# INSIGHT HEAD COIL (K061281)

_Insight Neuroimaging Systems, LLC · MOS · Jun 19, 2006 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K061281

## Device Facts

- **Applicant:** Insight Neuroimaging Systems, LLC
- **Product Code:** [MOS](/productcode/MOS.md)
- **Decision Date:** Jun 19, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Insight Head Coil is a quadrature transmitting and receiving device used to produce magnetic resonance images of the head regions that can be interpreted by a trained radiologist. This device provides a closed, ergonomic package with many patient friendly features including a shorter coil length so that no part of the coil will be over the patient's eyes, nose or mouth.

## Device Story

Insight Head Coil; quadrature transmit/receive device for MRI of head/neck. Tubular design (13" OD x 9.9" ID x 10" L) with 60-degree cutout; ergonomic, shorter length to avoid covering patient eyes, nose, or mouth; reduces claustrophobia via attachable mirror and open-front design. Used in clinical MRI suites; operated by radiology technicians/physicians. Device slides on rails to position patient head. Provides high SNR images for radiologist interpretation; aids in diagnostic decision-making by minimizing motion artifacts and improving patient comfort.

## Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation processes confirmed image quality comparable to predicate devices.

## Technological Characteristics

Quadrature 2-channel design; multiple elements surrounding head tissue. Materials: painted plastic polymer (biologically inert). Dimensions: 13" OD x 9.9" ID x 10" L. Compatible with Siemens and GE 3T MRI systems. Cleaning: mild bleach or alcohol.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Magnetom Trio Head Coil ([K021330](/device/K021330.md))
- AIR Head Coil ([K023929](/device/K023929.md))
- TEM 3000 Head Coil ([K040937](/device/K040937.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 19 2006

7

# 510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is submitted as part of the PreMarket Notification in accordance with the requirements of 21 CFR Part 807, Subpart E and Section 807.92.

## 1. Identification of Submitter:

| Submitter:     | Insight Neuroimaging Systems LLC        |
|----------------|-----------------------------------------|
| Address:       | 11 Canterbury St.<br>Worcester MA 01610 |
| Phone:         | (508) 799-6464                          |
| Fax:           | (508) 799-6030                          |
| Contact:       | Patricia A. Milbank                     |
| Title:         | Regulatory Consultant                   |
| Phone:         | 425-894-9733                            |
| Fax:           | 425-822-3648                            |
| Date Prepared: | May 5, 2006                             |

## 2. Identification of Product:

| Trade Name:        | Insight Head Coil                                                            |
|--------------------|------------------------------------------------------------------------------|
| Regulatory Number: | 21 CFR 892.1000                                                              |
| Regulation Name:   | Magnetic Resonance Diagnostic Device                                         |
| Common Name:       | Coil, Magnetic Resonance Specialty                                           |
| Regulatory Class:  | Class II                                                                     |
| Product Code:      | 90 MOS                                                                       |
| Manufacturer:      | Insight Neuroimaging Systems LLC.<br>11 Canterbury St.<br>Worcester MA 01610 |

#### 4. Indications for Use

The Insight Head Coil is a quadrature transmitting and receiving device used to produce magnetic resonance images of the head regions that can be interpreted by a trained radiologist. This device provides a closed, ergonomic package with many patient friendly features including a shorter coil length so that no part of the coil will be over the patient's eyes, nose or mouth.

{1}------------------------------------------------

#### 5. Device Description:

The Insight Head Coil is a quadrature transmitting and receiving device for use in magnetic resonance imaging of the head regions. This device incorporates advanced radio frequency technology housed in a closed, ergonomic package with many patient friendly features including a shorter coil length so that no part of the coil will be over the patients eyes, nose or mouth. The materials of construction and components are all identical to the predicate devices.

The tubular shaped coil is 13" OD x 9.9" ID x 10" L with a section removed from the bottom that is 60° of the coil circumference and 2" deep. The coil is attached to 2 rails that allow it to slide back and away from the patient, and then to be positioned with the patients head inside the tubular device. The back and sides of the coil will be placed over the center of the ears of the patient while the cut out front edge will be placed parallel with the eye brows.

The outer surfaces of the coil are manufactured from painted plastic polymer materials. These materials have been found to be biologically inert. These materials are routinely used in medical device construction and can be cleaned using mild bleach or alcohol as the customers will be instructed to do so.

The outer surfaces of the coil are manufactured from materials that are routinely used in medical device construction and can be cleaned using mild bleach or alcohol.

## Features of the Insight Head Coil:

#### Imaging

- . 2-channel quadrature design uses multiple elements that surround the head tissue
- . Optimized for high speed imaging applications
- . Compatible with Siemens and GE3T MRI systems
- . High SNR

#### Ergonomics

- . Large ID to accommodate a major percentage of patient population
- . Ergonomically designed for patient comfort to minimize motion artifact
- . Positioning accessories include an adjustable head support and retraction mechanism
- . Attachable mirror to reduce patient claustrophobic response
- . Complete lack of material in front of the patients face to reduce claustrophobic response.

{2}------------------------------------------------

## 6. Comparison with Legally Marketed Devices

The Insight Head Coil is substantially equivalent to the legally marketed devices listed below:

| Model:          | Magnetom Trio Head Coil   |
|-----------------|---------------------------|
| Manufacturer:   | USA Instruments           |
| 510 (k) Number: | K021330                   |
| Model:          | AIR Head Coil             |
| Manufacturer:   | Advanced Imaging Research |
| 510 (k) Number: | K023929                   |
| Model:          | TEM 3000 Head Coil        |
| Manufacturer:   | MR Instruments Inc.       |
| 510 (k) Number: | K040937                   |

All of these head coils are intended for use in conjunction with a magnetic resonance scanner to produce diagnostic images of the head and neck tissues that can be interpreted by a radiologist.

The predicate devices are designed for use with various MR scanners as is the Insight Head Coil.

The Insight Head Coil is a quadrature transmit and receive coil for imaging the tissues of the head and neck. All of these head coils have an open coil design to optimize imaging of heads of various sizes.

#### 7. Conclusions

The Insight Head Coil is substantially equivalent to the identified legally marketed devices. The potential hazards have been studied and controlled as part of the product development process, including risk analysis, test and design considerations, and planned verification and validation testing processes. The Insight Head Coil provides images comparable to the predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

JUN 19 2006

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

Ms. Patricia A. Milbank Regulatory Consultant Insight Neuroimaging Systems LLC 11 Canterbury Street WORCHESTER MA 01610

Re: K061281

Trade/Device Name: Insight Head Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: May 5, 2006 Received: May 9, 2006

Dear Ms. Milbank:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/3/Picture/10 description: The image is a circular logo with the text "1906-2006 FDA Centennial" in it. The letters "FDA" are in large, bold font in the center of the logo. There are three stars below the word "Centennial". The logo is surrounded by a circular border with text, but the text is too small to read.

Protecting and Promoting Public Health

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrlv/industry/support/index.html.

Sincerely vours.

Nancy C. hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

6

## Indication(s) for Use Statement

510(k) Number:

To be assigned by FDA

K061281

Device Name:

Insight Head Coil

Indications for Use:

The Insight Head Coil is a quadrature transmitting and receiving device used to produce magnetic resonance images of the head regions that can be interpreted by a trained radiologist. This device provides a closed, ergonomic package with many patient friendly features including a shorter coil length so that no part of the coil will be over the patient's eyes, nose or mouth.

Prescription Use
(Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Lyson

(Division Sign-Off) Division of Reproductive. Ab and Radiological Devic 510(k) Numbe

Original 510(k) PreMarket Notification Insight Neuroimaging Systems LLC., Insight Head Coll

---

**Source:** [https://fda.innolitics.com/device/K061281](https://fda.innolitics.com/device/K061281)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com