to the surgical light CHROMOPHARE X 65. Any difference that exists between the CHROMOPHARE X65 and the il.ED has no negative effect on safety or effectiveness and actually enhances the usefulness in the chosen application. Main Difference: The light source LightEmittingDiode, used by iLED, has a life time, which is 4 times the life time of a gas-discharge lamp. The CHROMOPHARE has two light sources whereas the iLED use 36 or 37 light sources per module. A failure of one light source of the iLED reduces the light intensity scarcely. PAGE 12 of 137 PAGES {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 22 2006 Trumpf Kreuzer Medizin System GmbH % Underwriters Laboratories, Inc. Mr. Jeffrey D. Rongero Senior Project Engineer 12 Laboratory Drive Research Triangle Park, North Carolina 27709 Re: K061317 Trade/Device Name: iLED Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulation Class: II Product Code: FSY Dated: June 30, 2006 Received: June 30, 2006 Dear Mr. Rongero: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Jeffrey D. Rongero forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrf/dsma/dsmamain.html Sincerely vours. C. Herbert Lehman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 4. Indication for use Statement 510(k) Number: KO61317 Device Name: ilED Indications for use: The surgical light iLED is intended to illuminate locally the operating site on the patient's body with a high intensity, shadow fee, "cold" light. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The counter Use (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WIRTE BELOW THIS LINE – CONTINUE ONANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office or Device Evaluation (ODE) (Division ign-Off Division of General, Restorative。 and Neurological Devices 510(k) Number PAGE 10 of 137 PAGES
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...