MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH VANCOMYCIN (0.25-64 UG/ML)
K060312 · Dade Behring, Inc. · LON · Mar 23, 2006 · Microbiology
Device Facts
Record ID
K060312
Device Name
MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH VANCOMYCIN (0.25-64 UG/ML)
Applicant
Dade Behring, Inc.
Product Code
LON · Microbiology
Decision Date
Mar 23, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.1645
Device Class
Class 2
Intended Use
Vancomycin at concentrations of 0.25 to 64 µg/mL on the MicroScan® Synergies plus ™ Gram-Positive MIC/Combo Panel is intended for use with MicroScan® Synergies plus ™ Panels read on the WalkAway® -SI System (including upgraded WalkAway® -40 or WalkAway® -96 to meet WalkAway® SI equivalence). MicroScan® panels are designed for use in testing colonies, grown on solid media, of rapidly growing aerobic and facultative anaerobic gram-positive cocci; the panels also provide quantitative and/or qualitative antimicrobial agent susceptibility for staphylococci and enterococci.
Device Story
MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels are miniaturized broth dilution susceptibility tests; dehydrated antimicrobial agents are rehydrated with Synergies plus™ Pos Broth after inoculation with standardized organism suspensions. Panels are incubated in WalkAway® SI System (or equivalent) for 4.5–18 hours. Device uses rapid fluorogenic identification and dried overnight AST technologies. Instrumentation reads minimum inhibitory concentration (MIC) by determining lowest antimicrobial concentration showing growth inhibition. Used in clinical laboratories by trained personnel to guide antimicrobial therapy for gram-positive infections. Benefits include rapid, quantitative susceptibility results for vancomycin (0.25–64 µg/ml) to inform clinical decision-making.
Clinical Evidence
Bench testing only. Performance evaluated using fresh and stock efficacy isolates and challenge strains. Compared against frozen reference panel. Overall essential agreement >97% for vancomycin. Reproducibility and precision confirmed using turbidity inoculum preparation and WalkAway® SI System. Quality control testing yielded acceptable results.
Technological Characteristics
Miniaturized broth dilution panels; dehydrated antimicrobial agents in water/buffer/broth. Sensing principle: fluorogenic identification and growth inhibition detection. Energy source: electrical (WalkAway® system). Connectivity: compatible with WalkAway® instrumentation. Sterilization: not specified. Software: automated reading of MIC via instrumentation.
Indications for Use
Indicated for testing Staphylococcus and Enterococcus spp. using vancomycin at 0.25–64 µg/mL concentrations at <16 hours or >16 hours (overnight).
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
K071268 — MICROSCAN SYNERGIES PLUS GRAM POSITIVE MIC/COMBO PANELS WITH TETRACYCLINE · Dade Behring, Inc. · Jun 26, 2007
K071076 — SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS · Dade Behring, Inc. · May 21, 2007
K071317 — MICROSCAN SYNERGIES PLUS GRAM POSITIVE MIC/COMBO PANELS NEW ANTIMICROBIAL-GENTAMICIN SYNERGY SCREEN · Dade Behring, Inc. · Aug 20, 2007
K051350 — SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS · Dade Behring, Inc. · Jul 25, 2005
K070860 — MICROSCAN SYNERGIES PLUS GRAM-POSITIVE MIC/COMBO PANELS WITH CLINDAMYCIN · Dade Behring, Inc. · May 11, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
MAR 2 3 2006
K060312
## 510(k) Summary
### 510(k) Submission Information:
| Device Manufacturer: | Dade Behring Inc. |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Contact name: | Robert Eusebio, Regulatory Affairs Manager |
| Fax: | 916-374-3144 |
| Date prepared: | January 24, 2006 |
| Product Name: | Microdilution Minimum Inhibitory Concentration (MIC) Panels |
| Trade Name: | MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels |
| Intended Use: | To determine antimicrobial agent susceptibility |
| 510(k) Notification: | New antimicrobial - Vancomycin |
| Predicate device: | MicroScan® Synergies plus™ Gram-Negative MIC/Combo Panels<br>and MicroScan® Dried Gram-Positive Panels |
#### 510(k) Summary:
MicroScan® Synergies plus Gram-Positive MIC/Combo Panels, utilizing both the MicroScan® Rapid Fluorogenic Identification and Dried Overnight Antimicrobial Susceptibility Testing (AST) technologies, are designed for use in determining quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive staphylococci and enterococci.
Note: Efficacy total for the Rapid read method equaled five hundred and seven (507) isolates. Results for forty-two (42) isolates were not reported due to read times ≥16 hours. Performance claims are based on the Rapid read method. Overnight Instrument and Manual read method(s) Essential Agreement & Categorical Agreement percentages were of equal, or greater, values than the Rapid read method.
The antimicrobial susceptibility tests are miniaturizations of the broth dilution susceptibility test that have been diluted in water and dehydrated. Various antimicrobial agents are diluted in water, buffer, or minute concentrations of broth, to concentrations bridging the range of clinical interest. Panels are rehydrated with Synergies plus™ Pos Broth, after inoculation with a standardized suspension of the organism. After incubation in the WalkAway® SJ System, or equivalent, for 4.5 - 18 hours, the minimum inhibitory concentration (MIC) for the test organism is read by determining the lowest antimicrobial concentration showing inhibition of growth.
The proposed MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel demonstrated substantially equivalent performance when compared with a frozen Reference Panel, as defined in the FDA document "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA", dated February 5, 2003. The Premarket Notification (510[k]) presents data in
{1}------------------------------------------------
support of the MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel with vancomycin.
The external evaluation was conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed Synergies plus™ Gram-Positive Panel by comparing its performance with a frozen Reference panel. Challenge strains were compared to Expected Results determined prior to the evaluation. The Synergies plus™ Gram-Positive Panel demonstrated acceptable performance with an overall Essential Agreement of >97% for vancomycin when compared with the frozen Reference panel.
Instrument reproducibility testing demonstrated acceptable reproducibility and precision for vancomycin, with Turbidity inoculum preparation method and the WalkAway® SI System or equivalent.
Quality Control testing demonstrated acceptable results for vancomycin.
{2}------------------------------------------------
DEPARTMENT OF HEAL TH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the symbol. The caduceus is depicted with three lines forming the wings and a stylized staff. The text is in all caps and is evenly spaced around the circle.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
MAR 2 3 2006
Mr. Robert Eusebio Regulatory Affairs Manager Dade MicroScan Inc. 1584 Enterprise Boulevard West Sacramento, CA 95691-9972
k060312 Re:
Trade/Device Name: MicroScan® Synergies plus TM Gram-Positive MIC/Combo Panels Vancomycin (0.25 - 64 µg/ml)
Regulation Number: 21 CFR 866.1645 Regulation Name: Fully automated short-term incubation cycle antimicrobial
susceptibility system.
Regulatory Class: Class II Product Code: LON, LRG, JWY, LTT Dated: February 6, 2006 Received: February 8, 2006
Dear Mr. Eusebio:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
#### Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) rins letter will and my you to organization of substantial equivalence of your device to a legally prematics notification " "I a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the I ou may octurity of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Sale, a Form
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): 长060312-
Device Name: MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panels with Vancomycin (0.25 - 64 ug/ml)
#### Indications For Use:
The MicroScan® Synergies plus™ Gram-Positive MIC/Combo Panel is used to determine quantitative and/or qualitative antimicrobial agent susceptibility of colonies grown on solid media of rapidly growing aerobic and facultative anaerobic gram-positive staphylococci and enterococci. After inoculation, panels are incubated for 4.5 - 18 hours at 35℃ +/- 1℃, in a WalkAway® SI, or equivalent, and read by the MicroScan® Instrumentation. Additionally, the panels may be incubated in a non-CO, incubator and the Antimicrobial Susceptibility Testing (AST) portions can be read visually, according to the Package Insert.
This particular submission is for the addition of the antimicrobial vancomycin, at concentrations of 0.25 to 64 µg/ml, to the test panel.
The gram-positive organisms which may be used for vancomycin susceptibility testing in this panel are:
- Enterococcus spp. (e.g., Enterococcus faecalis) �
- Staphylococcus spp. .
- Staphylococcus aureus (including methicillin-resistant strains) .
- Staphylococcus epidermidis (including methicillin-resistant strains) .
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) | |
|-------------------------------------------------------------------|--|
| Division Sign-Off | |
| Va 510k.doc | Office of In Vitro Diagnostic Device Evaluation and Safety | Page 1 of 1 |
|-------------|------------------------------------------------------------|-------------|
| 2/6/2006 | 510(k) K060512 | |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.