DIMENSION IRON FLEX REAGENT CARTRIDGE (IRON-DF85)

K060264 · Dade Behring, Inc. · JIY · Mar 8, 2006 · Clinical Chemistry

Device Facts

Record IDK060264
Device NameDIMENSION IRON FLEX REAGENT CARTRIDGE (IRON-DF85)
ApplicantDade Behring, Inc.
Product CodeJIY · Clinical Chemistry
Decision DateMar 8, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1410
Device ClassClass 1

Intended Use

The IRON method for the Dimension® clinical chemistry system is an in vitro diagnostic test intended to quantitatively measure iron in human serum and plasma. Iron measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia and other disorders of iron metabolism.

Device Story

The Dimension® IRON Flex® reagent cartridge is an in vitro diagnostic device used on the Dade Behring Dimension® clinical chemistry system. It quantitatively measures total iron in human serum and heparinized plasma. The device uses a chromatic assay principle: iron is released from transferrin in acidic conditions, reduced to Fe++ by ascorbic acid, and complexed with Ferene®. The system performs a bi-chromatic endpoint measurement at 600 and 700 nm. The absorbance is directly proportional to the iron concentration. The device is intended for professional use in clinical laboratories. Results assist healthcare providers in diagnosing and monitoring iron-related metabolic disorders. The system provides automated quantification, reducing manual processing requirements.

Clinical Evidence

Bench testing only. Precision/reproducibility evaluated per CLSI EP5-A2 (20 days, duplicate testing). Linearity confirmed (r=0.999) across 5.0–1000 µg/dL range. Analytical sensitivity is 5 µg/dL. Interference testing performed per CLSI EP7-A. Method comparison (n=147) against predicate showed high correlation (r=0.9996, slope 0.980). Matrix comparison (n=129) confirmed equivalence between serum and heparinized plasma.

Technological Characteristics

In vitro diagnostic reagent cartridge. Principle: Bi-chromatic endpoint spectrophotometry (600/700 nm). Reagents: Ferene® chromophore, thiourea, ascorbic acid. Standardization: NIST SRM 937. Connectivity: Integrated with Dimension® clinical chemistry system. Standards: CLSI EP5-A2, EP7-A, ISO 14971, ISO 15223.

Indications for Use

Indicated for quantitative measurement of iron in human serum and plasma to aid in the diagnosis and treatment of iron deficiency anemia and other iron metabolism disorders.

Regulatory Classification

Identification

An iron (non-heme) test system is a device intended to measure iron (non-heme) in serum and plasma. Iron (non-heme) measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia, hemochromatosis (a disease associated with widespread deposit in the tissues of two iron-containing pigments, hemosiderin and hemofuscin, and characterized by pigmentation of the skin), and chronic renal disease.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### MAR 8 2006 # 510(k) Summary for the Dimension® IRON Flex® reagent cartridge (DF85) A. 510(k) Number: K060264 B. Analyte: Total iron C. Type of Test: Quantitative # D. Applicant: Dade Behring Inc., P.O. Box 6101, Newark, DE 19714-6101 Contact: Andrea M. Tasker, Regulatory Affairs and Compliance Manager (302) 631-9454 Date of Preparation: January 30, 2006 #### E. Proprietary and Established Names: Dimension® Iron Flex® reagent cartridge (IRON-DF85) ## F. Regulatory Information: - Regulation section: 21 CFR §862.1410 Iron (non-heme) test system 1. - Classification: Class I 2. - Product Code: JIY 3. - Panel: Chemistry (75) 4. ### G. Intended Use: l. Intended for Use: The IRON method for the Dimension® clinical chemistry system is an in vitro diagnostic test intended to quantitatively measure iron in human serum and plasma. 2. Indications for Use: The IRON method for the Dimension® clinical chemistry system is an in vitro diagnostic test intended to quantitatively measure iron in human serum and plasma. Iron measurements are used in the diagnosis and treatment of diseases such as iron deficiency anemia and other disorders of iron metabolism. 3. Special condition for use statement(s): none - 4. Special instrument Requirements: Dimension® clinical chemistry system {1}------------------------------------------------ #### H. Device Description: The Dimension® IRON Flex® reagent cartridge (DF85) is an in vitro diagnostic device that consists of prepackaged reagents in a plastic eight well cartridge for use on the Dade Behring Dimension® clinical chemistry system for the quantitative determination of iron in serum and plasma. #### I. Substantial Equivalence Information: 1. Predicate Device: Dimension® IRN Iron Flex® reagent cartridge (DF49A) - 2. Predicate K Number(s): K944093 - 3. Comparison with Predicate: | Similarities | | | |-----------------------|--------------------------------------------------------------------------------------------------|------------| | Item | Device | Predicate | | Intended Use | Quantitative determination of<br>total iron | same | | Reagent<br>components | Ferene® (chromophore)<br>Thiourea (prevent Cu<br>interference)<br>Ascorbic acid (reducing agent) | same | | Measurement<br>method | Bi-chromatic endpoint<br>measurement (600 and 700<br>nm) | same | | Calibration | Three point linear calibration | same | | Analytical Range | 0 to 1,000 µg/dL | same | | Standardization | NIST SRM 937 | same | | Differences | | | | Item | Device | Predicate | | Matrix | Serum or heparinized plasma | Serum only | #### J. Standard/Guidance Document Referenced 1. Guidance; Guidance for Industry and FDA Staff - Use of Symbols on Labels and in Labeling of In Vitro Diagnostic Devices Intended for Professional Use, 11/30/2004 Guidance for Industry and FDA Staff; Replacement Reagent and Instrument Family Policy, 12/11/2003 Format for Traditional and Abbreviated 510(k)s - Guidance for Industry and FDA Staff, 08/12/2005 {2}------------------------------------------------ #### 2. Standards; | GP22-A | Continuous Quality Improvement Essential Management Approaches | |----------------|------------------------------------------------------------------------------------------------------| | ISO 15223 | Medical devices -- Symbols to be used with medical device labeling<br>and information to be supplied | | ISO 14971-2000 | Application of risk analysis to Medical devices | | EP7-A | Interference Testing in Clinical Chemistry | | EP5-A2 | Evaluation of Precision Performance of Clinical Chemistry Devices | | CEN 13640 | Stability testing of In-Vitro Diagnostic Devices | #### K. Test Principle: The automated Dimension® IRON method is an adaptation of direct iron assays using the chromophore Ferene ® . Under acidic conditions, iron (Fe+++) bound to the protein transferrin is released. In the presence of the reducing agent ascorbic acid, (Fe+++) is reduced to (Fe++). (Fe ++) forms a blue complex with 3-(2-pyridyl)-5,6-bis-2-(5-furyl sulfonic acid)-1,2,4-triazine, disodium salt (Ferene®). The absorbance of the complex, measured using a bichromatic (600, 700 nm) endpoint technique, is directly proportional to the concentration of transferrin-bound iron in the serum. #### L . Performance Characteristics: 1. Precision/Reproducibility: Reproducibility testing was done in accordance with the CLSI/NCCLS Approved Guideline for Evaluation of Precision Performance of Clinical Chemistry Devices (EPS-A2). Commercial controls and serum and plasma pools were analyzed in duplicate twice a day for 20 days. The repeatability and within-lab standard deviations were calculated by analysis of variance method. {3}------------------------------------------------ | | Mean | | Standard Deviation (%CV) | | |------------------------------------------|-------|----------|--------------------------|------------------| | Material | µg/dL | [µmol/L] | Repeatability | Within-lab | | Plasma pool | 101 | 18 | 0.5 [0.09] (0.5) | 0.7 [0.13] (0.7) | | Serum pool 1 | 95 | 16.9 | 0.5 [0.09] (0.5) | 0.6 [0.11] (0.6) | | Serum pool 2 | 316 | 56.6 | 1.5 [0.27] (0.5) | 3.5 [0.63] (1.1) | | Serum pool 3 | 533 | 95.4 | 2.4 [0.43] (0.5) | 4.2 [0.75] (0.8) | | BioRad Lyphochek® control Level 1 | 231 | 41.3 | 1.3 [0.23] (0.5) | 1.6 [0.29] (0.7) | | BioRad Lyphochek® control Level 2 | 50 | 8.9 | 0.5 [0.09] (1.1) | 0.9 [0.16] (1.9) | | BioRad Lyphochek® Anemia control Level 1 | 26 | 4.7 | 0.3 [0.05] (1.3) | 0.5 [0.09] (1.9) | | Reduced Sample Volumec | | | | | | Serum pool 1 | 103 | 18.5 | 0.7 [0.13] (0.6) | 1.0 [0.18] (0.9) | | Serum pool 2 | 316 | 56.6 | 1.5 [0.27] (0.5) | 3.5 [0.63] (1.1) | | Serum pool 3 | 530 | 94.9 | 2.9 [0.52] (0.5) | 4.2 [0.75] (0.8) | | BioRad Lyphochek® Anemia control Level 1 | 32 | 5.7 | 0.3 [0.05] (1.3) | 0.5 [0.09] (1.9) | | BioRad Multiqual® control Level 3 | 231 | 41.3 | 1.6 [0.29] (0.7) | 2.2 [0.39] (0.9) | # Reproducibilityª,b Lyphochek® and Multiqual® are registered trademarks of Bio-Rad Laboratories, Irvine, CA 92618. - Reproducibility testing was done in accordance with the CLSI/NCCLS Approved a. Guideline for Evaluation of Precision Performance of Quantitative Measurement Methods (EP5-A2, 2004). - Specimens at each level were analyzed in duplicate, twice a day, for 20 days. The b repeatability andwithin-lab standard deviations were calculated by analysis of variance method. - Using reduced sample size (25 uL). C. #### 2. Linearity/assay reportable range: Solutions of NIST SRM 937 were prepared by sequential mixing to create equally spaced samples ranging from 0 to 2000 µg/dL iron. The linearity of iron on the Dimension® RxL by the Dimension® IRON assay was evaluated by comparing observed versus expected values across the expected range. A linear regression analysis was performed on the data and plotted. The observed linearity across the reportable range has a correlation coefficient of 0.999 (using Pearson correlation calculation), slope of 0.999, and an intercept of 0.178. The assay range claim is 5.0 µg/dL to1000 µg/dL. {4}------------------------------------------------ 3. Traceability (controls, calibrators, or method): The recommended reference material for the Dimension® IRON method is the Dimension® IRON Calibrator (Cat. No. DC85). The assigned values of this product are standardized to SRM NIST 937 Iron Metal Clinical Standard which is a Standard Reference Material of the National Institute of Standards & Technology. #### 4. Detection limit: The analytical sensitivity of Dimension® IRON is 5 µg/dL [0.9 µmol/L]. The analytical sensitivity represents the lowest concentration of iron that can be distinguished from zero. This sensitivity is defined as the mean value (n=20) plus two standard deviations of the low level (0 ug/dL [0 umol/L] IRON Calibrator. #### 5. Analytical specificity: Interference testing was performed according to the CLSI/NCCLS Protocol EP-7A. Systematic inaccuracies (bias) due to these substances are less than 10% at an iron concentration of 26 to 38 µg/dL [4.7 to 6.8 µmol/L] and 118 to 136 µg/dL [21.1 to 24.3 umol/L]. A summary of potential interfering substances and the substances that do not interfere with the Dimension® IRON method when present in serum at the concentrations indicated can be found in the package insert. #### 6. Method comparison with predicate device: A total of 147 samples were tested using the Dimension® IRON and IRN assays on a Dimension® RxL clinical chemistry system. 99 individual patient serum samples and 48 individual patient serum samples that were spiked with various amounts of NIST-937 reference iron material were tested. Studies were performed using routine methods for quality control, maintenance and calibration as described in the instrument instructions for use. Linear regression analysis was performed using the SAS® System. The model equation for regression statistics is: Result of Dimension® system = (Slope x comparative method result) + Intercept. The range of IRON values in the correlation study was: 9 to 963 ug/dL. The linear regression statistics are presented in the table below. #### Method Comparison | Comparative<br>Method | Slope | Intercept<br>ug/dL | Correlation<br>Coefficient | n | |---------------------------|-------|--------------------|----------------------------|-----| | Dimension®<br>IRN (DF49A) | 0.980 | -0.488 | 0.9996 | 147 | {5}------------------------------------------------ #### 7. Matrix comparison 129 matched specimens of serum and heparinized plasma were drawn and processed by the IRON method on the Dimension® clinical chemistry system. Linear regression analysis showed excellent agreement between serum, lithium heparin plasma and sodium heparin plasma specimens. No clinically significant difference was observed between serum and plasma samples. Sodium heparin plasma versus serum: n = 129, y = 0.988x + 0.804, r = 0.999 Lithium heparin plasma versus serum: n = 129, y = 0.985x + 1.42, r = 0.999 Lithium heparin plasma verus sodium heparin plasma: n = 129, y = 0.997x + 0.666, r = 0.999 8. Reference Interval Males: 65-175 µg/dL [11.6-31.3 µmol/L]1 Females: 50-170 µg/dL [9.0- 30.4 µmol/L]1 Normal reference intervals can differ by as much as 35% between commercial iron methods2, therefore it is advised that each laboratory establish its own expected values for iron as performed on the Dimension® system. - Kaplan LA, Psece AJ. Clinical Chemistry Theory, Analysis, and Correlation, 3th ed. St. 1. Louis: Mosby, Inc., 1996: p 699, 713-714. - Burtis, CA, Ashwood ER, Bruns DE. Tietz Textbook of Clinical Chemistry and Molecular 2. Diagnostics, 4th ed. St. Louis: Elsevier Saunders, 2005, 1186-1193. {6}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service MAR 8 2006 Food and Drug Administratio 2098 Gaither Road Rockville MD 20850 Ms. Andrea Tasker Regulatory Affairs and Compliance Manager Dade Behring Inc. Glasgow Business Community P.O. Box 6101, Bldg. 500, M/S 514 Newark, DE 19714-6101 Re: k060264 K000201 Trade/Device Name: Dimension® Iron Flex® reagent cartridge (IRON-DF85) Regulation Number: 21 CFR§862.1410 Regulation Name: Iron (non-heme) test system Regulatory Class: Class I Product Code: JIY Dated: January 30, 2006 Received: February 1, 2006 Dear Ms. Tasker: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviews your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou mayy ators provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {7}------------------------------------------------ Page 2 -- This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally prarketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, rr you stions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto Guts Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ # Indications For Use Statement 510(k) Number (if known): Device Name: Dimension® Iron Flex® reagent cartridge (IRON - DF85) Indications for Use: The IRON method for the Dimension® clinical chemistry system is an in The IRON method for the Ditton and to quantitatively measure iron in human serum and plasma. Iron measurements are used in the diagnosis and scrum and plasma. "I on measing deficiency anemia and other disorders of iron metabolism. Concurrence of CDRH, Office of -In Vitro Điagnostic Devices (OIVD) **Prescription Use** (Per 21 CFR 801 Subpart D) AND/OR Over-the-counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANGTHER PLANTES NEEDED) Avision pign-ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ Office of In Vitro Diagnostic Device Evaluation and Safety 060264
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...