← Product Code [DXN](/productcode/DXN) · K060504

# IDEAL LIFE BP-MANAGER, MODEL BPM 0001; IDEAL LIFE POD, MODEL ILP 0001 (K060504)

_Ideal Life, Inc. · DXN · Mar 14, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060504

## Device Facts

- **Applicant:** Ideal Life, Inc.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Mar 14, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The IDEAL LIFE BP-MANAGER™ is a non-invasive blood pressure monitor intended for the measurement of systolic and diastolic blood pressure and heart rate (pulse rate) using the oscillometric technique. The IDEAL LIFE BP-MANAGER™ should be used in adults (individuals aged 18 and older) in a non-clinical environment such as in the home. The IDEAL LIFE BP-MANAGER™ should not be used on infants or children. The end user of this device should not have common arrhythmias, such as atrial or ventricular premature beats or atrial fibrillation. This device makes no interpretation, evaluation, medical judgments or recommendations for treatment. Clinical judgment and experience are required for interpretation of information. This device is not intended as a substitute for medical care. The IDEAL LIFE POD™ is an optional accessory to the IDEAL LIFE BP-MANAGER™ and is a transmitter that is intended to transmit data obtained by the IDEAL LIFE BP-MANAGER™ to the internet via common telephone lines from the user's home setting. The IDEAL LIFE POD™ is an optional accessory designed to assist in the management of user information. This device makes no interpretation, evaluation, medical judgments or recommendations for treatment. Clinical judgment and experience are required to check or interpret information transmitted. This device is not intended as a substitute for medical care.

## Device Story

BP-MANAGER™: oscillometric blood pressure monitor; measures systolic/diastolic BP and heart rate; stores up to 70 readings; displays results on integrated screen. POD™: optional wireless router accessory; transmits stored data from BP-MANAGER™ to internet via telephone lines. Used in home settings by patients; no clinical interpretation provided by device; requires physician consultation for result evaluation. Benefits: facilitates remote data management for patient health tracking.

## Clinical Evidence

Performance testing conducted per 'Non-Invasive Blood Pressure (NIBP) Monitor Guidance' (1997). Includes electromagnetic compatibility, electrical safety, and performance testing in human volunteers. No specific clinical trial metrics (e.g., sensitivity/specificity) provided in summary.

## Technological Characteristics

Oscillometric blood pressure measurement; software-controlled; includes blood pressure cuff; BP-MANAGER™ stores 70 readings; POD™ acts as RF transmitter/wireless router for internet connectivity via telephone lines. Class II device.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Omron Automatic Blood Pressure Monitor with Intellisense™, Model HEM-773AC ([K021682](/device/K021682.md))
- IN TOUCH® Diabetes Management Software System ([K984527](/device/K984527.md))
- Carematix Wellness System ([K031840](/device/K031840.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

IDEAL LIFE INC.

# MAR 1 4 2006 IDEAL LIFE BP-MANAGER™ 510(k) IDEAL LIFE POD™ 510(k)

# SUMMARY OF SAFETY AND EFFECTIVENESS

#### GENERAL INFORMATION l

#### 1.1 Submitter and Owner of the 510(k)

Jason Goldberg IDEAL LIFE INC. 110 Eglinton Avenue West Suite 303 Toronto, Ontario Canada M4R 1A3

#### 1.2 Official Correspondent

Diane Mandell, Ph.D., RAC THE WEINBERG GROUP INC. 1220 19th Street N.W., Suite 300 Washington, D.C. 20036 Telephone: 202.730.4130 Facsmile: 202.833.7057 Electronic mail: dima@weinnberggroup.com

### 1.3 Date of Preparation

February 6, 2006

#### 2 NAME OF THE DEVICES

#### Trade/Proprietary Names 2.1

IDEAL LIFE BP-MANAGER™, Model BPM 0001; IDEAL LIFE POD™, Model ILP 0001

#### 2.2 Common/Usual Names

Noninvasive Blood Pressure Measurement System: Radiofrequency Physiological Signal Transmitter and Receiver

#### 2.3 Classification Information

Classification Name: System, Measurement, Blood-Pressure, Noninvasive Classification Regulation:21 CFR §870.1130 Class: I I Product Code: DXN Panel: Circulatory System Devices

{1}------------------------------------------------

IDEAL LIFE INC.

| Classification Name:       | Transmitters and Receivers, Physiological Signal,<br>Radiofrequency |
|----------------------------|---------------------------------------------------------------------|
| Classification Regulation: | 21 CFR §870.2910                                                    |
| Class:                     | II                                                                  |
| Product Code:              | DRG                                                                 |
| Panel:                     | Circulatory System Devices                                          |

#### 3 PREDICATE DEVICES

- 1. Omron Automatic Blood Pressure Monitor with Intellisense™, Model HEM-773AC (K021682),
- 2. IN TOUCH® Diabetes Management Software System, K984527; and
- 3. Carematix Wellness System, K031840.

#### 4 DESCRIPTION OF THE DEVICES

The IDEAL LIFE BP-MANAGER™ is a single unit software-controlled device with a blood pressure cuff, and operation is automatic. The IDEAL LIFE BP-MANAGFER™ can bc used alone without any other accessories. Blood pressure and heart rate results are displayed on the screen of the IDEAL LIFE BP-MANAGER™. Up to 70 blood pressure readings are stored in the blood pressure computer processor. The device is not intended for self-diagnosis of discase and it is expected that individuals using the IDEAL LIFE BP-MANAGER™ will consult with a physician for interpretation of results.

If an individual desires automated graphical presentation of their blood pressure and heart rate data, the IDEAL LIFE POD™ may be purchased separately and used with the IDEAL LIFE BP-MANAGER™ as an optional accessory. The IDEAL LIFE POD™ is not required for operation of the IDEAL LIFE BP-MANAGER™ . The IDEAL LIFE POD™ is a simple transmitter (i.e., a wireless router) that is intended to transmit data obtained by the IDEAL LIFE BP-MANAGER™ to the internet via common telephone lines from the user's home setting.

## v. INDICATIONS FOR USE AND INTENDED USE

The IDEAL LIFE BP-MANAGER™ is a non-invasive blood pressure monitor intended for the measurement of systolic and diastolic blood pressure and heart rate (pulse rate) using the oscillometric technique.

The IDEAL LIFE BP-MANAGER™ should be used in adults (individuals aged 18 and older) in a non-clinical environment such as in the home. The IDEAL LIFE BP-MANAGER™ should not be used on infants or children. The end user of this device should not have common arrhythmias, such as atrial or ventrie ular ance or mis cats or atrial fibrillation. This device makes no interpretation, evaluation, medical judgments or

{2}------------------------------------------------

KOLOSC4 13/3

recommendations for treatment. Clinical judgment and experience are required for interpretation of information. This device is not intended as a substitute for medical care. The IDEAL LIFE POD™ is an optional accessory to the IDEAL LIFE BP-MANAGER™ and is a transmitter that is intended to transmit data obtained by the IDEAL LIFE BP-MANAGER™ to the internet via common telephone lines from the user's home setting.

The IDEAL LIFE POD™ is an optional accessory designed to assist in the management of user information. This device makes no interpretation, evaluation, medical judgments or recommendations for treatment. Clinical judgment and experience are required to check or interpret information transmitted. This device is not intended as a substitute for medical care.

#### 6 SUBSTANTIAL EQUIVALENCE

The IDEAL LIFE BP-MANAGER™ (Model No. BPM 0001) is substantially equivalent to the K021682, Omron Automatic Blood Pressure Monitor with Intellisense™, Model HEM-773AC. These devices have similar intended uses and technology comparisons.

The IDEAL LIFE POD™ (Model No. ILP 0001) is substantially equivalent to two predicate devices: K984527, IN TOUCH® Diabetes Management Software System (similar in intended use); and K031840, Carematix Wellness System (similar in technological comparison). Any differences between these devices and the IDEAL I.IFLi devices do not affect safety and effectiveness of the IDEAL LIFE devices, as demonstrated through bench and clinical testing of the IDEAL LIFE devices.

## 7 PERFORMANCE TESTING

This 510(k) provided performance testing data in accordance with the guidance document entitled "Non-Invasive Blood Pressure (NIBP) Monitor Guidance" dated March 10, 1997. These performance data included electromagnetic compatibility, electrical testing, performance testing in human volunteers, as well as other performance tests. Results of these tests confirm the appropriate performance of both the IDEAL LIFE BP-MANAGER™ and the IDEAL LIFE POD™, and the substantial equivalence of these devices to the predicates.

# 8 CONCLUSIONS

This 510(k) submission demonstrates that the IDEAL LIFE BP-MANAGER™ and the IDEAL LIFE POD™ are substantially equivalent to the three predicate devices.

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

MAR 1 4 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ideal Life, Inc. c/o Diane Mandell, Ph.D., RAC The Weinberg Group, Inc. 1220 19th Street N. W., Suite 300 Washington, DC 20036

Re: K060504

Trade Name: IDEAL LIFE BP-MANAGER™, Model BPM 0001 and IDEAL LIFE POD™, Model ILP 0001 Regulation Number: 21 CFR 870.1130 and 21 CFR 870.2910 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN and DRG Dated: February 06, 2006 Received: February 27, 2006

Dear Dr. Mandell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Diane Mandell, Ph.D., RAC

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Blyminman for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K060504

Attachment 1b

Page 1 of 1

510(k) Number (if known): K060504
Device Name: IDEAL LIFE BP-MANAGERT™ and IDEAL LIFE PODTM

Indications for Use:

The IDEAL LIFE BP-MANAGER™ is a non-invasive blood pressure monitor intended for the measurement of systolic and diastolic blood pressure and heart rate (pulse rate) using the oscillometric technique.

The IDEAL LIFE BP-MANAGER™ should be used in adults (individuals aged 18 and older) in a non-clinical environment such as in the home. The IDEAL LIFE BPMANAGER ™ should not be used on infants or children. The end user of this device should not have common arrhythmias, such as atrial or ventricular premature beats or atrial fibrillation. This device makes no interpretation, evaluation, medical judgeners or recommendations for treatment. Clinical judgment and experience are required for interpretation of information. This device is not intended as a substitute for reedical care.

The IDEAL LIFE POD™ is an optional accessory to the IDEAL LIFE BPMANAGER TM and is a transmitter that is intended to transmit data obtained by the IDEAL LIFE BP-MANAGER™ to the internet via common telephone lines from the user's home setting. The IDEAL LIFE POD™ is an optional accessory designed to ssist in the management of user information. This device makes no interpretation, evaluation, medical judgments or recommendations for treatment. Clinical judgment and experience are required to check or interpret information transmitted. This device is not intepded as a substitute for medical care.

(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Atymauma
(Division Sign-Off)

| Prescription Use   |  |
|--------------------|--|
| (Per 21CFR801.109) |  |

()[ર

Over-the-Counter Use - X

---

**Source:** [https://fda.innolitics.com/device/K060504](https://fda.innolitics.com/device/K060504)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
