← Product Code [DIH](/productcode/DIH) · K061093

# CHOLINESTERASE GEN.2 TEST SYSTEM (K061093)

_Roche Diagnostics Corp. · DIH · Jul 3, 2006 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K061093

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [DIH](/productcode/DIH.md)
- **Decision Date:** Jul 3, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3240
- **Device Class:** Class 1
- **Review Panel:** Clinical Toxicology

## Indications for Use

The Cholinesterase Gen.2 Test System is an in vitro test for the quantitative determination of the catalytic activity of cholinesterase (EC 3.1.1.8; acylcholine acylhydrolase) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders.

## Device Story

In vitro diagnostic test for quantitative determination of cholinesterase catalytic activity in serum and plasma; utilizes butyrylthiocholine method. Principle: cholinesterase hydrolyzes butyrylthiocholine to thiocholine and butyrate; thiocholine reduces yellow hexacyanoferrate III to colorless hexacyanoferrate II. Decrease in color measured spectrophotometrically on COBAS Integra analyzers. Used in clinical laboratory settings by trained technicians. Results interpreted by physicians in conjunction with patient medical history to diagnose and manage cholinesterase inhibition disorders.

## Clinical Evidence

Bench testing only. Precision studies (within-run and between-day) performed on Integra 800 analyzer showed CVs ranging from 0.45% to 4.32%. Linearity validated from 200 to 14000 U/L (extended to 28000 U/L). Method comparison against predicate (N=107) yielded correlation coefficients of 0.99. Interference testing confirmed no significant interference (<10%) from common drugs, hemoglobin, bilirubin, or lipemia. No clinical studies were required.

## Technological Characteristics

In vitro diagnostic reagent system. Reagents: R1 (Pyrophosphate, Potassium hexacyanoferrate), R2 (Butyrylthiocholine, GOOD's buffer, stabilizers). Sensing principle: spectrophotometric measurement of hexacyanoferrate III reduction. Form factor: liquid reagent for automated COBAS Integra analyzers. Standardized against reference method using manual butyrylthiocholine/hexacyanoferrate III method.

## Regulatory Identification

A cholinesterase test system is a device intended to measure cholinesterase (an enzyme that catalyzes the hydrolysis of acetylcholine to choline) in human specimens. There are two principal types of cholinesterase in human tissues. True cholinesterase is present at nerve endings and in erythrocytes (red blood cells) but is not present in plasma. Pseudo cholinesterase is present in plasma and liver but is not present in erythrocytes. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders (e.g., insecticide poisoning and succinylcholine poisoning).

## Predicate Devices

- Cholinesterase test system on the Integra Analyzer (k951595)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k061093

B. Purpose for Submission:
New Device

C. Measurand:
Cholinesterase

D. Type of Test:
Quantitative

E. Applicant:
Roche Diagnostics Corp.

F. Proprietary and Established Names:
Cholinesterase Gen.2

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  DIH | I, reserved | 21 CFR 862.3240 | 91 (Tox)  |

H. Intended Use:

1. Intended use(s):
The Cholinesterase Gen.2 Test System is an in vitro test for the quantitative determination of the catalytic activity of cholinesterase (EC 3.1.1.8; acylcholine acylhydrolase) in serum and plasma.

2. Indication(s) for use:
The Cholinesterase Gen.2 Test System is an in vitro test for the quantitative determination of the catalytic activity of cholinesterase (EC 3.1.1.8; acylcholine

{1}

acylhydrolase) in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of cholinesterase inhibition disorders.

3. Special conditions for use statement(s):

For prescription use only.

4. Special instrument requirements:

In this submission, the Cholinesterase Gen.2 test system is being applied to the previously cleared COBAS INTEGRA family of analyzers including the COBAS INTEGRA 400/400 plus; COBAS INTEGRA 700, and COBAS INTEGRA 800 analyzers. The COBAS INTEGRA family of analyzers was cleared in k951595.

I. Device Description:

The Cholinesterase Gen.2 Test System is an in vitro test for the quantitative determination of the catalytic activity of cholinesterase in serum and plasma. The test consists of two ready-to-use liquid reagents. The calibrator is the Calibrator for automated systems (C.f.a.s) that was cleared under k033501. The recommended control materials are Precinorm U or Precinorm U Plus; and Precipath U or Precipath U plus which were cleared under k041227 and k042389 respectively. In this submission the reagents are being applied to the COBAS INTEGRA 400, 400 plus, 700 and 800 analyzers.

J. Substantial Equivalence Information:

|  Predicate | k951595 - Cholinesterase test system on the Integra Analyzer  |
| --- | --- |
|  |   |
|  Similarites  |   |
|  Both the Cholinesterase and the Cholinesterase Gen.2 test systems are intended for the in vitro quantitative determination of the catalytic activity of cholinesterase in serum and plasma.  |   |
|  Both test systems are run on the COBAS INTEGRA family of analyzers.  |   |
|  The test principles are similar; both rely on spectrophotometric determination of reaction product after cholinesterase hydrolyzes a butrylthiocholine derivative.  |   |
|  Differences  |   |
|  Cholinesterase Gen.2 test system measures the decrease in color while the predicate device measures an increase in color.  |   |
|  The reagents for the new test system are provided in liquid ready-to-use format while the predicate device test system reagents require reconstitution  |   |
|  There are slight differences in reagent stability, measuring range, and limit of detection.  |   |

K. Standard/Guidance Document Referenced (if applicable):

{2}

None referenced.

## L. Test Principle:

The Cholinesterase Gen.2 Test System is an in vitro test for the quantitative determination of the catalytic activity of cholinesterase in serum and plasma. The test is based on the butyrylthiocholine method. Cholinesterase catalyzes the hydrolysis of butyrylthiocholine to thiocoline and butyrate. Thiocholine reduces the yellow substrate hexacyanoferrate III to the almost colorless hexacyanoferrate II. The decrease in color is measured spectrophotometrically.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Within-run (n=21) and between-day precision was conducted for the Cholinesterase Gen.2 test system using two controls and two human serum pools on the Integra 800 Analyzer. The between-day precision samples were run in triplicate and the second of the three results were used for the precision determination. The precision summaries are listed in the tables below.

|  Within-run | Control 1 | Control 2 | Serum Pool 1 | Serum Pool 2 | Serum Pool 3  |
| --- | --- | --- | --- | --- | --- |
|  N | 21 | 21 | 21 | 21 | 21  |
|  Mean (U/L) | 6263 | 6015 | 4028 | 6374 | 561.4  |
|  SD | 36.06 | 33.69 | 35.54 | 28.75 | 24.27  |
|  CV% | 0.58 | 0.56 | 0.88 | 0.45 | 4.32  |
|  Min | 6204 | 5948 | 3959 | 6316 | 510.0  |
|  Max | 6331 | 6075 | 4094 | 6423 | 613.0  |
|  Between-day | Control 1 | Control 2 | Serum Pool 1 | Serum Pool 2  |
| --- | --- | --- | --- | --- |
|  N | 21 | 21 | 21 | 21  |
|  Mean (U/L) | 6213 | 5964 | 3152 | 6675  |
|  SD | 70.34 | 55.66 | 39.69 | 95.92  |
|  CV% | 1.13 | 0.93 | 1.26 | 1.44  |
|  Min | 6093 | 5858 | 3065 | 6500  |
|  Max | 6341 | 6053 | 3246 | 6838  |

#### b. Linearity/assay reportable range:

The measuring range for this assay is 200 – 14000 U/L.

{3}

The sponsor conducted two serial dilution linearity studies for the Cholinesterase Gen.2 test system on the Integra 800 analyzer. The samples were prepared by spiking a saline solution with pure cholinesterase from human plasma and diluting with saline. One study measured samples ranging from 404 – 4801 U/L (theoretical spiked concentration). The second study measured samples ranging from 2762 – 27621 U/L (theoretical spiked concentration). All samples were recovered within +/- 10% of the theoretical concentration. The extended measuring range from 14000 to 28000 (based on post dilution factor of 2) was also validated in the second set of samples.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The calibrators and controls for use with this device have been previously cleared. Their composition has not been modified for use with the Cholinesterase Gen.2 test system. The calibrators and controls are standardized against and traceable to a reference method using a manual application of the butyrlthiocholine/ hexacyanoferrate (III) method on a photometer and the published molar absorptivity of hexacyanoferrate (III)

Stability studies for the device were described. Studies supported the sponsor’s 4-week on-board stability claim. The sponsor also described real-time and an accelerated studies to support the 15 month closed stability claim.

d. Detection limit:

Analytical sensitivity (Limit of Blank) was defined as the concentration at three standard deviations above the mean of repeated measurements (N=21) of the zero calibrator. The mean + (3xSD) was calculated to be 122.8 U/L. The sponsor claims a lower detection limit of 200 U/L.

e. Analytical specificity:

Human serum pools that contained 5817 – 9143 U/L cholinesterase was evaluated for endogenous interference by the Cholinesterase Gen.2 test system. The sponsor defined interference as percent recovery that exceeded +/- 10% of the expected 100% recovery. Human serum pools that contained approximately 4700 U/L cholinesterase were tested by the Cholinesterase Gen.2 test system for drug interference. The sponsor defined drug interference as results that deviated more than 10% from the reference value. A list of compounds tested can be found in the device labeling. There was no interference (&lt;10%) from the 18 drug pharmaceuticals tested and also from the following substances:

Hemoglobin up to 347 mg/dL,

{4}

Bilirubin: conjugated up to 65 mg/dL and unconjugated up to 86 mg/dL and Lipemia up to 997 U/L.

f. Assay cut-off:

Not applicable.

## 2. Comparison studies:

a. Method comparison with predicate device:

The sponsor conducted two sample method comparisons between the Integra cholinesterase test system and the Cholinesterase Gen.2 test system from samples collected from hospitals and blood donor clinics. See the table below for regression statistics (least squares linear regression).

|   | N | Range | Equation | Correlation coefficient  |
| --- | --- | --- | --- | --- |
|  Study #1 | 51 | 1144 to 13527 | Y =0.965(X) + 152.5 | 0.99  |
|  Study #2 | 56 (5 spiked) | 171 to 12995 | Y=0.967(X) +98.4 | 0.99  |

Y = Integra Cholinesterase test system and X = Cholinesterase Gen.2 test system

b. Matrix comparison:

K2 EDTA and K3 EDTA plasma samples (N=27 of each) and lithium heparin (N=54) samples were compared to paired serum samples in a matrix comparison study with the Cholinesterase Gen.2 test system on the Integra 800. The linear regression equations, ranges and correlation constants are listed in the table below.

|  Comparator against serum | N | Range (U/L) | Slope | Intercept | R  |
| --- | --- | --- | --- | --- | --- |
|  K2-EDTA Plasma | 27 | 5751 to 11115 | 0.964 | 156 | 0.995  |
|  K3-EDTA Plasma | 27 | 6342 to 9789 | 0.922 | 29.5 | 0.994  |
|  Li-Heparin Plasma | 54 | 5457 to 11129 | 0.970 | 29.8 | 0.995  |

{5}

6

3. Clinical studies:

a. Clinical Sensitivity:
Not Applicable

b. Clinical specificity:
Not Applicable

c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable

4. Clinical cut-off:
Not Applicable

5. Expected values/Reference range:

The expected values/ reference intervals for cholinesterase were provided from literature as follows:

Children, men, women aged 40 and over:
5320-12920 U/L

Women age 16 to 39 that are not pregnant or using hormonal contraceptives:
4260-11250

Women age 18 to 41 that are pregnant or taking contraceptives:
3650-9120 U/L.

Reference:

denBlauween DH et al.. Cholinesterase (EC 3.1.1.8) with Butyrulthiocholine-iodide as substrate: reference values with dependence on age and sex, and with regards to hormonal influence and pregnancy. J. Clin Chem Clin Biochem 1983; 21:381-386.

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

{6}

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda.innolitics.com/device/K061093](https://fda.innolitics.com/device/K061093)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
