The Lapdoc navigation system is a computerized navigational system intended to provide assistance in both the preoperative planning and the intra-operative surgical phases of dental implantation surgery. The system provides precise navigational guidance of surgical instruments such as a dental bur, with regard to the pre-operative planning in the dental implantation procedure.
Device Story
Lapdoc is a computer-assisted surgical navigation system for dental implantology. Input: 3D medical image data (CT) and real-time spatial tracking data. Operation: Surgeon performs pre-operative planning by placing virtual implant models on CT images and marking sensitive structures (e.g., nerves). Intra-operatively, an infrared-based stereo camera tracks rigid bodies attached to the patient (via dental splint) and the surgical bur (via cone nut connector). The system calculates real-time deviation between actual and planned bur position/orientation. Output: Visual and audible guidance via a mini-display near the patient's mouth. Surgeon manually adjusts bur position to match the plan. Benefits: Increased precision; reduced complication rates; avoidance of injury to sensitive structures. Used in clinics by dental implantologists. Does not replace human judgment; surgeon retains ultimate responsibility.
Clinical Evidence
Bench testing only. No clinical data presented. Performance validation focused on system accuracy (+/- 1.0 mm) compared against state-of-the-art technology, expert opinion, and predicate device specifications.
Technological Characteristics
Optical infrared-based stereo camera system; rigid body trackers with reflectors; dental splint and cone nut connector. Materials: Stainless steel (ASTM E889, A182), Titanium (ASTM F8265, F67). Connectivity: Computer-based system. Software: Programmable electrical medical system. Sterilization: Not specified. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-1, IEC 60601-1-4, ISO 10993-5/10, ISO 14971.
Indications for Use
Indicated for patients undergoing dental implantation surgery requiring precise navigational guidance of surgical instruments (e.g., dental bur) relative to pre-operative planning. No specific age or gender contraindications stated.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Ko61224
Name
510(k) Notification Lapdoc
SECTION E - 510(k) SUMMARY
SEP - 5 2006
Datum
| Section E - 510(k) Summary | | Erstelli<br>Geprüft | | |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--|--|
| Applicant´s Name: | RoboDent GmbH<br>Kurfürstendamm 213<br>10719 Berlin<br>Tel: +49 (30) 49 500 820<br>Fax: +49 (30) 49 500 822 | Genehmigt<br>Freigegeben | | |
| Contact Person : | Prof. Dr. mult. h.c. Jürgen Bier<br>RoboDent GmbH<br>Kurfürstendamm 213<br>10719 Berlin<br>Tel: +49 (30) 49 500 823<br>Fax: +49 (30) 49 500 822<br>i.bier@robodent.com | | | |
| Date prepared : | April 2006 | | | |
| Trade Name | Lapdoc | | | |
| Common name | Image Guided Surgery System, CAS / Stereotaxic Instrument | | | |
| Classification name | Stereotaxic Instrument (per 21 CFR section 880.4560 HAW) | | | |
| Regulatory Class: | Class II | | | |
| Predicate Devices: | IGI-System (K023424)<br>OrthoPilot (K003347)<br>Vectorvision Cranial / ENT (K023651)<br>Vectorvision2 (K983831) | | | |
Classification: The FDA has classified "stereotaxic Instruments" devices as class II, pursuant to 21 CFR 882.4560 (product code: HAW).
Description of the Device: The Lapdoc System is a computer based assistance system for dental implantologists. It supports surgeons during planning and execution phase of dental implant procedures by providing highly precise quiding assistance during the implantation process within the human jaw. The systems supports the surgical implantation process by providing visual guidance for position and orienting of a dental bur relative to a planned implant position.
The Lapdoc system is designed to support clinicians by decreasing complication rates of surgical interventions in the field of dental implantology. Using the system will allow precise orientation of the surgeon within the human jaw both in the medical image model of the jaw during the planning phase as well as in real anatomy during the execution phase. This is achieved by objective measuring the spatial positions of surgical instruments and the patients anatomy. Based on three dimensional medical image data, the surgeon plans an intervention by virtually placing models of implants within the graphical depiction (of the medical image set) of the jaw. The surgeon views and interprets the medical images which are graphically augmented with the computer models of the implants. It is further possible to manually mark the outlines of sensitive anatomical structures (i.e. nerves) within
510(k) Summary printed by: Dirk Mucha Print date: 2005-04-04 File printed: W.'SRL. Verwallung IBZWM PDF PROJEKTAKTEN2004 FDA LapDocl2006 FDA files'06-04-03 Summary.doo
> complete file Page 15 of 616 Pages
P. 1 of 5
{1}------------------------------------------------
510(k) Notification Lapdoc
#### SECTION E - 510(k) SUMMARY
the medical images. To perfectly understand the geometric dimensions of the patients anatomy and the later intervention, a three dimensional model is created from isodense surfaces in the CT images and displayed on the computer screen. Lapdoc uses the geometric information of the planned implants within the medical images to generate highly precise information to guide the surgeon's instrument directly to the intended implant position and orientation.
An optical measurement system consisting of an infrared-based stereo camera system and a set of two rigid structures (also known as rigid bodies or trackers) with attached reflectors is used to measure the spatial positions of the patient and the instruments. Therefore, the trackers are reproducible attached to the patient (using a dental splint) and to the surgical bur (using a cone nut connector). The stereo camera system comprises an internal position measurement processor and is connected to the Lapdoc computer. It senses the positions of the instruments and the patients tracker and communicates it to the Lapdoc computer. The computer calculates in real time the deviation between the actual and the desired (from the planning phase) transformation of the positions of both trackers. The actual bur position and the planned implant position and orientation are graphically displayed to the surgeon using a mini-display is placed within the vicinity of the surgeon and near the patient's mouth. The surgeon minimizes this deviation during the drill procedure by moving the bur in a way, that position and orientation of both the dental bur and planned implant are congruent. The system supports the surgeon by actieving a highly precise surgical outcome according to the planned scenario.
Lapdoc system will graphically and audibly signal to the surgeon, if the plan is fulfilled and the surgical bur has reached its designated position and orientation. It will further warn the surgeon by reaching a minimal proximity value close to sensitive anatomical structures (i.e. nerves) and therefore avoids injury of such structures. Lapdoc is specifically designed to minimize the interaction between the user and the system. The application of the system follows simple ergonomic rules. The usage and working principle is easy to understand by the surgeon. The clinical workflow is not altered using Lapdoc.
To describe the patients anatomy and the medical image model in one common coordinate system, it is necessary to perform an image registration process. This registration process comprises a navigation bow of known geometry and an algorithm to detect and measure the position of the navigation bow within the medical image data set. Lapdoc correlates the positions of the navigation bow within the image data with its measured spatial position. Since the navigation bow is attached reproducible to the patients jaw by click-snaping it onto the patient's teeth, positions on or within the patients anatomy can also be described within the medical image data set using the registration function.
510(k) Summary printed by: Dirk Mucha Print date: 2006-04-04 File printed: W.'SRL. Verwallung!BZWM PDF PROJEKTAKTEN2004 FDA LapDocl2006 FDA files\06-04-03 Summary.doc
> complete file Page 16 of 616 Pages
/2 of 5
{2}------------------------------------------------
#### SECTION E - 510(k) SUMMARY
The Lapdoc provides accurate guidance during the process of dental implantation. Surgical instruments are guided according to a pre-operative planning. The surgeon is assisted by avoiding the risk of causing damage to critical anatomical structures. Lapdoc is defined as a clinical assistance system, to support the surgeon during the implantation process with the ability of precise determination of position and orientation of surgical instruments within the patients anatomy. Lapdoc does not replace the human judgment. It is always the surgeons choice to define specific locations and angulations of the implant and to pursue this decision within the clinical treatment phase. Thus. the surgeon will not be released from his/her sole and ultimate clinical responsibility.
Intended use: The Lapdoc navigation system is a computerized navigational system intended to provide assistance in the preoperative planning and the intra-operative surgical phases of dental implantation surgery. The system provides precise navigational guidance of surgical instruments such as a dental bur, according to the preoperative planning in the dental implantation procedure.
Substantial Equivalence: Lapdoc shares technological and clinical features and an identical Intended use with different FDA-cleared tracking devices. The DenX-IGI (K023424), the Orthopilot (K003347), the VectorVision Cranial / ENT® (K023651) and the VectorVision 2 ® (K983831) system. Together, all this devices share some important features (not exclusively):
- All devices serve as assistance systems by providing visual support during the freehand ﻪ positioning and orientation process of surgical instruments relative to the patients anatomy;
- All devices measure positions of the human anatomy and surgical instruments by means of optical infrared-based measurement systems consisting of stereo camera systems;
- All devices calculate the position and orientation of surgical instruments within the medical image data set by using the position measurement data;
- All devices use sophisticated, yet easily understandable software tools to visualize medical ・ image data in different manners (slices and three dimensional models) and the positions and orientation of surgical instruments;
- All devices reduce the risk of harming or damaging anatomical structures by visualization of ﻴﺮ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻘﻠﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍ such with respect to the actual positions of surgical tools;
- All devices provide navigational accuracy of +/- 1,0 mm (overall system accuracy), which can not be achieved by brut human visual inspection.
The DenxIGP (K023424) system is a system for computer aided navigation of surgical instruments, whose purpose is to position a dental bur on the human jaw to optimally place and drill implant cavities. The system is based on the planning of such implant position within a three dimensional medical image data set (CT) and to pursue this position and angulation during the implantation process using an integrated position measurement device. The Orthopilot® (K003347) is
{3}------------------------------------------------
### SECTION E - 510(k) SUMMARY
a computer aided system to navigate surgical instruments and to optimize the positions of cutting templates/guides for total knee replacement surgery and providing intra-operative measurements of bone alignment. The VectorVision Cranial / ENT® (K023651) and the VectorVision 2 °(K983831) are computer-aided systems for open or percutaneous surgery. They are indicated for any medical condition, where a reference to a rigid anatomical structure, such as the skull, a long bone or vertebra can be identified relative to a CT, MR or X-ray based model of anatomy.
Lapdoc and its predicate devices provide precise and accurate guidance of surgical tools and instruments within the patient's anatomy, near to organs or parts of it. Guidance is achieved by measuring positions of both the instruments and the anatomy and objective visualization of proportions and/or coherences of this spatial relationship.
All described devices do not intend to replace the surgeons expert knowledge and to release the clinicians responsibility.
The following navigation systems are used for the same body organs:
| System | organ / body part / clinical application |
|---------------------|------------------------------------------|
| DenX IGI® (K023424) | Navigation in dental implantology |
| Lapdoc | Navigation in dental implantology |
Since the general principle of surgical navigation can be applied to any surgical field the following described navigation systems are used on different body organs:
| System | organ / body part / clinical application |
|-----------------------------|-----------------------------------------------------------------------|
| DenX IGI® (K023424) | Navigation in dental implantology |
| Orthopilot® (K003347) | Navigation for knee replacement surgery and bone alignment surgery |
| VectorVision 2 ® (K983831) | Navigation in open/percutaneous surgery (skull/long-bones/vertebraes) |
| VectorVision Cranial / ENT® | Navigation in open/percutaneous surgery (skull/long-bones/vertebraes) |
| (K023651) | |
| Lapdoc | Navigation in dental implantology |
The effectiveness of all navigation support systems is based on their accuracy during the actual spatial guiding process for surgical instruments. The underlying principles of measuring positions of parts of the human anatomy and the positions of instruments, the registration of medical image data to the anatomy does not change with respect to the specific field of application remain the same. It can be assumed, that the safety and performance assumptions are not altered, when the principle of navigation is applied to the field of dental navigation. Summarized, the LapDoc-system is substantially equivalent to the declared predicate devices.
P. 4 of 5
{4}------------------------------------------------
#### SECTION E - 510(k) SUMMARY
Conclusions: Data presented above clearly demonstrate that the Lapdoc-System is substantially equivalent to its predicate devices in its technological features, clinical intended uses, its safety and effectiveness.
#### Performance standards:
510(k) Notification
Lapdoc
| Standards<br>No. | Standards<br>Organization | Standards Title | Version | Date |
|------------------|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|---------|
| 601-1 | IEC | Medical electrical equipment - Part 1: General requirements<br>for safety | German version<br>EN 60601-:1990 +<br>A1:1993 +.A2:1995 | 03/1996 |
| 60601-1-2 | IEC | Equipment<br>-<br>Part<br>Medical<br>Electrical<br>1-2:<br>General<br>Requirements<br>Safety -<br>Collateral<br>for<br>Standard:<br>Electromagnetic Compatibility Requirements and Tests. | German version<br>EN 60601-1-2:2001 | 10/2002 |
| 60601-1-1 | IEC | Medical electrical equipment - Part 1-1: General requirements<br>for safety - Collateral standard: Safety requirements for<br>medical electrical systems | German version<br>EN 60601-1-1:2001 | 08/2002 |
| 60601-1-4 | IEC | Medical electrical equipment Part 1-4: General requirements<br>for safety - Collateral standard: Programmable electrical<br>medical systems | German version<br>IEC 60601-1-<br>4:1996+A1:1999 | 04/2001 |
| 60601-1/A13 | EN | Medical electrical equipment - Part 1: General requirements<br>for safety | German version<br>EN 60601-1 : 1990/<br>AM 13 :1996 | 10/1996 |
| 15223 | ાકભ | Medical Devices - Symbols to be used with medical device<br>labels, labelling and information to be supplied | 2nd edition | 04/2000 |
| 10993-5 | ાટળ | Biological Evaluation of medical devices<br>Part 5: Tests for cytotoxicity: in-vitro methods | German Version<br>EN ISO 10993-<br>5:1999 | 11/1999 |
| 10993-10 | ાકેર | Biological Evaluation of medical devices<br>Part 10: Tests for irritation and delayed-type hypersensitivity | German version<br>EN ISO 10993-<br>10:2002 | 02/2003 |
| 14791 | ાકર | Medical devices - Application of risk management to medical<br>devices | German version<br>EN ISO<br>14971:2001 | 03/2001 |
| 880 | EN | Graphical symbols for use in the labelling of medical devices | German version<br>EN 980:2003 | 08/2003 |
| 1041 | EN | Information supplied by the manufacturer with medical devices | German version<br>EN 1041:1998 | 04/1998 |
| E889 | ASTM | Standard Specification for Stainless Steel for Surgical<br>Instruments | 02 | 04/2002 |
| A182 | ASTM | Forged or Rolled Alloy and Stainless Steel Pipe Flanges,<br>Forged Fittings, and Valves and Parts for High-Temperature<br>Service | 04 | 10/2004 |
| 8265 | ASTM | Standard Specification for Titanium and Titanium Alloy Strip | 03 | 11/2003 |
| F67 | ASTM | Standard Specification for unalloyed Titanium for surgical<br>Implant Applications | 00 | 10/2000 |
Accuracy and functional performance standard: As of today, there are no specific FDA Guidance documents regarding the Image-Guided surgery systems for oral and maxillofacial surgery. The level of accuracy, defined to be acceptable in dental implantation navigation, was +/- 1,0 mm. This level of accuracy was defined based on review of state of the art technology (see Appendix documents No. 8 to No. 12), opinions by experts in the field (see Appendix documents No. 6 to No. 8) and the accuracy specified for the FDA cleared predicate devices (see Appendix documents No. 13 to No. 16).
510(k) Summary printed by: Dirk Mucha Print data: 2006-04-04 File printed: W:'SRL. Vervallung'BZMM PDF PROJEKTAKTE:N2004 FDA LapDocl2006 FDA files'06-04-03 Summary.coc
Ps.ofs
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human figures connected by flowing lines, representing the department's mission to protect the health of all Americans and provide essential human services.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## SEP - 5 2006
Robodent GMBH % Mr. Juergen Bier Kurfurstendamm 213 Berlin, Germany 10719
Re: K061224
Trade/Device Name: Lapdoc Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulation Class: Class II Product Code: HAW Dated: August 14, 2006 Received: August 21, 2006
Dear Mr. Bier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Juergen Bier
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
fra
Mark N. Mekerson
Mark N. Mekerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K061224
Device Name: Lapdoc
Indications For Use:
The Lapdoc navigation system is a computerized navigational system intended to provide assistance in both the preoperative planning and the intra-operative surgical phases of dental implantation surgery. The system provides precise navigational guidance of surgical instruments such as a dental bur, with regard to the pre-operative planning in the dental implantation procedure.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
CHIVIS ign-Off
Division of General, Restorative, and Neurological Devices
510(k) Number.
Page 1 of
complete file Page 13 of 616 Pages
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.