The CAREO Safety Syringe 1cc/mL is a sterile, single-use, disposable and non-reusable, retractable safety syringe which is intended to provide a safe and reliable method for intramuscular and subcutaneous injection of medication into a patient. The CARFO Safety Syringe 1cc/mL is also intended to prevent needlestick injuries. In addition, when the user breaks the plunger, reuse of the syringe is prevented.
Device Story
CAREO Safety Syringe 1cc/mL is a sterile, single-use, disposable piston syringe with a permanently attached needle. Used by clinicians for intramuscular or subcutaneous medication delivery. Device features a retractable needle mechanism; after injection, user manually retracts needle-plunger into barrel, snaps off plunger rod, and discards components to prevent needlestick injuries and syringe reuse. Operates as a standard manual syringe; no electronic or automated components.
Clinical Evidence
Bench testing only. Includes biocompatibility, packaging integrity, standards conformity, and simulated use testing. Simulated use testing demonstrated performance according to specifications.
Technological Characteristics
Piston syringe with permanently attached single lumen hypodermic needle. Sterile, single-use, disposable. Manual needle retraction and plunger-breaking mechanism for safety. No electronic components or software.
Indications for Use
Indicated for intramuscular and subcutaneous injection of medication into patients. Intended to prevent needlestick injuries and prevent syringe reuse.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
MAR 1 0 2006
K060205
l of 2
# 510(k) Premarket Notification for the Life-Shield Products, Inc. CAREO Safety Syringe 1cc/mL (per 21 CFR807.92)
# 1. Sponsor
| Life-Shield Products, Inc. | |
|---------------------------------|-----------------------|
| 3Fl., No. 10, Wuchiuan 7th Rd., | |
| Wugu Industrial Park, | |
| Taipei 248 | |
| Taiwan, R.O.C. | |
| Contact Person: | Mr. Hsiao, Chung-Chih |
| Telephone: | +886 2 2299 6033 |
Date Prepared: January 26, 2006
# 2. Device Name
| Proprietary Name: | CAREO Safety Syringe 1cc/mL |
|----------------------|--------------------------------------------------|
| Common/Usual Name: | Hypodermic Syringe (with needle) |
| Classification Name: | Piston syringe<br>Hypodermic single lumen needle |
### 3. Predicate Device
- 1cc SafePro* Safety Syringe (K022063)
- . CAREO Safety Syringe (3cc/mL, 5cc/mL, 10cc/mL) (K052397)
# 4. Device Description
The Life-Shield Products, Inc., CAREO Safety Syringe 1cc/mL is a sterile, single use and disposable, 1cc/mL piston syringe, provided with a permanently attached needle in fifteen product configurations. The CAREO Safety Syringe 1cc/mL is similar in appearance, size, materials operation, and purpose to the cited predicate device and other conventional single use, sterile, disposable syringes.
Appendix Q - page 1
January 26, 2006
{1}------------------------------------------------
#### Intended Use 5.
The CAREO Safety Syringe 1cc/mL is a sterile, single-use, disposable and non-reusable, retractable safety syringe which is intended to provide a safe and reliable method for intramuscular and subcutaneous injection of medication into a patient.
The CARFO Safety Syringe 1cc/mL is also intended to prevent needlestick injuries. In addition, when the user breaks the plunger, reuse of the syringe is prevented.
#### 6. Technological Characteristics and Substantial Equivalence
Life-Shield Products, Inc., makes a claim of substantial cquivalence of the CAREO Safety Syringe 1cc/mL to the 1cc SafePro* Safety Syringe (K022063) and CAREO Safety Syringe (3cc/mL, 5cc/mL, 10cc/mL) (K052397) based on similarities in intended use, design, technological and operational characteristics. All are indicated for injecting fluids into the body, while helping to reduce the risk of sharps injuries. All syringes are piston syringes that use permanently attached single lumen hypodermic needles. All syringes are provided sterile, single-use, and disposable. All syringes require the user to manually retract the needle-plunger into the syringe barrel, snap off the plunger rod, and discard the pieces.
#### 7. Testing
Testing provided in this premarket notification includes biocompatibility, packaging integrity, standards conformity, and simulated use testing. Side-by-side testing of the CAREO Safety Syringe Icc/mL and the 1cc SafePro* Safety Syringe shows that the two products are equivalent. Simulated use testing has demonstrates that the CAREO Safety Syringe performs according to specification. Testing also supports the claimed Indications for Use.
Appendix Q - page 2
Life-Shield Products, Inc., Special 510(k) CAREO Safety Syringe 1cc/mL
January 26, 2006
CONFIDENTIAL
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 0 2006
Mr. Chung-Chih Hsiao Senior Regulatory Affairs Consultant Life-Shield Products, Incorporated 3F, No. 10 & 12, Wuchiuan 7th Rd. Wugu Industrial Park Taipei, Taiwan 248 Republic of China
Re: K060208
Trade/Device Name: CAREO Safety Syringe 1 CC/ML Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: January 26, 2006 Received: February 8, 2006
Dear Mr. Hsiao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. Chung-Chih Hsiao
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Sydie Y. M. chae Ond.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital , Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications For Use
510(k) Number (if known):__K060208
Device Name:_Life-Shield Products, Inc., CAREO Safety Syringe 1cc/mL
Indications For Use:
The CAREO Safety Syringe 1cc/mL is a sterile, single-use, disposable and non-The CARLO Salery Syringe Toomie is intended to provide a safe and reliable reduable, Tetractable Baret) Symrge injection of medication into patient.
The CAREO Safety Syringe 1cc/mL is also intended to prevent needlestick injuries. In The OARCO Outley offinge is the plunger, reuse of the syringe is prevented
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cinta Keter
General Hospilli O Dental Devices
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.