HEALTHSTATS BPRO AMBULATORY BLOOD PRESSURE MONITORING SYSTEM
Applicant
Healthstats International Pte. , Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Apr 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The BPro® system is a noninvasive ambulatory blood pressure monitoring system based on arterial tonometry at the radial artery of the wrist. The system consists of a BPro wrist-mounted tonometric monitor, a BPCalibrator MC3000 oscillometric blood pressure monitor for calibration, and BProSoft software, a PC-based computer program. Before each measurement session, the wrist device is calibrated using the oscillometric monitor. Once calibrated, the BPro wrist device is capable of recording and displaying up to 96 measurements of systolic and diastolic blood pressure and pulse rate over a period of 24 hours. BProSoft PC software is used to provide data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to a 24-hour period. The BPro system is intended for use on patients who are eighteen (18) years and older and who have a palpable radial pulse. The BPro system is intended only for measurement, recording, and display. It makes no diagnosis. Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
Device Story
HealthSTATS BPro is a wrist-worn ambulatory blood pressure monitor using arterial tonometry at the radial artery. System includes BPro wrist monitor, BPCalibrator MC3000 (oscillometric) for pre-session calibration, and BProSoft PC software. Device records up to 96 measurements over 24 hours. Operated by patient during daily activity; data uploaded via RS-232 to PC for clinician review. Output provides 24-hour blood pressure profiles to assist clinicians in evaluating cardiovascular conditions. Device does not provide automated diagnosis; intended for measurement, recording, and display only.
Clinical Evidence
Clinical testing conducted in accordance with AAMI-SP10 (Electronic or Automated Sphygmomanometers). The device met requirements for stationary tests and recommendations for ambulatory tests. No specific sensitivity/specificity metrics provided; testing confirmed performance equivalent to predicate devices.
Technological Characteristics
Wrist-worn monitor with pressure transducer for arterial tonometry; oscillometric calibration via external monitor. Battery-operated with LCD display. Embedded microcontroller for measurement scheduling and data storage (up to 96 readings). RS-232 serial connectivity for PC data upload. Biocompatible skin-contact materials. Software verified and validated.
Indications for Use
Indicated for patients 18 years and older with a palpable radial pulse for noninvasive ambulatory blood pressure monitoring, recording, and display of systolic/diastolic pressure and pulse rate over 24 hours to assist qualified medical personnel in assessing cardiac health.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
CBM-3000/CBM-2000/CBM-7000 Blood Pressure System (K900247)
{0}------------------------------------------------
KOLO315
7 2006 APR
Page 1 of 4
510(k) Summary of Safety and Effectiveness in accordance with 21 CFR 807.92
| (a) (1) | Submitted by: | HealthSTATS International Pte. Ltd.<br>6 New Industrial Road #04-01/02<br>Hoe Huat Industrial Building<br>Singapore 536199<br><br>Phone: +65-6858 3248<br>Fax: +65-6858 0148<br><br>E-mail: cmting@healthstats.com.sg |
|---------|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person: | Dr. Ting Choon Meng, M.D. |
| | Position/Title: | President/CEO |
| | Date of Preparation: | February 4, 2006 |
| | Trade Name: | HealthSTATS BPro® Ambulatory Blood<br>Pressure Monitoring System |
| | Common/Classification Name: | System, Measurement, Blood-pressure,<br>Non-invasive |
| | Product Code: | 74 DXN, 21 CFR § 870.1130 |
| | Class: | Class II |
- Predicate Device(s): (3)
| K900247 | CBM-3000/CBM-2000/CBM-7000 Blood Pressure System,<br>Colin Medical Instruments Corp. |
|---------|---------------------------------------------------------------------------------------------------|
| K855127 | Ambulatory Blood Pressure Monitor #90202, Spacelabs, Inc. |
| K031479 | Ambulatory Blood Pressure (ABP) 92506 Report<br>Management System, (Spacelabs) Datex-Ohmeda, Inc. |
## Reason for Submission: New Device
## Description of Device: (4)
HealthSTATS BPro® is a noninvasive ambulatory blood pressure monitoring system based on arterial tonometry at the radial artery of the wrist. The system consists of three main elements:
- BPro/BPal wrist-mounted tonometric monitor. .
page 1 of 4
{1}------------------------------------------------
- . BProSoft report management software, a PC-based computer program.
- BPCalibrator MC3000 oscillometric blood pressure monitor [listed . device, K051546] for calibration of BPro/BPal monitors.
Once the BPro is calibrated using the BPcalibrator oscillometric device, it is capable of measuring and recording systolic pressure, diastolic pressure and pulse rate over a period of 24 hours.
### (5) Intended use:
Blood pressure is measured in millimeters of mercury (mmHg) and is represented by two values: systolic pressure and diastolic pressure. The systolic pressure represents the pressure in the blood vessels when the heart contracts (pumps), while the diastolic pressure is the pressure when the heart relaxes and blood fills the heart.
Blood pressure fluctuates throughout the day. Specific ranges are associated with normal, hyper- and hypotension. Ambulatory blood pressure measurement devices provide useful 24 hour blood pressure profiles. Evaluation of the profiles can assist with diagnosis of specific cardiovascular conditions.
## Indications for Use:
The BPro® system is a noninvasive ambulatory blood pressure monitoring system based on arterial tonometry at the radial artery of the wrist. The system consists of a BPro wrist-mounted tonometric monitor, a BPCalibrator MC3000 oscillometric blood pressure monitor for calibration, and BProSoft software, a PC-based computer program.
Before each measurement session, the wrist device is calibrated using the oscillometric monitor. Once calibrated, the BPro wrist device is capable of recording and displaying up to 96 measurements of systolic and diastolic blood pressure and pulse rate over a period of 24 hours.
BProSoft PC software is used to provide data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to a 24-hour period.
The BPro system is intended for use on patients who are eighteen (18) years and older and who have a palpable radial pulse.
The BPro system is intended only for measurement, recording, and display. It makes no diagnosis.
page 2 of 4
{2}------------------------------------------------
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
## (6) Technological Characteristics:
The BPro Ambulatory Blood Pressure Monitoring System combines two established measurement methods used by the referenced predicate devices:
- Blood pressure measurement at the wrist using arterial tonometry .
- . Oscillometric blood pressure measurement for calibration.
The BPro monitor is applied to the wrist with a pressure transducer placed over the radial artery. The monitor has a size and weight similar to a sports type wrist watch.
An embedded microcontroller supervises the actions of the monitor including scheduled measurements, on-demand measurements, and data storage of up to 96 systolic/diastolic/pulse rate readings in a 24 hour period.
The BPro monitor is battery operated and utilizes an LCD for display of the measurements. The BPro is capable of serial communication via an RS-232 port to upload ambulatory data to the BProSoft PC Report management software application. The BPal monitor is a reduced feature version of the BPro which does not support the serial data transmission.
The BPro/BPal devices measure a similar range of blood pressures as the predicate devices.
#### (b) (1) Non-Clinical Tests Submitted:
The BPro monitor devices were tested in accordance with applicable standards for medical device electrical safety, electromagnetic compatibility, shock and vibration, and environment (temperature and humidity). The devices passed the tests.
Materials utilized in skin contact surfaces were reviewed for conformance with biocompatibility requirements. The materials met the requirements.
System level tests of the BPro monitor included measurement accuracy, intra-device variability, life, and extended range bench testing. The devices tested satisfied performance criteria.
The BPro monitor embedded software and BProSoft PC software were verified to requirements and validated to meet intended use. Risk and failure mode analysis was performed and residual risks were determined to be acceptable.
page 3 of 4
{3}------------------------------------------------
## (2) Clinical Tests Submitted:
The BPro Ambulatory Blood Pressure Monitoring System was clinically tested in accordance with AAMI-SP10, Electronic or Automated Sphygmomanometers. The requirements of the standard were met for stationary tests and the recommendations of the standard were met for ambulatory tests.
## (3) Conclusions from Tests:
As described in (b)(1) and (b)(2) above, the testing demonstrates that the BPro® Ambulatory Blood Pressure Monitoring System is as safe and effective as, and functions in a manner equivalent to the predicate devices.
page 4 of 4
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
7 2006 APR
HealthSTATS International Pte. Ltd. c/o Ting Choon Meng, MD President/CEO 6 New Industrial Road #04-01/02 Hoe Huat Industrial Building SINGAPORE 536199
Re: K060315
Trade Name: HealthSTATS BPro® Ambulatory Blood Pressure Monitoring System Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: October 25, 2005 Received: February 08, 2006
Dear Mr. Meng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ting Choon Meng, MD
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please " contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
B.fremmatorfor
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
**510(k) Number (if known):** KC660315
Device Name:
# Indications for use:
The BPro® system is a noninvasive ambulatory blood pressure monitoring system based on arterial tonometry at the radial artery of the wrist. The system consists of a BPro wrist-mounted tonometric monitor, a BPCalibrator MC3000 oscillometric blood pressure monitor for calibration, and BProSoft software, a PC-based computer program.
Before each measurement session, the wrist device is calibrated using the oscillometric monitor. Once calibrated, the BPro wrist device is capable of recording and displaying up to 96 measurements of systolic and diastolic blood pressure and pulse rate over a period of 24 hours.
BProSoft PC software is used to provide data to qualified medical personnel for the purpose of assessing the patient's cardiac health via blood pressure readings taken during daily activity for up to a 24-hour period.
The BPro system is intended for use on patients who are eighteen (18) years and older and who have a palpable radial pulse.
The BPro system is intended only for measurement, recording, and display. It makes no diagnosis.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician or other licensed practitioner.
Prescription Use メ (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use_ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhemmmuma
de Sian-Off) Division of Cardiovascular Devices 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.