DADE PFA-100 PLATELET FUNCTION ANALYZER AND REAGENTS

K060489 · Dade Behring, Inc. · JOZ · Jul 21, 2006 · Hematology

Device Facts

Record IDK060489
Device NameDADE PFA-100 PLATELET FUNCTION ANALYZER AND REAGENTS
ApplicantDade Behring, Inc.
Product CodeJOZ · Hematology
Decision DateJul 21, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5700
Device ClassClass 2
AttributesReal-World Evidence, Pediatric

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K060489 · Jul 21, 2006DADE PFA-100 PLATELET FUNCTION ANALYZER AND REAGENTSDade Behring, Inc.Published clinical literature; Retrospective clinical study cohorts; Prospective observational clinical study cohortsThe sponsor utilized multiple published clinical studies to support the performance of the PFA-100 system in pediatric populations and in the management of bleeding disorders (VWD, platelet defects) and therapy monitoring (DDAVP).Pediatric performance; Bleeding disorders; VWD; DDAVP therapy monitoring; Literature review

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Cariappa et al.; Observational studyPediatric patients with platelet defects and VWD; Sample Size: 52Not applicable for this studySensitivity, specificity, and efficiency of PFA-100 in detecting platelet defects and VWD
Lippi et al.; Observational studyPediatric patients; Sample Size: 52Normal adultsComparison of PFA closure times (CT) between pediatric patients and normal adults
Fressinaud et al.; Observational studyPatients with congenital VWD; Sample Size: 41Not applicable for this studyNormalization of PFA CT after DDAVP infusion
Koscielny et al.; Combined prospective and retrospective studyPatients with VWD, aspirin-induced platelet defects, and other platelet defects; Sample Size: 5649Patients with normal hemostasisPFA CT prolongation, normalization after DDAVP, and blood usage outcomes

Indications for Use

The Dade® PFA-100® Platelet Function Analyzer and associated reagents are in vitro diagnostic devices intended to aid in the detection of platelet dysfunction in citrated human whole blood.

Device Story

System simulates injured blood vessel under high shear stress to measure primary hemostasis; uses test cartridge with membrane coated in collagen/epinephrine or ADP. Instrument measures time from test start until platelet plug occludes aperture; reports result as Closure Time (CT). Used in clinical settings to aid detection of platelet dysfunction; assists clinicians in bleeding management and monitoring DDAVP therapy. Output provides quantitative time interval; informs clinical decision-making regarding platelet function status.

Clinical Evidence

No new clinical data provided. Performance characteristics rely on previously cleared submissions (K970505, K002885). Labeling modifications regarding pediatric reference intervals and DDAVP therapy management are supported by literature references.

Technological Characteristics

In vitro diagnostic system; automated platelet aggregation measurement. Uses collagen/epinephrine (CEPI) and collagen/ADP (CADP) cartridges. Operates via high shear stress simulation in aperture. Standalone benchtop analyzer.

Indications for Use

Indicated for the detection of platelet dysfunction in citrated human whole blood, including intrinsic platelet defects, von Willebrand factor functional deficiencies, or exposure to platelet inhibiting agents.

Regulatory Classification

Identification

An automated platelet aggregation system is a device used to determine changes in platelet shape and platelet aggregation following the addition of an aggregating reagent to a platelet-rich plasma.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: K060489 B. Purpose for Submission: To obtain a change in indications for the Dade® PFA-100® Platelet Function Analyzer and Dade® PFA-100® Reagents. C. Measurand: Platelet function D. Type of Test: Aggregation E. Applicant: Dade Behring Inc. F. Proprietary and Established Names: Dade® PFA-100® Platelet Function Analyzer and Dade® PFA-100® Reagents G. Regulatory Information: 1. Regulation section: 21 CFR 864.5700 2. Classification: Class II 3. Product code: JOZ 4. Panel: {2} 81 Hematology H. Intended Use: 1. Intended use(s): The Dade® PFA-100® Platelet Function Analyzer and Dade® PFA-100® Reagents are in vitro diagnostic devices intended to aid in the detection of platelet dysfunction in citrated human whole blood. 2. Indication(s) for use: To aid in the detection of platelet dysfunction in citrated human whole blood. 3. Special conditions for use statement(s): 4. Special instrument requirements: I. Device Description: The PFA-100® system provides a tool for clinicians to use in the detection of platelet dysfunction induced by intrinsic platelet defects, von Willebrand factor (vWF) functional deficiencies, or exposure to platelet inhibiting agents. The PFA-100® system simulates, under high shear stress, the interaction of platelets with an injured blood vessel. These conditions allow the PFA-100® system to measure in vitro platelet function as related to primary hemostasis. The PFA-100® instrument determines the time from the start of the test until the platelet plug occludes the aperture, and reports the time interval as the Closure Time (CT). J. Substantial Equivalence Information: 1. Predicate device name(s): Dade® PFA-100® Platelet Function Analyzer and Dade® PFA-100® Reagents 2. Predicate 510(k) number(s): K970505 & K002885 3. Comparison with predicate: The PFA-100® system was described in detail and cleared under 510(k) Premarket Notification Document Control No. K970505. Subsequently, the system was cleared for a new indication, testing pediatric populations, under 510(k) Premarket Control No. K002885. {3} | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | PFA-100® | PFA-100® (K970505 & K002885) | | Intended Use | Same | To aid in the detection of platelet dysfunction in citrated human whole blood | | Indications | Same | ■ Normal population ■ ASA-induced platelet dysfunction ■ vWD patients ■ Glanzmann’s thrombasthenia ■ Adult & pediatric populations | | Principle | Same | Measures time required for platelets to adhere and aggregate resulting in occlusion of an aperture under high shear | | Reagents | Same | A test cartridge unit containing a membrane coated with collagen and epinephrine or ADP | | Specimens | Same | Whole blood with sodium citrate anticoagulant | | Test Results | Same | Closure time in seconds; aperture occlusion | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | PFA-100® | PFA-100® (K970505 & K002885) | | Indications | Bleeding management | (listed above) | K. Standard/Guidance Document Referenced (if applicable): L. Test Principle: The Dade® PFA-100® system measures the time required for platelets to adhere and aggregate resulting in occlusion of an aperture under high shear conditions. M. Performance Characteristics (if/when applicable): * Note: Dade Behring, Inc. proposed modifications to the PFA-100® system based on literature references published by independent organizations. The performance characteristics of the instrument and reagents were not modified. The Clinical Performance section of the Dade® PFA-100® Reagent package insert has been modified as follows: - The existing pediatric indication was modified to include more information regarding the use of the device as a screening assay and to add information for reference intervals for pediatric populations. - A new indication is added for using the PFA-100® system as a testing aid in the management of desmopressin acetate (DDAVP) therapy. {4} The modifications of the proposed PFA-100® system are limited to labeling modification to the package insert which has been revised to include the new device indications. No modifications have been made to the device intended use or PFA-100® instrument Operator's Manual. 1. Analytical performance: Refer to previously cleared 510(k) submissions – K970505 & K 002885 a. Precision/Reproducibility: b. Linearity/assay reportable range: c. Traceability, Stability, Expected values (controls, calibrators, or methods): d. Detection limit: e. Analytical specificity: f. Assay cut-off: 2. Comparison studies: Refer to previously cleared 510(k) submissions – K970505 & K 002885 a. Method comparison with predicate device: b. Matrix comparison: 3. Clinical studies: Refer to previously cleared 510(k) submissions – K970505 & K 002885 a. Clinical Sensitivity: b. Clinical specificity: c. Other clinical supportive data (when a. and b. are not applicable): 4. Clinical cut-off: 5. Expected values/Reference range: N. Instrument Name: Dade® PFA-100® Platelet Function Analyzer and Dade® PFA-100® Reagents {5} O. System Descriptions: Refer to previously cleared 510(k) submissions – K970505 & K 002885 1. Modes of Operation: 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ________ or No ________ 3. Specimen Identification: 4. Specimen Sampling and Handling: 5. Calibration: 6. Quality Control: P. Other Supportive Instrument Performance Characteristics Data Not Covered In The “Performance Characteristics” Section above: Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...