K060266 · Dade Behring, Inc. · JIS · Mar 9, 2006 · Clinical Chemistry
Device Facts
Record ID
K060266
Device Name
DIMENSION IRON CALIBRATOR
Applicant
Dade Behring, Inc.
Product Code
JIS · Clinical Chemistry
Decision Date
Mar 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1150
Device Class
Class 2
Intended Use
The IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the IRON method for the Dimension® clinical chemistry system.
Device Story
Dimension® Iron Calibrator is an aqueous solution of iron wire dissolved in dilute HCl, provided in a 6-ampule kit across three concentration levels (0, 50, 1075 µg/dL). It is used in clinical chemistry laboratories to calibrate the IRON assay on Dimension® clinical chemistry systems. By establishing a calibration curve traceable to NIST SRM 937, the device enables the system to accurately quantify iron levels in patient samples. Healthcare providers use the resulting assay measurements to assess patient iron status and inform clinical decision-making. The calibrator ensures analytical accuracy and consistency across the assay range.
Clinical Evidence
No clinical data. Bench testing only. Stability testing performed over 13 months comparing product stored at 2-8°C vs -20°C. Stress testing included hot/cold shipping simulations and freeze/thaw cycles. Drift criteria established: ≤5 µg/dL (Level 1), ≤5% CV (Level 2), and ≤8% CV (Level 3).
Technological Characteristics
Aqueous solution of iron wire in dilute HCl. Traceable to NIST SRM 937. Three-level calibrator kit. Stability testing per CEN 13640. Risk management per ISO 14971:2000.
Indications for Use
Indicated for use as an in vitro diagnostic calibrator for the IRON method on the Dimension® clinical chemistry system. For prescription use only.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
Dimension® IRN/TIBC Calibrator (DC21) (k944093)
Related Devices
K983548 — FERRITIN CALIBRATOR · Dade Behring, Inc. · Oct 27, 1998
K994114 — DIMENSION TOTAL IRON BINDING CAPACITY (IBCT) CALIBRATOR · Dade Behring, Inc. · Jan 20, 2000
K061251 — DIMENSION VISTA TOTAL IRON BINDING CAPACITY CALIBRATOR · Dade Behring, Inc. · May 31, 2006
K052002 — IRON STANDARD · Roche Diagnostics Corp. · Aug 22, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
#### 2006 MAR 9
# 510(k) Summary for the Dimension® Iron Calibrator (IRON Cal - DC 85)
#### 长060266 A. 510(k) Number:
- B. Analyte: Iron Calibrator
- C. Type of Test: Calibrator Material
# D. Applicant:
Dade Behring Inc., P.O. Box 6101, Newark, DE 19714-6101 Contact: Andrea M. Tasker, Regulatory Affairs and Compliance Manager (302) 631-9454
Date of Preparation: January 30, 2006
# E. Proprietary and Established Names: Dimension® Iron Calibrator (IRON Calibrator- DC 85)
# F. Regulatory Information:
- 1. Regulation section: 21 CFR § 862-1150 -- CALIBRATOR
- 2. Classification: Class II
- 3. Product Code: JIS CALIBRATORS, PRIMARY
- 4. Panel: CLINICAL CHEMISTRY
## G. Intended Use:
- l . Intended use(s):
The IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the IRON method for the Dimension® clinical chemistry system.
- 2. Indication(s) for use:
The IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the IRON method for the Dimension® clinical chemistry system.
This product was designed to meet the needs of users to assure accurate results over the assay range of this method.
- 3. Special condition for use statement(s): none
- 4. Special instrument Requirements: none
{1}------------------------------------------------
# H. Device Description:
The IRON calibrator is an aqueous solution of iron wire dissolved in a dilute solution of HCl. The kit consists of 6 ampules, two at each of three levels.
# . Substantial Equivalence Information:
- 1. Predicate Device: Dimension® IRN/TIBC Calibrator (DC21)
# 2. Predicate K Number(s): K944093
# 3. Comparison with Predicate:
| Similarities | | |
|--------------------------|---------------------------------------------------------------------------------------------------------------|-------------------|
| Item | Device | Predicate |
| Intended Use | To calibrate the iron method for the<br>Dimension® clinical chemistry<br>system. | same |
| Traceability | NIST SRM 937 (NIST SRM: National<br>Institute of Standards and<br>Technology- Standard Reference<br>Material) | same |
| Matrix | Aqueous solution of iron wire<br>dissolved in a dilute solution of HCl | same |
| Number of<br>Levels | 3 | same |
| Differences | | |
| Item | Device | Predicate |
| Target<br>Concentrations | 0, 50, 1075 ug/dL | 50, 500 1000ug/dL |
# J. Standard/Guidance Document Referenced:
1. Guidance;
Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators; Final, 02/22/1999
Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use, 11/30/2004
2. Standards: Continuous Quality Improvement Essential Management Approaches GP22-A Stability testing of In-Vitro Diagnostic Devices CEN 13640
{2}------------------------------------------------
Medical devices -Application of risk management to medical devices ISO 14971:2000 Medical devices -- Symbols to be used with medical device labeling ISO 15223 and information to be supplied
## K. Test Principle:
The Dimension® Iron Calibrator is an in vitro diagnostic product intended to be used to calibrate the Dimension® IRON (Cat # DF85) method for the Dimension® clinical chemistry system.
# L. Performance Characteristics:
1. Stability
Target shelf life for the Dimension® IRON Calibrator is 12 months. Studies require 13 months of real time testing on three lots of product. Calibrator shelf life is determined by comparing results of the product stored at 4℃ with product stored at -20℃ to ensure that analytical system drift is dissociated with calibrator drift.
### 2. Traceability:
The assigned values of the IRON calibrator are standardized to NIST SRM 937 (NIST SRM: National Institute of Standards and Technology- Standard Reference Material). Six working Iron Standard Solutions of NIST SRM 937 (0, 25, 50, 75, 500, 1075 µg/dL) are prepared and used to assign each new lot of calibrator.
# 3. Value Assignment
The new calibrator levels are made by adding acid dissolved NIST 837 into base matrix. The new calibrator lots are assigned verses the six working standard solutions prepared from NIST SRM 937(Reference Lot). The new calibrator lot must have acceptable recovery verses the Reference Lot and a Control Calibrator Lot (Control Calibrator Lot = Any Approved Calibrator Lot).
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is positioned to the right of a circular text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
Public Health Service
#### 2006 MAR 9
Food and Drug Administratio 2098 Gaither Road Rockville MD 20850
Ms. Andrea M. Tasker Regulatory Affairs and Compliance Manager Dade Behring Inc. Glasgow Business Community P.O. Box 6101, Building 500, M/S 514 Newark, DE 19714-6101
Re: k060266
> Trade/Device Name: Dimension® IRON Calibrator (IRON Cal- DC85) Regulation Number: 21 CFR§ 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIS Dated: January 30, 2006 Received: February 1, 2006
Dear Ms. Tasker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html,
Sincerely yours,
Alberto Gutierrez, Ph.D.
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications For Use Statement
510(k) Number (if known):
Device Name:
Dimension® IRON Calibrator (IRON Cal- DC85)
# Indications for Use:
The IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the IRON method for the Dimension® clinical chemistry system.
Prescription Use 11 (Per 21 CFR 801 Subpart D)
AND/OR
Over-the-counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of -In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
313(x) k070266
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.