← Product Code [GEX](/productcode/GEX) · K060310

# FRAXEL II SR LASER SYSTEM AND ACCESSORIES (K060310)

_Reliant Technologies, Inc. · GEX · Mar 29, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060310

## Device Facts

- **Applicant:** Reliant Technologies, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Mar 29, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Fraxel II SR Laser System is intended for use in: Dermatological procedures requiring the coagulation of soft tissue; Treatment of periorbital wrinkles; Photocoagulation of pigmented lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), melasma and dyschromia; Skin resurfacing procedures.

## Device Story

Fraxel II SR Laser System uses fiber lasers with scanning and focusing optics to deliver thermal energy patterns to epidermis and dermis. Operated by clinicians in dermatological settings. System includes embedded processor to control laser delivery. Accessories include tip kits and pre-treatment solution. Output is controlled thermal energy for tissue coagulation and resurfacing. Clinical benefit includes wrinkle reduction and treatment of pigmented lesions.

## Clinical Evidence

Clinical analysis conducted on Non-Significant Risk and IDE Reliant studies. Data demonstrated device performs as intended with no new safety or effectiveness issues.

## Technological Characteristics

Fiber laser system; scanning and focusing optics for thermal energy delivery; embedded processor control. Class II device. No changes to laser wavelength or operating principle compared to predicate.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Fraxel SR Laser System ([K050841](/device/K050841.md))
- Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and Delivery Device Accessories ([K022060](/device/K022060.md))
- Fraxel SR Laser System ([K053047](/device/K053047.md))
- Reliant Laser System II ([K040617](/device/K040617.md))
- Reliant Laser System ([K031795](/device/K031795.md))

## Reference Devices

- [K042319](/device/K042319.md)
- [K040617](/device/K040617.md)
- [K040617](/device/K040617.md)
- [K031795](/device/K031795.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Reliant Technologies, Inc.

# MAR 2 9 2006

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

K060310

# REGULATORY AUTHORITY

Safe Medical Devices Act of 1990, 21 CFR 807.92

### COMPANY NAME/CONTACT

Heather Tanner Reliant Technologies, Inc. 464 Ellis St. Mountain View, CA 94043 650 641-5861 650 473-0537 fax htanner@reliant-tech.com

#### NAME OF DEVICE

Trade Name: Common Name: Regulation Number Product code: Device Panel: Device Classification: Fraxel II SR Laser System and Accessories Laser Surgical Instrument 878.4810 GEX General Surgery/Restorative Devices Class II

#### LEGALLY MARKETED PREDICATE DEVICES

Name: Fraxel SR Laser System and Accessories 510(k) #: K050841

Name: Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and Delivery Device Accessories 510(k) #: K022060

#### DEVICE DESCRIPTION

The Fraxel II SR Laser System consists of a set of fiber lasers, controlled by an embedded processor, to be used in dermatology. The laser system uses scanning and focusing optics to deliver a pattern of thermal energy to the epidermis and dermis. Device accessories include tip kits and pre-treatment solution.

{1}------------------------------------------------

#### INDICATION FOR USE STATEMENT

The Fraxel II SR Laser System is intended for use in:

Dermatological procedures requiring the coagulation of soft tissue;

Treatment of periorbital wrinkles;

Photocoagulation of pigmented lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots) and dyschromia;

Skin resurfacing procedures.

### SUBSTANTIAL EQUIVALENCE COMPARISON

#### Indications for Use

Substantial equivalence for the Fraxel II SR Laser System and Accessories is supported by the predicate devices listed in this submission, which have identical or similar indication statements.

#### Clinical Performance Data

Clinical analysis was conducted on Non-Significant Risk and IDE Reliant studies to support the clinical performance of the Fraxel SR Laser System. Sufficient safety data has been gathered to determine that the Fraxel II Laser System and Accessories performs as clinically intended and that no new issues of safety and effectiveness are introduced.

#### Technological Characteristics

Key technological characteristics of the Fraxel II SR Laser System, such as energy type and operating principle, are equivalent to the Fraxel SR Laser System as described in submission K050841, K042319, K040617 and K031795.

#### CONCLUSION

Based on the design, materials, function, intended use and clinical evaluation, the Fraxel II Laser System and Accessories is substantially equivalent to the devices currently marketed under the Federal Food, Drug and Cosmetic Act. No changes are being made in the laser wavelength or operating principle. Safety and effectiveness are reasonably assured, justifying 510(k) clearance.

{2}------------------------------------------------

| D |

# PREDICATE DEVICES

Name: Fraxel SR Laser System 510(k) #: K053047

Name: Reliant Laser System II 510(k) #: K040617

Name: Reliant Laser System 510(k) #: K031795

Name: Lumenis UltraPulse Encore Carbon Dioxide Surgical Laser and Delivery Device Accessories 510(k) #: K022060

# DEVICE DESCRIPTION

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features the department's name encircling a stylized eagle emblem. The eagle is depicted with three lines forming its body and wings, symbolizing health, human services, and well-being. The seal is presented in black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 9 2006

Reliant Technologies, Incorporated c/o Ms. Heather Tanner 464 Ellis Street Mountain View, California 94043

Re: K060310

Trade/Device Name: Fraxell II SR Laser System and Accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 2, 2006 Received: February 7, 2006

Dear Ms. Tanner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practicc requirements as set

{4}------------------------------------------------

Page 2 – Ms. Heather Tanner

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Halent Semen co

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): K 060310

Device Name: Fraxel II SR Laser System and accessories Indications For Use:

Dermatological procedures requiring the coagulation of soft tissue;

Treatment of periorbital wrinkles;

Photocoagulation of pigmented lesions, such as, but not limited to lentigos (age spots), solar lentigos (sun spots), melasma and dyschromia;

Skin resurfacing procedures."

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Helene Lemercier

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

Page 1 of

510(k) Number Kob6310

---

**Source:** [https://fda.innolitics.com/device/K060310](https://fda.innolitics.com/device/K060310)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
