← Product Code [IZI](/productcode/IZI) · K060867

# SPY INTRAOPERATIVE IMAGING SYSTEM (K060867)

_Novadaq Technologies, Inc. · IZI · May 3, 2006 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060867

## Device Facts

- **Applicant:** Novadaq Technologies, Inc.
- **Product Code:** [IZI](/productcode/IZI.md)
- **Decision Date:** May 3, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1600
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

For use in intraoperative visual assessment of the coronary vasculature and bypass grafts during coronary artery bypass (CABG) surgery.

## Device Story

SPY Intraoperative Imaging System (SP2000) provides real-time visualization of coronary vasculature and bypass grafts during CABG surgery. System uses laser light source to illuminate heart surface; Indocyanine Green (ICG) dye injected intravenously via central venous line, bypass pump, or cardioplegia line. Laser light excites dye, causing infrared emission; CCD camera captures resulting fluorescent images. Surgeon views, records, and replays images intraoperatively to evaluate graft integrity and vascular perfusion. Device intended for use by surgeons in operating room. Provides visual feedback to assist in surgical decision-making regarding graft patency and vascular integrity, potentially improving surgical outcomes.

## Clinical Evidence

Safety and effectiveness supported by animal studies and human experience in 1,485 CABG procedures as of February 2006. Testing confirmed: laser exposure (35 mW/cm2) below ANSI MPE (327 mW/cm2); no thermal damage to heart tissue; no adverse changes to ECG or arterial pressures; no acute/long-term cellular, renal, or hepatic effects. System demonstrated capability to acquire high-quality images of entire vascular bed.

## Technological Characteristics

Fluorescent angiographic system. Components: laser light source, CCD camera. Connectivity: standalone. Standards: IEC 60601-1, UL2601-1 (electrical); IEC 60601-1-2 (EMC); 21 CFR 1040, ANSI Z136.1, ANSI Z136.3 (laser safety).

## Regulatory Identification

An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Novadaq SPY® Intraoperative Imaging System ([K042961](/device/K042961.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary

CONFIDENTIAL

3 2006 MAY

Common Name: Fluorescent Angiographic System

Classification: 21 CFR 892.1600

Manufacturer: Novadaq Technologies Inc. 2585 Skymark Avenue Suite 306 Mississauga, Ontario Canada L4W 4L5 905.629.3822 ext. 240

- Contact Name: Allison Manners Vice President - Regulatory and Clinical Affairs
### Legally Marketed Predicate Devices:

The Novadaq SPY® Intraoperative Imaging System had received FDA 510(k) clearance in January 2005 (#K042961).

#### Device Description:

The SPY Intraoperative Imaging System: SP2000 is indicated for use for intra-operative visual assessment of the coronary vasculature and grafts during CABG surgery.

The SPY System provides the surgeon with the capability to view record and replay fluorescent images of blood vessels and bypass grafts of the A laser light source is used to illuminate the heart surface. heart. Indocyanine Green (ICG) is injected intravenously through the central venous line, bypass pump or cardioplegia line and, while coursing through the coronary arteries and grafts, the absorption of laser light causes excitation of the dye followed by emission of infrared energy. The result is a fluorescent image of the coronary vasculature. A CCD camera captures the image. These images are used to evaluate the integrity of the coronary vasculature and bypass grafts.

No changes or modifications have been made to the device since the 510(k) clearance of this device, #K042961. This 510(k) submission entails a labeling revision that does not alter the device's intended use or alter the devices fundamental scientific technology.

{1}------------------------------------------------

Animal studies, human experience and in vitro testing were conducted to support the safe and effective use of the SPY® System.

### In Vitro Testing:

Testing of the SPY System was completed in conformance with the following standards. The SPY System successfully met all of the requirements for these standards.

- 1. Electrical per IEC 60601-1 and UL2601-1
- 2. Electromagnetic Compatibility per IEC 60601-1-2
- 3. Light Emitting Laser Products per 21 CFR 1040
- 4. Safe Use of Lasers in Health Care Facilities per ANSI Z136.3
- 5. American National Standard for Safe Use of Lasers per ANSI Z136.1

#### In Vivo Testina:

Animal studies and human experience in 1485 CABG procedures, as of February 2006, were completed to demonstrate safety and effectiveness.

- 1. The exposure for the SPY System is 35 mWcm2 which is far below the MPE of 327 mW/cm2 established by ANSI for exposure to the skin.
- 2. Use of the SPY System does not cause any thermal damage to the heart tissue.
- 3. There were no changes to the electrocardiograms or arterial pressures during and following SPY use.
- 4. There were no acute or long-term cellular effects of using the SPY System.
- 5. There were no acute or long-term renal or hepatic effects of using the SPY System.
- 6. The SPY System was able to acquire excellent images of the entire vascular bed on each aspect of the heart.

#### Conclusions:

The above testing demonstrates that the SPY Intra-operative Imaging System is safe and effective in imaging the coronary vasculature and bypass grafts intra-operatively and is equivalent to the predicate devices.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

3 2006 MAY

Ms. Allison Manners Vice President, Regulatory and Clinical Affairs Novadaq Technologies Inc. 2585 Skymark Avenue Suite 306 Mississauga, Ontario CANADA L4W 4L5

Re: K060867

Trade Name: SPY® Intraoperative Imaging System Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic X-Ray System Regulatory Class: Class II (two) Product Code: IZI Dated: March 29, 2006 Received: March 30, 2005

Dear Ms. Manners:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Ms. Allison Manners

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Afummar for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known):

K_060867

Device Name:

SPY® Intra-operative Imaging System: SP2000

Indications for Use:

ﻪ « « « « « « «

For use in intraoperative visual assessment of the coronary vasculature and bypass grafts during coronary artery bypass (CABG) surgery.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use_ (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

Blumuma

n Sign-C

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K060867](https://fda.innolitics.com/device/K060867)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
