← Product Code [IYN](/productcode/IYN) · K060993

# GE LOGIQ P5 AND A5 (K060993)

_General Electric Co. · IYN · May 3, 2006 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060993

## Device Facts

- **Applicant:** General Electric Co.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** May 3, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: [Table of clinical applications including Fetal/Obstetrics, Abdominal, Pediatric, Small Organ, Neonatal/Adult Cephalic, Cardiac, Peripheral Vascular, Musculoskeletal, Urology, Transesophageal, Transrectal, Transvaginal, and Intraoperative].

## Device Story

GE LOGIQ P5/A5 is a mobile, general-purpose diagnostic ultrasound system. It acquires, processes, and displays ultrasound data via a console with a keyboard, specialized controls, and LCD display. It utilizes various transducers to perform B-mode, M-mode, PW/CW Doppler, Color Doppler, Power Doppler, and harmonic imaging. Operated by qualified physicians in clinical settings, the system provides real-time visualization of anatomy and blood flow. Output is viewed on the integrated display to assist in clinical diagnosis and patient management. The device is a modification of the GE LOGIQ 5 PRO, incorporating features from other GE systems to improve portability and convenience.

## Clinical Evidence

No clinical data required. Substantial equivalence is supported by non-clinical bench testing, including acoustic output measurements, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety testing. The device conforms to 21 CFR 820, ISO 9001, and ISO 13485 quality systems.

## Technological Characteristics

Mobile console (49x64x135-141 cm) with LCD display. Supports multiple transducer types (e.g., 3.5C, 8L, E8C, 4D3C). Modes: B, M, PW/CW Doppler, Color/Power Doppler, Harmonic Imaging, Coded Pulse. Connectivity: Digital acquisition/display. Sterilization: Compatible with standard cleaning/disinfection protocols. Software: Embedded system for image processing and control.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- GE LOGIQ 5 PRO (K03297)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K0604923

MAY 3 2006

Attachment B:

Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).

Image /page/0/Picture/5 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular shape. The letters are stylized and appear to be hand-drawn. The logo is black and white.

GE Healthcare

General Electric Company P.O. Box 414, Milwaukee, WI 53201

## Section a):

| 1. | Submitter:       | GE Healthcare<br>PO Box 414<br>Milwaukee, WI 53201                                                               |
|----|------------------|------------------------------------------------------------------------------------------------------------------|
|    | Contact Person:  | Allen Schuh,<br>Manager, Safety and Regulatory Engineering<br>Telephone: 414-721-3992; Fax: 414-721-3868         |
|    | Date Prepared:   | April 7, 2006                                                                                                    |
| 2. | Device Name:     | GE LOGIQ P5/A5 Diagnostic Ultrasound System<br>Ultrasonic Pulsed Doppler Imaging System, 21 CFR 892.1550, 90-IYN |
| 3. | Marketed Device: | GE LOGIQ 5 PRO diagnostic ultrasound system, 510(k) Numbers K03297-                                              |

currently in commercial distribution.

4. Device Description: The GE LOGIQ P5 & A5 are full-featured, general-purpose diagnostic ultrasound systems consisting of a mobile console approximately 49 cm wide, 64 cm deep and 135-141 cm high that provide digital acquisition, processing and display capability. The user interface includes a computer keyboard, specialized controls and color video LCD display. The modification combines features of the unmodified system with additional features from other systems currently marketed by GE Healthcare to provide users with a smaller, more convenient ultrasound system.

5. Indications for Use: The device is intended for use by a qualified physician for the evaluation of soft tissue and blood flow in the clinical applications: Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculoskeletal Conventional and Superficial; Urology (including prostate); Transesophageal (TE); Transrectal (TR); Transvaginal (TV); and Intraoperative (abdominal, thoracic, vascular and neurological).

6. Comparison with Predicate Device: The GE LOGIQ P5/A5 are of a comparable type and substantially equivalent to the current GE LOGIQ 5 PRO. They have the same technological characteristics, are comparable in key safety and effectiveness features, utilize similar design, construction, and materials. and have the same intended uses and basic operating modes as the predicate device.

### Section b):

1. Non-clinical Tests: The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.

- 2. Clinical Tests: None required.
3. Conclusion: Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms to 21 CFR 820, ISO 9001 and 13485 quality systems. The device conforms to applicable medical device safety standards and compliance is verified through independent evaluation with ongoing factory surveillance. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE LOGIO P5/A5 Diagnostic Ultrasound are substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

#### Public Health Service

#### MAY 3 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Allen Schuh Manager, GE Ultrasound Safety and Regulatory Engineering General Electric Company GE Medical Systems, Ultrasound and Primary Care Diagnostics, LLC 9900 Innovation Drive WEST MILWAUKEE WI 53219

Re: K060993

Trade Name: GE LOGIQ P5/A5 Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYN, IYO, and ITX Dated: April 7, 2006 Received: April 11, 2006

Dear Mr. Schuh:

· We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the GE LOGIQ P5/A5 Ultrasound System, as described in your premarket notification:

Image /page/1/Picture/11 description: The image is a circular logo for the FDA Centennial, celebrating 100 years from 1906-2006. The letters "FDA" are prominently displayed in the center of the logo. Below the letters, the word "Centennial" is written in a smaller font. Three stars are arranged beneath the word "Centennial."

Protesting and Promoting Public Health

{2}------------------------------------------------

## Transducer Model Number

| 3.5C         | ERB Biplane  | રક    |
|--------------|--------------|-------|
| 3.5CS        | i739 or t739 | ત્વાદ |
| বুটি         | 8L           | 8T    |
| રુદ્દ        | तेनि         | 4D3C  |
| 8C           | 10L          | 2D    |
| E8C          | 11L          | 6D    |
| E8Cs         | 12L          |       |
| BE9C Biplane | 35           |       |

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Sophie Paquerault at (301) 594-1212.

Sincerely yours,

David A. Segerson

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ P5/A5 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |         |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|---------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other** |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |         |
| Fetal / Obstetrics                                 | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              | N       |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              | N       |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              | N       |
| Small Organ [2]                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |         |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |         |
| Adult Cephalic                                     | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |         |
| Cardiac[3]                                         | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |         |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |         |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |         |
| Musculo-skeletal Superficial                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              |         |
| Other[4]                                           | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   | P              | N       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |         |
| Transesophageal                                    | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |         |
| Transrectal                                        | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |         |
| Transvaginal                                       | P                 | P | P             | P             | P                | P                  | P                | P                 | P                   |                |         |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |         |
| Intraoperative[5]                                  | P                 | P | P             | P             | P                | P                  | P                | P                 | N                   | P              |         |
| Intraoperative Neurological                        | P                 | P | P             | P             | N                | P                  | P                | P                 | N                   | P              |         |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |         |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |         |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic

[2] Small organ includes breast, testes, and thyroid.

[3] Cardiac is Adult and Pediatric.

[4] Other use includes Urology/Prostate

[5] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV).

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

[**] Other mode is 4D / Realtime 3D

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David G. Ingram

Prescription User (Per 21 CFR 801.109)

{5}------------------------------------------------

ﺴﻪ، ﻟ

Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ P5/A5 with 3.5C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                            |         |                |       |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|----------------------------|---------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined Harmonic<br>Modes | Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                | P                  | P                | P                          | P       | P              |       |  |
| Abdominal[1]                                       | P                 | P | P             |               | P                | P                  | P                | P                          | P       | P              |       |  |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Small Organ (specify)                              | P                 | P | P             |               | P                | P                  | P                | P                          | P       | P              |       |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                          | P       | P              |       |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Other[4]                                           | P                 | P | P             |               | P                | P                  | P                | P                          | P       | P              |       |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                            |         |                |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Benson

(Division Sign-Off) Division of Reproductive, A and Radiological Devices 510(k) Number

Prescription User (Per 21 CFR 801.109)

{6}------------------------------------------------

ﻟﻌﺪ ﺗ

Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ P5/A5 with 3.5CS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Small Organ (specify)                              | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                           | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel R. Hanson

(Division Sign-Off Division of Reproductive, I and Radiological Device 510(k) Number

Prescription User (Per 21 CFR 801.109)

{7}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ P5/A5 with 4C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal / Obstetrics                                 | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |       |  |
| Abdominal[1]                                       | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |       |  |
| Pediatric                                          | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |       |  |
| Small Organ (specify)                              |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Peripheral Vascular                                | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Other[4]                                           | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| l aparoscopic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN;

[4] Other use includes Urology;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Seligman

(Division Sign-Off) Division of Reproductive, Abde and Radiological Devices 510(k) Number

Prescription User (Per 21 CFR 801.109)

E-ર

{8}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ P5/A5 with 5Cs Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                     | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|
| Clinical Application<br>Anatomy/ Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse |
| Ophthalmic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Fetal / Obstetrics                                  | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Abdominal[1]                                        | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Pediatric                                           | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Small Organ (specify)                               |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Neonatal Cephalic                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Adult Cephalic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Cardiac                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Peripheral Vascular                                 | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Musculo-skeletal Conventional                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Musculo-skeletal Superficial                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Other [4]                                           | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              |
| Exam Type, Means of Access                          |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transesophageal                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transrectal                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transvaginal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Transuretheral                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative (specify)                            |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intraoperative Neurological                         |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Intravascular                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |
| Laparoscopic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

[4] Other use includes Urology/Prostate;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Ingram

(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number _

Prescription User (Per 21 CFR 801.109)

{9}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ P5/A5 with 8C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal[1]                                       | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Small Organ (specify)                              | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Neonatal Cephalic                                  | P                 | P | P             |               | P                | P                  | P                | P                 | P                   | P              |       |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Ingram

(Division Sign-Off) Division of Reproductive, Abdomit and Radiological Devices 510(k) Number .

Prescription User (Per 21 CFR 801.109)

{10}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ P5/A5 with E8C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow anaiysis of the human body as follows:

| Clinical Application<br>Anatomy/ Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
|-----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Ophthalmic                                          |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                  | P | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Abdominal[1]                                        | P | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Pediatric                                           |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Small Organ (specify)                               |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                   |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                             |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                 |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other[4]                                            | P | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Exam Type, Means of Access                          |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                     |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                         | P | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Transvaginal                                        | P | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |
| Transuretheral                                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                            |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                         |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

.

[4] Other use includes Urology/Prostate;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Stegmann

(Division Sign-Off) (Division Sign of Reproductive, Al Division of Hoprocess
and Radiological Devices 510(k) Number -

Prescription User (Per 21 CFR 801.109)

{11}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

## GE LOGIQ P5/A5 with E8Cs Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/ Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
|-----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Ophthalmic                                          |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                  | N | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Abdominal (1)                                       | N | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Pediatric                                           |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Small Organ (specify)                               |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                   |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac                                             |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                 |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other [4]                                           | N | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Exam Type, Means of Access                          |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                     |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                         | N | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Transvaginal                                        | N | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Transuretheral                                      |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                            |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                         |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                       |   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                        |   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

[4] Other use includes Urology/Prostate;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David A. Leggett

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number _

Prescription User (Per 21 CFR 801.109)

{12}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ P5/A5 with BE9C Biplane Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                     | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |  |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/ Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal / Obstetrics                                  | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Abdominal                                           | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Pediatric                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Small Organ (specify)                               |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Neonatal Cephalic                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Cardiac                                             |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Peripheral Vascular                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Conventional                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Musculo-skeletal Superficial                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Other (specify)                                     | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Exam Type, Means of Access                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                         | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Transvaginal                                        | P                 | P | P             |               | P                |                    | P                | P                 | P                   | P              |       |  |
| Transuretheral                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative Neurological                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intravascular                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes GYN/Pelvic;

[4] Other use includes Urology/Prostate;

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Ingram

---

(Division Sign-Off)

Division of Reproductive, Ab and Radiological Devices 510(k) Number_

Prescription User (Per 21 CFR 801.109)

{13}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ P5/A5 with ERB Biplane Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                     | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|
| Clinical Application<br>Anatomy/ Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |
| Ophthalmic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Fetal / Obstetrics                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Abdominal                                           | N                 | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Pediatric                                           |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Small Organ (specify)                               |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Neonatal Cephalic                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Adult Cephalic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Cardiac[3]                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Peripheral Vascular                                 |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Conventional                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Musculo-skeletal Superficial                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Other (specify)                                     | N                 | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Exam Type, Means of Access                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transesophageal                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transrectal                                         | N                 | N | N             |               | N                |                    | N                | N                 | N                   | N              |       |
| Transvaginal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Transuretheral                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative (specify)                            |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intraoperative Neurological                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Intravascular                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |
| Laparoscopic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |

N = new indication; P = previously cleared by FDA; E = added under Appendix E Notes: [3] Cardiac is Adult and Pediatric.

1

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel R. Slayton

(Division Sign-Off) Division of Reproductive, Abdom and Radiological Devices 510(k) Number____

Prescription User (Per 21 CFR 801.109)

{14}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

# GE LOGIQ P5/A5 with i739 or t739 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                     | Mode of Operation |   |               |               |                  |                    |                  |                   |                     |                |       |  |
|-----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|-------------------|---------------------|----------------|-------|--|
| Clinical Application<br>Anatomy/ Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes | Harmonic<br>Imaging | Coded<br>Pulse | Other |  |
| Ophthalmic                                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Fetal / Obstetrics                                  |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Abdominal                                           | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Pediatric                                           | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Small Organ (specify)                               | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Neonatal Cephalic                                   |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Adult Cephalic                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Cardiac [3]                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Peripheral Vascular                                 | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Musculo-skeletal Conventional                       | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Musculo-skeletal Superficial                        | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Other (specify)                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Exam Type, Means of Access                          |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transesophageal                                     |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transrectal                                         |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transvaginal                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Transuretheral                                      |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Intraoperative (specify)                            | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Intraoperative Neurological                         | N                 | N | N             | N             | N                | N                  | N                | N                 | N                   | N              | N     |  |
| Intravascular                                       |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |
| Laparoscopic                                        |                   |   |               |               |                  |                    |                  |                   |                     |                |       |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid.

[5] Intraoperative includes abdominal, thoracic, and vascular.

[*] Combined modes are B/M, B/PWD, B/Color/PWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Legnon

(Division Sign-Off) Division of Reproductive, Abdo…

---

**Source:** [https://fda.innolitics.com/device/K060993](https://fda.innolitics.com/device/K060993)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
