K060052 · Ge Medical Systems, LLC · JAK · Apr 19, 2006 · Radiology
Device Facts
Record ID
K060052
Device Name
GE LIGHTSPEED XTRA CT SCANNER SYSTEM
Applicant
Ge Medical Systems, LLC
Product Code
JAK · Radiology
Decision Date
Apr 19, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The GE LightSpeed Xtra CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. The system is capable of generating images for the guidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology. The system allows imaging of obese patients, up to and including the morbidly obese population (BMI > 40). When used in the LightSpeed RT Pro16 configuration: The system acquires CT anatomical images that are clinically useful in the simulation and planning of radiation therapy for the treatment of cancer.
Device Story
LightSpeed Xtra CT Scanner System comprises gantry, patient table, operator console, and PDU. System acquires X-ray data at multiple angles in axial/helical modes; computer reconstructs data into cross-sectional images. Features 80cm wide bore for large patients and radiation therapy immobilization devices. Operated by clinicians in clinical settings for diagnostic imaging, interventional guidance, and radiation therapy planning. Output images assist physicians in tumor/pathology localization, biopsy/ablation guidance, and cancer treatment simulation. Benefits include improved access for interventional procedures and accommodation of morbidly obese patients.
Clinical Evidence
No clinical data. Substantial equivalence based on bench testing and comparison of technological characteristics to predicate devices.
Technological Characteristics
Computed tomography X-ray system; 80cm wide bore; 16-slice acquisition; 0.5s rotation speed; 20mm axial coverage. Complies with 21 CFR 1020.30/1020.33, UL 60601-1, and IEC 60601-1. Includes gantry, table, console, and PDU.
Indications for Use
Indicated for head and whole-body CT imaging in patients including the morbidly obese (BMI > 40). Used for guidance of minimally invasive procedures (biopsy, ablation) and radiation therapy simulation/planning for cancer treatment.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a black and white logo of the company General Electric (GE). The logo is a circular shape with the letters "GE" in a stylized font in the center. The letters are connected and appear to be handwritten. The background of the circle is filled with a pattern of dots, giving it a textured appearance.
APR 1 9 2006
GE Healthcare Technologies
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
#### Submitter:
| Name: | GE Medical Systems, LLC (GE Healthcare) |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Address: | 3000 N. Grandview Blvd., W-1140<br>Waukesha, WI 53188 |
| Contact: | Glen E. Sabin<br>FCT Safety and Regulatory Engineer<br>Tel: 262-312-7703, Fax: 262-312-7144<br>e-mail: glen.sabin@med.ge.com |
| Date Prepared: | 11 November 2005 |
| Product Identification: | |
| Proprietary Device Names: | GE LightSpeed Vtra CT Scanner System |
Proprietary Device Names:
GE LightSpeed Xtra CT Scanner System Available also as GE LightSpeed RT Pro16
Common Name: Classification Name:
CT Scanner Computed Tomography X-ray System (21 CFR 892.1750, Product Code JAK)
# Predicate Device(s):
GE LightSpeed 5.0 CT Scanner System
K030420
# Device Description:
The LightSpeed Xtra CT Scanner System is composed of a gantry, patient table, operator console, power distribution unit (PDU), and interconnecting cables. The system includes image acquisition hardware, image acquisition and reconstruction software, and associated accessories.
The system generates images through the computer reconstruction of data acquired at different angles of the same axial plane. The gantry can rotate at up to 0.5 seconds per rotation, and can acquire up to 16 slices of data with a maximum total coverage of 20mm in the axial direction. The system may be operated in both axial and helical scan modes.
The system features an 80cm diameter wide bore to accommodate large patients and radiation therapy planning immobilization devices, to allow easy access during interventional procedures, and for ease of patient positioning.
E. Medical Systems, LLC
{2}------------------------------------------------
The LightSpeed Xtra system is also available as the LightSpeed RT ProJ6, which features a unique accessory package designed to assist planning of radiation therapy procedures.
#### Intended Use:
The GE LightSpeed Xtra CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications.
The system is capable of generating images for the quidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology.
The system allows imaging of obese patients, up to and including the morbidly obese population (BMI > 40).
When used in the LightSpeed RT Pro16 configuration:
The system acquires CT anatomical images that are clinically useful in the simulation and planning of radiation therapy for the treatment of cancer.
#### Comparison with Predicate:
The LightSpeed Xtra CT Scanner System represents a modification to the legally marketed LightSpeed 5.0 scanner (510lk) number K030420). The LightSpeed Xtra is an evolutionary change to the LightSpeed product line, and includes many features, functions, software, and hardware that are common to previous generations, including the LightSpeed 4.0 (K013561) 16 slice system. the LightSpeed 3.0 (K002978) 8 slice system, and LightSpeed 2.0 (K000300) 4 slice system. It has the same technological characteristics and operating principles, is comparable in key safety and effectiveness features, and uses the same basic design, construction, and materials.
ln the opinion of GE Healthcare, the LightSpeed Xtra CT Scanner System is of comparable type and is substantially equivalent to currently marketed head and whole body X-ray computed tomography systems that comply with the same or equivalent standards and have similar intended uses.
# Conclusion:
LightSpeed Xtra is an evolutionary modification to the LightSpeed 5.0 scanner (K030420), does not result in any new potential safety risks, and performs as well as or better than devices currently on the market. LightSpeed Xtra will be certified to comply with the X-ray requirements of 21CFR1020.30 and 1020.33, as well as the safety requirements of UL 60601-1, and IEC 60601-1 and associated collateral and particular standards. GE considers the LightSpeed Xtra CT Scanner System to be equivalent to other marketed devices with similar indications for use and meeting similar standards.
{3}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES .
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be a caduceus, a symbol often associated with healthcare, featuring a staff with a serpent entwined around it.
Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
APR 1 9 2006
GE Medical Systems, LLC (GE Healthcare) % Mr. Tamas Borsai Program Manager, Third Party Review Program TÜV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470
Re: K060052
Trade/Device Name: GE LightSpeed Xtra CT Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulation Number: 21 CFR 892.5840 Regulation Name: Radiation therapy simulation system Regulatory Class: II Product Code: JAK and KPQ Dated: April 3, 2006 Received: April 4, 2006
Dear Mr. Borsai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/2/Picture/10 description: The image shows the FDA Centennial logo, which includes the years 1906-2006. The logo is surrounded by text that reads "U.S. Department of Health & Human Services". Below the logo is the text "Protecting and Promoting Public Health" in a decorative font. The image appears to be a promotional or commemorative item related to the FDA's centennial anniversary.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Kobosz
Device Name:
GE LightSpeed Xtra CT Scanner System
# Indications for Use:
The GE LightSpeed Xtra CT Scanner System is indicated for head and whole body Xray Computed Tomography applications.
The system is capable of generating images for the guidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology.
The system allows imaging of obese patients, up to and including the morbidly obese population (BMI > 40).
When used in the LightSpeed RT Pro16 configuration:
The system acquires CT anatomical images that are clinically useful in the simulation and planning of radiation therapy for the treatment of cancer.
Prescription Use >> AND/OR (21 CFR 801 Subpart D)
Over-the-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K060052
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.