K061028 · Ge Medical Systems, LLC · JAA · Jun 7, 2006 · Radiology
Device Facts
Record ID
K061028
Device Name
PRECISION THUNIS 800+
Applicant
Ge Medical Systems, LLC
Product Code
JAA · Radiology
Decision Date
Jun 7, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1650
Device Class
Class 2
Indications for Use
The Precision THUNIS 800+ is a digital remote R&F system intended to generate fluoroscopic and spot-film images of the patient during diagnostic procedures. The system is indicated also for use in generating radiographic images of human anatomy in general purpose diagnostic procedures. The system is not intended for mammographic or dental applications.
Device Story
Precision THUNIS 800+ is a stationary remote radiographic and fluoroscopic (R&F) system. Inputs include X-ray radiation passed through patient anatomy; captured by 9-inch image intensifier with CCD video camera. System transforms X-ray signals into digital images via frame grabber. Components include tilting patient support table, high-voltage generator, X-ray tube, beam limiter, and control console with workstation/display. Operated by clinicians in protected control room; optional in-room controls available for specific procedures. Provides fluoroscopic, spot-film, and radiographic images for diagnostic visualization. Benefits include remote operation for reduced radiation exposure to staff and digital image acquisition replacing traditional film changers.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Stationary remote R&F system. Components: tilting table, high-voltage generator, X-ray tube, 9-inch image intensifier, CCD camera, digital frame grabber. Energy: 3-Phase Mains (380/400/415/440/480Vac, 50/60Hz). Standards: UL60601-1:2003, IEC60601 series. No patient-contact materials. Digital image acquisition via frame grabber.
Indications for Use
Indicated for general purpose diagnostic radiographic and fluoroscopic imaging of human anatomy. Not for mammographic or dental applications.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized font, enclosed within a circular frame. The letters and the frame are black, while the background is white, creating a high-contrast design.
Kobe 1028
Page 1 of 2
## GE Healthcare
## 510(k) Summary
JUN - 7 2006
This 510(k) summary is prepared in accordance with the requirements of 21 CFR Part 807.87(h) and Part 807.92.
#### (1) Identification of Submitter:
| Submitter:<br>Address:<br>Tel.<br>Contact Person:<br>Summary Prepared: | GE Healthcare<br>3000 N. Grandview Blvd. Waukesha, WI 53118<br>(262) 544-3894<br>Larry A. Kroger, Ph.D. (Senior Regulatory Programs Manager)<br>04/10/2006 |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| (2) Information of Device: | |
| GE Model Name:<br>Generic Name:<br>Classification Name:<br>Device Class:<br>Product Code: | Precision THUNIS 800+<br>Stationary Diagnostic Radiographic and Fluoroscopic X-ray System (Remote)<br>Image-intensified fluoroscopic x-ray system (21CFR § 892.1650)<br>Class II<br>JAA |
| (3) Predicate Device: | |
| 510(k) Number: | K043805 |
| 510(k) Number: | K943805 |
|----------------------|--------------------------------------------------------------------------|
| GE Model Name: | Prestige V, VH & VHD |
| Generic Name: | Stationary Diagnostic Radiographic and Fluoroscopic X-ray System (Remote |
| Classification Name: | Image-intensified fluoroscopic x-ray system (21CFR § 892.1650) |
| Device Class: | Class II |
| Product Code: | JAA |
#### (4) Device Description and Specification
Precision THUNIS 800+ is a Radiographic and Fluoroscopic system which consists of a tilting patient support table with both local and remote motion controls, a High-Voltage generator, a touch screen X-ray control console in control room (protected area), a X-ray tube assembly, a beam limiting device, a 9" inch Image Intensifier with Charge Coupled Device (CCD) video Camera, a removable 14"X17" film cassette tray, a Power Distribution Unit for whole system, an integrated control console including a workstation, Image Display, remote controls of tilting patient support table, collimator controls, spot film control and Fluoroscopic exposure control.
An in-room image display and in-room Fluoroscopic control foot switch are provided as an option for certain procedure which may need local X-ray loading and imaging.
No traditional spot film device using a film changer is provided with this system, spot-film image acquisition is fulfilled through digital image frame grabber and video camera.
Precision THUNIS 800+
SECTION 5
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a handwritten text that appears to be a signature or a short note. The text is written in a cursive style, with some characters overlapping and connecting to each other. The writing is in black ink on a white background, and the overall impression is that of a quick, informal inscription.
For an illustration of the whole system and detailed product structure, please refer to Size titled "Device Description".
#### (5) Statement of Intended Use
The Precision THUNIS 800+ is a digital remote R&F system intended to generate fluoroscopic and spot-film images of the patient during diagnostic procedures.
The system is indicated also for use in generating radiographic images of human anatcmy in general purpose diagnostic procedures.
The system is not intended for mammographic or dental applications.
See SECTION 4 for a formal declaration of "Indications for Use".
See SECTION 12 for a comparison of Intended uses with the identified predicate device.
#### (6) Technological Characteristics Comparison
Precision THUNIS 800+ uses the same technological characteristics as the predicate device. Below is a brief summary.
- Precision THUNIS 800+ use the same X-ray imaging technology, same kind of film -Design: screen cassettes for Radiographic applications, same kind of video camera (Charge Coupled Device) and Image Intensifier for Fluoroscopic applications.
- All construction materials are compliant with latest UL60601-1:2003 and applicacle Material: IEC60601 series standards which are comparable with the old UL 187 and IEC standards.
#### Chemical Composition:
Chemical composition of materials is not critical for this kind of device, because no material is implanted or inserted into the patient body or orifice to contact body tissues or fluids. Patient contact is not required for proper functioning of the device. See SECTION 15 for a statement of patient support tabletop material.
Energy Source: Precision THUNIS 800+ system uses the same kind of energy source as its predicate device: 3-Phase Mains Supply with Grounding, rated voltages can be set as any one of 380/400/415/440/480Vac, 50/60Hz.
#### Conclusions:
GE Healthcare considers the Precision THUNIS 800+ to be substantially equivalent to the identified predicate device which has the same indications for use and meets the similar standards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
### JUN -7 2006
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems, LLC 3000 N. Grandview Blvd. W-440 WAUKESHA WI 53188
Re: K061028
Trade/Device Name: Precision THUNIS 800+ Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: April 10, 2006 Received: April 13, 2006
Dear Dr. Kroger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows the FDA Centennial logo. The logo is a circular seal with the words "1906-2006" at the top and "Centennial" in the center. The words "Protecting and Promoting Public Health" are written in a decorative font below the logo. The logo is black and white.
{3}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) 240-276-0115 | |
|-----------------|------------------------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology). | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, which has decorative swirls or flourishes at various points around the circumference.
# GE Healthcare
## Indications For Use Statement
510(k) Number (if known):
K061028
Device Name:
Precision THUNIS 800+
Indications for Use:
The Precision THUNIS 800+ is a digital remote R&F system intended to generate fluoroscopic and spot-film images of the patient during diagnostic procedures.
The system is indicated also for use in generating radiographic images of human anatomy in general purpose diagnostic procedures.
The system is not intended for mammographic or dental applications.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Brogdon
(Division Sign-Off))
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K061028
Precision THUNIS 800+
SECTION 4
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.