← Product Code [IYO](/productcode/IYO) · K061215

# VOLCANO S51 FAMILY OF IMAGING SYSTEMS (K061215)

_Volcano Corporation · IYO · Aug 10, 2006 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K061215

## Device Facts

- **Applicant:** Volcano Corporation
- **Product Code:** [IYO](/productcode/IYO.md)
- **Decision Date:** Aug 10, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

Volcano s5i Family of Imaging Systems is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures. ChromaFlo is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.

## Device Story

System consists of imaging catheter, patient interface module, and console. Gathers/displays high-resolution intraluminal images for qualitative/quantitative analysis. Used in cardiac catheterization labs by physicians as adjunct to interventional therapies (e.g., balloon angioplasty). ChromaFlo feature overlays 2D color map of relative blood flow onto grayscale images. In-Line Digital option provides 2D, 360-degree rotation, and longitudinal views. Software enables connectivity with third-party fluoroscopic systems to synchronize patient demographics (DICOM) and provide remote access to IVUS functions from the fluoroscopic UI. Output assists clinicians in vessel analysis and procedural guidance.

## Clinical Evidence

Bench testing only. Performance data evaluated modifications to the system; results met all pre-defined test plans and protocols, establishing safety and effectiveness.

## Technological Characteristics

Ultrasonic pulsed echo imaging system (21 CFR 892.1560). Components: imaging catheter, patient interface module, system console. Features: ChromaFlo (color flow mapping), In-Line Digital (360-degree/longitudinal imaging), and third-party fluoroscopic system connectivity. Software-based image processing and workflow integration.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Cathscanner III Imaging System ([K944004](/device/K944004.md))
- ColorFlo Option for the Oracle InVision Intravascular Ultrasound Imaging System ([K963290](/device/K963290.md))
- Resolve Option for the Oracle InVision Intravascular Ultrasound Imaging System ([K965223](/device/K965223.md))
- In Vision Imaging System ([K031148](/device/K031148.md))
- Volcano s5 Imaging System ([K051920](/device/K051920.md))

## Reference Devices

- Visions catheter ([K982329](/device/K982329.md))
- Avanar catheter ([K000820](/device/K000820.md))
- Eagle Eye Gold ([K031346](/device/K031346.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

149

(06/21 2

Volcano s5i Family of Imaging System Special 510(k) Premarket Notification

# 510(k) SUMMARY

The 510(k) Summary is submitted as required by Section 807.92(a)

| Submitter Name:                               | Volcano Corp.                                                               | AUG 10 2006 |
|-----------------------------------------------|-----------------------------------------------------------------------------|-------------|
| Contact Person:                               | Michelle J. Badal, RAC<br>Manager, Regulatory Affairs                       |             |
| Address:                                      | 2870 Kilgore Road<br>Rancho Cordova, CA 95670                               |             |
| Phone Number:                                 | 916-231-4497                                                                |             |
| Fax Number:                                   | 916-638-2647                                                                |             |
| Date Prepared:                                | May 1, 2006                                                                 |             |
| Device Trade Name:                            | Volcano s5i Family of Imaging Systems                                       |             |
| Device Common Name:                           | Ultrasonic imaging system                                                   |             |
| Classification Name,<br>Number, Product Code: | Ultrasonic pulsed echo imaging system<br>21 CFR 892.1560, Product Code: IYO |             |

## Predicate Device:

Submitter:

Volcano Corp.

Cathscanner III Imaging System cleared under K944004; ColorFlo Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K963290; Resolve Option for the Oracle InVision Intravascular Ultrasound Imaging System cleared under K965223; In Vision Imaging System cleared under K031148, and Volcano s5 Imaging System cleared under K051920.

## Device Description:

The Volcano s5i Family of Imaging Systems consists of the imaging catheter, the patient interface module, and the system console. The system console gathers and displays highresolution intraluminal images that can be analyzed both qualitatively and quantitatively. In addition to supplying diagnostic information, the Volcano s5i Imaging Systems can be an adjunct to interventional therapies, such as balloon angioplasty. With ChromoFlo® a twodimensional color map of relative blood flow is overlaid on the grayscale image, providing additional information for vessel analysis. The In-Line Digital option displays a twodimensional, 360 rotations and longitudinal view of the Volcano s5i Imaging System software provides an option for connectivity with third party fluoroscopic imaging systems. The software option simplifies the Cardiac Catherization Lab workflow with the

# 1/2

{1}------------------------------------------------

| Submitter:    | Volcano s5i Family of Imaging System  |
|---------------|---------------------------------------|
| Volcano Corp. | Special 510(k) Premarket Notification |

fluoroscopic imaging systems by, 1) automatically synchronizing the patient demographic information (patient name, date of birth, DICOM attributes, etc.) from the fluoroscopic imaging system with the Volcano s5i Imaging System, 2) providing a remote access to commonly used IVUS functions from the fluoroscopic imaging system user interface.

The imaging catheters are all marketed under separate premarket notifications; Visions catheter K982329, Avanar catheter K000820 and Eagle Eye Gold K031346.

## Intended Use:

Volcano s5i Family of Imaging Systems is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.

ChromaFlo is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.

#### Performance Data:

Applicable testing was performed to evaluate the modifications to the Volcano s5i Family of Imaging Systems. The test results were found to be acceptable as required by the respective test plans and protocols.

#### Conclusion:

The testing reported in this 510(k) establishes the device is safe and effective for its intended use and substantially equivalent to the predicate devices.

4061215

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 0 2006

Ms. Michelle J. Badal, RAC Manager, Regulatory Affairs Volcano Corp. 2870 Kilgore Road Rancho Cordova, CA 95670

Re: K061215

Trade/Device Name: Volcano s5i Family of Imaging Systems Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO Dated: July 7, 2006 Received: July 10, 2006

Dear Ms. Badal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and coometer fore, market the device, subject to the general controls provisions of the Act. The I ou may, increrers, misions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it rr your de rice to such additional controls. Existing major regulations affecting your device can may be subject to back address and egulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

# Page 2 - Ms. Michelle J. Badel, RAC

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bhimmer for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Indications for Use

510(k) Number (if known): _ K 061215

Device Name: Volcano s5i Family of Imaging Systems

Indications for Use:

The Volcano s5i Family of Imaging Systems is used for the qualitative and quantitative evaluation of vascular morphology in the coronary arteries and vessels of the peripheral vasculature. It is also indicated as an adjunct to conventional angiographic procedures to provide an image of vessel lumen and wall structures.

ChromaFlo® is indicated for qualitative blood flow information from peripheral and coronary vasculature; flow information can be an adjunct to other methods of estimating blood flow and blood perfusion.

Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

R.A. Montana

(Posted November 13, 2003)

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K06/215

---

**Source:** [https://fda.innolitics.com/device/K061215](https://fda.innolitics.com/device/K061215)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
