MODIFICATION TO ONTRAK TESTCUP II AND ONSITE CUPKIT
K060896 · Varian, Inc. · DJC · Jun 9, 2006 · Clinical Toxicology
Device Facts
Record ID
K060896
Device Name
MODIFICATION TO ONTRAK TESTCUP II AND ONSITE CUPKIT
Applicant
Varian, Inc.
Product Code
DJC · Clinical Toxicology
Decision Date
Jun 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.3610
Device Class
Class 2
Indications for Use
TesTcup II and OnSite CupKit products are in vitro diagnostics tests intended for professional use for the qualitative detection of drug or drug metabolite in urine at or above the stated cutoff concentrations. Cutoff Concentrations: Amphetamines: 1000 ng/mL Benzodiazepines: 200 ng/mL Cocaine metabolite: 300 ng/mL Methamphetamine: 300 ng/mL Methamphetamine: 500 ng/mL Morphine: 300 ng/mL Morphine (M2K): 2000 ng/mL Phencyclidine (PCP): 25 ng/mL Tetrahydrocannabinols: 50 ng/mL
Device Story
In vitro diagnostic test for qualitative detection of drugs of abuse in urine; utilizes immunochromatographic assay with microparticle capture inhibition. Urine collected in test cup; drug profile card inserted into lid holder; capillary action draws urine into card. Competition occurs between drug in urine and immobilized drug conjugate on membrane; absence of drug allows antibody-coated microparticles to bind conjugate, forming blue band. Presence of drug inhibits binding, resulting in absence of blue band (preliminary positive). Used in professional settings; visually read by operator. Provides preliminary results requiring confirmation by specific alternate chemical method. Benefits include rapid, point-of-care screening for drug presence.
Clinical Evidence
Bench testing only. Method comparison study performed using 100 negative and 50 positive clinical urine specimens. Negative samples confirmed by GC/MS. Positive samples evaluated against GC/MS values, including samples near the 300 ng/mL cutoff (75-125%). Precision study conducted with 378 devices across 6 concentrations (0-450 ng/mL) over 3 days. Analytical specificity tested against various methamphetamine-related compounds to determine cross-reactivity thresholds.
Technological Characteristics
Microparticle capture inhibition assay. Components: urine collection cup, drug profile card, antibody-coated microparticles, drug conjugate membrane. Visual readout via blue band formation. Standalone, non-electronic, manual test.
Indications for Use
Indicated for professional use for qualitative detection of drugs/metabolites in human urine. Target analytes: amphetamines, benzodiazepines, cocaine metabolite, methamphetamine, morphine, phencyclidine (PCP), and tetrahydrocannabinols (THC) at specified cutoff concentrations. Provides preliminary results requiring confirmation by alternate chemical methods.
Regulatory Classification
Identification
A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.
Special Controls
*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Predicate Devices
OnTrak TesTcup® II and OnSite CupKit™ (k033902)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k060896
B. Purpose for Submission:
Clearance of a device for drugs of abuse with a new cutoff (300 ng/mL) for methamphetamine.
C. Measurand:
Methamphetamine (d-Methamphetamine), Amphetamines, Benzodiazepines, Cocaine metabolites, Morphine, Phencyclidine (PCP), and Tetrahydrocannabinols
D. Type of Test:
Qualitative, visually read, immunochromatographic test
E. Applicant:
Varian, Inc.
F. Proprietary and Established Names:
OnTrak TesTcup® II Pro 5-AS and OnSite CupKit™ Pro 5-AS
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.3610, Methamphetamine Test System
21 CFR § 862.3100, Amphetamine Test System
21 CFR § 862.3170, Benzodiazepine Test System
21 CFR § 862.3250, Cocaine and Cocaine Metabolite Test System
21 CFR § 862.3640, Morphine Test System
Unclassified, Enzyme Immunoassay, Phencyclidine
21 CFR § 862.3870, Cannabinoids Test System
2. Classification:
Class II
3. Product code:
DJC, DKZ, JXM, DIO, DNK, LCM, and LDJ respectively.
4. Panel:
Toxicology (91)
{1}
2
H. Intended Use:
1. Intended use(s):
TesTcup II and OnSite CupKit products are in vitro diagnostics tests intended for professional use for the qualitative detection of drug or drug metabolite in urine at or above the stated cutoff concentrations.
Cutoff Concentrations:
Amphetamines: 1000 ng/mL
Benzodiazepines: 200 ng/mL
Cocaine metabolite: 300 ng/mL
Methamphetamine: 300 ng/mL
Methamphetamine: 500 ng/mL
Morphine: 300 ng/mL
Morphine (M2K): 2000 ng/mL
Phencyclidine (PCP): 25 ng/mL
Tetrahydrocannabinols: 50 ng/mL
2. Indication(s) for use:
See Intended Use above.
3. Special conditions for use statement(s):
The device is for in vitro diagnostic prescription use. TesTcup II and OnSite CupKit products provide only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result.
4. Special instrument requirements:
Not applicable. The device is a visually read single-use device.
I. Device Description:
The OnTrak TesTcup II and OnSite Cupkit assays are in vitro diagnostic tests intended for professional use for the qualitative detection of amphetamines (d,l-amphetamine 1000 ng/mL), benzodiazepines (oxazepamn 200 ng/mL), cocaine metabolite (benzoylecgonine 300 ng/mL), methamphetamine (d-methamphetamine 500 ng/mL), morphine (morphine 300 ng/mL), morphine 2000 (morphine 2000 ng/mL), PCP (phencyclidine 25 ng/mL), and THC (1 1-nor-A9-THC-9-carboxylic acid 50 ng/mL) and were previously cleared under k033902. The OnTrak TesTcup II and OnSite Cupkit product lines have several different types that use varying combinations of the above drugs of abuse test strips.
This submission (k060896) is for the clearance of methamphetamine (d-methamphetamine) at a 300 ng/mL cut-off to be packaged with the drugs of abuse test strips above. The methamphetamine assay with a 300 ng/mL cutoff has exactly the same antibody and antigen as the previously cleared 500 ng/mL methamphetamine
{2}
assay. The label system (polystyrene beads), nitrocellulose membrane, control-line antibody, and sample pad are also identical between the two. The difference between the two assays is a slight difference in the typical volume of polystyrene bead labeled antibody and capture antigen.
# J. Substantial Equivalence Information:
1. Predicate device name(s): OnTrak TesTcup® II and OnSite CupKit™
2. Predicate 510(k) number(s): k033902
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Intended Use | TesTcup II and OnSite CupKit products are in vitro diagnostics tests intended for professional use for the qualitative detection of drug or drug metabolite in urine at or above the stated cutoff concentrations. | TesTcup II and OnSite CupKit products are in vitro diagnostics tests intended for professional use for the qualitative detection of drug or drug metabolite in urine at or above the stated cutoff concentrations. |
| Cutoff | Amphetamines: 1000 ng/mL Benzodiazepines: 200 ng/mL Cocaine metabolite: 300 ng/mL Methamphetamine: 500 ng/mL Morphine: 300 ng/mL Morphine (M2K): 2000 ng/mL Phencyclidine (PCP): 25 ng/mL Tetrahydrocannabinols: 50 ng/mL | Amphetamines: 1000 ng/mL Benzodiazepines: 200 ng/mL Cocaine metabolite: 300 ng/mL Methamphetamine: 500 ng/mL Morphine: 300 ng/mL Morphine (M2K): 2000 ng/mL Phencyclidine (PCP): 25 ng/mL Tetrahydrocannabinols: 50 ng/mL |
| Sample Types | Urine | Urine |
| Test Method | Immunochromatographic | Immunochromatographic |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Cutoff | Methamphetamine: 300 ng/mL included | Methamphetamine: 300 ng/mL not included |
{3}
K. Standard/Guidance Document Referenced (if applicable):
None were identified by the applicant.
L. Test Principle:
The assays are based on the principle of microparticle capture inhibition. The test relies on the competition between drug, which may be present in the urine being tested, and drug conjugate immobilized on a membrane in the test chamber. Urine is collected directly in the test cup provided. The drug profile card is placed in the samples by inserting it into the lid holder, then securing the lid onto the cup. Urine is drawn in the profile card by capillary action and reacts with antibody coated microparticles and drug conjugate present on the membrane. In the absence of drug, the antibody is free to interact with the drug conjugate, causing the formation of a blue band.
When drug is present in the specimen, it binds to the antibody-coated microparticles. If sufficient drug is present, the microparticles are inhibited from binding the drug conjugate and no blue band is formed at the result window. A preliminary positive ("non-negative" result is the absence of a blue band).
An additional antibody/antigen reaction occurs at the "VALID" area. The "VALID" blue band forms when antibodies, which are imbedded in the reagent membrane, interact with and bind to the antigen on the blue microparticles.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The precision of the assay for methamphetamine (300 ng/mL cut-off) was determined by testing 63 replicates across three different days with 6 different concentrations: 0, 75, 150, 225, 375, and 450 ng/mL (21 replicates for each concentration per day, for a total of 378 devices tested) using a Methamphetamine urine standard containing drug at different concentrations. Results were read by the three operators. Samples were randomized and blind with respect to the operators. Results are summarized below.
| Methamphetamine (ng/mL) Cutoff = 300 ng/mL | Precision Results (%) | | | |
| --- | --- | --- | --- | --- |
| | Lot 1 | | Lot 2 | |
| | + | - | + | - |
| 0 | 0 | 100 | 0 | 100 |
| 75 | 0 | 100 | 0 | 100 |
| 150 | 1 | 100 | 1.6 | 98.4 |
| 225 | 17.5 | 82.5 | 60.3 | 39.7 |
| 375 | 100 | 0 | 98.4 | 1.6 |
| 450 | 100 | 0 | 100 | 0 |
{4}
b. Linearity/assay reportable range: Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The sponsor did not indicate any degree of traceability for their devices. Control materials are required but are not specifically identified in the labeling. No calibrators are required. The device is calibrated during the manufacturing process.
d. Detection limit:
Sensitivity of qualitative assays may be characterized by validating performance around the claimed cutoff concentration of the assay, and demonstrating the lowest concentration of drug that is capable of or consistently producing a positive result. This information appears in the precision section, a., above.
e. Analytical specificity:
Each substance was tested starting with a 100,000 ng/mL solution or a 1000 ng/mL solution in a negative urine control. Each compound was diluted and tested negative result appeared for methamphetamine. Three OnSite CupKit devices were tested per substance by pipetting approximately 100 to 200 uL of spiked urine onto the appropriate feedhole. Devices were color rated according to the QC color rating scale. If at least 2 out of 3 negative results were seen for methamphetamine, then the sponsor defines this as no interference at that level. The entire protocol was repeated for each of the two lots. Results are summarized below.
| Methamphetamine-related Compounds | Minimum Concentration Required to give a Positive result (ng/mL) |
| --- | --- |
| d,l-Methamphetamine | 500 |
| 3,4-Methylenedioxymethamphetamine (MDMA) | 1000 |
| Propylhexedrine | 2000 |
| Fenfluramine | 2000 |
| I-Methamphetamine | 3000 |
| p-Hydroxymethamphetamine | 5000 |
| 3,4-Methylenedioxyethylamphetamine (MDEA) | 10,000 |
| Ranitidine | 25,000 |
| d-Amphetamine | 50,000 |
| I-Phenylephrine | 50,000 |
{5}
| Methamphetamine-related Compounds | Minimum Concentration Required to give a Positive result (ng/mL) |
| --- | --- |
| B-Phenethylamine | 50,000 |
| d,I-Amphetamine | 50,000 |
| d,I-Ephedrine | 100,000 |
| 3,4-Methylenedioxyamphetamine (MDA) | >100,000 |
| I-Amphetamine | >100,000 |
| I-Pseudoephedrine | >100,000 |
| d-Pseudoephedrine | >100,000 |
f. Assay cut-off:
See Intended Use above.
2. Comparison studies:
a. Method comparison with predicate device:
The methamphetamine strip within the TesTcup II /CupKit was evaluated in a SAMHSA certified laboratory using clinical specimens. The clinical negative samples were screened negative by automated immunoassay and reported as negatives according to SAMSHA guidelines. Clinical specimens screened positive by automated immunoassay were subsequently analyzed by GC/MS. Ten samples were diluted once with negative human urine to achieve the appropriate test range.
A total of 100 negative urine samples were tested, 10% of these specimens, chosen at random were confirmed negative by GC/MS. A total of 50 positive urine samples were tested. 10% of the samples were between 75% and 100% of the analyte cutoff; another 10% of the samples were between 100% and 125% of the cutoff. The remainders of the samples were at concentrations greater than 125% of the cutoff. Results are summarized below.
| TesTcup II/CupKit Methamphetamine Cutoff = 300 ng/mL | | Negative Samples | GC/MS Values ng/mL | | |
| --- | --- | --- | --- | --- | --- |
| | | | Near Cutoff | | >125% of Cutoff |
| | | | 75% - 100% | 100% - 125% | |
| | + | 0 | 0 | 3 | 39 |
| | - | 100 | 5 | 3 | 0 |
b. Matrix comparison:
Not applicable.
{6}
3. Clinical studies:
a. Clinical Sensitivity: Not applicable.
b. Clinical specificity: Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable.
4. Clinical cut-off: Not applicable.
5. Expected values/Reference range: No elicit drugs should be present in urine.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.