CELLTRACKS ANALYZER II

K060110 · Immunicon Corp. · NQI · Mar 16, 2006 · Immunology

Device Facts

Record IDK060110
Device NameCELLTRACKS ANALYZER II
ApplicantImmunicon Corp.
Product CodeNQI · Immunology
Decision DateMar 16, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.6020
Device ClassClass 2

Indications for Use

The Immunicon CellTracks® Analyzer II is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. The product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that are legally marketed for in vitro diagnostic use with this device.

Device Story

Semi-automated fluorescence microscope for rare cell analysis; used with CellTracks AutoPrep System and ferrofluid-based immunomagnetic capture reagents. Input: biological fluid samples (e.g., whole blood) processed into cartridges. Operation: MagNest device creates magnetic field to align labeled cells on cartridge surface; system scans surface with fluorescence filters. Output: cell images compiled in gallery format for user review/classification. Used in clinical labs/research institutions by trained personnel. Software facilitates image acquisition, analysis, and reporting. Benefits: enables identification and enumeration of rare target cells for clinical decision-making.

Clinical Evidence

Bench testing only. Software validation performed using GAMP 4 Guide for Validation of Automated Systems. Testing included verification against software requirements specifications (SRS), design documents, and fault testing for power interruptions. Trace Matrix used to ensure requirements coverage.

Technological Characteristics

Semi-automated fluorescence microscope; utilizes immunomagnetic capture with ferro-fluid-based reagents. MagNest magnetic cell presentation device. Connectivity: Standalone workstation. Software: Mandrake Linux OS with custom GUI. Standards: ISO 14971, IEC 61010-1, BS EN61326, EP5-A, EP9-A.

Indications for Use

Indicated for use by CLIA-compliant, trained testing personnel to enumerate circulating tumor cells (CTC) in biological fluids, including whole blood, for patients requiring analysis of rare cells.

Regulatory Classification

Identification

An immunomagnetic circulating cancer cell selection and enumeration system is a device that consists of biological probes, fluorochromes, and other reagents; preservation and preparation devices; and a semiautomated analytical instrument to select and count circulating cancer cells in a prepared sample of whole blood. This device is intended for adjunctive use in monitoring or predicting cancer disease progression, response to therapy, and for the detection of recurrent disease.

Special Controls

The device is classified as Class II under regulation 21 CFR 866.6020 with special controls. The special control guidance document "Immunomagnetic Circulating Cancer Cell Selection and Enumeration System" is available at https://www.fda.gov/cdrh/oivd/guidance/1531.pdf

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Immunomagnetic Circulating Cancer Cell Selection and Enumeration System.” See § 866.1(e) for availability of this guidance document.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k060110 B. Purpose for Submission: Software architecture in CellTracks® Analyzer II operating system was changed. The operating system MS Windows XP has been changed to Mandrake Linux. This new operating system (Linux) is not compatible with the current Graphical User Interface (GUI) and a new GUI program was also developed. C. Manufacturer and Instrument Name: Immunicon Corporation’s CellTracks Analyzer II D. Type of Test or Tests Performed: A semi-automated, qualitative immunomagnetic-capture immunofluorescent detection image analysis fluorescence microscope used to aid a CLIA-compliant, CellTracks® trained testing personnel in the enumeration of circulating tumor cells (CTC) (cells appearing like tumor cells with epithelial cell markers and no lymphocyte marker on their surfaces). E. System Descriptions: 1. Device Description: The CellTracks® Analyzer II is a semi-automated fluorescence microscope. The product consists of the CellTracks® Analyzer II, a dedicated computer loaded with CellTracks® software, monitor, keyboard and mouse. The CellTracks® Analyzer II is for analysis of rare cells that are isolated from biological fluids including whole blood. It is used in conjunction with CellTracks® AutoPrep System, which automates, standardizes and optimizes the sample preparation with specific reagent kits. 2. Principles of Operation: The CellTracks® Analyzer II is used in conjunction with the CellTracks® AutoPrep System and reagent kits that contain a ferro-fluid-based capture reagent and immunofluorescent reagents for detection and characterization of the captured cells. The ferrofluid reagent consists of nano-particles with a magnetic core surrounded by a polymeric layer coated with antibodies targeting the cells of interest. After immunomagnetic capture and enrichment, fluorescent reagents are added for identification and enumeration of the target cells. The processed reagent/sample mixture is dispensed by the CellTracks® AutoPrep System into a cartridge that is inserted into a MagNest® cell presentation device. The strong magnetic field of the MagNest® device causes the magnetically-labeled target cells to move to the surface of the cartridge. The cartridge is then placed on the CellTracks® Analyzer II for data acquisition and analysis. The CellTracks® Analyzer II scans the entire surface of the cartridge with a series of fluorescence filters that are defined for the assay. Cell images from the filter are compiled and presented in a gallery format for final cell classification by the user. 3. Modes of Operation: Semi-automated, one-at-a-time MagNest® analysis. {1} 4. Specimen Identification: Identification information entered individually for each patient sample. 5. Specimen Sampling and Handling: The processed reagent/sample mixture is dispensed by the CellTracks AutoPrep system into a cartridge that is inserted into a MagNest® cell presentation device. The MagNest® is not mixed or pierced, but put into place manually and then the surface is automatically scanned. The CLIA-compliant, CellTracks® trained testing personnel then manually interprets the results. 6. Calibration: The analyzer is calibrated using the CellTracks® System Verification Cartridge which contains a strip of material impregnated with dyes. System verification checks the optical performance and chamber "skew" which ensures proper scanning. There is no recognized reference material or method. 7. Quality Control: For the circulating tumor cell assay the CellSearch® Circulating Tumor Cell Control Kit (available from Veridex, LLC, a Johnson and Johnson company) should be run once per day of patient testing as a patient sample following each user's quality assurance program. 8. Software: FDA has reviewed applicant's Hazard Analysis and Software Development processes for this line of product types: Yes ☐ X ☑ or No ☐ Frederick Prevo, PhD, (OSEL/DESE) has reviewed the software submission and found it to be acceptable for a moderate hazard level. Software validation and traceability of the requirements related to RUO and Administration were completed on the IVD aspects of the application only. Requirements related to other functionality were to be tested and traced prior to launch. Immunicon Corporation will submit to FDA the validation, traceability documentation and software report upon completion. F. Regulatory Information: 1. Regulation section: 21 CFR 866.6020-Immunomagnetic Circulating Cancer Cell Selection and Enumeration System 2. Classification: Class II 3. Product code: NQI, System, Immunomagnetic, Circulating Cancer Cell, Enumeration 4. Panel: Immunology (82) G. Intended Use: 1. Indication(s) for Use: The Immunicon CellTracks® Analyzer II is a semi-automated fluorescence microscope used to enumerate fluorescently labeled cells that are immunomagnetically selected and aligned. The product is for in vitro diagnostic use when used in tandem with specimen preparation equipment and reagents that 2 {2} are legally marketed for in vitro diagnostic use with this device. 2. Special Conditions for Use Statement(s): For prescription use only H. Substantial Equivalence Information: 1. Predicate Device Name(s) and 510(k) numbers: CellTracks® Analyzer II (k050145) 2. Comparison with Predicate Device: | Similarities | | | | --- | --- | --- | | Item | New Device | Predicate | | Name | CellTracks® Analyzer II | CellTracks® Analyzer II | | Sample | Processed via MagNest® fixture | Same | | Available channels for analysis | 4 | Same | | Sample Holder | MagNest | Same | | Power | 100-240 Vac 50-60 Hz. | Same | | Differences | | | | Item | New Device | Predicate | | Computer Operating system^{1} | Mandrake Linux | MS Windows XP | | Graphical User Interface^{2} | Developed to be Linux compatible | MS Windows XP | ¹The Operating system (OS) performs basic tasks such as recognizing input from the keyboard, sending output to the display screen, keeping track of files and directories on the disk, and controlling peripheral devices such as disk drives and printers. All the algorithms associated with image acquisition, analysis, cell selection, review, reporting and archiving; the logic and interface to the PC remain the same with the replacement of the OS. ² The Graphical User Interface (GUI) is a program interface that uses a computer's graphic capabilities to make the program easier to use. Graphical interfaces use a pointing device to select objects, including icons, menus, text boxes, etc. A GUI includes standard formats for representing text and graphics. This allows users to compile defined data in a manner more usable to the operator. It does not allow them to alter the basic cell definition, count or image quality. I. Special Control/Guidance Document Referenced (if applicable): The CellTracks was developed in conformance to the following standards. ISO 14971 Medical Devices- Application of Risk Management to Medical Devices FDA Guidance: General Principles of Software Validation EP5-A NCCLS document: Evaluation of Precision Performance of Clinical Chemistry {3} Devices EP9-A NCCLS document: Method comparison and Bias Estimation Using Patient Samples All requirements for these standards were met. EP9-A testing was performed using donor spiked samples rather than actual cancer patient samples. In addition, the CellTracks Analyzer II will conform to the following two standards prior to marketing. IEC 61010-1 IEC: Safety requirements for Electrical Equipment for Measurement, Control and Laboratory Use Part 1 BS EN61326 Electrical Equipment for Measurement, Control and Laboratory Use - EMC requirements. # J. Performance Characteristics: # 1. Analytical Performance: # a. Accuracy Accuracy was demonstrated by a recovery study. $7.5\mathrm{mL}$ of pooled donor blood was pipetted into 12 AutoPrep sample tubes. Six of the 12 tubes were spiked with 50 SKBR-3 cells/ $50\mu \mathrm{L}$ and the other 6 were spiked with the 300 SKBR-3 cells/ $100\mu \mathrm{L}$ . The Veridex LLC CellSearch™ Epithelial Cell kit/Cell Spotter Analyzer, (k031588) was used to assay the samples. Three of the 50 cell count samples were scanned first with the new OS version Linux 1. The same samples were rescanned with the predicate OS version 1.2 (XP). The other three 50 cell count samples were first scanned with the predicate OS version 1.2 (XP) and rescanned with new OS version Linux 1. The same process was performed with the 300 cell count samples. Equivalent recovery was obtained with both operating systems (see table below). | | New device | | Predicate | | | --- | --- | --- | --- | --- | | Cell counts | 50 | 300 | 50 | 300 | | Mean | 53 | 304 | 53 | 304 | | SD | 9.2 | 12.5 | 8.5 | 13.7 | | % CV | 17.3 | 4.1 | 16.0 | 4.5 | Substantial equivalence was also demonstrated by comparing bead counts of high and low controls between the two devices. Six high control samples and 6 low control samples were pipetted into 12 AutoPrep sample tubes. Each control sample cartridge was scanned twice. Half of the low control samples were first scanned with the predicate version 1.2 (XP) and rescanned with the new device version Linux 1. The other half of the low control samples were first scanned with the new version Linux 1 and rescanned with predicate version 1.2(XP). The same process was done with the high control samples. According to the control product labels, the control cell ranges were: High control 632-922 beads Low control 15-65 beads {4} The following results were obtained: Low control comparison (Linux) = 1.0588 (1.2 XP) - 1.4118 $\mathrm{R}^2 = 0.9608$ High control comparison (Linux) = 1.0748 (1.2 XP) - 45.652 $\mathrm{R}^2 = 0.9939$ b. Precision/Reproducibility: Precision was measured using a single bead cartridge (approximately 100 beads) scanned 10 times with version 1.2 (XP) software and 10 times with version 1 Linux software on two different CellTracks® II Analyzers. The fluorescent beads do not photo-bleach and can be used to compare multiple scans. Results showed precisions seen with both operating systems are substantially equivalent. Summary of Precision Analyses | | New device | Predicate | | --- | --- | --- | | Mean | 98 | 98 | | SD | 0.49 | 0.48 | | % CV | 0.49 | 0.49 | c. Linearity Linear regression analysis was performed using five cartridges each containing the following number of fluorescent beads: 0, 5, 99, 559 and 950 into $320~\mu \mathrm{L}$ of $60\%$ glycerol as diluent. 1000-2000 glacial blue beads were added to each cartridge to provide enough UV events for focusing. Only glacial blue beads were added to the 0 cartridge. Fluorescent bead count for each cartridge was verified using the CellSpotter system (Veridex LLC CellSearch™ Epithelial Cell kit/Cell Spotter Analyzer, k031588). Each cartridge were scanned with OS version 1.2 (XP) and rescanned with OS version1 (Linux). The following results were obtained: $\mathrm{y} = 1.0009\mathrm{x} + 0.0958$ $\mathrm{R}^2 = 1.0$ Conclusion: The new device version is substantially equivalent to the predicate device and linear to 950 fluorescent beads in 320uL diluent. c. Carryover: Refer to predicate device k050145 for carryover studies. The changes made to the instrument were OS related. This should have no effect on carryover. d. Interfering Substances: Refer to predicate device k050145 for interference studies. The changes made to the instrument were OS related and there should be no effect on interference. Thus, no new interfering substances studies were needed. {5} K. Proposed Labeling: The data indicate substantial equivalence between the new and the old operating systems. Thus, no changes need to be made to the labeling. The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. L. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 6
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...