← Product Code [LAR](/productcode/LAR) · K060773

# ONLINE TDM PROCAINAMIDE (K060773)

_Roche Diagnostics Corp. · LAR · Aug 23, 2006 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060773

## Device Facts

- **Applicant:** Roche Diagnostics Corp.
- **Product Code:** [LAR](/productcode/LAR.md)
- **Decision Date:** Aug 23, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3320
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The ONLINE TDM Procainamide assay is for the quantitative determination of procainamide in human serum or plasma on Roche automated clinical chemistry analyzers. Measurements are used in the diagnosis and treatment of procainamide overdose and in monitoring levels of procainamide to ensure proper therapy.

## Device Story

ONLINE TDM Procainamide is an in vitro diagnostic assay for quantitative measurement of procainamide in human serum or plasma. Used on Roche automated clinical chemistry analyzers (Hitachi 911, 912, 917, Modular P) in clinical laboratory settings. Operates via homogeneous enzyme immunoassay; competition between sample drug and G6PDH-labeled drug for antibody binding sites. Enzyme activity inversely proportional to drug concentration; active enzyme converts NAD to NADH, measured spectrophotometrically. Results assist clinicians in monitoring therapeutic levels and managing overdose. Benefits include automated, rapid quantification of drug levels to ensure appropriate patient dosing.

## Clinical Evidence

No clinical studies performed. Evidence consists of analytical bench testing: precision (CV 1.5-9.3%), linearity (0.33–14 µg/mL), analytical specificity (no significant interference from common drugs or endogenous substances), and method comparison studies against predicate devices (n=51 and n=48 samples, r=0.9975 and r=0.997 respectively).

## Technological Characteristics

Homogeneous enzyme immunoassay; competitive binding principle using bacterial (Leuconostoc mesenteroids) G6PDH enzyme. Measures absorbance change spectrophotometrically. Designed for use on Roche automated clinical chemistry analyzers (Hitachi 911, 912, 917, Modular P).

## Regulatory Identification

A digoxin test system is a device intended to measure digoxin, a cardiovascular drug, in serum and plasma. Measurements obtained by this device are used in the diagnosis and treatment of digoxin overdose and in monitoring levels of digoxin to ensure appropriate therapy.

## Predicate Devices

- COBAS Integra Procainamide (k951595)

## Reference Devices

- Preciset TDM II Calibrators (k031856)
- TDM Control Set (k060429)
- EMIT 2000 Procainamide (Roche/Hitachi 911)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k060773

B. Purpose for Submission:
New device

C. Measurand:
Procainamide

D. Type of Test:
Enzyme immunoassay

E. Applicant:
ROCHE DIAGNOSTICS CORP.

F. Proprietary and Established Names:
ONLINE TDM PROCAINAMIDE

G. Regulatory Information:

|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  Enzyme Immunoassay, Procainamide (LAR) | Class II | 21 CFR 862.3320, Digoxin test system. | 91 CLINICAL TOXICOLOGY (TX)  |

H. Intended Use:

1. Intended use(s):
See Indications for Use below.

2. Indication(s) for use:
The ONLINE TDM Procainamide assay is for the quantitative determination of procainamide in human serum or plasma on Roche automated clinical chemistry analyzers. Measurements are used in the diagnosis and treatment of procainamide overdose and in monitoring levels of procainamide to ensure proper therapy.

3. Special conditions for use statement(s):
For prescription use only

{1}

2

4. Special instrument requirements:

Evaluations represented in the 510(k) were performed on the Hitachi 917.

I. Device Description:

The device consists of two ready-to-use reagents ('working solutions'). R1 contains a mouse anti-procainamide monoclonal antibody, glucose-6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in buffer while R2 contains procainamide labeled with glucose-6-phosphate dehydrogenase (G6PDH) in buffer.

J. Substantial Equivalence Information:

|  Predicate | k951595, COBAS Integra Procainamide  |
| --- | --- |
|  Describe the item being compared  |   |
|  The Roche ONLINE TDM Procainamide assay is substantially equivalent to the currently marketed Roche COAS INTEGRA Procainamide (k951595).  |   |
|  Similarites  |   |
|  The ONLINE TDM Procainamide and the COBAS INTEGRA Procainamide assays are both indicated for the quantitative determination of Procainamide in human serum or plasma on Roche automated clinical analyzers.  |   |

K. Standard/Guidance Document Referenced (if applicable):

|  GUIDANCE  |   |   |   |
| --- | --- | --- | --- |
|  Document Title | Office | Division | Web Page  |
|  Guidance for Industry and FDA Staff; Replacement Reagent and Instrument Family Policy | OIVD |  | http://www.fda.gov/cdrh/oivd/guidance/950.html  |

L. Test Principle:

The assay is a homogeneous enzyme immunoassay technique used for the quantitative analysis of procainamide in human serum or plasma. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for antibody binding sites. Enzyme activity decreases upon binding to the antibody, so the drug concentration in the sample can be measured in terms of enzyme activity. Active enzyme converts oxidized nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that is measured spectrophotometrically. Endogenous serum G6PDH does not interfere because the coenzyme functions only with the bacterial (Leuconostoc mesenteroides) enzyme employed in the assay.

M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

Reproducibility was determined using three levels of controls and two levels

{2}

of a human sera pool (HSP) with a modified version of CLSI EP5-T2 where 3 replicates were tested once a day for 21 days (n = 63). Results are shown in the table below:

Precision: Roche ONLINE TDM Procainamide Assay

|   |  | Within-Run |   | Total Precision  |   |
| --- | --- | --- | --- | --- | --- |
|  Sample | Mean | Std Dev | CV | Std Dev | CV  |
|   | ng/mL | ng/mL | % | ng/mL | %  |
|  Control 1 | 1.7 | 0.03 | 1.5 | 0.04 | 2.6  |
|  Control 2 | 7.1 | 0.07 | 1.0 | 0.27 | 3.8  |
|  Control 3 | 11.6 | 0.28 | 2.4 | 1.08 | 9.3  |
|  Low HSP | 3.6 | 0.04 | 1.1 | 0.06 | 1.5  |
|  High HSP | 8.8 | 0.14 | 1.6 | 0.45 | 5.1  |

b. Linearity/assay reportable range:

The reportable range of the assay is 0.33 – 14 µg/mL.

Assay linearity was evaluated by measuring percent recovery of expected values in a dilution series using a procainamide-spiked serum pool diluted with a non-spiked serum pool. Results are summarized below:

Linearity of the ONLINE TDM Procainamide Assay

|  Dilution of Sample | Theoretical Value (ug/mL) | Measured Value (ug/mL) | % Recovery  |
| --- | --- | --- | --- |
|  neat | 19.20 | 31.59 | 165  |
|  0.9 | 17.28 | 17.57 | 102  |
|  0.8 | 15.36 | 15.26 | 99  |
|  0.7 | 13.44 | 12.50 | 93  |
|  0.6 | 11.52 | 10.44 | 91  |
|  0.5 | 9.60 | 8.88 | 93  |
|  0.4 | 7.68 | 7.24 | 94  |
|  0.3 | 5.76 | 5.57 | 97  |
|  0.2 | 3.84 | 3.94 | 103  |
|  0.1 | 1.92 | 1.81 | 94  |

The sponsor also performed post-dilution studies that supported their recommendation for dilution of samples above the measuring range of the device.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): Stability: The shelf life of the Roche ONLINE TDM Procainamide Assay is claimed as 18 months based on accelerated and real time stability testing of

{3}

three lots of product. Onboard (opened and refrigerated) stability is claimed as 60 days.

**Calibrators:** The Preciset TDM II Calibrators are prepared to contain known quantities of procainamide in normal human serum and are traceable to USP reference standards. These calibrators are used to establish a standard curve from which the quantity of drug in unknown specimens can be determined. Please see 510(k) k031856 for more information.

**Controls:** The TDM Control Set contains liquid controls based on human serum and are traceable to USP reference standards. Please see 510(k) k060429 for more information.

d. Detection limit:

The manufacturer's claimed a lower assay limit of 0.33 ug/mL. This claim was determined by measuring the absorbance values of 21 replicates of the zero calibrator and adding two standard deviations to the mean of measured values.

e. Analytical specificity:

The assay was evaluated on the Hitachi 917 for interference from drugs, and endogenous compounds:

**Cross-reactivity:**

The following compounds were tested for cross-reactivity in serum samples containing approximately 4.5 ug/mL procainamide. Results were compared to those of control samples without cross-reactant. All compounds tested had undetectable cross-reactivity (ND) at the listed concentration, defined by the manufacturer as a difference in values less than the sensitivity of the assay.

|  Cross Reactant | Concentration (ug/ml)  |
| --- | --- |
|  Acetaminophen | 100  |
|  Desethyl-N-acetylprocainamide | 100  |
|  Digoxin | 0.1  |
|  Diphenylhydantoin | 100  |
|  Disopyramide | 100  |
|  Ephedrine | 100  |
|  Furosemide | 100  |
|  Glycinexylidide | 100  |
|  Hydrochlorothiazide | 100  |
|  Isoproterenol hydrochloride | 100  |
|  Lidocaine | 100  |
|  Monoethylglycinexylidide | 100  |
|  N-Acetylprocainamide | 40  |

{4}

5

|  N-(2-Diethylaminoethyl)isonicotinamide | 100  |
| --- | --- |
|  p-Acetamidobenzoic acid | 100  |
|  p-Aminobenzoic Acid | 100  |
|  Procaine | 100  |
|  Propanolol | 100  |
|  Quinidine | 100  |
|  Tocainide | 100  |

Other drugs:

The drugs listed below were spiked into normal human serum pools containing 2.5 ug/mL procainamide, and the drug levels shown below. The manufacturer indicated that the drug levels evaluated were at a concentration greater than what would be expected for a maximum daily dose. Control samples consisted of serum spiked with procainamide, without any additional added drug. Less than 10% interference was observed at the concentrations shown.

|  Drug (max concentration tested)  |   |
| --- | --- |
|  Acetyl cysteine (150 ug/mL) | Metronidazole (200 ug/mL)  |
|  Ampicillin-Na (1000 ug/mL) | Phenylbutazone (400 ug/mL)  |
|  Ascorbic acid (300 ug/mL) | Doxycycline HCl (50 ug/mL)  |
|  Ca-Dobesilate (200 ug/mL) | Acetylsalicylic acid (1 mg/mL)  |
|  Cyclosporine (5000 ng/mL) | Rifampicin (60 ug/mL)  |
|  Cefoxitin-Na (2500 ug/mL) | Acetaminophen (200 ug/mL)  |
|  Heparin (5 ug/mL) | Ibuprofen (500 ug/mL)  |
|  Levodopa (20 ug/mL) | Theophylline (100 ug/mL)  |
|  Methyldopa + 1,5 (20 ug/mL) |   |

Endogenous substances:

Interference by endogenous substances was assessed by determining the recovery of procainamide from samples spiked with procainamide (about 4 ug/mL) and different concentrations of interferents. There was no significant (&lt; ±10%) interference from:

- I index of 1-30 (approximate conjugated and unconjugated bilirubin concentration: 30 mg/dL)
- H index of 0-800 (approximate hemoglobin concentration: 800 mg/dL)
- Lipemic index of 0-1500 (0-1500 mg/dL intralipid).
- HAMA 1 and HAMA 2 samples
- Rheumatoid factors up to 100 IU/mL.
- Total protein 2-12 g/dL.
- Triglycerides up to 1000 mg/dL

{5}

f. Assay cut-off:
Not applicable.

2. Comparison studies:

a. Method comparison with predicate device:
A comparison of the ONLINE TDM Procainamide assay on a Roche/Hitachi 917 analyzer (y) with a commercially available method (COBAS FP Procainamide on a COBAS INTEGRA 700 analyzer) (x) gave the following correlation ($\mu \mathrm{g} / \mathrm{mL}$) using 51 human serum samples:
Passing/Bablok:
$$
\begin{array}{l}
y = 1.007 x + 0.247 \\
r = 0.9975 \\
\mathrm{SD} \, (\mathrm{md} \, 95) = 0.311 \\
\end{array}
$$
The sample concentrations were between 0.3 and $10.6~\mu \mathrm{g/mL}$.

A comparison of the ONLINE TDM Procainamide assay on a Roche/Hitachi 917 analyzer (y) with a commercially available method (EMIT 2000 Procainamide on a Roche/Hitachi 911 analyzer) (x) gave the following correlation ($\mu \mathrm{g} / \mathrm{mL}$) using 48 human serum samples:
Passing/Bablok:
$$
\begin{array}{l}
y = 0.981 x + 0.02 \\
r = 0.997 \\
\mathrm{SD} \, (\mathrm{md} \, 95) = 0.241 \\
\end{array}
$$
The sample concentrations were between 1.06 and $11.18~\mu \mathrm{g/mL}$.

b. Matrix comparison:
The sponsor found serum and plasma equivalent with the following anticoagulants: heparin (sodium or lithium), potassium (K2 or K3) EDTA, citrate, and oxalate.

3. Clinical studies:

a. Clinical Sensitivity:
Not applicable.

b. Clinical specificity:
Not applicable.

c. Other clinical supportive data (when a. and b. are not applicable):

4. Clinical cut-off:
Not applicable.

5. Expected values/Reference range:
The sponsor states that the commonly accepted therapeutic range for procainamide is $4 - 10~\mu \mathrm{g/mL}$ (16.9 - 42.3 $\mu \mathrm{mol/L}$) and the sum of procainamide

{6}

and its primary active metabolite, N-acetlyprocainamide (NAPA) is 5 - 30 µg/mL (21.2 - 126.9 µmol/L). Procainamide is thought to be toxic at levels greater than 16 µg/mL (&gt;67.7 µmol/L).

The factors that can influence the relationship between procainamide serum or plasma concentrations and clinical response include: renal and circulatory function, rate of acetylation, the severity and type of cardiac arrhythmia, general state of health, and use of other drugs. The concentration of procainamide in serum or plasma depends on the time of the last drug dose; mode of administration; concomitant drug therapy; sample condition; time of sample collection; and individual variations in absorption, biotransformation, distribution, and excretion. These parameters must be considered when interpreting results. Each laboratory should investigate the transferability of the expected values to its own patient population and if necessary determine its own reference range.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda.innolitics.com/device/K060773](https://fda.innolitics.com/device/K060773)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
