K060216 · Energist, Ltd. · GEX · Mar 16, 2006 · General, Plastic Surgery
Device Facts
Record ID
K060216
Device Name
ENERGIST ULTRA VPL INTENSE PULSE LIGHT SYSTEM
Applicant
Energist, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 16, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Energist ULTRA™ VPL Intense Pulsed Light System is intended for permanent hair reduction and the treatment of mild to moderate inflammatory Acne Vulgaris. It is also indicated for photocoagulation of dermatological vascular lesions, photothermolysis of blood vessels and the treatment of benign pigmented lesions. The 530 - 950nm intense pulsed wavelengths are indicated for: The treatment of mild to moderate inflammatory Acne Vulgaris. The treatment of benign pigmented epidermal and cutaneous lesions including warts, scars and striae. The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea. melasma, angiomas and spider angiomas, poikiloderma of Civatte, leg veins. facial veins and venous malformations. The 610 - 950nm intense pulsed wavelengths are indicated for: The removal of unwanted hair from all skin types and to effect stable longterm or permanent hair reduction in skin types I - V through selective targeting of melanin in hair follicles.
Device Story
Energist ULTRA VPL is a compact, self-contained intense pulsed light (IPL) system; delivers non-ionizing radiation via removable handpiece with integrated lamp, filter, and glass coupling block. System includes control console, display panel, power supply, and cooling system. Operates by emitting pulsed light at 530-950nm or 610-950nm wavelengths to target melanin in hair follicles or vascular/pigmented structures via photothermolysis/photocoagulation. Used in clinical settings by trained operators. Output allows clinicians to perform dermatological procedures; benefits include hair reduction and treatment of vascular/pigmented lesions and acne.
Clinical Evidence
Bench testing only. No clinical data provided. Performance data demonstrated the system provides necessary outputs to achieve required treatment parameters.
Technological Characteristics
IPL system; wavelengths 530-950nm and 610-950nm. Features control console, display, power supply, cooling system, and removable handpiece with integrated switch, lamp, filter, and glass coupling block. Software-controlled high-voltage component.
Indications for Use
Indicated for permanent hair reduction (skin types I-V) and treatment of mild-to-moderate inflammatory Acne Vulgaris. Also indicated for photocoagulation/photothermolysis of benign pigmented epidermal/cutaneous lesions (warts, scars, striae) and benign cutaneous vascular lesions (port wine stains, hemangiomas, telangiectasias, rosacea, melasma, angiomas, spider angiomas, poikiloderma of Civatte, leg/facial veins, venous malformations).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# Section 5 - 510(k) Summary
This 510(k) summary of safety and effectiveness for the Energist ULTRA variable pulsed light system has been prepared in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows relevant Office of Device Evaluation (ODE) guidance concerning the organisation and content of a 510(k) summary.
### 5.1 Applicant
Energist Ltd
### 5.2 Address
2 Clos Llyn Cwm , Enterprise Park, Swansea SA6 8QY, UK
#### 5.3 Contact Person
Peter Smith, Engineering & Regulatory Affairs Manager Tel: +44 (0)1792 798768 Fax: +44 (0)1792 762099
### 5.4 Preparation Date
January 19, 2006
#### 5.5 Device Trade Name
Energist ULTRA Variable Pulsed Light (VPL™) System
### 5.6 Device Common Name
Intense Pulsed Light System
#### 5.7 Device Classification
Laser surgical instrument for use in General and Plastic Surgery and Dermatology Product code: GEX 21 CFR 878.4810
#### 5.8 Legally marketed Predicate Devices
McCue Energist ULTRA™, K040659 Cyden IFL Professional System, K050165
#### 5.9 Device Description
The Energist ULTRA VPL™ System is a light based medical device that delivers a beam of pulsed non-ionising radiation in the region of 530nm. The system is designed to be compact and self contained and includes the following features:
- . Control console unit
- Display panel .
- Power supply ◆
- Cooling system .
- . Removable handpiece with integrated switch, lamp, filter and glass coupling block
#### 5.10 Intended Use
The Energist ULTRA™ VPL Intense Pulsed Light System is intended for permanent hair reduction and the treatment of mild to moderate inflammatory Acne Vulgaris. It is also indicated for photocoagulation of dermatological vascular lesions, photothermolysis of blood vessels and the treatment of benign pigmented lesions.
- . Intense Pulsed Light Energy / wavelengths (530 - 950nm)
{1}------------------------------------------------
The 530 - 950nm intense pulsed wavelengths are indicated for:
The treatment of mild to moderate inflammatory Acne Vulgaris.
The treatment of benign pigmented epidermal and cutaneous lesions including warts, scars and striae.
The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea. melasma, angiomas and spider angiomas, poikiloderma of Civatte, leg veins, facial veins and venous malformations.
- Intense Pulsed Light Energy / wavelengths (610 950nm) .
The 610 - 950nm intense pulsed wavelengths are indicated for:
The removal of unwanted hair from all skin types and to effect stable longterm or permanent 1 hair reduction in skin types I - V through selective targeting of melanin in hair follicles.
1 Permanent hair reduction is defined as a long-term stable reduction in the number of hairs regrowing after a treatment regimen.
#### 5.11 Comparison of Technological Characteristics
The Energist ULTRA VPL System has an identical intended use to the McCue predicate system and identical technological characteristics except for modifications to the device software and high voltage component. These differences do not result in differences in performance or raise new questions of safety and efficacy. The Energist ULTRA VPL System also has the same principles of operation, mode of action and equivalent energy outputs that are used by the Cyden predicate device to achieve the treatment of Acne Vulgaris. Performance data was provided to demonstrate that the system is capable of providing the outputs necessary to achieve its required treatment parameters. The Energist ULTRA VPL ™ System is therefore concluded to be substantially equivalent to both of the above named predicate devices and minor differences to the systems do not raise additional concerns of safety and efficacy.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings, facing right. The eagle is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in capital letters.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 16 2006
Energist, Ltd. c/o Mr. Peter Smith Engineering & Regulatory Affairs Manager 2 Clos Llyn Cwm Enterprise Park Swansea City and County of Swansea SA6 AQY United Kingdom
Re: K060216
Trade/Device Name: Energist ULTRA™ VPL Intense Pulse Light System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: January 19, 2006 Received: January 27, 2006
Dear Mr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
## Page 2 - Mr. Peter Smith
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your dryice to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
060216 510(k) Number (if known): A
Device Name: ___ Energist ULTRA™ VPL Intense Pulse Light System
Indications for Use:
The Energist ULTRA™ VPL Intense Pulsed Light System is intended for permanent hair reduction and the treatment of mild to moderate inflammatory Acne Vulgaris. It is also indicated for photocoagulation of dermatological vascular lesions, photothermolysis of blood vessels and the treatment of benign pigmented lesions.
- Intense Pulsed Light Energy / wavelengths (530 950nm) .
The 530 - 950nm intense pulsed wavelengths are indicated for:
The treatment of mild to moderate inflammatory Acne Vulgaris.
The treatment of benign pigmented epidermal and cutaneous lesions including warts, scars and striae.
The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea. melasma, angiomas and spider angiomas, poikiloderma of Civatte, leg veins. facial veins and venous malformations.
Intense Pulsed Light Energy / wavelengths (610 - 950nm) .
The 610 - 950nm intense pulsed wavelengths are indicated for:
The removal of unwanted hair from all skin types and to effect stable longterm or permanent 1 hair reduction in skin types I - V through selective targeting of melanin in hair follicles.
1 Permanent hair reduction is defined as a long-term stable reduction in the number of hairs regrowing after a treatment regimen.
(Over)
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Image /page/4/Picture/19 description: This image shows a document with the title "Concurrence of CDRH, Office of Device Evaluation (ODE)". It also states "Division of General, Restorative, and Neurological Devices". The document includes the section number and page number, with "Section 4" and "Page 2" written at the bottom. The 510(k) number is K060216.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.