The APC Integrated Filter Probes are intended for use in Argon Plasma Coagulation (APC). The devices are used to treat many conditions in endoscopy for various surgical procedures.
Device Story
APC Integrated Filter Probes and Adapter are accessories for ERBE Argon Plasma Coagulator (APC 300/APC 2) and ESU. Probes are flexible, single-use, sterile devices with integrated 0.45 µm membrane filters to prevent backflow contamination. Probes feature various lengths (1.5m-3m), diameters (1.5mm-3.2mm), and tip configurations (axial, lateral, circumferential) for endoscopic tissue treatment. Adapter is a reusable, non-sterile silicone/plastic cable connecting probe to coagulator. Physician threads probe through endoscope working channel to target tissue; activates system via footswitch. High-frequency voltage creates electrically conductive argon plasma in gas stream, enabling contact-free, uniform coagulation. Integrated EPROM-based instrument recognition system automatically sets conservative default parameters on compatible coagulators. Adapter includes resistor-based recognition to ensure safe default settings when direct probe recognition is unavailable. System benefits include reduced contamination risk, precise tissue coagulation, and automated safety settings.
Clinical Evidence
Bench testing only. Validation performed using Recognized Standards to demonstrate functionality, reliability, and biocompatibility of modified materials and structural designs. No clinical data provided.
Technological Characteristics
Materials: plastics, silicone, nickel-coated brass, polyurethane, PTFE, stainless steel, ceramic. Sensing: EPROM-based or resistor-based instrument recognition. Energy: Electrosurgical current and argon gas. Form factor: Flexible probes (1.5-3m length, 1.5-3.2mm diameter) and 13.5-inch adapter cable. Sterilization: Ethylene oxide (probes). Connectivity: Proprietary connection to ERBE APC 300/APC 2 coagulators.
Indications for Use
Indicated for Argon Plasma Coagulation (APC) during endoscopic surgical procedures to treat various conditions. Prescription use only.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
EB 1 6 2006 26018:
# 510K SUMMARY
Submitted By:
ERBE USA. Inc. 2225 Northwest Parkway Marietta, GA 30067
Tel: 770-955-4400 Fax: 770-955-2577
John Tartal Contact Person:
01/23/06 Date Prepared:
Argon Plasma Coagulation (APC) Integrated Filter Probes Common Name: and Adapter
ERBE APC Integrated Filter Probes and Adapter Trade/Proprietary Name:
and device and coagulation Electrosurgical cutting Classification Name: accessories (21CFR878.4400)
GEI Product Code:
Legally Marketed Device: APC Connector Hose and Probes submitted in the ERBE 510(k) Number K013348 as well as APC Membrane Filter (as an Accessory) in ERBE 510(k) Number K024047
The APC Integrated Filter Probes and Adapter are being submitted in this Note: premarket notification are modified from the accessories in the above 510(k)s. The modifications were determined to require a 510(k) submission.
# Device Description:
The APC Integrated Filter Probes incorporate a filter similar to the APC Membrane Filter and the hose (specifically the length) of an APC Connector Hose into one device. See Section IV, Appendix 1, Product Pictures (IV-2). The Probes are flexible. They are provided in various lengths, diameter sizes, and tip configurations to accommodate the various size/types of endoscopes for a variety of applications (i.e., the treatment of various target tissues inside a patient). The working lengths (i.e., the segment of the Probe that would go down a scope) of the APC Integrated Filter Probes are 4.9ft. (1.5m) to 9.8ft. (3m). The working diameters of the Probes are 4.5 French (1.5mm) to 9.6 The Probes have either an axial (straight fire), lateral (side fire), or French (3.2mm). circumferential opening allowing the energy to be delivered straight, at a 45 to 90 degree angle, or directly to target tissue regardless of the probe's position. The three types of openings for the APC Integrated Filter Probes allow the physician a choice in the direction or the option not to have to consider the direction (just proximity) of delivering the argon plasma to the treatment site which is based upon tissue resistance.
{1}------------------------------------------------
K060183
Page 2 of 5
Note: Since the last 510(k), two (2) additional APC Probes became apart of the product Note: "Since the last of of(i), two (=) ticant. One involved a different outer diameter and line. The modificulture were within the ranges of dimensions that were already included in approved Probes. The other involved modifying the tip of a side fire probe included in application. Besides the modification to the tip shape, the bonding for circumici application. "Backlower, the material difference in the tip did not raise any biocompatibility issues.
For each Probe, the blue connector/filter end is made of plastics (including the ron Cath Probo, the brass contact pins coated with nickel. The filter is 0.45 µm and mombrane filter) was a Coagulator (the Equipment) from contamination if there is a protote the Arger. I have other fluid contaminant. The hose part of the Probe is made of baoklow of brood of any the tube portion of the Probe is made of polyurethane, PTFE rond a stainless steel wafer shape electrode close to the tip for terminating electrical current. Various glues/adhesives are used to secure all of the Probe's parts. Ceramic tips are bonded on the APC Integrated Filter Side Fire and Circumferential Probes. Functionality and reliability of the Probes were demonstrated through validation testing using Recognized Standards. All added or modified materials in the APC Integrated Filter Probes were evaluated for biocompatibility with no problem found (See Section III, Biocompatibility, III-10 to III-19).
The APC Integrated Filter Probes are provided sterile by means of ethylene oxide and are disposable (Single Use) [Note: The sterilization cycle has been validated. See Section III, Sterilization Information, III-20.].
The APC Integrated Filter Probe Adapter is a flexible connecting cable. See Section IV, Appendix 1, Product Pictures (IV-3). It connects the APC Integrated Filter Probe to an ERBE Argon Plasma Coagulator. The Adapter is a conduit for both electrosurgical current and argon gas. It is approximately 13-1/2" in length. The connector ends of the Adapter are made of plastics. The cable portion of the Adapter is silicone. The Adapter's performance was validated via Recognized Standards. The APC Integrated Filter Probe Adapter is provided non-sterile and is reusable (Note: The Adapter only needs to be externally cleaned/disinfected unless it is contaminated.).
The APC Integrated Filter Probes and Adapter are accessories of the ERBE Argon Plasma Coagulator Models APC 300 and APC 2 [Note: The Coagulator is used with an associated ERBE ElectroSurgical Unit (ESU)]. The APC Integrated Filter Probe Adapter attaches to gas as well as electrical inputs of the Coagulator and then to an APC Integrated Filter Probe [Note: It is possible to connect an APC Integrated Filter probe directed to ERBE Coagulator APC 2 Model if it has an FiAPC Receptacle). An endoscope is manipulated inside the patient to locate tissue that requires treatment. Upon finding the target tissue, the Probe is threaded into the working channel of the endoscope, until the tip of the APC Integrated Filter Probe slightly protrudes from the end of the scope. The opening of the Probe is positioned towards/in close proximity of the area to be treated (Note: The APC Integrated Filter Probes have depth marker rings close to the tip for positioning purposes.). The APC/ESU system is activated via a
{2}------------------------------------------------
K060183
Page 3 of (5)
footswitch. When high frequency voltage reaches the critical level and the proximity to tissue is close enough, electrically conductive argon plasma forms in the gas stream. tissue is close chough, colochiany contractly contract of the tissue. Current density upon Trills allows the burrent to how asses coagulation. The application of the energy to the tissue is uniform and contact free.
#### Intended Use:
The APC Integrated Filter Probes are intended for use in Argon Plasma Coagulation (APC). The devices are used to treat many conditions in endoscopy for various surgical procedures.
Note: The APC Integrated Filter Probe Adapter is just the attachment piece to connect Note: "The Ar of the Probe to an ERBE Coagulator Model APC 300. The Adapter is only needed for the ERBE APC 2 Model if a FiAPC Receptacle is not installed.
Similarities and Differences of the Modified Device to the Current Device (Predicate Comparison/Substantial Equivalence):
APC Integrated Filter Probes and Adapter
#### Similarities
The APC Integrated Filter Probes have the same performance, working dimensional and tip specifications (i.e., working diameter and lengths as well as tip configurations), intended use, labeling, and packaging specifications as the predicate APC Probes. The Interabled Filter Probes incorporate a filter similar to and performs like the predicate APC Membrane Filter. Also like the predicates, the Filter Probes are provided sterile by ethylene oxide and are single use. Furthermore, the hose length of the APC Integrated Filter Probes is the same as the APC Connector Hose. Like the predicate Integrated the APC Integrated Filter Probes have an instrument recognition system. If attached directly into an ERBE Argon Plasma Coagulator Model APC 2 via an FiAPC receptacle, the Coagulator recognizes the Probe diameter size and automatically sets conservative default settings.
The APC Integrated Filter Probe Adapter is basically the connectors of the APC Connector Hose. The connectors that attach to an ERBE Coagulator are the same as the connectors on the APC Connector Hose. The APC Integrated Filter Probe Adapter has the same type of materials (plastics and silicone), performance specifications, intended use, labeling, and packaging specifications as the predicate APC Connector Hose. Also like the predicate, it is provided non-sterile and is reusable.
# Differences
The APC Integrated Filter Probes have a different connector end as well as incorporates a filter and length of hose (Note: Due to having the filter at the end of the Probes. the connector was modified. This was also done so that it only fits its designated Adapter.). The structural modifications of changing the connector end didn't
{3}------------------------------------------------
K060183
Page 4 of 5
raise any safety or efficacy issues. The connector fits snugly on the designated Adapter raise any saicty of emodor locator). Also no negative impact was found with the (f.e., the condunt to the ERDE obagana filter and the hose in the Probes. The filter in the diffiensional modification of installing a filter stand alone APC Membrane Filter size being 0.2 I robe is 0.45 pm as ound to be sufficient in keeping backflow fluid from pin. The farger port of Coachator. The slight materials changes (i.e., in plastics, the readling the Nadptor or Sougalaton be acceptable in producing a functionally reliable dror in grace/ nore recognition system in the APC Integrated Filter Probes are Probe. chipped (EPROM) based as compared to the predicate APC Probes which are chipped (LI (Chi) based as semped based recognition system is more specific and recognized 'na rolletereration. Finally; slightly different plastics, metals, and adheswes can provide more information. This and Membrane Filter, Probes, and Connector Hose. The materials were found via validation Membrano Filter Recognized Standards to work well in the APC Integrated Filter testing and the materials are biocompatible. See Section III, Biocompatibility and Summary of Design Control Activities.
The APC Integrated Filter Probe Adapter is different that the APC Connector Hose in that it doesn't have the hose portion and the connector to the Probe was modified so it only attaches to the APC Integrated Filter Probes. The dimensional modification of not having the hose portion and the structural modification to the end that connects to the neving the need portions any safety or efficacy concerns for the Adapter. The APC Integrated Filter Probe Adapter connects snugly to the Filter Probes. It was also found to function properly through validated testing and using Recognized Standards. The Adapter is also different that the predicate APC Connector Hose in that the Adapter has a resistor instrument recognition system. If the Adapter is used, it overrides specific probe recognition. However, no safety or efficacy issues are expected because the Adapter's resistor sets the lowest/most conservative default settings (i.e., gas flow and wattage) on an ERBE Coagulator. As a result for the Probe being used, the settings adiustments on the ERBE Coagulator would only be upward. Setting maximums for the gas flow rate and wattage are specified for each APC Probe in the Notes On Use (NOU) for the Probes. In the "How to Use" section of the NOU, the user is also directed to setting limitations and instructed not to exceed maximums for the APC Probe being used. If the gas flow rate limit is appreciably exceeded when using the Adapter and the smallest diameter APC Probe, then the Coagulator errors preventing activation. With the Adapter and the largest diameter APC Probe the maximum gas flow rate is limited: however, no problem with in the clinical environment is expected. Typically, desired settings are established during test firing of the Probe. Therefore, no problems in regards to not having specific probe recognition are expected. Biocompatibility testing was not necessary because the Adapter doesn't come in contact with the patient.
{4}------------------------------------------------
K060 183
Image /page/4/Picture/1 description: The image shows the handwritten words "Page 5 of 5". The word "page" is written in cursive, and the numbers are written in a more standard font. The last number 5 is circled.
Note: Incorporating the filter in the Probes eliminates the need to reprocess the Adapter Note. Incorporating the meet in the APC Connector Hose because backflow between "Sacob" and reach the Adapter or Coagulator.
All the changes have been verified or validated in design control.
#### Conclusion:
The APC Integrated Filter Probes and Adapter have the same intended use, principles The AFC Integration, and technological characteristics as the accessories in the previously of operation, and technological onations were found to have a cicaled predioute action. The APC Integrated Filter Probes and Adapter.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, which is a traditional symbol of medicine and healthcare. The symbol features a staff with two snakes entwined around it, and a pair of wings at the top.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 6 2006
Mr. John Tartal Manager Quality Assurance and Regulatory Affairs ERBE USA, Inc. 2225 Northwest Parkway Marietta, Georgia 30157
Re: K060183
Trade/Device Name: ERBE APC Integrated Filter Probes and Adapter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: January 23, 2006 Received: January 25, 2006
Dear Mr. Tartal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. Tartal
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K060183
Device Name: ERBE APC Integrated Filter Probes and Adapter
Indications for Use:
Argon Plasma Coagulation (APC)
Prescription Use _ X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRII. Office of Device Evaluation (ODE)
---
ivis ் 201-: 11: Jivision of General, Restorative, and Neurological Devices
Page 1 of 1
316(k) Number_________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.