FEATHERLITE AQUA TOCODYNAMOMETER

K061044 · Ventrex, Inc. · HFM · Jul 12, 2006 · Obstetrics/Gynecology

Device Facts

Record IDK061044
Device NameFEATHERLITE AQUA TOCODYNAMOMETER
ApplicantVentrex, Inc.
Product CodeHFM · Obstetrics/Gynecology
Decision DateJul 12, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.2720
Device ClassClass 2

Indications for Use

The FeatherLite Aqua™Toco dynamometer is intended for use in conjunction with telemetry systems for standard fetal monitors for the evaluation of external uterine activity during antepartum and intrapartum periods. Is should only be used in a clinical setting. It may be fully submerged in water for water birthing or monitoring in a whirlpool. Caution: FeatherLite Aqua™Toco is not intended for home monitoring of pre-term labor. Placement over the fundal area of the uterus to the left or right of the midline is suitable for most patients, It is held on the surface of the abdomen using a standard button hole or Velcro™ type belt. The product is supplied non-sterile and is intended for single use only.

Device Story

External tocodynamometer (toco) for fetal monitoring; uses flexible nipple and silicone gel to transmit uterine wall muscle tone changes to internal resistive strain gauge sensor; held against abdomen via elastic belt. Input: mechanical pressure from uterine contractions. Output: electrical signal representing uterine activity. Used in clinical settings (including water birth/whirlpools) with telemetry systems; operated by clinicians. Includes adapter with digital filtering to minimize noise and gain switch for monitoring obese patients. Single-use device; non-sterile. Provides continuous uterine activity data to fetal monitor to assist clinicians in assessing labor progress and fetal well-being.

Clinical Evidence

No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to predicate devices.

Technological Characteristics

Resistive strain gauge sensor; rigid thermoplastic housing; silicone gel/nipple interface. Cable-based atmospheric venting. Adapter-integrated digital filtering and gain switch. Single-use; non-sterile. Designed for water submersion. Connects to telemetry-based fetal monitors.

Indications for Use

Indicated for evaluation of external uterine activity during antepartum and intrapartum periods in clinical settings. Suitable for use in water birthing or whirlpool monitoring. Not indicated for home monitoring or pre-term labor.

Regulatory Classification

Identification

An external uterine contraction monitor (i.e., the tokodynamometer) is a device used to monitor the progress of labor. It measures the duration, frequency, and relative pressure of uterine contractions with a transducer strapped to the maternal abdomen. This generic type of device may include an external pressure transducer, support straps, and other patient and equipment supports.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for Ventrex, Inc. The logo features a stylized "V" made of diamond shapes, followed by the word "Ventrex" in a bold, sans-serif font with a line above it. Below the company name is the tagline "Vision in Design and Manufacturing" in a smaller font. JUL 1 2 2006 # 510(K) SUMMARY FeatherLite™Aqua Tocodynamometer | Date: | April 10, 2006 | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer: | Ventrex Inc.<br>3007 Bunsen Avenue, Suite K<br>Ventura,CA 93003<br>FDA Registration # 2029412 | | Owner/Operator: | Ventrex Inc.<br>3007 Bunsen Avenue, Suite K<br>Ventura,CA 93003 | | Contact Person: | George Austria<br>Director of Quality Assurance<br>Ventrex, Inc.<br>3007 Bunsen Avenue, Suite K<br>Ventura,CA 93003 | | Device Trade Name: | FeatherLite Aqua™Tocodynamometer | | Common Name: | Toco | | Classification Name: | Class II, 85HFM, External Uterine Contraction<br>Monitor and Accessories | | Regulatory Reference: | 884.2720 | | Predicate Device: | FeatherLite™Tocodynamometer under 510k#<br>K013477 and Corometrics 2264 under<br>K982651. | | Description and<br>Intended Use: | The FeatherLite Aqua™ tocodynamometer<br>(toco) is intended for use in conjunction with<br>telemetry systems for standard fetal monitors<br>for the evaluation of external uterine activity<br>during antepartum and intrapartum periods in a<br>clinical setting. The fetal monitor is not<br>intended for home monitoring or pre-term<br>labor. It can be used in both dry and fully<br>submerged environments It | {1}------------------------------------------------ is held in place on the surface of the abdomen using an elastic belt. It is sold non-sterile and is intended for single use only. ## Summary of Similarities and Differences: ### Similarities: Ventrex FeatherLite™ Toco under 510(k) K013477 (Predicate Device) and FeatherLite Aqua™ Toco (New Device): - Both are used in conjunction with fetal monitors for evaluation of external . uterine activity. - Same performance characteristics. . - Both use a flexible nipple and silicone gel to transmit the changes in uterine . wall muscle tone to the strain gauge pressure sensor housed inside. - The sensor for both FeatherLite™ and FeatherLite Aqua™ are the same. . (resistive type) - . The housing materials are the same . (molded from rigid thermoplastics) - The same attachment methods. - ◆ Both devices cannot be serviced. - Both use an adaptor to connect the universal connector of the cable to the . specific fetal monitor. ### Differences: - Atmospheric pressure referencing on the predicate device is achieved through . a vent hole on top of the housing while the venting on new device is achieved through a lumen in the cable. - The FeatherLite Aqua™ Toco may be used in a water environment. . - The cover is sealed to the housing using a water proof sealant for the new . device to prevent water from entering. - The adapter includes a digital filter to minimize unwanted noise without . interfering with performance. - The adapter has a switch to allow the user to increase the toco gain for more . effective monitoring of obese patients. - The FeatherLite Aqua™ is intended for single use only. . - The FeatherLite Aqua™ shall ONLY be used with telemetry units. . # Similarities: GE/Corometrics 2264 Tocodynamometer under K982651. - Same principle of sensing the change in muscle tone of the uterus as it . contracts by firmly holding a nipple-like appendage against the abdominal wall utilizing an elastic belt as the holding means. - . Both devices can be used in either water or non-water environments. - Both devices require a telemetry unit when used as a submersible device. . {2}------------------------------------------------ ## Differences: - Monitor connector is hard wired onto the cable of the predicate device. . - Predicate device has an electronic filtering circuitry in the toco. New device has . in the adaptor. ## CONCLUSION: - The proposed FeatherLite Aqua™ (New Device) is comparable to its predicate . devices in safety and effectiveness. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an eagle. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 JUL 1 2 2006 Mr. George Austria Director of Quality Assurance Ventrex, Inc. 3007 Bunsen Avenue, Suite K VENTURA CA 93003 Re: K061044 Trade/Device Name: FeatherLite Aqua™ Tocodynamometer Regulation Number: 21 CFR 884.2720 Regulation Name: External uterine contraction monitor and accessories Regulatory Class: II Product Code: HFM Dated: April 10, 2006 Received: April 17, 2006 #### Dear Mr. Austria: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Booken o o(t) for device is substantially equivalent (for the indications for relevelocu above and nave acterimined warketed predicate devices marketed in interstate commerce use stated in the encreater to regally man of the Medical Device Amendments, or to devices that prov to way 20, 1970, and characterity the provisions of the Federal Food, Drug, and Cosmetic have bech reclaismed in accordances inst a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, mative all device, bagon weequirements for annual registration, listing of devices, good controls provibitions or und prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket If your devise is exactive (500 are hadditional controls. Existing major regulations affecting your Apployally, it the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/3/Picture/10 description: The image shows a logo with the text "FDA Centennial" and the years "1906-2006". The logo is circular and contains stars. The image also contains the words "Protecting and Promoting Public Health" in a decorative font. {4}------------------------------------------------ Page 2 - Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |----------------|----------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Nancy C.Bogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications For Use Statement Ventrex, Inc. Applicant: 510(k) Number: K061044 Device Name: FeatherLite Aqua™ #### Indication For Use The FeatherLite Aqua™Toco dynamometer is intended for use in conjunction with telemetry systems for standard fetal monitors for the evaluation of external uterine activity during antepartum and intrapartum periods. Is should only be used in a clinical setting. It may be fully submerged in water for water birthing or monitoring in a whirlpool. Caution: FeatherLite Aqua™Toco is not intended for home monitoring of pre-term labor. Placement over the fundal area of the uterus to the left or right of the midline is suitable for most patients, It is held on the surface of the abdomen using a standard button hole or Velcro™ type belt. The product is supplied non-sterile and is intended for single use only. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODF) Prescription Use (Per 21 CFR 801.109) √ ଠାର Over-The Counter Use _ David R. Leggmm (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number. K061044
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...