The NeuroMetrix NC-stat is intended to stimulate and measure neuromuscular signals that are useful in diagnosing and evaluating systemic and entraponent neuropathies.
Device Story
NC-stat is a nerve conduction testing system; stimulates and measures neuromuscular signals. Components: handheld instrument, docking station, disposable biosensors, onCall Information System. Instrument controls nerve conduction study; determines supramaximal stimulus intensity; acquires/analyzes motor and sensory waveforms; displays latency, conduction velocity, amplitude, and response parameters on LCD. Biosensors (single-use) contain electrodes for stimulation, response detection, and skin temperature measurement. Docking station downloads data via infra-red link to onCall Information System via analog phone line. onCall system generates hardcopy reports comparing results to normal ranges; reports sent via fax or email. Used by clinicians to diagnose neuropathies; provides objective data for clinical decision-making; benefits patients through standardized, automated nerve conduction assessment.
Clinical Evidence
Clinical data submitted demonstrates that nerve conduction measurements obtained using the NC-stat are comparable to those obtained using conventional nerve conduction measurement equipment.
Technological Characteristics
System includes handheld instrument, docking station, and disposable biosensors. Biosensors include electrodes for stimulation, response detection, and skin surface temperature measurement. Connectivity via infra-red link and analog phone line. Instrument features LCD readout for real-time data display. Software controls stimulus intensity, waveform acquisition, and analysis.
Indications for Use
Indicated for patients requiring diagnosis and evaluation of systemic and entrapment neuropathies.
Regulatory Classification
Identification
A nerve conduction velocity measurement device is a device which measures nerve conduction time by applying a stimulus, usually to a patient's peripheral nerve. This device includes the stimulator and the electronic processing equipment for measuring and displaying the nerve conduction time.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a rectangular box with the text 'K060584' written inside. The text is written in a bold, handwritten style. The box is outlined with a thick black line.
Page 6 of 3
# 510(k) Summary for NeuroMetrix NC-stat
JUL 3 1 2006
#### 1. SPONSOR
NeuroMetrix, Inc. 62 Fourth Avenue Waltham, MA 02451
Contact Person: Rainer Maas Telephone: (781) 314-2781
Date Prepared: July 26, 2006
#### 2. DEVICE NAME
| Proprietary Name: | NC-stat |
|----------------------|--------------------------------------------|
| Common/Usual Name: | Nerve Conduction Testing System |
| Classification Name: | Nerve Conduction Velocity Measuring Device |
#### 3. PREDICATE DEVICES
- NC-stat (K982359, K000565, K003508, K013459 and K041320) .
- TECA TD-10/TD-20 (K802637) .
#### 4. INTENDED USE
The NeuroMetrix NC-stat is intended to stimulate and measure neuromuscular signals that are useful in diagnosing and evaluating systemic and entraponent neuropathies.
#### ട. DEVICE DESCRIPTION
The NC-stat system consists of the following four components:
- An instrument that contains the electronic circuitry and software required 5.1 to control a nerve conduction study, including determination of the supramaximal stimulus intensity, acquisition and analysis of motor and -
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "page 2 of 3" written in cursive. The number 2 is circled. The text appears to be handwritten.
sensory waveforms, and display of latency/conduction velocity, amplitude and response configuration parameters. The instrument saves patient and study data for optional transmission to a docking station for real-time hard copy report generation. The instrument displays nerve conduction data in real-time on the LCD readout, including the distal motor latency (DML), Compound Muscle Action Potential (CMAP) amplitude, CMAP duration, CMAP area, motor conduction velocity (CV), F-wave latency, distal sensory latency (DSL), the Sensory Nerve Action Potential (SNAP) amplitude, SNAP duration, sensory CV, limb indicator (left or right), low battery indicator, the memory slot being used to store the test data, and user messages (menu selections, sensor serial numbers, device status, operator instructions, and error conditions).
- A stationary docking station is used to download the test data from the 5.2 instrument via an infra-red communication link to the onCall Information System via an analog phone line. The docking station is non-transportable and connected to both a power outlet and an analog phone line.
- Single-use, disposable biosensors are available for the median motor, 5.3 ulnar motor, median motor & sensory, ulnar motor & sensory, deep peroneal, posterior tibial, and sural nerves. The biosensors contain electrodes for nerve stimulation, response detection, and skin surface temperature measurement.
- The onCall Information System for automatic generation of the hardcopy 5.4 patient test report, which includes test results (DML, Compound Muscle Action Potential - CMAP- amplitude, DSL, SNAP amplitude, conduction velocity, F-wave latency, and associated response waveforms) and comparison of patient results to normal ranges. Reports are sent to the user by facsimile or e-mail.
#### BASIS FOR SUBSTANTIAL EQUIVALENCE 6.
The NC-stat that is the subject of this 510(k) Premarket Notification is substantially equivalent to the NC-stat as previously cleared for marketing and to the TECA TD-10/TD-20 EMG and nerve conduction velocity measurement device. The purpose of this 510(k) is to describe changes made to the NC-stat, including:
- Addition of the modified median motor-sensory biosensor for use in 6.1 diagnosing neuropathies affecting the median motor-sensory nerves.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "Page (3) of 3" written in cursive. The number 3 is circled. The text appears to be handwritten, and it is likely part of a document or report indicating the page number.
### 6.2 Modifications to the labeling.
None of the changes to the NC-stat have altered the intended use of the device, the basic device design or the device operation. Clinical data submitted in the 510(k) demonstrates that nerve conduction measurements obtained using the NCstat are comparable to those obtained using conventional nerve conduction measurement equipment.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the words "U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three wavy lines above them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 3 1 2006
NeuroMetrix, Inc. % Rainer Maas Director, Quality Assurance and Regulatory Affairs Associate 62 Fourth Avenue Waltham, Massachusetts 02451
Re: K060584 Trade/Device Name: NC-stat Regulation Number: 21 CFR 882.1550 Regulation Name: Nerve conduction velocity measurement device Regulatory Class: II Product Code: JXE, IKN Dated: July 5, 2006 Received: July 6, 2006
Dcar Rainer Maas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{5}------------------------------------------------
### Page 2 - Rainer Maas
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely vours.
Helmut Lachenmann
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Intended Use
510(k) Number (if known): K060584
Device Name: NC-stat
Intended Use:
The NeuroMetrix NC-stat is intended to to stimulate and measure neuromuscular signals that are useful in diagnosing and evaluating systemic and entrapment neuropathies.
Prescription Use _____________________________________________________________________________________________________________________________________________________________
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE If NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hulbert Brewer
(Division Simmal 317 Division of General, Restorative, and Neurologica. Devices
510(k) Number Ko 60584
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.