← Product Code [MEB](/productcode/MEB) · K060258

# AUTOFUSER (K060258)

_Ace Medical Us, LLC · MEB · May 30, 2006 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060258

## Device Facts

- **Applicant:** Ace Medical Us, LLC
- **Product Code:** [MEB](/productcode/MEB.md)
- **Decision Date:** May 30, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, subcutaneous, intra-arterial and epidural, and into intra-operative (soft tissue/body cavity) sites. Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button. General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery.

## Device Story

Autofuser is an ambulatory elastomeric infusion pump; delivers medication via continuous infusion or patient-controlled bolus button. Used in clinical or home settings for regional anesthesia and pain management. Device consists of pump reservoir and integrated administration set; may be provided in convenience kits. Modification adds new flow rates, reservoir sizes, and kit components. Healthcare providers select appropriate pump model/flow rate for patient; patient may self-administer bolus doses if equipped. System provides consistent drug delivery for pre-operative, perioperative, and postoperative surgery; benefits include improved pain management and mobility.

## Clinical Evidence

Bench testing only. Design verification and validation activities performed to confirm modified device meets specifications and is equivalent to predicate devices.

## Technological Characteristics

Elastomeric infusion pump; mechanical flow control; integrated administration set. Materials and components are substantially equivalent to legally marketed predicate devices. ISO 14971 risk management applied.

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Predicate Devices

- Autofuser ([K041585](/device/K041585.md))
- Woo Young/McKinley Medical Accufuser System ([K033039](/device/K033039.md))
- I-Flow Homepump/Eclipse/OnQ Painbuster system ([K980558](/device/K980558.md), [K982946](/device/K982946.md), [K944692](/device/K944692.md), [K896546](/device/K896546.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K060258

## Modifications to the Ace Medical Autofuser System

19 January 2006

Submitted by: Ace Medical US, LLC, 9102 Turfway Bend Drive, Powell, OH 43065-8497 Contact for Questions: Andy Lamborne, 303-443-7500 x.228, alamborne.co@netzero.net Trade Name: Autofuser Common Name: Elastomeric Infusion Pump & Procedure Kit Classification Name: Elastomeric Infusion Pump Device Classification: Class II Regulation Number: 880.5725 Classification Panel: 80, General Hospital and Personal Use Device Product code: MEB - Elastomeric Infusion Pump Original cleared 510(k): K041585 Establishment Registration: pending assignment from FDA Owner/Operator Number: 9082449

## Summary of Safety and Effectiveness of the Autofuser System 5.

- This premarket notification submission is intended to notify the Food and Drug 5.1 Administration that Ace Medical US, LLC intends to market a modification to an existing device (K041585), the Autofuser system.
- The cleared indications for use for the Autofuser system are: 5.2

The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, subcutaneous, intra-arterial and epidural, and into intra-operative (soft tissue/body cavity) sites.

Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button.

General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery.

- 5.3 Summary of the Modification
	- 5.3.1 The modification to the existing device is the addition of new flow rates, new reservoir size and procedure kit components.
	- 5.3.2 A risk analysis for the modified Autofuser device has been performed. incorporating risk analysis, risk evaluation and risk control in accordance with ISO 14971.

510(k) Ace Medical Autofuser System

510(k) Summary p. 1 of 2

{1}------------------------------------------------

- The risk control measures are verified and/or validated, as ಡಿ. appropriate.
- b. The design verification and validation activities have been performed and the results demonstrate that the predetermined acceptance criteria were met. These results demonstrate that the modified device is substantially equivalent to the cleared original device.

1060258

- 5.3.3 The kit components are legally marketed-either pre-amendment devices. devices exempt from premarket notification, or devices found to be substantially equivalent through the premarket notification process for the intended use of the kit. The kit components are substantially equivalent to the components included in pain management kits for the original Autofuser, the Accufuser and the OnQ/Painbuster predicate devices.
- 5.4 Conclusion: The modified Autofuser system does not raise any new safety and efficacy concerns when compared to the original Autofuser device that is already legally marketed. The modified Autofuser system is substantially equivalent to the original cleared Autofuser (K041585). The modified Autofuser system also demonstrates substantial equivalence to the Woo Young/McKinley Medical Accufuser System (K033039), and to the I-Flow Homepump/Eclipse/OnQ Painbuster system (K980558, K982946, K944692, K896546).

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" are arranged in a circular pattern around the caduceus.

Public Health Service

MAY 30 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Andrew N. Lamborne Technical Advisor Ace Medical US, LLC 9102 Turfway Bend Drive Powell, Ohio 43065-8497

Re: K060258

Trade/Device Name: Autofuser Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEB Dated: March 20, 2006 Received: March 28, 2006

Dear Mr. Lamborne:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

## Page 2 - Mr. Lamborne

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): K060258

Device Name: Autofuser

Indications For Use:

The Autofuser family of ambulatory infusion pumps with integrated administration set, either separately or as part of a convenience kit, is intended for general infusion use. Routes of administration include intravenous, percutaneous, subcutaneous, intra-arterial and epidural, and into intra-operative (soft tissue/body cavity) sites.

Within the Autofuser family are pump models intended for patient-controlled infusion using the integrated bolus button.

General infusion uses include continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative surgery.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anting D, ma

ology, General Hospita
on Centre, Lental Devices

501 4258

---

**Source:** [https://fda.innolitics.com/device/K060258](https://fda.innolitics.com/device/K060258)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
