SPACELABS MEDICAL BISPECTRAL INDEX (BISX) ANALYSIS MODULE 91482 AND ACCESSORIES MODEL-91842
K060900 · Spacelabs Medical, Inc. · OLW · Jun 23, 2006 · Neurology
Device Facts
Record ID
K060900
Device Name
SPACELABS MEDICAL BISPECTRAL INDEX (BISX) ANALYSIS MODULE 91482 AND ACCESSORIES MODEL-91842
Applicant
Spacelabs Medical, Inc.
Product Code
OLW · Neurology
Decision Date
Jun 23, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Intended Use
The Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482 is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals. The Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482 may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
Device Story
The BISx Module 91482 is a microprocessor-based, two-channel EEG monitor used in hospitals to assess brain state. It acquires EEG signals via disposable sensors, a patient interface cable, and a BISx pod (manufactured by Aspect Medical Systems). The module processes these signals to monitor the effects of anesthetic agents. Output is displayed on the Spacelabs Medical Ultraview monitor for healthcare providers. By providing real-time data on the depth of anesthesia, the device assists clinicians in guiding anesthetic administration, potentially reducing the incidence of intraoperative awareness with recall in adult patients.
Clinical Evidence
No clinical data provided in the summary; substantial equivalence is based on technological characteristics and intended use consistent with existing BIS monitoring devices.
Technological Characteristics
Microprocessor-based, two-channel EEG monitor; includes BISx pod, patient interface cable, and disposable sensors. Integrates with Spacelabs Medical Ultraview system for display. Connectivity via external serial data connections.
Indications for Use
Indicated for adult and pediatric patients in hospital/medical settings to monitor brain state via EEG signal acquisition; used as an aid in monitoring anesthetic agent effects and reducing incidence of awareness with recall during general anesthesia and sedation.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Related Devices
K040183 — BISX · Aspect Medical Systems, Inc. · Feb 25, 2004
K031694 — MODIFICATION TO EEG MONITOR WITH BIS, MODEL A-2000 · Aspect Medical Systems, Inc. · Aug 29, 2003
K012466 — BIS/EEG MODULE · General Electric Medical Systems Information Techn · Oct 1, 2001
K973596 — SPACELABS MEDICAL 90482 BISPECTRAL INDEX ANALYSIS (BIS) MODULE · Spacelabs Medical, Inc. · Feb 3, 1998
K052981 — BIS EEG MONITOR, MODEL BISX4 · Aspect Medical Systems, Inc. · Nov 1, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Premarket Notification K060900 Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482 and Accessories Summary of Safety and Effectiveness
May 15, 2006
The SIO(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 1992.
| Subject: | 510(k) Summary of Safety and Effectiveness Information for the Spacelabs<br>Medical Bispectral Index (BISx) Analysis Module 91482 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Medical, Inc.<br>PO Box 7018<br>Issaquah, WA 98029 |
| | David J. Geraghty<br>Phone: 1 425 657 7200<br>Fax: 1 425 657 7207<br>david.geraghty@slmd.com |
| Proprietary<br>Name: | Spacelabs Medical Bispectral Index (BISx) Analysis Module, model 91482 |
| Common<br>Name and<br>Classification: | Electroencephalograph (EEG) Monitor<br>(84 GWQ, §882.1400, Class II) |
| Device<br>Description: | The Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482<br>(BISx Module 91482) is an easy-to-use, slim, single module in the Spacelabs<br>Medical family of Spacelabs Medical Ultraview System (Ultraview) modules.<br>The BISx Module 91482 is a microprocessor based, two-channel EEG unit<br>designed for use on adult and pediatric patients within a hospital or medical<br>facility. Its system configuration includes the BISx Module 91482 with<br>connectors for external serial data connections, a BISx pod, a patient interface |
cable, disposable sensors, and printer options. The BISx pod, patient interface cable and disposable sensors are manufactured by Aspect Medical Systems, and distributed by Spacelabs Medical for use with the BISx Module 91482, The Spacelabs Medical Ultraview monitor provides the display capabilities for the
BISx Summary of Safety and Effectiveness Addendum to 510(k), K060900
care provider.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing segments, representing the department's mission to protect the health of all Americans and provide essential human services. The department's name is written in a circular fashion around the eagle. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Spacelabs Medical Inc. c/o Mr. David J. Geraghty Manager, Regulatory and Quality P.O. Box 7018 Issaquah, Washington 98027
APR - 9 2012
Re: K060900
Trade/Device Name: Spacelabs Medical Bispectral Index (BIS) Analysis Module 91482 and Accessories
Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: II Product Code: OLW, OMC, OLT, ORT Dated (Date on orig SE ltr): May 23, 2006 Received (Date on orig SE Itr): May 25, 2006
Dear Mr. Geraghty:
This letter corrects our substantially equivalent letter of June 23, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
文
Kesia Alexander
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications for Use
KOGO900
510(k) Number (if known):
Device Name:
Indications for Use:
Spacelabs Medical Bispectral Index (BIS) Analysis Module 91482, and Accessories
The Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482 is intended for use under the direct supervision of a licensed healthcare practitioner or by personnel trained in its proper use. It is intended for use on adult and pediatric patients within a hospital or medical facility providing patient care to monitor the state of the brain by data acquisition of EEG signals.
The Spacelabs Medical Bispectral Index (BISx) Analysis Module 91482 may be used as an aid in monitoring the effects of certain anesthetic agents. Use of BIS monitoring to help guide anesthetic administration may be associated with the reduction of the incidence of awareness with recall in adults during general anesthesia and sedation.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR Over-The-Counter Use
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH2 Office of Device Evaluation (ODE)
Helmut Werner
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Ko 60900 510(k) Number.
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.