The PMP4 BP Pro is used for self-testing by patients and by healthcare professionals at home and at medical settings to monitor blood pressure (systolic and diastolic) and pulse rate from the upper arm. The device consists of a table top monitor and a receiving program - the PMP4 Medical Application. The table top monitor obtains the blood pressure measurement through the oscillometric method, i.e. the blood movement through the artery in the upper arm are detected and converted into digital readings. For evaluation, the stored measurement results can be transmitted by Bluetooth interface to wirelyss mobile platform, e.g. PDA or a static platform, i.e., PC with the PMP4 Medical Application installed. The PMP4 Medical Application interfaces to the table-top monitor and is used to receive for Pro the test results and other medical data, to process and save these test results, and to synchroized and test results with the PMP4 Medical Center. The PMP4 Medical Application is also designed to receive ECG, spirometric, glucometric, and oximetric parameters from other Card Guard medical testing accessories in addition to BP+HR. When the table top monitor is not used to measure blood pressure, it displays time and date. The intended product users are Non-Medical Persons (NMP), i.e. persons expected to understand only basic medical data.
Device Story
PMP4 BP Pro is an oscillometric blood pressure monitor for upper arm use. Device detects arterial blood movement, converts to digital readings; transmits data via Bluetooth to mobile platforms (PDA) or PCs running PMP4 Medical Application. Application processes, stores, and synchronizes data with PMP4 Medical Center; also supports ECG, spirometry, glucometry, and oximetry data from other accessories. Inflation logic uses envelope curve approximation to minimize discomfort, inflating to just above systolic pressure; subsequent measurements use auto-feedback logic (previous systolic + 15 mmHg). Used by patients (self-testing) or healthcare professionals in home/clinical settings. Output allows providers to review historical data, facilitating clinical decision-making and patient monitoring.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Oscillometric blood pressure measurement system. Features Bluetooth wireless connectivity for data transmission to mobile/static platforms. Software-controlled inflation logic with auto-feedback. Complies with EN 1060 (non-invasive sphygmomanometers), AAMI/ANSI SP10, and EN 60601-1/1-2/1-4 safety and EMC standards. Quality management per ISO 9001/13485.
Indications for Use
Indicated for monitoring blood pressure and heart rate from the upper arm in patients, including non-medical persons, for home or clinical use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
APR 4 2006
CARD GUADD
## PMP4 BP Pro Blood Pressure Monitor 510(k) Summary of Safety and Effectiveness
### 1. General
| Submitter | Card Guard Scientific Survival Ltd., |
|-----------------|------------------------------------------------|
| Address | 2 Pekeris St. P.O.B. 527 Rehovot 76100, Israel |
| Contact: | Alex Gonorovsky, RA Manager |
| Phone: | 972-8-9484019 Fax: 972-8-9484044 |
| E-mail: | galex@cardguard.com |
| Device | |
| Trade Name: | PMP4 BP Pro Blood Pressure Monitor |
| Classification: | Noninvasive Blood Pressure Measurement System |
| Product Code: | DXN |
| Regulation No: | 21 CFR 870.1130 |
| Class: | II |
### 2. Definition and Intended Use
The PMP4 BP Pro is used for self-testing by patients and by healthcare professionals at home and at medical settings to monitor blood pressure (systolic and diastolic) and pulse rate from the upper arm. The device consists of a table top monitor and a receiving program - the PMP4 Medical Application. The table top monitor obtains the blood pressure measurement through the oscillometric method, i.e. the blood movement through the artery in the upper arm are detected and converted into digital readings. For evaluation, the stored measurement results can be transmitted by Bluetooth interface to wirelyss mobile platform, e.g. PDA or a static platform, i.e., PC with the PMP4 Medical Application installed. The PMP4 Medical Application interfaces to the table-top monitor and is used to receive for Pro the test results and other medical data, to process and save these test results, and to synchroized and test results with the PMP4 Medical Center.
The PMP4 Medical Application is also designed to receive ECG, spirometric, glucometric, and oximetric parameters from other Card Guard medical testing accessories in addition to BP+HR.
When the table top monitor is not used to measure blood pressure, it displays time and date. The intended product users are Non-Medical Persons (NMP), i.e. persons expected to understand only basic medical data.
### 3. Referenced Standards
- MDD 93/42/EEC Medical Device Directive Council Directive 93/42/EEC; June 14, 1993 -
- MPG The Act on Medical Devices (Medizinproduktegesetz), rev 2; Germany; Dec 2001 2.
- EN 475: Medical devices Electrically-generated alarm signals ; April 1995 నా
- EN 980: Graphical symbols for use in the labeling of medical devices; August 2003 র্বা
- EN 1041: Terminology, Symbols and Information provided with Medical Devices; Information supplied by the 5. manufacturer with medical devices; April 1998
- EN 1060: Non-invasive sphygmomanometers, Part 1: General requirements, Dec. 95 (EN 1060-1/A1: Changes, Sept. 02), 6. Part 3: Supplements for electro-mechanical blood pressure measuring systems; Sept. 1997. Change
- EN ISO 9001 : Quality management systems Requirements; December 2000 7.
- EN ISO 13485: Quality systems Medical devices; August 2000 8.
- EN ISO 14971: Medical devices application of risk management to medical devices, March 2001 9.
- 10. EN ISO 10993 Biological evaluation of medical devices Part 1. Evaluation and testing; Dec. 1997
- 11. EN 60601-1: Medical electrical equipment; Part 1: General requirements for safety; Boo. 2002
- 12. EN 60601-1-2: Medical electrical equipment; Part 1: 2. Collateral standard: EMC ; requirements and tests; 2001
- 13. EN 60601-1-4: Medical electrical equipment; Part 1: 4. Collateral Std. Programmable electric medical systems; Apr. 01
- 14. AAMI/ANSI SP10: Electronic or automated sphygmomanometers; October 1992
{1}------------------------------------------------
Image /page/1/Figure/0 description: The image shows the logo for "CARD GUARD Scientific Survival LTD". The text is in bold, sans-serif font. To the right of the text is a graphic that resembles an EKG readout.
### PMP4BP Pro Blood Pressure Monitor 510(k) Summary of Safety and Effectiveness
60703
#### 4. Method of Operation
### Inflating the Cuff
To lessen patient discomfort, the cuff does not always inflate to a defined predetermined value, but by means of an approximation of the envelope curve during the inflation phase attempts to inflate only up to a value just above the systolic pressure. To do this the pressure is increased in defined predetermined pressure steps (85, 100, 120, 140, 160 ... mmHg) and a single pressure pulse is measured at each stage. An increase in pulse amplitude to a maximum and the subsequent decrease is determined during this type of inflation. The systolic pressure is then regarded as definitely attained and the inflation terminated when a pulse amplitude of <52% of the maximum is measured after exceeding the maximum.
### Inflation with Auto Feedback Logic
After the first measurement, the device recognizes the systolic pressure. During the next inflation, the cuff inflates up to the most recent systolic value plus ~ 1 5mmHg instead of inflating stepwise to a predefined value. The advantage is the reduced pumping time and therefore the reduced discomfort for the patient.
#### Substantial Equivalence 5.
Substantial equivalence (SE) is claimed to the following predicate devices:
- Card Guard's BP Pro K053395 for physical and functional identity and close similarity in IU. 1.
- 2. Card Guard's CG-7000DX-BT K052556 - for the remote communication functionality,
- Card Guard's PMP Spiro Pro K050853 for including the PMP4 Medical Application. 3.
- Card Guard's PMP* Medical Web Center K050940 for multiple-format medical data feature. 4.
#### 6. PMP4 Medical Application
The PMP4 Medical Application is designed for wireless mobile platforms, e.g. PDA and for static platforms, i.e., PC. It is used to receive from the BP Pro and other Card Guard's medical testing accessories, the test results and other medical data, to process and save these test results, and synchronize data and test results with the PMP4 Medical Center. The Application is a part of a personal medical system solution. The PMP4 Medical Application performs the following activities:
- Receives medical test inputs from the external accessories. 1.
- Collects medical test data and other related information as defined for each test. 2.
- 3. Accesses historical test and related data stored on the device.
- Transmits medical test data and additional information to Center for professional evaluation or backup. বা
- ડાં Receives data from Center.
- Enables configuring GPRS data connection (based on mobile phone GPRS/CDMA capabilities), 6. changing user name and password.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a logo with a circular border containing text, surrounding a stylized symbol. The symbol consists of three abstract human figures in profile, arranged in a row. The figures are depicted with simple, curved lines, giving them a modern and minimalist appearance. The text in the circular border is small and difficult to read, but it appears to be arranged in a clockwise direction around the symbol.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 2006
APR
Card Guard Scientific Survival Ltd. c/o Mr. Alex Gonorovsky Regulatory Affairs Manager 2 Pekeris St. P. O. Box 527 Rehevot 76100 ISRAEL
Re: K060703
Trade Name: PMP4 BP Pro Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: February 21, 2006 Received: March 16, 2006
Dear Mr. Gonorovsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. Alex Gonorovsky
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blumuman for
Bran D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): Device Name: PMP4 BP Pro Blood Pressure Monitor
Indications for Use:
The PMP BP Pro Blood Pressure Monitor is intended to be used for monitoring Blood Pressure (BP) and Heart Rate (HR) from the upper arm. BP and HR can be recorded and transferred to a remote hand held device/PC for viewing and processing.
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use _ 区 AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
(Posted November 13, 2003)
Bhumanna
510(k) Nu
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.