The ONLINE TDM Tobramycin assay is for the quantitative determination of tobramycin in human serum or plasma on Roche automated clinical chemistry analyzers. Measurements obtained by this device are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy.
Device Story
ONLINE TDM Tobramycin is an in vitro diagnostic assay for quantitative determination of tobramycin in human serum or plasma. Used on Roche automated clinical chemistry analyzers (Hitachi 911, 912, 917, and Modular P). Operates by measuring drug concentration to assist clinicians in monitoring peak and trough levels; critical for preventing nephrotoxicity and hearing impairment associated with aminoglycoside therapy. Results enable dosage adjustment based on individual patient renal function and drug clearance. Intended for clinical laboratory use by trained personnel.
Clinical Evidence
No clinical trials were performed. Evidence consists of analytical bench testing, including precision (CLSI EP5-T2), linearity (0.33-10.0 ug/mL), analytical specificity (cross-reactivity and interference testing), and method comparison. Method comparison against a commercially available assay using 55 clinical samples showed a Passing-Bablok slope of 1.006-1.060 and intercept of 0.1572-0.2436. A second comparison (n=52) yielded y = 0.92x + 0.01, r = 0.983.
Technological Characteristics
In vitro diagnostic reagent kit for automated clinical chemistry analyzers. Quantitative assay for tobramycin in serum/plasma. Utilizes automated photometric/turbidimetric measurement principles standard to Roche Hitachi/Modular P platforms. No specific materials of construction or software algorithm architecture details provided beyond standard automated analyzer integration.
Indications for Use
Indicated for the quantitative measurement of tobramycin in human plasma and serum to aid in the diagnosis and treatment of tobramycin overdose and to monitor drug levels to ensure appropriate therapy in patients receiving aminoglycoside antibiotic treatment.
Regulatory Classification
Identification
A tobramycin test system is a device intended to measure tobramycin, an aminoglycoside antibiotic drug, in plasma and serum. Measurements obtained by this device are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy.
Predicate Devices
COBAS INTEGRA Tobramycin (k964457)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k060853
B. Purpose for Submission:
New device
C. Measurand:
Tobramycin
D. Type of Test:
Quantitative, homogeneous microparticles agglutination immunoassay
E. Applicant:
Roche Diagnostics Corp.
F. Proprietary and Established Names:
ONLINE TDM Tobramycin
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| Radioimmunoassay, Tobramycin (KLB) | Class II | 21 CFR 862.3900, Tobramycin test system. | 91 CLINICAL TOXICOLOGY (TX) |
H. Intended Use:
1. Intended use(s):
The ONLINE TDM Tobramycin assay is for the quantitative determination of tobramycin in human serum or plasma on Roche automated clinical chemistry analyzers.
2. Indication(s) for use:
The ONLINE TDM Tobramycin assay is for the quantitative determination of tobramycin in human serum or plasma on Roche automated clinical chemistry analyzers. Measurements obtained by this device are used in the diagnosis and treatment of tobramycin overdose and in monitoring levels of tobramycin to ensure appropriate therapy.
3. Special conditions for use statement(s):
{1}
For prescription use only
4. Special instrument requirements:
Roche Hitachi 911/912/917 and MODULAR P analyzers
I. Device Description:
The test consists of ready-to-use reagents containing anti-tobramycin antibody (sheep polyclonal), glucose-6-phosphate, nicotinamide adenine dinucleotide, and bovine albumin in serum.
J. Substantial Equivalence Information:
| Predicate | k964457, COBAS INTEGRA Tobramycin |
| --- | --- |
| Describe the item being compared | |
| The Roche ONLINE TDM Tobramycin assay is substantially equivalent to other products in commercial distribution intended for similar use. Most notably, it is substantially equivalent to the currently marketed Roche COBAS INTEGRA Tobramycin (k964457).
The ONLINE TDM Tobramycin assay is for the quantitative determination of tobramycin in human serum or plasma on Roche automated clinical chemistry analyzers. The proposed labeling indicates the Roche Hitachi 911, 912, 917 and Modular P analyzers can be used with the Roche ONLINE TDM Tobramycin reagent kits. | |
| Similarities | |
| The new device and the predicate are both indicated for the quantitative determination of Tobramycin In human serum or plasma on Roche automated clinical analyzers. | |
| Differences | |
| The ONLINE TDM Tobramycin uses a homogeneous microparticles agglutination immunoassay. In Contrast, the Roche COBAS INTEGRA tobramycin assay utilizes a fluorescence polarization immunoassay (FPIA) system. | |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-T2, Evaluation of Precision Performance of Quantitative Measurement Methods
Guidance for Industry and FDA Staff; Replacement Reagent and Instrument Family Policy, December 11, 2003
L. Test Principle:
The assay is based on a homogeneous enzyme immunoassay technique used for the quantitative analysis of tobramycin in human serum or plasma. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the sample can be measured in terms of enzyme activity. Active enzyme converts oxidized nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that is measured spectrophotometrically. Endogenous serum G6PDH does not
{2}
interfere because the coenzyme functions only with the bacterial (Leuconostoc mesenteroides) enzyme employed in the assay.
# M. Performance Characteristics (if/when applicable):
# 1. Analytical performance:
# a. Precision/Reproducibility:
The precision study was conducted on control material and spiked human serum pools using a modified version of CLSI-EP5-T2. The evaluation included using triplicate samples, 3 runs per day over 21 days $(n = 63)$ . The results of the study met the sponsor's acceptance criteria for within-run precision (SD no greater than 0.14 up to a concentration of $2\mu \mathrm{g / mL}$ or $\mathrm{CV} \leq 7\%$ at higher concentrations) and total and between-day precision (SD no greater than 0.18, up to a concentration of $2\mu \mathrm{g / mL}$ or $\mathrm{CV} \leq 9\%$ at higher concentrations). The results are tabulated below:
| Specimen | Low spike | High spike | Control 1 | Control 2 | Control 3 |
| --- | --- | --- | --- | --- | --- |
| Total mean (ug/ml) | 3.22 | 8.64 | 1.31 | 4.15 | 7.10 |
| Within-run SD (ug/ml) | 0.0255 | 0.0614 | 0.0514 | 0.0374 | 0.0412 |
| Within-run %CV | 0.8 | 0.7 | 3.9 | 0.9 | 0.6 |
| Total SD (ug/ml) | 0.0553 | 0.1523 | 0.0685 | 0.0685 | 0.0928 |
| Total %CV | 1.7 | 1.8 | 5.2 | 1.7 | 1.8 |
| Between-day SD (ug/ml) | 0.0491 | 0.1393 | 0.0453 | 0.0574 | 0.0832 |
| Between-day %CV | 1.5 | 1.6 | 3.5 | 1.4 | 1.2 |
# b. Linearity/assay reportable range:
To assess the linearity of the assay, an 11-level dilution series was prepared using tobramycin spiked human serum pool diluted with a tobramycin negative human serum pool. The percent recoveries within the reportable range (0.33 - 10.0 ug/mL) were 94.2-101.8% and met the sponsor's acceptance criteria of % recovery within ± 10% of the theoretical value up to 10 μg/mL. Results were evaluated by linear regression. The linearity results are tabulated below:
| % High Sample | Theoretical Value (μg/mL) | Assayed Value (μg/mL) | % Recovery | Recovery range limit |
| --- | --- | --- | --- | --- |
| 100 | 13.3 | 11.78 | 88.6 | Out of |
| 100 | 13.3 | 11.78 | 88.6 | Out of |
{3}
| % High Sample | Theoretical Value (μg/mL) | Assayed Value (μg/mL) | % Recovery | Recovery range limit |
| --- | --- | --- | --- | --- |
| | | | | range |
| 90 | 11.97 | 11.01 | 92.0 | Out of range |
| 80 | 10.64 | 10.02 | 94.2 | In range |
| 70 | 9.31 | 9.19 | 98.7 | In range |
| 60 | 7.98 | 8.12 | 101.8 | In range |
| 50 | 6.65 | 6.65 | 100.0 | In range |
| 40 | 5.3 | 5.36 | 100.8 | In range |
| 30 | 3.99 | 3.82 | 95.7 | In range |
| 20 | 2.66 | 2.66 | 100.0 | In range |
| 10 | 1.33 | 1.35 | 101.5 | In range |
The sponsor evaluated the performance of the TDM Preciset Multianalyte Calibrator diluent by comparing results of the diluted samples with neat samples (8.6 and 9.3 ug/mL) and the results of the diluted samples with spiked samples (12 and 17 ug/mL). The sponsor's acceptance criteria for these studies: percent (%) difference ±10% of the values of the neat samples and the spiked samples.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The 18 month shelf life claim for the Online TDM Tobramycin reagent was initially assessed by accelerated stability experiments and followed up with real time stability experiments. The 60 days on board instrument stability claim was assessed with real time studies. The sponsor's acceptance criteria was reviewed and found to be acceptable.
Calibrators and controls were previously cleared (k060429 and k031856) and are sold separately.
d. Detection limit:
The lower detection limit claim of 0.33 ug/ml is based on the mean and two standard deviations of 21 determinations of zero calibrator material. A linear interpolation model, based on these determinations of the zero calibrator material well as five replicate determinations of Calibrator B (1.0 ug/mL), was used to calculate the concentration equivalent to the mean plus two standard deviations.
e. Analytical specificity:
{4}
To evaluate potential interference from drugs and metabolites, serum pools were spiked with potential interferents and approximately $3\mathrm{ug / ml}$ tobramycin. Drugs found to cross-react at high concentrations were run in a series of dilutions. Percent cross reactivity for drugs and metabolites were defined as: $\left[\left(\mathrm{Da - Dt}\right) / \mathrm{C}\right]$ x100, where $\mathrm{Dt} =$ the measured concentration of the control analyte, $\mathrm{Da} =$ measured concentration of the control analyte + cross-reactant and $\mathrm{C} =$ known concentration of cross-reactant. Calculated cross-reactivities are based on the median of triplicate determinations Observed cross-reactivities with the compounds tested are tabulated below:
| Compound | Concentration (ug/ml) | % cross-reactivity |
| --- | --- | --- |
| Amikacin | 100 | 1.55 |
| Kanamycin | 100 | 5.32 |
No cross-reactivity was detected for the following drugs (concentrations in $\mathrm{ug / ml}$ ): chloramphenicol (1000), erythromycin (1000), tetracycline (1000), sulphamethoxazole (600), trimethoprin (25), carbenicillin (1000), cephalothin (1000), clindamycin (1000), gentamicin (25), gentamicin (100), neomycin (100), netilmicin (100), penicillin G (1000), sisomycin (100), streptomycin (100) and vancomycin (200).
Sixteen common drugs were tested for interference on the Hitachi 917 in normal human serum pools spiked with tobramycin at $2.5\mu \mathrm{g / mL}$ . Sponsor's acceptance criteria: recovery $+ / - 10\%$ of the control value. No significant interference was observed for the following drugs at the concentrations tested (concentrations in $\mathrm{ug / ml}$ ):
acetylcysteine (150), ampicillin (1000), ascorbic acid (300), K-Dobesilate (200), methyldopa (20), Doxycycline (50), cyclosporine (5), levodopa (20), metronidazole (200), phenylbutazone (400), acetylsalicylic acid (1000), rifampicin (60), acetaminophen (200), ibuprofen (500), cefoxitin (2500), and theophylline (100).
To evaluate interference from endogenous compounds, a series of dilutions containing varying levels of the endogenous compounds was prepared from spiked pooled serum. Testing was in the presence of $3\mathrm{ug / ml}$ tobramycin (except for intralipid, conjugated and unconjugated bilirubin in which samples contained $7\mathrm{ug / ml}$ tobramycin.) Percent recovery was calculated relative to control samples containing tobramycin without spiked endogenous
{5}
compounds. The median of triplicate determinations was used in calculations of recovery. Concentrations ranges of endogenous compound in which assay recovery $\geq 90\%$ is observed are tabulated below:
| Compound/sample condition tested | Concentration range within recovery criteria (+/- 10% bias) |
| --- | --- |
| Bilirubin (conjugated and unconjugated) | I index up to 30 (approximately equivalent to 30 mg/dL) |
| Hemoglobin | H index up to 800 (approximately equivalent to 800 mg/dL) |
| Lipemia | L index up to 750 (no significant interference from triglycerides up to 750 mg/dL intralipid) |
| Rheumatoid factor | Up to 100 IU/ml |
| Samples containing HAMA 1/HAMA 2 | 100.2 %/100.8 % |
| Total protein | Range: 2-12 g/dL |
f. Assay cut-off: Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
Fifty five clinical samples with concentrations below, within and above the medical decision points were analyzed in singlicate using the new device and a commercially available tobramycin assay. Sample values ranged from $0.2 - 9.0\ \mathrm{ug/ml}$ as measured by the predicate device. Results of the sponsor's analysis based on the Passing-Bablok model are shown below:
| | Lower CI | Upper CI |
| --- | --- | --- |
| Slope | 1.006 | 1.060 |
| Intercept | 0.1572 | 0.2436 |
Fifty two clinical samples with concentrations below, within and above the medical decision points were analyzed in singlicate using the new device and a commercially available tobramycin assay. Sample values ranged from $0.3 - 9.3\ \mathrm{ug/ml}$. Results of the sponsor's analysis based on the Passing-Bablok model are shown below:
$$
y = 0.92x + 0.01,\ r = 0.983,\ SD\ (\mathrm{md}\ 95) = 0.680.
$$
{6}
7
b. Matrix comparison:
To evaluate the effect of plasma anticoagulants, comparisons of serum samples versus samples containing EDTA, sodium heparin, sodium fluoride/potassium oxalate, citrate, and lithium heparin were conducted. Thirty samples were included for fluoride/potassium oxalate and citrate anticoagulants; fifteen samples were included for EDTA, sodium heparin, and lithium heparin. No significant bias due to these anticoagulants was observed.
3. Clinical studies:
a. Clinical Sensitivity: Not applicable
b. Clinical specificity: Not applicable
c. Other clinical supportive data (when a. and b. are not applicable): Not applicable
4. Clinical cut-off: Not applicable
5. Expected values/Reference range:
The therapeutic range for this specific assay was not determined. The sponsor stated that similar diagnostic technologies have shown that in most adults, a peak therapeutic response is achieved with tobramycin concentration ranges of 6-10 µg/mL and trough concentration ranges of 0.5-2.0 µg/mL. The sponsor stated in the package insert that each laboratory should investigate the transferability of the expected values to its own patient population and if necessary determine its own reference range. Test findings should always be assessed in conjunction with patient's medical history, clinical examination and other medical findings. The following ranges for peak and trough were taken from the literature and included in the package insert:
| Investigator | Peak µg/mL | Trough µg/mL |
| --- | --- | --- |
| Baselt and Cravey | 6-10 | 1.1-4.3 |
| Sande and Mandell | 5-8 | 2.1-4.3 |
| Dipersio | 4-8 | 2.1-4.3 |
{7}
8
| Lew | 5-10 | - |
| --- | --- | --- |
Baselt RC, Cravey RH. Disposition of Toxic Drugs and Chemicals in Man. 3rd ed, 1990:805-807.
Sande MA, Mandell GL. Antimicrobial agents, the aminoglycosides. In: Gilman AG, Goodman LS, Gilman A, eds. The Pharmacological Basis of Therapeutics. NEW, NY: MacMillan 1980:1162-1180.
Dipersio JR. Gentamicin and other aminoglycosides. In: Pesce AJ and Kaplan LA eds. Methods in Clinical Chemistry. St. Louis, MO: CV Mosby Co. 1987.
Lew M. Interpretation of aminoglycoside serum levels. Hosp. Pharm 1979; 14:465-472.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.