NAKED

K060797 · Veraxis Int'L, Inc. · HIS · Apr 7, 2006 · Obstetrics/Gynecology

Device Facts

Record IDK060797
Device NameNAKED
ApplicantVeraxis Int'L, Inc.
Product CodeHIS · Obstetrics/Gynecology
Decision DateApr 7, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.5300
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Veraxis Int'l Inc. Naked™ condom is to be used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases).

Device Story

Naked™ condom is a physical barrier device designed for contraception and STI prevention. Features include a reservoir tip for semen containment, an integral base ring for a secure seal, and a relaxed-fit body design to improve sensation and comfort compared to standard condoms. Available in multiple sizes to enhance fit and usage compliance. Used by individuals during sexual activity; functions as a mechanical barrier. Benefits include reduced risk of pregnancy and disease transmission through improved user adherence facilitated by better fit and comfort.

Clinical Evidence

Bench testing only. The device conforms to ISO 10993 (biocompatibility), ISO 4074 (condom requirements and test methods), and ASTM D3492 (standard specification for rubber contraceptives). No clinical trial data presented.

Technological Characteristics

Natural rubber latex condom; reservoir tip; integral base ring; relaxed-fit body design; multiple sizes. Conforms to ISO 10993, ISO 4074, and ASTM D3492. Non-electronic, mechanical barrier device.

Indications for Use

Indicated for over-the-counter use by sexually active individuals for contraception and prevention of sexually transmitted diseases.

Regulatory Classification

Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 7 2006 APR # 3. 510(k) Summary (21 CFR §807.92) | Sponsor: | Veraxis Int'l Inc. (or "Veraxis")<br>9663 Santa Monica Blvd., Suite 2500<br>Beverly Hills, CA 90210<br>(646) 285-7714 | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Jules T. Mitchel, President<br>Target Health Inc.<br>261 Madison Avenue, 24th Floor<br>New York, NY 10016<br>Tel: (212) 681-2100<br>Fax: (212) 682-2105<br>JulesMitchel@TargetHealth.com | | Date: | February 23, 2006 | | Name of the<br>Device: | Trade Name: Naked™ | | | Common Name: Condom | | | Classification Name: Condom 884.5300 Class II (Product Code<br>HIS). | | Substantial<br>Equivalence: | The Veraxis Naked™ condom is equivalent in intended use and<br>technological characteristics to Oceans Seven, Intn'l The Official<br>Condom of the 21st Century (They-Fit™) condoms and Church &<br>Dwight, Trojan® Magnum XL condoms. | | Device<br>Description: | By creating an actual physical barrier between partners engaged in<br>sexual activity, the condom represents the best available solution<br>to two of the world's major health challenges: sexually transmitted<br>diseases/AIDS and unwanted pregnancy. | | | The design of the condom is essential to its effectiveness. The<br>closed end has a reservoir tip to contain semen. The condom<br>described herein has an integral ring at its base and a snug fit<br>along its lower portion in order to ensure a safe, secure seal and<br>barrier. | | | Although standard condoms are one of the few over-the-counter,<br>non-prescription contraceptives, condoms have many deficiencies<br>that limit their usage among sexually active populations: (i) they are<br>uncomfortable, (ii) they do not transmit sensation well, and (iii) theyAIDS epidemic, and the high risk involved in unprotected sex, a<br>significant portion of sexually active populations decline to use<br>condoms or use them only occasionally. | {1}------------------------------------------------ The design of the Naked™ condom described herein addresses these deficiencies so that condom usage will significantly improve among sexually active populations. The condom described herein provides a slightly relaxed fit along the main body of the product between the snug base and reservoir tip. This relaxed fit provides a more natural feel during intercourse, an easier application process, and improved sensation, while maintaining a safe, secure seal and barrier. This product is designed to fully conform to ISO 10993 and ISO 4074. It also conforms to ASTM D3492. The Veraxis Int'l Inc. Naked ™ condom is to be used for Intended Use: contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases). > The Veraxis Int'l Inc. Naked ™ condom is indicated for Over-The-Counter Use (21 CFR 801 Subpart C) Technological Characteristics: This condom has the same technical characteristics as the predicate condoms identified above. The product described herein features a slightly relaxed fit along the main body of the product. The relaxed fit at the closed end of the condom provides a greater lay flat width than the width of the snug base and safety seal it creates at the open end of the condom. The nature and value of the relaxed fit is well expressed by Ansell Incorporated, the second largest manufacturer of condoms in the world, in an article in Mens Journal in April of 1998: "For 20 years this industry has been telling people that thinner condoms are better, more sensitive, but really, you can't tell the difference between the standard thickness of .07 mm and the thin condom of .05 mm. The real difference is looseness (the relaxed fit), which allows for friction between the penis and the latex and the vaginal wall." The Naked ™ condom comes in a range of sizes that provide increased customization for each individual which makes for a safer, more secure, and better fit. These two technical characteristics (relaxed fit and various sizes) maximize product effectiveness and sensation, and thereby increase product usage among the sexually active populations. As stated in the product description, this product is designed to fully conform to ISO 10993 and ISO 4074. It also conforms to ASTM D3492. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES" written around the top half of the circle. In the center of the circle is a stylized image of an eagle with three lines representing its wings. The logo is black and white and appears to be a seal or emblem. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 7 2006 Veraxis Int'l Inc. c/o Mr. Jeffrey D. Rongero Senior Project Engineer Conformity Assessment Services Underwriters Laboratories, Inc. Research Triangle Park Division 12 Laboratory Drive P.O. Box 13995 RESEARCH TRIANGLE PARK NC 27709-3995 Re: K060797 Trade/Device Name: Naked™ Condom Regulation Number: 21 CFR §884.5300 Regulation Name: Condom Regulatory Class: II Product Code: HIS Dated: March 17, 2006 Received: March 24, 2006 Dear Mr. Rongero: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # KO6797 ### 2. Indication for Use ## 510(k) Number (if known): NHA KO60797 Device Name: Naked™ male natural rubber latex condom Indications For Use: The Veraxis Int'l Inc. Naked™ condom is to be used for contraceptive and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases). Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use × (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number Page 1 of 007
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...