FINGER AND EAR SENSORS AND NEONATAL-ADULT AND ADULT/PEDIATRIC ADHESIVE SENSORS FOR USE WITH NELLCOR R-CAL SPO2
K061191 · Measurement Specialties, Inc. · DQA · May 22, 2006 · Cardiovascular
Device Facts
Record ID
K061191
Device Name
FINGER AND EAR SENSORS AND NEONATAL-ADULT AND ADULT/PEDIATRIC ADHESIVE SENSORS FOR USE WITH NELLCOR R-CAL SPO2
Applicant
Measurement Specialties, Inc.
Product Code
DQA · Cardiovascular
Decision Date
May 22, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Finger sensor: The Finger Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 20 kg. Ear sensor: The Ear Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 10 kg. Neonatal-Adult sensor: The Neonatal-Adult Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, noninvasive functional arterial oxygen saturation and pulse rate monitoring in adult, pediatric, and neonatal patients. Adult/Pediatric Sensor: The Adult/Pediatric Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, noninvasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 20 kg.
Device Story
Electro-optical sensors for pulse oximetry; non-invasive; no skin penetration or electrical contact. Sensor body contains two LEDs (light sources) and one photodiode (detector). Connects to Nellcor® R-Cal compatible pulse oximeters via 9-pin D-connector. Finger and Ear sensors are reusable with spring-loaded structures; Neonatal-Adult and Adult/Pediatric sensors are disposable with adhesive tape attachment. Used in clinical settings to monitor arterial hemoglobin saturation and pulse rate. Output displayed on connected patient monitor to assist clinicians in patient status assessment and clinical decision-making.
Clinical Evidence
SpO2 accuracy testing performed on healthy adult volunteers during induced hypoxia (no motion). Accuracy measured +/-3 digits over 70-100% SpO2 range compared to co-oximeter blood samples. Bench testing verified pulse rate accuracy and sensor surface temperature safety.
Technological Characteristics
Electro-optical sensors; two LEDs and one photodiode; 9-pin D-connector interface. Reusable sensors use spring-loaded mechanical structure; disposable sensors use adhesive tape. Compatible with Nellcor® R-Cal pulse oximeters.
Indications for Use
Indicated for continuous, non-invasive monitoring of functional arterial oxygen saturation and pulse rate. Finger/Adult-Pediatric sensors: patients >= 20 kg. Ear sensor: patients >= 10 kg. Neonatal-Adult sensor: neonatal, pediatric, and adult patients.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K06 1191" in a handwritten style. Below this text, the date "MAY 22 2006" is printed in a bold, sans-serif font. The date indicates that the image or document is related to May 22, 2006.
Image /page/0/Picture/1 description: The image shows the logo for Measurement Specialties. The logo consists of the word "measurement" in bold, sans-serif font, with the word "SPECIALTIES" in a smaller, sans-serif font underneath. To the right of the word "measurement" is a series of dots that get progressively larger.
movation performance reliability
# Section 10 - Premarket Notification 510(k) Summary
Premarket Notification 510(k) Summary As required by section 807.92
SpO2 Sensors for use with Nellcor® R-Cal SpO2 (Finger sensor, reusable; Ear sensor, reusable; Neonatal-Adult sensor, disposable; Adult/Pediatric sensor, disposable)
### GENERAL COMPANY INFORMATON as required by 807.92(a)(1)
### COMPANY NAME/ADDRESS/PHONE/FAX:
Measurement Specialties, Inc 1000 Lucas Way Hampton, VA 23666 Telephone: 757-766-1500 Fax: 757-766-4347
NAME OF CONTACT: Susan Zaks, Product Line Manager, Optical products
DATE: April 25, 2006
### DEVICE NAME as required by 807.92(a)(2)
### TRADE NAME(S):
Finger Sensor for use with Nellcor® R-Cal SpO2 Ear Sensor for use with Nellcor® R-Cal SpO2 Neonatal-Adult Adhesive Sensor for use with Nellcor® R-Cal SpO2 Adult/Pediatric Adhesive Sensor for use with Nellcor® R-Cal SpO2
### COMMON NAME:
SpO2 Sensor (accessory to pulse oximeter and ear oximeter)
# CLASSIFICATION NAME:
수수
The following Class II classification appear applicable DQA 21 CFR 870.2700, Finger Sensor for use with Nellcor® R-Cal SpO2 DQA 21 CFR 870.2700, Neonatal-Adult Adhesive Sensor for use with Nellcor® R-Cal SpO2
> 1000 Lucas Way. Hampton, VA 23666 Toll Free: 800 745 8008 ● Tcl: 757 766 1500 ● Fax: 757 766 4297
www.meas-spec.com
,
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "measurement SPECIALTIES". The word "measurement" is in bold, black, sans-serif font. Below it, "SPECIALTIES" is in a smaller, less bold font. To the right of the text, there is a curved line of black dots that decrease in size as they move away from the text.
nnovation performance reliability
### DQA 21 CFR 870.2700, Adult/Pediatric Adhesive Sensor for use with Nellcor® R-Cal SpO2 DPZ 21 CFR 870.2710. Ear Sensor for use with Nellcor® R-Cal SpO2
# NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)
SpO2 sensors for use with Nellcor® R-Cal SpO2 (Finger sensor, reusable; Ear sensor, reusable; Neonatal-Adult sensor, disposable; Adult/Pediatric sensor, disposable) are substantially equivalent to the predicate SpO2 sensors:
K042704 Flexi-Stat(tm) Finger clip SpO2 sensor K042675 Flexi-Stat(tm) Ear sensor K042705 All-fit Flexi-Stat(tm) Disp SpO2 sensor (Nellcor-compatible) K041522 Flexi-Stat(tm) Disposable Adhesive SpO2 sensor (Nellcor-compatible)
# DEVICE DESCRIPTION as required by 807.92(a)(4)
SpO2 sensors are electro-optical sensors that function without skin penetration, electrical contact, or heat transfer. The sensor uses optical means to determine the light absorption of functional arterial hemoglobin by being connected between the patient and the oximeter. The sensor body contains three optical components: two light emitting diodes (LED) that serve as light sources and one photodiode that acts as a light detector. The sensor body is connected to a cable that is terminated with a 9-pin D-connector for connecting the sensor to a patient monitor or an interconnect cable.
The Finger and Ear sensors are durable, multiple-patient-use sensors. The sensor structure has a spring structure that easily and comfortably accomodates the sensor to the application site. The Neonatal-Adult and Adult/Pediatric sensors are disposable single-patient-use sensors. The attachment mechanism of these sensors is a soft adhesive tape that is wrapped around the application site.
# INTENDED USE as required by 807.92(a)(5)
### Finger sensor:
The Finger sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 20 kg.
### Ear sensor:
.
The Ear sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 10 kg.
### Neonatal-Adult sensor:
1000 Lucas Way, Hampton, VA 23666 Toll Free: 800 745 8008 ● Tel: 757 766 1500 ● Fax: 757 766 4297 www.meas-spec.com
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the words "measurement" in large bold font on the top line and "SPECIALTIES" in smaller font on the second line. The word "measurement" is the primary focus of the image. There are some decorative dots on the right side of the image.
novation performance reliability
The Neonatal-Adult sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, noninvasive functional arterial oxygen saturation and pulse rate monitoring in adult, pediatric, and neonatal patients.
### Adult/Pediatric Sensor:
The Adult/Pediatric sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, noninvasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 20 kg.
# SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)
The Finger sensor, Ear sensor, Neonatal-Adult sensor and Adult/Pediatric sensor use the same theory and principle of operation as the predicate devices. Design characteristics are equivalent in terms of safety and effectiveness, as demonstrated by validation tests.
# SUMMARY OF THE PERFORMANCE TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)
SpO2 accuracy testing was conducted during induced hypoxia studies on healthy adult volunteer subjects during no motion conditions conducted in an independent research laboratory. The measured arterial hemoglobin saturation value (SpO2) of the sensors was compared to arterial hemoglobin oxygen value determined from blood samples with a laboratory co-oximeter. The accuracy of the sensors in comparison to the co-oximeter samples was measured to be +/-3 digits over the SpO2 range 70-100%.
Bench testing with a simulator and the same patient monitors was performed to verify pulse rate measurement accuracy.
Patient interface temperature (Sensor surface temperature) testing was conducted to ensure the sensor surface temperature during the use of the sensors is not higher than for the predicate devices.
### CONCLUSION:
The summary above shows that there are no new questions of safety and effectiveness for the SpO2 sensors Finger Sensor for use with Nellcor® R-Cal SpO2, Ear Sensor for use with Nellcor® R-Cal SpO2, Neonatal-Adult Adhesive Sensor for use with Nellcor® R-Cal SpO2 and Adult/Pediatric Adhesive Sensor for use with Nellcor® R-Cal SpO2 as compared to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, a symbol often associated with the United States.
Public Health Service
MAY 2 2 2006
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Measurement Specialties, Incorporated C/O Ms. Krista Oakes Amica Solutions 2300 McDermott Road, #200-207 Plano, Texas 75025
Re: K061191
Trade/Device Name: Finger Sensor for use with Nellcor R-cal SpO2, Ear Sensor for use with Nellcor R-cal SpO2, Neonatal-Adult Sensor for use with Nellcor R-cal SpO2, Adult/Pediatric Sensor for use with Nellcor R-cal SpO2 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DOA, DPZ Dated: April 28, 2006 Received: April 28, 2006
Dear Ms. Oakes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Ms. Oakes
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sylite y. Michael Dm's
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
#### Enclosure
{5}------------------------------------------------
# Section 5 - Indications For Use Statement
Indications for Use
# 510(k) Number (if known): ΚΟ 6 || 91
Device Name(s): Finger Sensor for use with Nellcor R-cal SpO2, Ear Sensor for use with Nellcor Rcal SpO2, Neonatal-Adult Sensor for use with Nellcor R-cal SpO2, Adult/Pediatric Sensor for use with Nellcor R-cal SpO2.
#### Indications for Use:
#### Finger sensor:
The Finger Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 20 kg.
#### Ear sensor:
The Ear Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, non-invasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 10 kg.
#### Neonatal-Adult sensor:
The Neonatal-Adult Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, noninvasive functional arterial oxygen saturation and pulse rate monitoring in adult, pediatric, and neonatal patients.
#### Adult/Pediatric Sensor:
The Adult/Pediatric Sensor for use with Nellcor® R-cal SpO2 is indicated for continuous, noninvasive functional arterial oxygen saturation and pulse rate monitoring in patients weighing >= 20 kg.
Prescription Use XX (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.png"/> |
|-----------------------------------------------|--------------------------------------------|
| (Sign-Off) | |
| Division of Anesthesiology, General Hospital, | |
| Section Control, Dental Devices | |
| K Number: | K061191 |
| | 23 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.