← Product Code [MHX](/productcode/MHX) · K060221

# THE PHILLIPS, MODEL MP20,MP30,MP40,MP50,MP60,MP70,MP80,AND MP90 INTELLIVUE PATIENT MONITORS (K060221)

_Philips Medizin Systeme Boeblingen GmbH · MHX · Feb 17, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060221

## Device Facts

- **Applicant:** Philips Medizin Systeme Boeblingen GmbH
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Feb 17, 2006
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare facilities. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within healthcare facilities. ST Segment monitoring is restricted to adult patients only. The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

## Device Story

IntelliVue Patient Monitors (MP20-MP90) collect physiological data via sensors/electrodes; process signals for real-time monitoring, recording, and alarming; display data for clinician review. Used in hospital environments (including transport for MP20-MP50) by healthcare professionals. Output informs clinical decision-making regarding patient status. Release D.02 software update provides enhanced monitoring capabilities. Benefits include continuous patient surveillance and timely alerts for physiological changes.

## Clinical Evidence

Bench testing only. Performance, functionality, and reliability were established through system-level testing, pass/fail analysis, and safety testing compared against cleared predicate devices.

## Technological Characteristics

Multi-parameter physiological monitoring system. Includes arrhythmia detection, ST segment monitoring, blood pressure measurement, gas analysis, and thermodilution. Connectivity via network/communication systems. Software-based signal processing and alarm management. Class II device.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- [K014159](/device/K014159.md)
- [K021778](/device/K021778.md)
- [K030038](/device/K030038.md)
- [K032858](/device/K032858.md)
- [K040304](/device/K040304.md)
- [K040183](/device/K040183.md)
- [K040357](/device/K040357.md)
- [K041235](/device/K041235.md)
- K41741
- [K042845](/device/K042845.md)
- [K050141](/device/K050141.md)
- [K050762](/device/K050762.md)
- [K040337](/device/K040337.md)
- [K041259](/device/K041259.md)
- K1273522
- Respironics Novametrix Rostion, Rossister and Accessories PN 3468-3475 ([K053174](/device/K053174.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K060221
p1/3

## FEB 1 7 2006

## 510 (k) Summary 8.0

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe
is submitted in accordance with the requirements of the Safe 13 Submitted In accordinate of 1990 and 21 C.F.R. \$807.92.

1. The submitter of this premarket notification is: Herbert van Dyk Philips Medizin Systeme Boeblingen GmbH Hewlett-Packard-Str. 2 D-71034 Boeblingen, Germany - 71054 Bocking - 463-1734 - Fax: ++49 7031 463-2442 e-mail: herbert.van_dyk@philips.com

This summary was prepared on January 27, 2006.

- 2. The names of the devices are the Philips MP20, MP30, MP40,
 MP40, MP30, Lunco, Lunco, Intellilly, Pations Monitor The Mames Of the devices a 8990 IntelliVue Patient Monitors. Classification names are as follows:

| Device Panel                             | Classification | ProCode | Description                                                                 |
|------------------------------------------|----------------|---------|-----------------------------------------------------------------------------|
| Circulatory<br>System Devices<br>(12625) | \$870.1025, II | DSI     | Detector and alarm, arrhythmia                                              |
|                                          | \$870.1025, II | MLD     | Monitor, ST Segment with Alarm                                              |
|                                          | \$870.1025, II | MHX     | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) |
|                                          | \$870.1100, II | DSJ     | Alarm, Blood Pressure                                                       |
|                                          | \$870.1110, II | DSK     | Computer, Blood Pressure                                                    |
|                                          | \$870.1130, II | DXN     | System, Measurement, Blood-<br>Pressure, Non-Invasive                       |
|                                          | \$870.1435, II | DXG     | Computer, Diagnostic, Pre-<br>Programmed, Single-Function                   |
|                                          | \$870.1915, II | KRB     | Probe, Thermodilution                                                       |
|                                          | \$870.2060, II | DRQ     | Amplifier and Signal<br>Conditioner, Transducer Signal                      |
|                                          | \$870.2300, II | DRT     | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm)                   |
|                                          | \$870.2300, II | MSX     | System, Network and<br>Communication, Physiological<br>Monitors             |
|                                          | \$870.2340, II | DPS     | Electrocardiograph                                                          |
|                                          | \$870.2340, II | MLC     | Monitor, ST Segment                                                         |
|                                          | \$870.2350, II | DRW     | Electrocardiograph, Lead<br>Switching Adapter                               |
|                                          | \$870.2370, II | KRC     | Tester, Electrode, Surface,<br>Electrocardiograph                           |
|                                          | \$870.2450, II | DXJ     | Display, Cathode-Ray Tube,<br>Medical                                       |
|                                          | \$870.2600, I  | DRJ     | System, Signal Isolation                                                    |
|                                          | \$870.2700, II | DQA     | Oximeter                                                                    |
|                                          | \$870.2770, II | DSB     | Plethysmograph, Impedance                                                   |
|                                          | \$870.2800, II | DSH     | Recorder, Magnetic tape,<br>Medical                                         |
|                                          | \$870.2810, I  | DSF     | Recorder, Paper Chart                                                       |
|                                          | \$870.2850, II | DRS     | Extravascular Blood Pressure<br>Transducer                                  |

COMPANY CONFIDENTIAL

{1}------------------------------------------------

|                                                      | \$870.2900, I           | DSA            | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector               |
|------------------------------------------------------|-------------------------|----------------|------------------------------------------------------------------------------|
|                                                      | \$870.2910, II          | DRG            | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency       |
| Anesthesiology<br>and Respiratory<br>Therapy (12624) | \$868.1400, II          | CCK            | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase                              |
|                                                      | \$868.1500, II          | CBQ            | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                      | \$868.1500, II          | NHO            | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                      | \$868.1500, II          | NHP            | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)   |
|                                                      | \$868.1500, II          | NHQ            | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                      | \$868.1620, II          | CBS            | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                      | \$868.1700, II          | CBR            | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
|                                                      | \$868.1720, II          | CCL            | Analyzer, Gas, Oxygen, Gaseous-<br>Phase                                     |
|                                                      | \$868.1880, II          | BZC            | Data calculator Pulmonary-<br>function                                       |
|                                                      | \$868.2375, II          | BZQ            | Monitor, Breathing Frequency                                                 |
|                                                      | \$868.2480, II          | LKD            | Monitor, Carbon Dioxide,<br>Cutaneous                                        |
|                                                      | \$868.2500, II          | KLK            | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia           |
| General Hospital<br>and Personal Use<br>(12520)      | \$880.2910, II          | FLL            | Thermometer, Electronic,<br>Clinical                                         |
|                                                      | Neurological<br>(12513) | \$882.1400, II | GWR                                                                          |
|                                                      |                         | \$882.1420, I  | GWS                                                                          |

- The modified devices are substantially equivalent to 3. me modified dored Philips devices marketed pursuant to K014159, K021778, K030038, K032858, K040304, K040183 and K040357, K041235, K41741, K042845, K050141, 050762 and K040337, K041259, K1273522, and to Respironics Novametrix Rostion, Ro32001 , Rossister and Accessories "PN 3468-3475" (K053174)
- The modification is the introduction of Release D.02 software 4 . The MOGILLEACION ID che Internitor devices, MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90
- The modified devices have the same intended use as the 5. legally marketed predicate devices. They are intended for the regarry marketa ding, and alarming of multiple physiological Monicoring, recording, and trics, and neonates in hospital

COMPANY CONFIDENTIAL

{2}------------------------------------------------

environment and during transport within hospital environments .

- The modified devices have the same technological 6. The modified devices have the band elected predicate devices.
- 7. Verification, validation, and testing activities establish verformance, functionality, and reliability the performance, functionally, mevice with respect to the characteristics of the modified as a level tests, performance
predicate. Testing involved system levely - Rang (Fail predicate. Testing involved bybeard analysis. Pass/Fail tests, and safety testing from hassisions cleared for the criteria were babed on storesults showed substantial predicate device and other demonstrate that the Philips equivalence. The rebuild acments and reliability requirements and performance claims.

## COMPANY CONFIDENTIAL

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 1 7 2006

Philips Medizinsysteme Boeblingen GmbH c/o Mr. Herbert van Dyk Sr. Regulatory Affairs Engineer Philips Medical Services 3000 Minuteman Road Andover, MA 01810

Re: K060221

文

K 060221
Trade Name: The Philips MP20, MP30, MP30, MP60, MP70, MP80 and MP90
Trade Name: The Philips MP30, Dalam Dalam D 02 Intellivue Patient Monitors, Release D.02 Regulation Number: 21 CFR 870.1025 Regulation Name: 21 CFR 870.1023
Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: Class II (two) Product Code: MHX Dated: January 27, 2006 Received: January 30, 2006

Dear Mr. Van Dyk:

We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section 510(x) prematics in the manufated (for the indications referenced and nave decimined the device are devices marketed in interstate for use stated in the encrosule) to regally manced provice Americal Device Ameralments, or to commerce prior to May 20, 1970, the chaouthore with the provisions of the Federal Food, Drug, devices that have been reclassified in accordance what approval approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval applic and Cosmetic Act (Act) that do not require approvince approvisions of the Act. The Act. The You may, merelore, market the device, saloject to the greateration, listing of
general controls provisions of the Act include requirements for annual registration, listing of general controls provisions of the Are, Inc. Increase some in misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) me existing major regulations affecting your device can may be subject to such additional controlis. Entroling the 300 to 898. In addition, I'DA may be found in the Ood of Peachartergerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 – Mr. Herbert van Dyk

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I hat FDA has made a determination that your device complies with other requirements of the Act that i Drederal statutes and regulations administered by other Federal agencies. You must or uny I with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI K Part 607), laoomig (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control pro begin marketing your device as described in your Section 510(k) This letter will and w yourse FDA finding of substantial equivalence of your device to a legally marketed predicated. " caresults in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific as not any at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Bhimmar fa

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: The Philips MP20, MP30, MP40, MP50, MP60, MP70, MP80 and MP90 IntelliVue Patient Monitors, Release D.02

Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarmine of multiple physiological parameters of adults, pediatrics and memates in healthcare facilities. The MP20, MP30, MP40 and MP50 are additionally intended for use in transport situations within healthcare facilities.

ST Segment monitoring is restricted to adult patients only.

The transcutaneous gas measurement (tcp02 / tcpCO2) is restricted to neonatal patients only.

Prescription Use yes___________________________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

Over-The-Counter Use _ No (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Blummenstor

n Sign-

---

**Source:** [https://fda.innolitics.com/device/K060221](https://fda.innolitics.com/device/K060221)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
