BREAS VIVO BILEVEL SYSTEM, MODEL VIVO 30

K060546 · Breas Medical AB · MNS · May 26, 2006 · Anesthesiology

Device Facts

Record IDK060546
Device NameBREAS VIVO BILEVEL SYSTEM, MODEL VIVO 30
ApplicantBreas Medical AB
Product CodeMNS · Anesthesiology
Decision DateMay 26, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Breas Vivo 30 is designed for spontaneous breathing patients who require long-term support by mechanical ventilation during night and part of the day. The Breas Vivo 30 shall only be used by patients with spontaneous breathing. The Breas Vivo 30 is not intended for life-sustaining applications or for transport of critical care patients. The Breas Vivo 30 is intended for non-invasive use. The Breas Vivo 30 is intended for treatment of adult (who weigh more than 30 kg) patients. The Breas Vivo 30 is intended to be operated by qualified and trained personnel.

Device Story

Vivo 30 is a microprocessor-controlled, non-invasive bilevel ventilator; provides Pressure Controlled Ventilation (PCV), Pressure Support Ventilation (PSV), or CPAP. Inputs include patient respiratory effort; device regulates airflow via blower based on software-controlled pressure, flow, and time parameters. Output is delivered via single lumen tube to non-invasive interfaces (e.g., nasal masks). Used in clinical settings (hospitals, sleep labs) and home environments; operated by qualified personnel under physician prescription. Includes display indicators, audible alarms, and clinician-controlled lockout features. Benefits patients with chronic respiratory failure by ensuring sufficient gas exchange. Not for life support.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including waveform performance, work of breathing, pressure dynamic regulation, electrical safety (IEC 601-1), electromagnetic compatibility, and mechanical/environmental testing. Software validated per FDA 2005 guidance.

Technological Characteristics

Microprocessor-controlled blower; non-invasive; single lumen outlet. Software-based pressure, flow, and time regulation. Power: 100-240 VAC or 12.5/24 VDC. Dimensions: 7.2 x 8.9 inches; weight: 7.3 lbs. Standards: IEC 601-1, ISO 10651-6, ISO 17510-1. Connectivity: Standalone.

Indications for Use

Indicated for adult patients (>30 kg) with spontaneous breathing requiring long-term mechanical ventilation support during night and part of the day. Not for life-support or critical care transport.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K060546 510(k) Premarket Notification_________________________________________________________________________________________________________________________________________________ MAY 2 6 2006 Breas Vivo™ 30 System Breas Medical AB . # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS | Submitter | Breas Medical AB | |---------------------|--------------------------------------------------| | | Företagsvagen 1 | | | SE 435 33 Molnlycke | | | Sweden | | Contact Person | Karl- Johan Holm | | | Quality Assurance and Regulatory Affairs Manager | | | Phone: +46 31 868830 | | | Fax: +46 31 868810 | | Summary Date | February 13, 2006 | | Name of Device | Breas Vivo 30 System | | Common Name | Bilevel system | | Classification Name | Non- continuous ventilator (21 CFR 868.5895) | | Product Code | MNS | | Predicate Device | ResMed VPAP II ST-A (K974417) | {1}------------------------------------------------ ### Device Description: The Vivo 30 is a pressure-supported and pressure-controlled ventilator with a CPAP function intended for spontaneous breathing patients who require long-term support by mechanical ventilation during night and part of the day. In the treatment of chronic respiratory failure, positive airway pressure ventilation is well established and common practice as a mean to assure sufficient gas exchange. There are a number of devices legally marketed in the United States for this application. The Vivo 30 can be used in clinical settings (e.g., hospitals, sleep laboratories, sub-acute care institutions) and home environments. It must always be prescribed by a licensed physician. It is not intended for life support applications or for transport of critical care patients. The therapy delivered by the Breas Vivo 30 System can be either: - 1) Pressure Controlled Ventilation (PCV) or - 2) Pressure Support Ventilation (PSV) or - 3) Constant Positive Airway Pressure (CPAP) The Vivo 30 airflow is delivered via a single lumen outlet tube that may be connected to various non-invasive patient interfaces, such as nasal masks. To minimize CO2 rebreathing, masks or other interfaces permitting a leak flow of at least 12 liters/minute at the output pressure setting of 4 cmH20 are recommended. The Vivo 30 has an auto-switching power supply that facilitates use in conjunction with international travel (100 - 240 VAC). It can also be used with an external 12.5/ 24 VDC power source when AC mains line voltage is not available. The outer dimensions of the Vivo 30 housing are 7.2 × 8.9 inches, and the device weighs 7.3 pounds. ### Intended Use: The Breas Vivo 30 is designed for spontaneous breathing patients who require long-term support by mechanical ventilation during night and part of the day. The Breas Vivo 30 shall only be used by patients with spontaneous breathing. The Breas Vivo 30 is not intended for life-sustaining applications or for transport of critical care patients. {2}------------------------------------------------ The Breas Vivo 30 is intended for non-invasive use. The Breas Vivo 30 is intended for treatment of adult (who weigh more than 30 kg) patients. The Breas Vivo 30 is intended to be operated by qualified and trained personnel. ## Comparison of Use and Technological Characteristics: The Vivo 30 can be used in clinical settings (e.g., hospitals, sleep laboratories, sub-acute care institutions) and home environments. It must always be prescribed by a licensed physician. It is not intended for life support applications or for transport of critical care patients. As compared with the cited predicate device, the Breas Vivo 30 System has: Same intended uses Same environments of use Similar design (microprocessor-controlled blower as air source) Same technology (software based pressure-, flow- and time- regulation) The differences that do exist are minimal and involve primarily user preference features. The Breas Vivo 30 System has additional display indicator and audible alarm features as well as lockout features (to provide clinicians with optional means to control the ability of patients to change pressure settings). These features are described in labeling for the device that includes an Operator Manual. # Summary of Performance Testing: - 1. Non-clinical testing was conducted to verify that the Breas Vivo 30 System is capable of meeting its stated performance specifications and that all Risk Analysis issues have been appropriately addressed. The device passed all tests. - 2. Comparative testing to predicate devices was performed. This bench- testing confirmed that the Breas Vivo 30 System is substantial equivalent with regards to Wave-form performance as well as Work of Breathing and Pressure Dynamic regulation. - 3. Testing was conducted to demonstrate compliance with applicable requirements in the November 1993 draft "Reviewer Guidance for Premarket Notification Submissions" published by the FDA's Division of Cardiovascular, Respiratory, {3}------------------------------------------------ and Neurological Devices and the July 1995 "Draft Reviewer Guidance for Ventilators". The testing included but was not limited to: - Electrical Safety testing per IEC 601-1 . - Safety and Essential Performance testing per ISO 10651-6 � - . Safety and Performance testing per ISO 17510-1 - Electromagnetic Compatibility testing (EMC testing) - Mechanical Safety testing - . Environmental testing - Functional testing . - Particle matter testing The device passed all tests. - 4. All device ( the embedded software and the Calendar Analysis PC software ) softwares was documented and tested in accordance with the FDA's May 11, 2005 "Guidance for the content of Premarket Submissions for Software Contained in Medical Devices". The device passed all tests. - 5. Clinical studies were not required to support a substantial equivalence determination. #### Conclusions: The Breas Vivo 30 System meets its stated performance specifications and criteria outlined in the Reviewer Guidance publications referenced above. We conclude that the device is capable of operating safely in its intended environments and will be effective in fulfilling its intended use. {4}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three horizontal lines forming the body and head. Below the bird are curved lines representing water or waves. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # MAY 2 6 2006 Mr. Karl-Johan Holm Quality Assurance & Regulatory Affairs Mananger Breas Medical AB Företagsvagen 1 Molnlycke, Vastra Gotaland SWEDEN 'SE 435 33 Re: K060546 Trade/Device Name: Breas Vivo 30 System Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: MNS Dated: February 22, 2006 Received: March 1, 2006 Dear Mr. Holm: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ #### Page 2 – Mr. Holm Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # INDICATIONS FOR USE 510(k) Number (if known): K060546 Device Name: . Breas Vivo 30 System Indications For Use: The Breas Vivo 30 is designed for spontaneous breathing patients who require long-term support by mechanical ventilation during night and part of the day. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|---------| | | | | of Anesthesiology, General Hospital, | | | on Control, Dental Devices | | | | K060546 |
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%