← Product Code [GAT](/productcode/GAT) · K060721

# LARIAT LOOP APPLICATOR (K060721)

_Sentreheart · GAT · Jun 2, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060721

## Device Facts

- **Applicant:** Sentreheart
- **Product Code:** [GAT](/productcode/GAT.md)
- **Decision Date:** Jun 2, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5000
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The LARIAT Loop Applicator facilitates suture placement and knot tying for use in surgical applications where soft tissue are being approximated and/or ligated with a pre-tied polyester suture.

## Device Story

LARIAT Loop Applicator is a single-use, one-piece surgical instrument designed for suture delivery and deployment. Device features a pre-tied polyester suture loop and a central lumen for tissue aspiration and stabilization. Used by surgeons in clinical settings to approximate or ligate soft tissue. Operation involves positioning the device to deliver the suture loop to the target site, facilitating knot tying. Device provides mechanical assistance for tissue management during surgery, potentially improving procedural efficiency and patient outcomes by enabling secure tissue approximation.

## Clinical Evidence

Bench testing only. In-vitro performance testing demonstrated the device is safe and effective for its intended use. Biocompatibility confirmed via materials common to predicate devices.

## Technological Characteristics

Single-use, one-piece suture delivery device. Includes central lumen for aspiration/stabilization. Pre-tied polyester suture loop. Materials are biocompatible and consistent with predicate devices. No electronic components or software.

## Regulatory Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Ethicon Endosuture System ([K963329](/device/K963329.md))
- Genzyme Saph-Loop Ligating Lopic ([K022410](/device/K022410.md))
- HysteRx Liga-Loop Suture Applicator ([K993695](/device/K993695.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Ko6072/
510(k) Notification

1/2

# 510(k) Summary

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Date Prepared: March 10, 2006

510(k) number:

# Applicant Information:

SentreHeart 2468 Embarcadero Way Palo Alto, CA 94303

# Contact Person

| Russell Seiber |                |
|----------------|----------------|
| Phone Number:  | (650) 354-1200 |
| Fax Number:    | (650) 354-1204 |

### Device Information:

| Trade Name:          | LARIAT Loop Applicator            |
|----------------------|-----------------------------------|
| Classification:      | Class II                          |
| Classification Name: | Suture, Non-absorbable, Synthetic |

# Physical Description:

The LARIAT Loop Applicator is a one piece, single-use suture delivery and deployment device with a pre-tied polyester suture loop that is pre-loaded on the device. A central lumen within the LARIAT Loop Applicator is designed for aspiration and stabilization of tissue during the delivery of the LARIAT Suture Loop.

The suture is itself a cleared medical device as a part of Pre-Market Notification K021019.

## Intended Use:

The LARIAT Loop Applicator facilitates suture placement and knot tying for use in surgical applications where soft tissue are being approximated and/or ligated with a pre-tied polyester suture.

### Equivalent Device:

The subject device is substantially equivalent in intended use and/or method of operation to the Ethicon Endosuture System (K963329), the Genzyme Saph-Loop Ligating Lopic (K022410), and the HysteRx Liga-Loop Suture Applicator (K993695),

# Test Results:

#### Performance

Results of in-vitro testing demonstrate that the LARIAT Loop Applicator is safe and effective for its intended use.

SentreHeart

JUN - 2 2006

{1}------------------------------------------------

#### Biocompatibility

Bloompatibling biocompatible. The materials are used in the identified predicates and are also commonly used in other medical devices.

#### Summary:

Based on the intended use, product, performance and biocompatibility information provided in this notification, the subject device has been shown to be substantially equivalent to the currently marketed predicate devices.

SentreHeart

CONFIDENTIAL

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features an eagle-like emblem with three stylized lines representing the wings, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the emblem. The emblem is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 2 2006

SentreHeart™ % Mr. Robert Chin Regulatory Consultant 2468 Embarcadero Way Palo Alto, California 94303

Re: K060721

Trade/Device Name: LARIAT Loop Applicator Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly (ethylene terephthalate) surgical suture Regulation Class: II Product Code: GAT Dated: May 24, 2006 Received: May 25, 2006

Dear Mr. Chin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Robert Chin

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address .http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Hubert Lenoir us

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

LARIAT Loop Applicator

| 510(k) Number (if known): | K060721             |
|---------------------------|---------------------|
|                           | Indications for Use |

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

The LARIAT Loop Applicator facilitates suture placement and knot tying for use in surgical applications where soft tissue are being approximated and/or ligated with a pre-tied polyester suture.

Prescription Use V (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

![Signature](signature.png)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

| 510(k) Number | K060721 |
|---------------|---------|
|---------------|---------|

SentreHeart
CONFIDENTIAL
Page __ of __
Page 9

---

**Source:** [https://fda.innolitics.com/device/K060721](https://fda.innolitics.com/device/K060721)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
